ADVANIX BILIARY STENT WITH NAVIFLEX RX DELIVERY SYSTEM
Applicant
Boston Scientific Corp
Product Code
FGE · Gastroenterology, Urology
Decision Date
Jul 23, 2010
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 876.5010
Device Class
Class 2
Attributes
Therapeutic
Intended Use
Advanix™ Biliary Stent with NaviFlex™ RX Delivery System is intended for delivery of the stent to the biliary tract for drainage of the bile ducts, for splinting of a bile duct during healing, or for providing bile duct patency in a stricture or past a stone.
Device Story
Advanix Biliary Stent with NaviFlex RX Delivery System; plastic biliary stent and delivery catheter. Stent provides drainage of biliary duct/gallbladder; maintains patency; prevents closure. Stent shapes include center bend and duodenal bend; tapered leading end for papilla access; rounded trailing end. Delivery system includes guide catheter and outer push catheter with locking mechanism; radiopaque markers for visualization. Used in clinical setting by physicians; delivered to biliary tract. Stent constructed of polyethylene; guide catheter of polyvinylidene fluoride; push catheter of polyether block amide. Device facilitates bile flow; aids healing; manages strictures/stones.
Clinical Evidence
Bench testing only. Evaluated per EN ISO 10993-1:2009 for biocompatibility (cytotoxicity, sensitization, reactivity, systemic toxicity, genotoxicity, implantation). Performance testing included drainage lumen ID, stent dimensions, repositionability, deployment/trackability force, compatibility (guidewire, duodenoscope, caps), and mechanical integrity (tensile strength, buckling resistance, marker band strength). All tests passed.
Technological Characteristics
Stent: Polyethylene. Guide catheter: Polyvinylidene fluoride. Push catheter: Polyether block amide. Radiopaque markers included. Diameters: 7Fr, 8.5Fr, 10Fr. Lengths: 5cm-18cm. Delivery system working length: 202.5cm. Mechanical delivery via RX (rapid exchange) catheter system with locking mechanism. Biocompatibility per EN ISO 10993-1:2009.
Indications for Use
Indicated for patients requiring biliary drainage, bile duct splinting during healing, or maintenance of bile duct patency in the presence of strictures or stones.
Regulatory Classification
Identification
A biliary catheter and accessories is a tubular flexible device used for temporary or prolonged drainage of the biliary tract, for splinting of the bile duct during healing, or for preventing stricture of the bile duct. This generic type of device may include a bile collecting bag that is attached to the biliary catheter by a connector and fastened to the patient with a strap.
Special Controls
*Classification.* Class II (special controls). The device, when it is a bile collecting bag or a surgical biliary catheter that does not include a balloon component, is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 876.9.
Predicate Devices
Microvasive® Biliary Stent and Delivery System (K965147)
Related Devices
K203162 — Advanix Biliary Stent with NaviFlex RX Delivery System · Boston Scientific Corporation · Dec 18, 2020
{0}------------------------------------------------
K 101314
Page 1 of 2
SECTION 5 510(k) SUMMARY
#### 510(k) SUMMARY
## 1. Submitter:
Boston Scientific Corporation 100 Boston Scientific Way Marlborough, MA 01752 Telephone: 508-683-4347 Fax: 508-683-5939
Contact: Elena Nieves Senior Regulatory Affairs Specialist Date Prepared: May 10, 2010
### 2. Proposed Device:
Trade Name: Advanix™ Biliary Stent with NaviFlex™ RX Delivery System Classification Name: Biliary Catheters and Accessories Regulation Number: 876.5010 Product Code: FGE Classification: Class II
#### 3. Predicate Device:
Trade Name: Microvasive® Biliary Stent and Delivery System Manufacturer and Clearance Number: Boston Scientific Corporation, K965147 Classification Name: Biliary Catheters and Accessories Regulation Number: 876.5010 Product Code: FGE Classification: Class II
# 4. Proposed Device Description:
The proposed Advanix™ Biliary Stent with NaviFlex™ RX Delivery System consists of biliary plastic stents and delivery system catheters. They are sold separately or in a pre-loaded configuration, in which a stent comes attached to the catheter via a suture. The system is comprised of two (2) primary components: stent and delivery catheter with a locking mechanism. The Advanix Biliary Stent allows for drainage of the biliary duct and gall bladder by preventing closure and maintaining the patency of the biliary duct. These biliary stents are provided in center bend and duodenal bend shapes and have leading and trailing bars, a tapered leading end tip to facilitate access through the papilla, and a rounded tailing end to match the profile of the push catheter portion of the delivery system.
The Advanix Biliary Stent will be offered in 7Fr. 8.5Fr, and 10Fr diameters. The stent leneths for each diameter vary from 5cm-18cm stent lengths. The Advanix biliary stents are constructed of a Polyethylene material.
The NaviFlex RX Delivery System will be offered in 202.5cm working length. The delivery system consists of a guide catheter and outer push catheter. The guide catheter is constructed of Polyvinylidene fluoride material. The guide catheter also includes a radiopaque marker. The push catheter is composed of polyether block amide material and also has a Nylon repositioning suture.
Special 510(k) Premarket Notification, Advanix™ Biliary Stent with NaviFlex™ RX Delivery System Proprietary and Confidential Information of Boston Scientific Corporation
JUL 2 3 2010
{1}------------------------------------------------
# 5. Intended Use:
Advanix™ Biliary Stent with NaviFlex™ RX Delivery System is intended for delivery of the stent to the biliary tract for drainage of the bile ducts, for splinting of a bile duct during healing, or for providing bile duct patency in a stricture or past a stone.
