Advanix Biliary Stent with NaviFlex RX Delivery System
K203162 · Boston Scientific Corporation · FGE · Dec 18, 2020 · Gastroenterology, Urology
Device Facts
Record ID
K203162
Device Name
Advanix Biliary Stent with NaviFlex RX Delivery System
Applicant
Boston Scientific Corporation
Product Code
FGE · Gastroenterology, Urology
Decision Date
Dec 18, 2020
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 876.5010
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
Advanix™ Biliary Stent with NaviFlex™ RX Delivery System is intended for delivery of the stent to the biliary tract for drainage of the bile ducts, for splinting of a bile duct during healing, or for providing bile duct patency in a stricture or past a stone.
Device Story
Biliary plastic stent and delivery system; used for biliary drainage, splinting, or maintaining patency in strictures/stones. System includes stent and delivery catheter with locking mechanism; stent features center or duodenal bend shapes, tapered leading tip, and rounded trailing end. Operated by clinicians in clinical settings. Stent prevents duct closure; delivery system facilitates placement via papilla. Device is provided sterile; material change implemented for specific 7F and 8.5F configurations.
Clinical Evidence
Bench testing only. Biocompatibility testing performed: MEM Elution Cytotoxicity, Guinea Pig Maximization, Intracutaneous Reactivity, Acute Systemic Injection, and Implantation. Bench performance testing included: drainage lumen ID, stent dimensions, shape, repositionability, deployment force, trackability, guidewire/duodenoscopy compatibility, barb flap cover compatibility, tensile strength (aged in bile), and lumen patency. All tests met acceptance criteria.
Technological Characteristics
Biliary plastic stent and delivery catheter. Materials evaluated per ISO 10993. Sterilization via Ethylene Oxide (ISO 11135-1, ISO 10993-7). Configurations: 7F and 8.5F, lengths 5-18cm, center or duodenal bend. Mechanical delivery system with locking mechanism.
Indications for Use
Indicated for patients requiring biliary tract drainage, bile duct splinting during healing, or maintenance of bile duct patency in the presence of a stricture or stone.
Regulatory Classification
Identification
A biliary catheter and accessories is a tubular flexible device used for temporary or prolonged drainage of the biliary tract, for splinting of the bile duct during healing, or for preventing stricture of the bile duct. This generic type of device may include a bile collecting bag that is attached to the biliary catheter by a connector and fastened to the patient with a strap.
Special Controls
*Classification.* Class II (special controls). The device, when it is a bile collecting bag or a surgical biliary catheter that does not include a balloon component, is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 876.9.
Predicate Devices
Advanix™ Biliary Stent with NaviFlex™ RX Delivery System (K101314)
Submission Summary (Full Text)
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which consists of the letters "FDA" in a blue square, followed by the words "U.S. FOOD & DRUG" in blue, and then the word "ADMINISTRATION" in a smaller font size below that.
December 18, 2020
Boston Scientific Corporation Elena Hennessey Fellow, Regulatory Affairs Specialist 100 Boston Scientific Wav Marlborough, MA 01752
Re: K203162
Trade/Device Name: Advanix Biliary Stent with NaviFlex RX Delivery System Regulation Number: 21 CFR 876.5010 Regulation Name: Biliary catheter and accessories Regulatory Class: Class II Product Code: FGE Dated: October 21, 2020 Received: October 23, 2020
# Dear Elena Hennessey:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part
{1}------------------------------------------------
801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4. Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
For Thelma I. Valdes, Ph.D. Acting Assistant Director DHT3A: Division of Renal, Gastrointestinal, Obesity and Transplant Devices OHT3: Office of GastroRenal, ObGyn, General Hospital and Urology Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
# Indications for Use
510(k) Number (if known)
K203162
#### Device Name
Advanix™ Biliary Stent with NaviFlex™ RX Delivery System
#### Indications for Use (Describe)
Advanix™ Biliary Stent with NaviFlex™ RX Delivery System is intended for delivery of the stent to the biliary tract for drainage of the bile ducts, for splinting of a bile duct during healing, or for providing bile duct patency in a stricture or past a stone..
