WALLFLEXTM BILIARY RX STENT SYSTEM (UNCOVERED, PARTIALLY COVERED, AND FULLY COVERED) WALLFLEXTM BILIARY TRANSHEPATIC STE
Applicant
Boston Scientific Corporation
Product Code
FGE · Gastroenterology, Urology
Decision Date
Sep 28, 2012
Decision
SESU
Submission Type
Traditional
Regulation
21 CFR 876.5010
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
The WallFlex™ Biliary RX Stent System Uncovered, WallFlex™ Biliary RX Stent System Partially Covered, and WallFlex™ Biliary RX Stent System Fully Covered are indicated for use in the palliative treatment of biliary strictures produced by malignant neoplasms.
Device Story
Self-expanding metal biliary stent system; includes uncovered, partially covered, and fully covered versions. Stent wires feature radiopaque core; delivery system includes radiopaque marker bands for imaging guidance during deployment. Accommodates 0.035 in (0.89 mm) guidewire. Used by clinicians for palliative treatment of malignant biliary strictures. Device provides mechanical support to maintain biliary patency. This submission updates labeling to include MR Conditional safety status for 1.5 and 3.0 Tesla environments.
Clinical Evidence
No clinical data. Bench testing performed as part of the WallFlex™ Biliary Transhepatic System submission (K112543).
Technological Characteristics
Self-expanding metal biliary stent; uncovered or covered (Permalume™). Radiopaque wire core. Delivery system with radiopaque marker bands. Compatible with 0.035 in guidewire. MR Conditional for 1.5T and 3.0T.
Indications for Use
Indicated for palliative treatment of biliary strictures caused by malignant neoplasms in patients requiring biliary stenting.
Regulatory Classification
Identification
A biliary catheter and accessories is a tubular flexible device used for temporary or prolonged drainage of the biliary tract, for splinting of the bile duct during healing, or for preventing stricture of the bile duct. This generic type of device may include a bile collecting bag that is attached to the biliary catheter by a connector and fastened to the patient with a strap.
Special Controls
*Classification.* Class II (special controls). The device, when it is a bile collecting bag or a surgical biliary catheter that does not include a balloon component, is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 876.9.
Predicate Devices
WallFlex™ Biliary RX Stent System Uncovered (K061231, K081733)
WallFlex™ Biliary RX Stent System Partially Covered (K083374)
WallFlex™ Biliary RX Stent System Fully Covered (K083627)
{0}------------------------------------------------
K122072
pg 1 of 2
SECTION 5 510(k) SUMMARY
#### 510(k) SUMMARY
# 1. Submitter:
SEP 2 8 2012
Boston Scientific Corporation 100 Boston Scientific Way Marlborough, MA 01752 Telephone: 508-683-4872 Fax: 508-683-5939
Contact: Laurie Pannella, RAC Regulatory Affairs Specialist Date Prepared: July 13, 2012
# 2. Proposed Device:
Trade Name: WallFlex™ Biliary RX Stent System Uncovered WallFlex™ Biliary RX Stent System Partially Covered WallFlex™ Biliary RX Stent System Fully Covered Classification Name: Catheter, Biliary, Diagnostic Regulation Number: 876.5010 Product Code: FGE Classification: Class II
## 3. Predicate Device:
Trade Name: WallFlexTM Biliary RX Stent System Uncovered Manufacturer and Clearance Number: Boston Scientific Corporation, K061231, K081733 Classification Name: Catheter, Biliary, Diagnostic Regulation Number: 876.5010 Product Code: FGE Classification: Class II
Trade Name: WallFlex™ Biliary RX Stent System Partially Covered Manufacturer and Clearance Number: Boston Scientific Corporation, K083374 Classification Name: Catheter, Biliary, Surgical Regulation Number: 876.5010 Product Code: FGE Classification: Class II
Trade Name: WallFlex™ Biliary RX Stent System Fully Covered Manufacturer and Clearance Number: Boston Scientific Corporation, K083627 Classification Name: Catheter, Biliary, Surgical Regulation Number: 876.5010 Product Code: FGE Classification: Class II
{1}------------------------------------------------
K122072
pg 20 f 2
Trade Name: WallFlex™ Biliary Transhepatic System Manufacturer and Clearance Number: Boston Scientific Corporation, K112543 Classification Name: Catheter, Biliary, Surgical Regulation Number: 876.5010 Product Code: FGE Classification: Class II
## 4. Proposed Device Description:
The WallFlex™ Biliary RX Stent System Uncovered, WallFlex™ Biliary RX Stent System Partially Covered, and WallFlex 114 Biliary RX Stent System Fully Covered consist of a flexible delivery system preloaded with a self-expanding biliary metal stent.
The stent is offered uncovered or covered. The covered stents are offered as fully covered, or partially covered with a Permalume™ stent covering. The stent wires have a radiopaque core to improve radiopacity.
The stent is preloaded onto the delivery system. which has radiopaque marker bands used to aid in imaging during deployment of the stent. The delivery system accommodates a 0.035 in (0.89 mm) guidewire.
