WELLFLEX BILIARY RX STENT SYSTEM

K061231 · Boston Scientific Corp · FGE · Sep 1, 2006 · Gastroenterology, Urology

Device Facts

Record IDK061231
Device NameWELLFLEX BILIARY RX STENT SYSTEM
ApplicantBoston Scientific Corp
Product CodeFGE · Gastroenterology, Urology
Decision DateSep 1, 2006
DecisionSESU
Submission TypeTraditional
Regulation21 CFR 876.5010
Device ClassClass 2
AttributesTherapeutic

Indications for Use

The proposed WallFlex Biliary RX Uncovered Stent System is indicated for use in the palliative treatment of biliary strictures produced by malignant neoplasms.

Device Story

The WallFlex Biliary RX Uncovered Stent System is a self-expanding metal stent used for the palliative treatment of malignant biliary strictures. The device consists of a stent made of Platinum-cored Nitinol wires wound into a cylindrical shape with proximal and distal flares. It is mounted on a coaxial sheath delivery catheter. The physician deploys the stent by retracting the exterior tube of the delivery system, allowing the stent to expand within the biliary duct. The device is intended for use by clinicians in a clinical setting. It provides mechanical support to maintain patency in obstructed biliary ducts, thereby relieving symptoms associated with malignant strictures.

Clinical Evidence

Bench testing only. Comparative performance testing included dimensional evaluation, radial force, deployment and reconstrainment force, and bond integrity.

Technological Characteristics

Self-expanding metal stent; Platinum-cored Nitinol wire construction; coaxial sheath delivery catheter; non-implantable delivery system; mechanical deployment mechanism.

Indications for Use

Indicated for palliative treatment of biliary strictures caused by malignant neoplasms.

Regulatory Classification

Identification

A biliary catheter and accessories is a tubular flexible device used for temporary or prolonged drainage of the biliary tract, for splinting of the bile duct during healing, or for preventing stricture of the bile duct. This generic type of device may include a bile collecting bag that is attached to the biliary catheter by a connector and fastened to the patient with a strap.

Special Controls

*Classification.* Class II (special controls). The device, when it is a bile collecting bag or a surgical biliary catheter that does not include a balloon component, is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 876.9.

Predicate Devices

Submission Summary (Full Text)

