ADVANIX PANCREATIC STENT AND NAVIFLEX RAPID EXCHANGE PANCREATIC DELIVERY SYSTEM AND PUSHERS
K133700 · Boston Scientific Corporation · FGE · May 14, 2014 · Gastroenterology, Urology
Device Facts
Record ID
K133700
Device Name
ADVANIX PANCREATIC STENT AND NAVIFLEX RAPID EXCHANGE PANCREATIC DELIVERY SYSTEM AND PUSHERS
Applicant
Boston Scientific Corporation
Product Code
FGE · Gastroenterology, Urology
Decision Date
May 14, 2014
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 876.5010
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
The Advanix™ Pancreatic Stent and NaviFlex™ Rapid Exchange (RX) Pancreatic Delivery System and Pushers are intended for delivery of the stent to the pancreatic duct (PD): · Used to drain pancreatic ducts
Device Story
Device consists of pancreatic stents and delivery systems/pushers for pancreatic duct (PD) drainage. Stents are straight or single pigtail, featuring radiopaque (RO) polymer construction, optional barbs, and markers for fluoroscopic/endoscopic visualization. Delivery system includes guide catheter, pull-wire, and locking mechanism for controlled deployment; pushers facilitate stent advancement. Used by physicians in clinical settings during endoscopic procedures. Device provides mechanical drainage of PD; radiopacity assists physician in accurate placement under fluoroscopy. Benefits include effective ductal drainage and ease of deployment/removal.
Clinical Evidence
Bench testing only. Evaluated per EN ISO 10993-1:2009 for biocompatibility (cytotoxicity, irritation, sensitization, acute systemic toxicity, subchronic toxicity, genotoxicity, implant, USP physicochemical, latex). Mechanical performance testing included drainage lumen ID, stent length/diameter, stiffness, tensile strength, pusher/delivery system working length, and deployment/trackability forces.
Technological Characteristics
Stents and pushers constructed of radiopaque (RO) polymer. Stent sizes 3F-10F; pusher OD 3F-5F. Features include straight or single pigtail shapes, optional leading/trailing barbs, and endoscopic/fluoroscopic markers. Delivery system utilizes guide catheter, pull-wire, and locking mechanism. Sterilization method not specified. No software or electronic components.
Indications for Use
Indicated for patients requiring pancreatic duct drainage via stent placement.
Regulatory Classification
Identification
A biliary catheter and accessories is a tubular flexible device used for temporary or prolonged drainage of the biliary tract, for splinting of the bile duct during healing, or for preventing stricture of the bile duct. This generic type of device may include a bile collecting bag that is attached to the biliary catheter by a connector and fastened to the patient with a strap.
Special Controls
*Classification.* Class II (special controls). The device, when it is a bile collecting bag or a surgical biliary catheter that does not include a balloon component, is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 876.9.
Advanix™ Biliary Stent with NaviFlex™ RX Delivery System (K101314)
Submission Summary (Full Text)
{0}------------------------------------------------
# MAY 1 4 2014
## SECTION 5 510(k) SUMMARY
#### 1. Submitter
Boston Scientific Corporation 100 Boston Scientific Way Marlborough, MA 01752
Contact: Elena Nieves Principal, Regulatory Affairs Specialist Telephone: 508-683-4430 Fax: 508-683-5939
Date Prepared: November 26, 2013
### 2. Device
Trade Name:
Classification Name: Product Code: Device Class and panel: Classification Regulation:
### 3. Predicate Devices
Trade Name: Trade Name: Manufacturer and Clearance Number: Classification Name: Product Code: Device Class and panel: Classification Regulation:
Trade Name:
Manufacturer and Clearance Number: Classification Name: Product Code: Device Class and panel: Classification Regulation:
Advanix™ Pancreatic Stent and NaviFlex™ Rapid Exchange (RX) Pancreatic Delivery System and Pushers Biliary Catheters and Accessories FGE - Biliary Catheter and accessory Class II, Gastroenterology/Urology CFR 876.5010
Zimmon Endoscopic Pancreatic Stent Geenen Pancreatic Stent Wilson-Cook, K900923 Biliary Catheters and Accessories FGE - Biliary Catheter and accessory Class II, Gastroenterology/Urology CFR 876.5010
Advanix™ Biliary Stent with NaviFlex™ RX Delivery System Boston Scientific Corporation, K101314 Biliary Catheters and Accessories FGE - Biliary Catheter and accessory Class II, Gastroenterology/Urology CFR 876.5010
### 4. Device Description
The Advanix™ Pancreatic Stent and NaviFlex™ Rapid Exchange (RX) Pancreatic Delivery System and Pushers product consists of pancreatic stents and delivery systems or stent pushers.