#### 5. Intended Use:
The WallFlex™ Biliary RX Stent System Uncovered, WallFlex™ Biliary RX Stent System Partially Covered, and WallFlex™ Biliary RX Stent System Fully Covered are indicated for use in the palliative treatment of biliary strictures produced by malignant neoplasms.
## 6. Technological Characteristics:
There are no differences in the technological characteristics between the proposed and predicate devices. The purpose of the this Traditional 510(k) is to request a labeling claim that the proposed WallFlex:™ Biliary RX Stent System Uncovered, WallFlex™ Biliary RX Stent System Partially Covered, and WallFlex™ Biliary RX Stent System Fully Covered stents are MR Conditional safe and compatible for both 1.5 and 3.0 Tesla.
#### 7. Performance Data:
In-vitro testing on the biliary metal stent was performed as part of the WallFlex™ Billary Transhepatic System Tradition 510(k) submission (K112543).
#### 8. Conclusion:
Boston Scientific Corporation has demonstrated that the proposed WallFlex™ Biliary RX Stent System Uncovered, WallFlex™ Biliary RX Stent System Partially Covered, and WallFlex™ Biliary RX Stent System Fully Covered are substantially equivalent to Boston Scientific Corporation's currently marketed WallFlex™ Biliary RX Stent System (K061231, K081733, K083374, and K083627), and the WallFlex™ Biliary Transhepatic Stent System (K112543).
Traditional 510(k) Premarket Notification, WallFlex 131 Biliary RX Stent System Uncovered, WallFlex™ Biliary RX Stent System Partially Covered, and WallFlex™ Biliary RX Stent System Fully Covered
{2}------------------------------------------------
# DEPARTMENT OF HEALTH & HUMAN SERVICES
Image /page/2/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a stylized eagle with three tail feathers, representing the department's commitment to health, human services, and well-being. The words "DEPARTMENT OF HEALTH & HUMAN SERVICES • USA" are arranged in a circular pattern around the eagle.
Food and Drug Administration 10903 New Hampshire Avenue Document Control Room -WO66-G609 Silver Spring, MD 20993-0002
Ms. Laurie Pannella Regulatory Affairs Specialist Boston Scientific Corporation 100 Boston Scientific Way MARLBOROUGH MA 01752
# SEP 2 8 2012
Re: K122072
Trade/Device Name: WallFlex™ Biliary RX Stent System Uncovered WallFlex™ Biliary RX Stent System Partially Covered WallFlexTM Biliary RX Stent System Fully Covered Regulation Number: CFR 21§ 876.5010
Regulation Name: Biliary catheter and accessories Regulatory Class: II Product Code: FGE Dated: July 13, 2012 Received: July 16, 2012
Dear Ms. Pannella:
We have reviewed your Section 510(k) premarket notification of intent to market the devices referenced above and have determined the devices are substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments; or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act). You may, therefore, market the device, subject to the general controls provisions of the Act and the limitations described below. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
The Office of Device Evaluation has determined that there is a reasonable likelihood that these devices will be used for an intended use not identified in the proposed labeling and that such use could cause harm. Therefore, in accordance with Section 513(i)(1)(E) of the Act, the following limitation must appear in the Warnings section of the devices' labeling:
The safety and effectiveness of this device for use in the vascular system has not been established.
Furthermore, the indication for palliative treatment of biliary strictures produced by malignant neoplasms must be prominently displayed in all labeling, including pouch box, and carton labels, instructions for use, and other promotional materials, in close proximity to the trade name, of a similar point size, and in bold print.
{3}------------------------------------------------
# Page- 2 - Ms. Laurie Pannella
Please note that the above labeling limitations are required by Section 513(i)(1)(E) of the Act. Therefore, a new 510(k) is required before these limitations are modified in any way or removed from the device's labeling.
The FDA finding of substantial equivalence of your devices to legally marketed predicate devices results in a classification for your devices and permits your devices to proceed to the market. This letter will allow you to begin marketing your devices as described in your Section 510(k) premarket notification if the limitation statement described above is added to your labeling.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your devices comply with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your devices on our labeling regulation (21 CFR Part 801), please go to http://www.fda.gov/AboutFDA/CentersOffices/CDRH/CDRHOffices/ucm115809.htm for the Center for Devices and Radiological Health's (CDRH's) Office of Compliance. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm.
Sincerely yours,
Christy Yeoman
Christy Foreman Director Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{4}------------------------------------------------
# Indications for Use
510(k) Number: K122072
# Device Name: WallFlex™ Biliary RX Stent System Uncovered WallFlex™ Biliary RX Stent System Partially Covered WallFlex™ Biliary RX Stent System Fully Covered
Indications For Use: For use in the palliative treatment of biliary strictures produced by malignant neoplasms
Prescription Use _ X (Part 21 CFR 801 Subpart D) AND/OR
Over-The-Counter Use (21 CFR 807 Subpart C)
(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
Tony. K. Wh
Page 1 of 1
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Report an issue or suggest a feature
Spot a broken citation, a wrong extraction, a missing document, or something you wish this app did? Tell us — no account needed.
or drag and drop files here
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.