{0}------------------------------------------------ Boston Scientific Corporation August 7, 2006 KD61231 21 of 2 # SECTION 5 510(K) SUMMARY #### 510(K) SUMMARY # 1. Submitter: Boston Scientific Corporation 100 Boston Scientific Way Marlborough, MA 01760 Telephone: 508-683-4356 Fax: 508-683-5939 Contact: Allyson Barford, RAC Regulatory Affairs Specialist Date Prepared: August 7, 2006 ### 2. Device: Trade Name: WallFlex™ Biliary RX Uncovered Stent System Common Name: Biliary Stent Classification Name: Biliary Catheter and Accessories Regulation Number: 876.5010 Product Code: FGE Classification: Class II # 3. Predicate Device: Boston Scientific Corporation's Wallstent Biliary Endoprosthesis, K000308, K993232, K982184, K964119 and K925406 Boston Scientific Corporation's Wallstent RX Biliary Endoprosthesis, K012752 and K030107 The predicate devices are class II devices per 21 CFR 876.5010 #### 4. Device Description: The proposed WallFlex™ Biliary RX Uncovered Stent System consists of a selfexpanding metal stent and a delivery catheter. The proposed stent consists of Platinum cored Nitinol wires wound together to form a cylinder with both proximal and distal flares. The proposed stent is mounted onto a sheath delivery system. The groposed delivery system is a coaxial tubing assembly that constrains the stent onto the delivery catheter shaft until the stent is released by retracting the exterior tube. {1}------------------------------------------------ # 5. Intended Use: The proposed WallFlex Biliary RX Uncovered Stent System is indicated for use in the palliative treatment of biliary strictures produced by malignant neoplated 1. # 6. Technological Characteristics: The proposed WallFlex Biliary RX Uncovered Stent System has similar technological characteristics to the currently marketed Wallstent Biliary Endoprosthesis (K000308, K993232, K982184, K964119 and K925406) and Wallstent RX Biliary Endoprosthesis (K012752 and K030107). ## 7. Performance Data: Comparative performance testing was completed to establish substantial equivalence between the proposed WallFlex™ Biliary RX Uncovered Stent System and the predicate devices. This testing included but was not limited to dimensional evaluation, radial force, deployment and reconstrainment force and bond integrity. ## 8. Conclusion: Boston Scientific Corporation has demonstrated that the proposed WallFlex Biliary RX Uncovered Stent System is substantially equivalent to Boston Scientific Corporation's currently marketed Wallstent RX Biliary Endoprosthesis and Wallstent Boildy Endoprosthesis. {2}------------------------------------------------ Image /page/2/Picture/1 description: The image shows the logo for the U.S. Department of Health and Human Services. The logo consists of a stylized eagle with three lines representing its body and wings. The eagle is facing right. The text "DEPARTMENT OF HEALTH AND HUMAN SERVICES - USA" is arranged in a circular pattern around the eagle. Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850 Ms. Allyson Barford, R.A.C. Regulatory Affairs Specialist Boston Scientific Corporation Endoscopy 100 Boston Scientific Way MARLBOROUGH MA 01752 SEP - 1 2006 Re: K061231 Trade/Device Name: WallFlexTM Biliary RX Uncovered Stent System Regulation Number: 21 CFR §876.5010 Regulation Name: Biliary catheter and accessories Regulatory Class: II Product Code: FGE Dated: August 7, 2006 Received: August 8, 2006 Dear Ms. Barford: We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act). You may, therefore, market the device, subject to the general controls provisions of the Act and the limitations described below. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. The Office of Device Evaluation has determined that there is a reasonable likelihood that this device will be used for an intended use not identified in the proposed labeling and that such use could cause harm. Therefore, in accordance with Section 513(i)(1)(E) of the Act, the following limitation must appear in the Warnings section of the device's labeling: The safety and effectiveness of this device for use in the vascular system have not been established. Furthermore, the indication for biliary use must be prominently displayed in all labeling. including pouch, box, and carton labels, instructions for use, and other promotional materials, in close proximity to the trade name, of a similar point size, and in bold print. {3}------------------------------------------------ # Page 2 - Ms. Allyson Barford Please note that the above labeling limitations are required by Section 513(i)(1)(E) of the Act. Therefore, a new 510(k) is required before these limitations are modified in any way or removed from the device's labeling. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and permits your device to proceed to the market. This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification if the limitation statement described above is added to your labeling. If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register. Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050. If you desire specific information about the application of other labeling requirements to your device (21 CFR Part 801), please contact the Office of Compliance at (240) 276-0115. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International, and Consumer Assistance at its toll-free number (800) 638-2041 or (240) 276-3150 or at its Internet address http://www.fda.gov/cdrh/dsma/dsmamain.html. . Sincerely yours, Donna-Bea Tillman, Ph.D., M.P.A. Director Office of Device Evaluation Center for Devices and Radiological Health Enclosure {4}------------------------------------------------ # Indications for Use 510(k) Number: K061231 Device Name: WallFlex™ Biliary RX Uncovered Stent System FDA's Statement of the Indications for Use for the device: The WallFlex Biliary RX Uncovered Stent system is indicated for use in the palliative treatment of biliary strictures produced by malignant neoplasms. Prescription Use X (Part 21 CFR 801 Subpart D) AND/OR Over-The-Counter Use (21 CFR 807 Subpart C) (PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED) Concurrence of CDRH, Office of Device Evaluation (ODE) David Ch. Loymm (Division Sign-Off) Division of Reproductive, Abdominal, and Radiological Devices 510(k) Number. Page 1 of 1
Innolitics
Guidance Recalls

Panel 1

/
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...