RUSCH SIMPLASTIC FOLEY CATHETERIZATION SET
Device Facts
| Record ID | K963993 |
|---|---|
| Device Name | RUSCH SIMPLASTIC FOLEY CATHETERIZATION SET |
| Applicant | Rusch, Inc. |
| Product Code | FCM · Gastroenterology, Urology |
| Decision Date | Dec 17, 1996 |
| Decision | SESE |
| Submission Type | Traditional |
| Regulation | 21 CFR 876.5130 |
| Device Class | Class 2 |
| Attributes | Therapeutic |
Intended Use
The Rüsch Simplastic Foley Catheterization Set contains: - one Rüsch Simplastic Foley Catheter - one pack of lubricating jelly. The Rüsch Simplastic Foley Catheter is intended to be used to pass fluids to and from the urinary tract. A pack of lubricating jelly is intended to assist insertion of the catheter through the urethra.
Device Story
Set includes polyvinyl chloride (PVC) Foley catheter and lubricating jelly packet. Catheter features radiopaque BaSO4-loaded PVC stripe, adhesive-bonded latex balloon, funnel-connected main lumen, and luer-activated valve. Available in 2-way/3-way formats, sizes 12-26 F.G., various hardnesses, balloon sizes, and tip/eye configurations. Used by clinicians to facilitate fluid passage to/from urinary tract; jelly assists urethral insertion. Device provides drainage/irrigation access for patient urinary management.
Clinical Evidence
Bench testing only. Performed per FDA Draft Guidance for Conventional and Antimicrobial Foley Catheters: drainage flow rate, balloon rupture resistance, pullout resistance, inflation maintenance, manufacturing tolerances, deflation performance, and ISO 10993 biocompatibility.
Technological Characteristics
Materials: Polyvinyl chloride (PVC) tube, BaSO4-loaded PVC radiopaque stripe, latex balloon. Adhesive bonding. Luer-activated valve. Sizes: 12-26 F.G. Sterilization: Not specified (standard medical device processing).
Indications for Use
Indicated for patients requiring urinary tract fluid drainage or irrigation. No specific age or gender restrictions stated.
Regulatory Classification
Identification
A urological catheter and accessories is a flexible tubular device that is inserted through the urethra and used to pass fluids to or from the urinary tract. This generic type of device includes radiopaque urological catheters, ureteral catheters, urethral catheters, coudé catheters, balloon retention type catheters, straight catheters, upper urinary tract catheters, double lumen female urethrographic catheters, disposable ureteral catheters, male urethrographic catheters, and urological catheter accessories including ureteral catheter stylets, ureteral catheter adapters, ureteral catheter holders, ureteral catheter stylets, ureteral catheterization trays, and the gastro-urological irrigation tray (for urological use).
Predicate Devices
- Rusch Simplastic Foley Catheter (Preamendment)
- Urotec/Franklin Soft Simplastic Catheters (K851684)
- Inmed Foley Catheterization Tray (K832363)
Related Devices
- K233094 — Wellead® Latex Foley Catheter · Well Lead Medical (Hainan) Co., Ltd. · Jun 6, 2024
- K181616 — PSM 3-Way Silicone Foley Catheter · Pathway, LLC · Sep 21, 2018
- K091767 — 3 WAY SILICONE FOLEY BALLOON CATHETER · Cook Urological, Inc. · Oct 9, 2009
- K162989 — Rusch Simplastic Foley Catheters · Teleflexmedical, Inc. · Feb 24, 2017
- K173340 — Yikang Latex Foley Catheter · Jiangxi Yikang Medical Instrument Group Co., Ltd. · Jul 13, 2018