EMBEDDING TOOL, MODEL TE-1000
K060897 · Medigroup, Inc. · FJS · Jul 18, 2006 · Gastroenterology, Urology
Device Facts
| Record ID | K060897 |
| Device Name | EMBEDDING TOOL, MODEL TE-1000 |
| Applicant | Medigroup, Inc. |
| Product Code | FJS · Gastroenterology, Urology |
| Decision Date | Jul 18, 2006 |
| Decision | SESE |
| Submission Type | Traditional |
| Regulation | 21 CFR 876.5630 |
| Device Class | Class 2 |
| Attributes | Therapeutic |
Intended Use
The Embedding TM Tool is used to "embed" about 30cm of the external portion of a peritoneal dialysis catheter into the subcutaneous tissue immediately following the initial tunneling step of catheter implantation in order to be retrieved at a later time to begin peritoneal dialysis. After this secondary tunneling step, the titanium tip is unscrewed from the tool handle and functions as a plug at the catheter end. The titanium cap is then screwed onto the open end of the titanium tip to close it off. This Embedding™ Tool is intended to be used by the same physician at the same time and setting when he is implanting the catheter initially. As such, it can be used with any patient who is a suitable candidate for delaved peritoneal dialysis catheter utilization.
Device Story
Embedding™ Tool facilitates secondary tunneling of peritoneal dialysis (PD) catheters. Device consists of rigid PVC handle and detachable titanium tip/cap. Used by physicians in clinical setting immediately following initial catheter implantation. Tool embeds ~30cm of external catheter portion into subcutaneous tissue for future retrieval. Upon completion, titanium tip is unscrewed from handle to serve as catheter plug; titanium cap is screwed onto tip to seal. Enables delayed PD treatment initiation. Benefits patient by allowing catheter placement in advance of dialysis need while maintaining sterility and patency.
Clinical Evidence
Clinical evaluation performed at a dialysis center demonstrated the tool functioned as intended. Bench testing performed to demonstrate mechanical integrity.
Technological Characteristics
Rigid PVC handle; detachable titanium tip and cap. Manual surgical instrument. Non-powered. Dimensions: designed to embed ~30cm of catheter.
Indications for Use
Indicated for subcutaneous embedding of the external portion of peritoneal dialysis (PD) catheters in patients who are candidates for PD and delayed onset of PD treatment. Procedure must be performed immediately following successful catheter implantation, provided catheter patency is established and the catheter is suitable for embedding.
Regulatory Classification
Identification
(1) A peritoneal dialysis system and accessories is a device that is used as an artificial kidney system for the treatment of patients with renal failure or toxemic conditions, and that consists of a peritoneal access device, an administration set for peritoneal dialysis, a source of dialysate, and, in some cases, a water purification mechanism. After the dialysate is instilled into the patient's peritoneal cavity, it is allowed to dwell there so that undesirable substances from the patient's blood pass through the lining membrane of the peritoneal cavity into this dialysate. These substances are then removed when the dialysate is drained from the patient. The peritoneal dialysis system may regulate and monitor the dialysate temperature, volume, and delivery rate together with the time course of each cycle of filling, dwell time, and draining of the peritoneal cavity or manual controls may be used. This generic device includes the semiautomatic and the automatic peritoneal delivery system.(2) The peritoneal access device is a flexible tube that is implanted through the abdominal wall into the peritoneal cavity and that may have attached cuffs to provide anchoring and a skin seal. The device is either a single use peritioneal catheter, intended to remain in the peritoneal cavity for less than 30 days, or a long term peritoneal catheter. Accessories include stylets and trocars to aid in the insertion of the catheter and an obturator to maintain the patency of the surgical fistula in the abdominal wall between treatments. (3) The disposable administration set for peritoneal dialysis consists of tubing, an optional reservoir bag, and appropriate connectors. It may include a peritoneal dialysate filter to trap and remove contaminating particles. (4) The source of dialysate may be sterile prepackaged dialysate (for semiautomatic peritoneal dialysate delivery systems or “cycler systems”) or dialysate prepared from dialysate concentrate and sterile purified water (for automatic peritoneal dialysate delivery systems or “reverse osmosis” systems). Prepackaged dialysate intended for use with either of the peritoneal dialysate delivery systems is regulated by FDA as a drug.
Special Controls
*Classification.* Class II (special controls). The following accessories are exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 876.9: A catheter finger grip that is non-patient contacting and intended for single use with a peritoneal catheter; a continuous ambulatory peritoneal dialysis (CAPD) belt; and a catheter stand that does not include weigh scales.
Predicate Devices
Related Devices
- K213602 — Argyle Peritoneal Dialysis Cathetor and Kit, Ritus Peritoneal Dialysis Catheter Tunneling Tool Kit · Covidien, LLC · Jan 14, 2022
- K130441 — FALLER STYLET · Medionics International, Inc. · Apr 19, 2013
- K071167 — FLEX-NECK EXXTENDED PERITONEAL DIALYSIS CATHETER AND ACCESSORIES, MODEL CF-5560 · Medigroup, Inc. · Aug 31, 2007
- K121383 — PS X SERIES (BASIC COMPONENTS AND CATHETER), PD X SERIES (CATHETER WITH ALL COMPONENTS) · Medical Components, Inc. · Sep 5, 2012
- K120130 — FALLER TROCAR · Medigroup, Inc. · May 21, 2012
Submission Summary (Full Text)
{0}------------------------------------------------
K060897 Page 1 of 2
Image /page/0/Picture/1 description: The image shows the logo for MEDIGROUP, Inc. The logo consists of a stylized heart shape formed by two interlocking shapes, with diagonal lines filling the lower portion of the heart. Below the heart is the text "MEDIGROUP, Inc." in a serif font.