{1}------------------------------------------------
The pancreatic stents are provided in straight or single pigtail shape. The straight shape stents have trailing barbs and/or leading barbs depending on application, a rounded or tapered leading end tip to facilitate access through the papilla, and a rounded trailing end to abut the push catheter portion of the delivery system or stent pusher. The single pigtail stent may or may not have leading end barbs depending on application. Some stents have lateral drainage holes in the pigtails, a tapered or rounded leading end tip, and a rounded trailing end. All stents have either an endoscopic marker, fluoroscopic marker, or both on the trailing or leading end of the stent to assist with depth of placement in the pancreatic duct. The location and presence of an endoscopic or fluoroscopic marker is dependent on the length and diameter size of the stent. Some codes have side port holes in the body of the stent.
## 5. Indication for Use:
The Advanix™ Pancreatic Stent and NaviFlex™ Rapid Exchange (RX) Pancreatic Delivery System and Pushers is intended for delivery of the stent to the pancreatic duct (PD): . · Used to drain pancreatic ducts
### 6. Technological Characteristics:
The Advanix Pancreatic Stent is designed for pancreatic duct (PD) drainage. The stent is constructed of a radiopaque (RO) polymer. The leading end has a tapered or rounded tip and the trailing end of the stent has an endoscopically visible marker to aid in placement. Stent sizes 4F and 5F contain a RO marker while 3F stents do not. The stent is available in two different shapes: straight stent and single pigtail. 3F and 4F stents have no-leading barb; 5F stents are available with and without a leading barb. Stent sizes 7F and 10F contain an RO marker except for the 3cm length stents. The 7F stent is available in two different shapes: straight stent and single pigtail. The 10F stent is only available in the straight stent shape.
The pusher is constructed of RO polymer. The trailing end of the pusher has a white indicator to indicate the non-working end. There are 2 pusher options: RX and longwire. The RX pusher has a guidewire exit marker on the leading end to indicate the guidewire exit location. The wire for the longwire pusher exits at the trailing end of the pusher.
The delivery system consists of a guide catheter connected to a pull-wire, and a push catheter connected to a locking mechanism. The locking mechanism is designed to keep the guide catheter in the position desired while being delivered to the deployment site. The stent can be deployed by disengaging the locking mechanism and retracting the guide catheter.
The delivery system, stent pushers and stent for this product have the following technological characteristics:
- . Range of delivery systems and pushers tailored to the stents
- Range of stents that are deliverable and removable through the endoscope .
- Stents with a pigtail, and trailing and/or leading barbs .
- Radiopaque stent, pusher and delivery system designed to be easily identifiable under . fluoroscopy.
{2}------------------------------------------------
When compared with the predicate devices [Boston Scientific Advanix Biliary Stent with NaviFlex RX Delivery System (K101314), Wilson-Cook Zimmon Pancreatic Stent (K900923), and Wilson-Cook Geenen Pancreatic Stent (K900923)] the proposed Advanix Pancreatic Stent and NaviFlex RX Pancreatic Delivery System and Pushers has the same the technological characteristics as follows: identical indication statement; offered in the same stent lengths that range from 2cm -18cm; offered in the same stent shape configurations (straight and single pigtail; leading and no-leading barb); offered in the same stent outer diameter that range from 3F to 10F; offered in the same Pusher outer diameter (OD) that range from 3F to 5F; offered in the delivery system OD that range from 7F to 10F; and offered in the same delivery system working length of 202.5cm. The proposed devices technological characteristics that are similar in comparison to the predicate device are the stent and delivery system materials, and pusher working length.