| Type of Use (Select one or both, as applicable) |
|-------------------------------------------------|
|-------------------------------------------------|
| <span style="text-decoration: underline;">X</span> Prescription Use (Part 21 CER 801 Subpart D) | <span style="text-decoration: underline;"> </span> Over-The-Counter Use (21 CFR 801 Subpart C) |
|-------------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------|
|-------------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------|
## CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
## *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
K203162 Page 1 of 2
## SECTION 5 510(K) SUMMARY
#### 510(k) SUMMARY
1. Submitter:
Boston Scientific Corporation 100 Boston Scientific Way Marlborough, MA 01752
Contact: Elena Hennessev Fellow Regulatory Affairs Specialist Tel: 508-683-4347 Date Prepared: December 15, 2020
## 2. Proposed Device:
Trade Name: Advanix™ Biliary Stent with NaviFlex™ RX Delivery System Common Name: Stents, Drains And Dilators For The Biliary Ducts Regulation Name: Biliary Catheters and Accessories Regulation Number: 876.5010 Product Code: FGE Classification: Class II
## 3. Predicate Device:
### Predicate Device:
Trade Name: Advanix™ Biliary Stent with NaviFlex™ RX Delivery System Common Name: Stents, Drains And Dilators For The Biliary Ducts Regulation Name: Biliary Catheters and Accessories Regulation Number: 876.5010 Product Code: FGE Classification: Class II 510(k) Clearance Number: K101314
#### 4. Device Description:
The Advanix™ Biliary Stent with NaviFlex™ RX Delivery System consists of biliary plastic stents and delivery system catheters. They are sold separately or in a pre-loaded configuration, in which a stent comes attached to the catheter via a suture. The system is comprised of two (2) primary components: stent and delivery catheter with a locking mechanism. The Advanix Biliary Stent allows for drainage of the biliary duct by preventing closure and maintaining the patency of the biliary duct. These biliary stents are provided in center bend and duodenal bend shapes and have leading and trailing bars, a tapered leading end tip to facilitate access through the papilla, and a rounded tailing end to match the profile of the push catheter portion of the delivery system. The Advanix Biliary Stent with NaviFlex RX Delivery System is currently cleared under K101314. This proposed material change will only impact the Advanix Biliary 7F and 8.5F stent singles and the 7F and 8.5F pre-loaded configurations as shown below in Table 4-1:
{4}------------------------------------------------
| | Advanix Biliary Stent | |
|---------------------|--------------------------------------------------------|--|
| Shape | • Center Bend | |
| | • Duodenal Bend | |
| Sizes (Fr) | • 7F | |
| | • 8.5F | |
| Lengths (cm) | • 5cm | |
| | • 7cm | |
| | • 9cm | |
| | • 12cm | |
| | • 15cm | |
| | • 18cm | |
| Single or Stent Set | • Center Bend Single Stent (no delivery system) | |
| | • Duodenal Bend Single Stent (no delivery system) | |
| | • Center Bend Pre-loaded (stent and delivery system) | |
| | • Duodenal Bend Pre-loaded (stent and delivery system) | |
## Table 4-1 - Advanix Biliary Stent Material Change Stent Configurations:
## 5. Indications for Use:
The proposed Advanix™ Biliary Stent with NaviFlex™ RX Delivery System is intended for delivery of the stent to the biliary tract for drainage of the bile ducts, for splinting of a bile duct during healing, or for providing bile duct patency in a stricture or past a stone.
## 6. Technological Characteristics:
The proposed Advanix™ Biliary Stent with NaviFlex™ RX Delivery System is identical to the predicates in design, intended use and all materials with the exception of the stent material.
## 7. Performance Data:
The proposed device meets the requirements of ISO 10993 "Biological Evaluation of Medical Devices – Part 1: Evaluation and Testing within a risk management process", ISO 11135-1 "Sterilization of Health Care products - Ethylene Oxide - Part 1: Requirements for Development, Validation, and Routine Control of Sterilization processes for Medical Devices", and ISO 10993-7 "Biological evaluation of medical devices -Part 7: ethylene oxide sterilization residuals".
The following Biocompatibility tests were performed on the Advanix™ Biliary Stent with NaviFlex™ RX Delivery System: MEM Elution Cytotoxicity; Guinea Pig Maximization; Intracutaneous Reactivity; Acute Systemic Injection; and Implantation. In addition, the results of chemical extractable studies are provided which meet FDA's Guidance document, entitled, "Use of International Standard ISO 10993-1, "Biological evaluation of medical devices – Part 1: Evaluation and testing within a risk management process."
The proposed device bench testing is in alignment with FDA guidance document, "Recommended Content and Format of Non-Clinical Bench Performance Testing Information in Premarket Submissions". The following bench tests were performed on the Advanix™ Biliary Stent with NaviFlex™ RX Delivery System: Drainage Lumen ID; Stent Length; Stent OD; Stent Shape; Repositionability (pre-loaded system only); Deployment Force; Device Trackability Force; Guidewire Compatibility; Duodenoscopy Compatibility; Barb Flap Cover Compatibility; Tensile: Aged in Bile; and Lumen Patency. All tests met required acceptance criteria, therefore supporting a finding of substantial equivalence.
## 8. Conclusion:
The information Boston Scientific Corporation provided in this submission demonstrates that the proposed Advanix Biliary Stent with NaviFlex RX Delivery System is substantially equivalent to the currently cleared Advanix Biliary Stent with NaviFlex RX Delivery System (K101314).
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.