JUL 18 2006
#### 510(k) Summary
#### Basic Information
Submitter:
Medigroup, Inc. 14 A Stonehill Road Oswego, IL 60543
Establishment Registration Number: #1450420 Contact:
Telephone: Fax: E-mail Date of Submission John A. Navis, President (630) 554-5533 (630) 554-5535 inavis@medigroupinc.com March 30, 2006
#### Device Information
Trade Name: Common Name: Classification Name: Class: II
Embedding TM Tool. Catheter Tunneling Device. 78 FJS, accessory.
#### Predicate Devices
510(k) 823331 Tunnelor® Tool, issued January 26, 1983.
#### Product Description
The Embedding™ Tool consists of a gently curved handle portion made of rigid PVC with a detachable titanium tip and a separate titanium cap.
#### Intended Use
The Embedding TM Tool is used to "embed" about 30cm of the external portion of a peritoneal dialysis catheter into the subcutaneous tissue immediately following the initial tunneling step of catheter implantation in order to be retrieved at a later time to begin peritoneal dialysis. After this secondary tunneling step, the titanium tip is unscrewed from the tool handle and functions as a plug at the catheter end. The titanium cap is then screwed onto the open end of the titanium tip to close it off.
This Embedding™ Tool is intended to be used by the same physician at the same time and setting when he is implanting the catheter initially. As such, it can be used with any patient who is a suitable candidate for delaved peritoneal dialysis catheter utilization.
#### Substantial Equivalence
The "handle" or tunneling part of this Embedding™ Tool is made of the same rigid polyvinyl compound (RPVC) and has a similar configuration as the Tunnelor® Tool, predicate device 510(k) 823331. The configuration of the tip of the new device is similar to the shape of the tip in the predicate device, but made of titanium instead of RPVC.
#### MEDIGROUP, Inc.
(Division of Janin Group, Inc.)
7-1
Image /page/0/Picture/26 description: The image contains three logos related to quality assurance. The first logo on the left is the Lloyd's Register Quality Assurance logo with the letter R in the center and the text "ISO 13485" at the bottom. The middle logo is the UKAS Quality Management logo with the number 001 at the bottom. The third logo on the right is the Lloyd's Register Quality Assurance logo with the letter R in the center and the text "ISO 9001" at the bottom.
{1}------------------------------------------------
### 510(k) Summary
## Testing
Functional testing has been performed to demonstrate mechanical integrity. Clinical evaluation at a dialysis center showed the Embedding™ Tool functioned as intended.
## Conclusions
The Embedding™ Tool works as designed and intended. Its use is intended for use by physicians familiar with proper catheter tunneling techniques.
{2}------------------------------------------------
Image /page/2/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a stylized eagle with its wings spread, and the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" arranged in a circular pattern around the eagle. The logo is black and white.
Food and Drug Administration 9200 Corporate Blvd. Rockville MD 20850
## JUL 18 2006
Mr. John A. Navis President Medigroup, Inc. 14 A Stonehill Road OSWEGO IL 60543-9400
Re: K060897
Trade/Device Name: TE-1000 Embedding™ Tool Regulation Number: 21 CFR §876.5630 Regulation Name: Peritoneal dialysis system and accessories Regulatory Class: II Product Code: FJS Dated: June 26, 2006 Received: June 27, 2006
Dear Mr. Navis:
We have reviewed your Section 510(k) premarket notification of intent to market the device we nave reviewed your bection 910(t) produce is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce use stated in the encreater to regally that of the Medical Device Amendments, or to devices that prior to way 20, 1770, the onloanent and the provisions of the Federal Food, Drug, and Cosmetic liave bech toolassified in accordation was a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general therefore, mainer the device, salgos to annual registration, listing of devices, good Controls provibions of allabeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into either class II (Special Controls) or class III (Premarket IT your device is elassified (500 accreen additional controls. Existing major regulations affecting your Apploval, it that be sacreer to additions, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Image /page/2/Picture/10 description: The image shows the FDA Centennial logo, which includes the years 1906-2006. Below the logo is the text "Protecting and Promoting Public Health". The logo is circular and contains the letters FDA in a stylized font.
{3}------------------------------------------------
Page 2 -
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Office of Compliance at one of the following numbers, based on the regulation number at the top of this letter:
| 21 CFR 876.xxx | (Gastroenterology/Renal/Urology | 240-276-0115 |
|----------------|---------------------------------|--------------|
| 21 CFR 884.xxx | (Obstetrics/Gynecology) | 240-276-0115 |
| 21 CFR 894.xxx | (Radiology) | 240-276-0120 |
| Other | | 240-276-0100 |
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (240) 276-3150
or at its Internet address http://www.fda.gov/cdrl/industry/support/index.html.
Sincerely yours,
Nancy C. Brogdon
Nancy C. Brogdon Director, Division of Reproductive, Abdominal, and Radiological Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{4}------------------------------------------------
# Indications for Use
510(k) Number (if known): K060897
Device Name: TE-1000 Embedding™ Tool
Indications For Use:
This device can be used to embed most known brands and styles of PD catheters when the nephrologist determines this action is in the best interest of the patient, however, only immediately after successful catheter implantation.
The Embedding™ Tool is indicated for embedding the external portion of most PD catheters subcutaneously in anticipation of future retrieval of that part of the catheter, provided that:
- The embedding procedure is done immediately following PD . catheter implantation.
- Catheter patency has been completely established. .
- The normally external part of the PD catheter can be embedded. .
- The patient is a candidate for delayed onset of PD treatment. .
- The patient is a candidate for PD. .
Prescription Use X (Part 21 CFR 801 Subpart D) AND/OR
Over-The-Counter Use (21 CFR 801 Subpart C)
(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
Nancy C brogdon
510(k) Number
Page 1 of 1