### 7. Performance Data:
The proposed Boston Scientific Advanix™ Pancreatic Stent and NaviFlex™ Rapid Exchange (RX) Pancreatic Delivery System and Pushers were evaluated in accordance with EN ISO 10993-1:2009- Evaluation and Testing within a risk management system. The following tests were performed on the stent: Cytotoxicity, Irritation, Sensitization, Acute Systemic Toxicity, Subchronic Toxicity, Genotoxicity, Implant, USP Physicochemical, and Latex.
The delivery system and pushers were evaluated in accordance with EN ISO 10993-1:2009. The following tests were performed on the delivery system and pushers: Cytotoxicity, Sensitization, Irritation, USP Physicochemical and Latex.
The following tests were conducted on the Advanix™ Pancreatic Stent and NaviFlex™ Rapid Exchange (RX) Pancreatic Delivery System and Pushers: Drainage Lumen ID, Stent Length, Stent Outer Diameter, Pigtail Stent Shape, Pigtail Stent Diameter, No-Leading Barb Stent Stiffness, Leading Barb Stent Stiffness, Stent Tensile, Pusher Working Length, Delivery System Working Length, Pusher OD, Delivery System Push Catheter OD, Pusher Deployment Force. Delivery System Deployment Force, Pusher Trackability Force, and Delivery System Trackability Force.
#### 8. Conclusion:
Boston Scientific Corporation has demonstrated that the proposed Boston Scientific AdvanixTM Pancreatic Stent and NaviFlex™ Rapid Exchange (RX) Pancreatic Delivery System and Pushers is substantially equivalent to the currently marketed Boston Scientific Corporation Advanix('M Biliary Stent with NaviFlex™ RX Delivery System (K101314), Wilson-Cook Zimmon Endoscopic Pancreatic Stent (K900923) and Wilson-Cook Geenen Pancreatic Stent (K900923).
{3}------------------------------------------------
Image /page/3/Picture/0 description: The image shows the logo for the Department of Health & Human Services - USA. The logo consists of a circular seal with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" around the perimeter. Inside the circle is a stylized symbol that resembles an abstract caduceus or a representation of human services.
Public Health Service
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002
May 14, 2014
Boston Scientific Corporation Elena Nieves Principle, Regulatory Affairs Specialist 100 Boston Scientific Way Marlborough, MA 01752
Re: K133700
> Trade/Device Name: Advanix™ Pancreatic Stent and NaviFlex™ Rapid Exchange (RX) Pancreatic Delivery System and Pushers Regulation Number: 21 CFR§ 876.5010 Regulation Name: Biliary catheter and accessories Regulatory Class: II Product Code: FGE Dated: April 17, 2014 Received: April 21, 2014
Dear Elena Nieves,
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies.
{4}------------------------------------------------
Page 2 - Elena Nieves
You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please go to http://www.fda.gov/AboutFDA/CentersOffices/CDRH/CDRHOffices/ucm115809.htm for the Center for Devices and Radiological Health's (CDRH's) Office of Compliance. Also, please note the regulation entitled. "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm.
Sincerely yours,
Image /page/4/Picture/7 description: The image shows the name "Benjamin Fisher-S". The name is written in a simple, sans-serif font. The letters are black and the background is white. The letters "FDA" are in the middle of the name.
Benjamin R. Fisher, Ph.D. Director Division of Reproductive, Gastro-Renal, and Urological Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{5}------------------------------------------------
#### SECTION 4 INDICATIONS FOR USE STATEMENT
Indications for Use:
510(k) Number (if known): K133700
Device Name: Advanix™ Pancreatic Stent and NaviFlex™ Rapid Exchange (RX) Pancreatic Delivery System and Pushers
Indications for Use:
The Advanix™ Pancreatic Stent and NaviFlex™ Rapid Exchange (RX) Pancreatic Delivery System and Pushers are intended for delivery of the stent to the pancreatic duct (PD): · Used to drain pancreatic ducts
Prescription Use X (Part 21 CFR 801 Part D) AND/OR
Over-The-Counter Use (21 CFR 807 Subpart C)
(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
Image /page/5/Picture/11 description: The image contains the name "Benjamin R. Fisher -S" on the top line. The second line contains the date "2014.05" and the time "-04'00'". The text appears to be a signature or stamp.
Premarket Notification, Advanix Pancreatic Stent
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.