# 6. Technological Characteristics:
The proposed Advanix™ Biliary Stent with NaviFlex™ RX Delivery System has the same technological characteristics as the predicate Microvasive® Biliary Stent and Delivery System (K965147).
The proposed device has the same intended use and is placed using the same methodology as the predicate device. Both the proposed and predicate device function in the same manner allowing for biliary drainage through the lumen.
The shape of the proposed Advanix Biliary Stent is the same as the predicate Microvasive Biliary Stent and Delivery System and also includes an additional center bend stent shape.
The materials of the proposed Advanix Biliary Stent have been modified to the polymer materials mentioned above in part 4 of this 510(k) Summary.
## 7. Performance Data:
In-vitro testing has been performed and all components, subassemblies, and/or full devices met the required specifications for the completed tests.
The proposed Advanix™ Biliary Stent was evaluated in accordance with EN ISO 10993-1:2009. The following tests were performed on the stent: Cytotoxicity, Sensitization, Intracutaneous Reactivity, Systemic Toxicity -Acute Systemic Toxicity, Subacute Toxicity - Intravenous and Intraperitoneal, Genotoxicity - Ames Assay and Mouse Lymphoma, Implantation, and USP Physicochemical.
The NaviFlex™ RX Delivery System was evaluated in accordance with EN ISO 10993-1:2009. The following tests were performed on the delivery system: Cytotoxicity, Sensitization, Intracutaneous Reactivity, and USP Physicochemical.
The following tests were conducted on the Advanix™ Biliary Stent with NaviFlex™ RX Delivery System: Drainage Lumen ID, Stent Length, Stent OD, Stent Shape, Repositionability, Deployment Force, Trackability Force, Guidewire Compatibility, Duodenoscope Compatibility, Barb Flap Cover Compatibility, Biopsy Cap Compatibility, Delivery System Working Length, Delivery System OD, Guide Catheter Pushability, RX Port Buckling Resistance, Locking Mechanism Retention Ability, Locking Mechanism to Dual Lumen Catheter Tensile Strength, Dual Lumen Catheter to Single Lumen Catheter Tensile Strength, Pullwire Assembly to Guide Catheter Tensile Strength, Pullwire Tensile Strength, Guide Catheter RO Marker Band Strength, and Push Catheter RO Marker Band Strength.
#### 8. Conclusion:
All biocompatibility tests conducted on the Advanix™ Biliary Stent with NaviFlex™ RX Delivery System passed. Therefore, the Advanix Biliary Stent with NaviFlex RX Delivery System is considered biocompatible for its intended use.
All device bench test results were acceptable. The data demonstrate that the Advanix™ Biliary Stent with NaviFlex™ RX Delivery System sufficiently meets the design specifications and is suitable for the intended use.
Boston Scientific Corporation has demonstrated that the proposed Advanix™ Biliary Stent with NaviFlex™ RX Delivery System is substantially equivalent to Boston Scientific Corporation's currently marketed Microvasive® Biliary Stent and Delivery System (K965147).
{2}------------------------------------------------
Image /page/2/Picture/0 description: The image shows a circular seal with the words "DEPARTMENT OF HEALTH & HUM. S." written around the edge. Inside the circle is a stylized image of an eagle with its wings spread. The image is black and white and appears to be a scan of a document.
Food and Drug Administration 10903 New Hampshire Avenue Document Mail Center - WO66-G609 Silver Spring, MD 20993-0002
Ms. Elena Nieves Sr. Regulatory Affairs Specialist Boston Scientific Corporation 100 Boston Scientific Way MARLBOROUGH MA 01752
JUL 2 3 2010
Re: K101314
Trade/Device Name: Advanix™ Biliary Stent with NaviFlex™ RX Delivery System Regulation Number: 21 CFR& 876.5010 Regulation Name: Biliary catheter and accessories Regulatory Class: II Product Code: FGE Dated: May 10, 2010 Received: May 11, 2010
Dear Ms. Nieves:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related
ः ।
{3}------------------------------------------------
Page 2 -
adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the aurerse overns (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please rryou desire specific da 100 too too too too too fices/CDRH/CDRHOffices/ucm115809.htm for the Center for Devices and Radiological Health's (CDRH's) Office of Compliance. Also, please the Concer 101 Donece and the beginne by reference to premarket notification" (21CFR Part note the regalation regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Tou may other of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm.
Sincerely yours,
Janine M. Morris
Acting Director, Division of Reproductive, Abdominal, and Radiological Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{4}------------------------------------------------
# INDICATIONS FOR USE STATEMENT
510(k) Number (if known):
Device Name:
Indications for Use:
K101314
AdvanixTM Biliary Stent with NaviFlex™ RX Delivery System
Advanix™ Biliary Stent with NaviFlex™ RX Delivery System is intended for delivery of the stent to the biliary tract for drainage of the bile ducts, for splinting of a bile duct during healing, or for providing bile duct patency in a stricture or past a stone.
Prescription Use X (Part 21 CFR 801 Subpart D) Over-The-Counter Use (21 CFR 807 Subpart C)
(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)
AND/OR
Concurrence of CDRH, Office of Device Evaluation (ODE)
Hebert Demir
(Division Sign-Off) (Division Sign-Ont)
Division of Reproductive, Abdominal, and Radiological Device 510(k) Number
Panel 1
/
Sort by
Ready
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.