STEMPRO MSC SFM

K103302 · Life Technologies Corporation · NDS · Feb 18, 2011 · Gastroenterology, Urology

Device Facts

Record IDK103302
Device NameSTEMPRO MSC SFM
ApplicantLife Technologies Corporation
Product CodeNDS · Gastroenterology, Urology
Decision DateFeb 18, 2011
DecisionSESE
Submission TypeTraditional
Regulation21 CFR 876.5885
Device ClassClass 2

Indications for Use

StemPro® MSC SFM Medium is a liquid tissue culture media product intended for human ex vivo tissue and cell culture processing applications.

Device Story

StemPro® MSC SFM Medium is a chemically defined, serum-free liquid tissue culture medium consisting of a basal medium and a supplement. It is used in laboratory settings by trained personnel for the ex vivo expansion and maintenance of human mesenchymal stem cells (MSCs). The medium supports cell growth and expansion over multiple passages while preserving the multipotential phenotype of the MSCs, allowing for subsequent differentiation into osteogenic, chondrogenic, or adipogenic lineages. It provides a consistent, controlled environment for cell culture compared to traditional serum-supplemented media. The product is aseptically processed and manufactured under quality system regulations to ensure purity and biocompatibility.

Clinical Evidence

Bench testing only. Evidence includes performance assays demonstrating support of human MSC growth and maintenance, Limulus Ameobocyte (LAL) testing for endotoxin/pyrogen levels (USP <85>), and validation of aseptic processing to achieve a Sterility Assurance Level (SAL) ≥ 10^-3. Stability testing confirmed a 12-month shelf life at 2°–8°C, verifying pH stability and lack of cytotoxicity.

Technological Characteristics

Serum-free, chemically defined liquid medium consisting of basal medium and supplement. Manufactured in accordance with QSR requirements. Aseptically processed. Tested for chemical purity using USP, ACS, FCC, GIBCO, or cell culture requirements. Biocompatible with human cells.

Indications for Use

Indicated for human ex vivo tissue and cell culture processing applications to support the growth or maintenance of human tissue or cells in culture.

Regulatory Classification

Identification

Tissue culture media for human ex vivo tissue and cell culture processing applications consist of cell and tissue culture media and components that are composed of chemically defined components (e.g., amino acids, vitamins, inorganic salts) that are essential for the ex vivo development, survival, and maintenance of tissues and cells of human origin. The solutions are indicated for use in human ex vivo tissue and cell culture processing applications.

Special Controls

*Classification.* Class II (special controls): FDA guidance document, “Class II Special Controls Guidance Document: Tissue Culture Media for Human Ex Vivo Processing Applications; Final Guidance for Industry and FDA Reviewers.”

Predicate Devices

Related Devices

Submission Summary (Full Text)

{0}------------------------------------------------ Pay 113 K103302 | 510(k) Summary | | FEB 18 2011 | |--------------------------------------------------------------------------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------|-------------| | Life Technologies Corporation - StemPro® MSC SFM Medium | | | | Device Name: | StemPro® MSC SFM Medium | | | Common/Usual Name: | StemPro® MSC SFM | | | Classification Name: | Tissue culture media for human <i>ex vivo</i> tissue and cell<br>culture processing applications (per 21 CFR § 876.5885) | | | Product Code: | NDS | | | Submitter: | Life Technologies Corporation<br>3175 Staley Road<br>Grand Island, New York 14072 | | | Telephone: | (716) 774-3122 | | | Facsimile: | (716) 774-0230 | | | Email: | kelli.tanzella@lifetech.com | | | Contact: | Kelli Tanzella, Ph.D. | | | Date Prepared: | October 11, 2010 | | | Predicate Device: | | | | Trade Name | Manufacturer | 510(k) | | Knockout™ SR Medium | Life Technologies Corporation | K100616 | | Intended Use | | | | StemPro® MSC SFM Medium is a liquid tissue culture media product intended for<br>human <i>ex vivo</i> tissue and cell culture processing applications. | | | # Substantial Equivalence The tissue culture media presented in this 510(k) submission, StemPro® MSC SFM The tissue ontal views provalent to Knockout™ SR Medium S10(k) K100616 manufactured by Life Technologies Corporation. Knockout™ SR Medium received 510(k) clearance from the FDA on May 20, 2010 as a Class II Device, and is listed under Product Code 78 NDS (Tissue culture media for human ex vivo tissue and cell culture processing applications). The 510(k) for Knockout™ SR Medium was reviewed by the gastroenterology control panel. : : . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . ". . . .; : Kansa : Cara : {1}------------------------------------------------ K103302 - A. These tissue culture products (StemPro® MSC SFM and Knockout™ SR Medium) are used for human ex vivo tissue and cell culture processing applications and have the same principles of operation, technological characteristics, efficacy (generic cellular growth and maintenance) and safety (consistency in chemical content and formulation, biocompatibility with cells. and purity). : . . # Intended Uses StemPro® MSC SFM Medium is a tissue culture product intended for human ex vivo tissue and cell culture processing applications. This device is a chemically defined tissue culture media used to support the growth or maintenance of human tissue or cells in culture. # B. Principles of Operation and Technological Characteristics StemPro® MSC SFM is a serum-free medium (SFM) specially formulated for the growth and expansion of human mesenchymal stem cells (MSCs). StemPro® MSC SFM enables human MSC growth and increased consistency compared to classical serum-supplemented medium. In addition, human MSCs can be expanded for multiple passages while maintaining their multipotential phenotype (i.e. ability to differentiate into osteogenic, chondrogenic, adipogenic lineages). StemPro® MSC SFM contains two components: StemPro® MSC SFM Basal Medium and StemPro® MSC SFM Supplement. $$\text{The } \mathcal{L}_{\omega} \text{-representation } \mathcal{L}_{\omega} \text{ are defined by } \omega \text{ and } \omega_{\omega} \text{ are defined by } \omega_{\omega} \text{ and } \omega_{\omega}$$ ## C. Pre-Clinical Testing .. Performance standards under Section 514 of the Federal Food, Drug, and Cosmetic Act have been established in Guidance Document "Class II Special Controls Guidance Document: Tissue Culture Media for Human Ex Vivo Tissue and Cell Culture Processing Applications; Final Guidance for Industry and FDA Reviewers," issued May 16, 2001. The specific assay tests and Life Technologies Corporation's equivalent tests are described below in greater detail. | Special Control Objective | Life Technologies Corporation<br>Knockout™ SR Medium | |------------------------------------------------------------------------------------------------------------------------------------------------|-----------------------------------------------------------------------| | Demonstrate lack of potential<br>toxicity of materials in the media to<br>cells or tissue and demonstrate<br>support of tissue and cell growth | StemPro® MSC SFM Performance<br>Assay | | Demonstrate lack of endotoxin or<br>pyrogen contamination | Limulus Ameobocyte (LAL) test (25<br>USP Monograph <85>) | | Validation of Aseptic Processing and<br>Sterility Assurance Level (SAL) | Determination of SAL to be ≥ 10-3<br>compliance with GMP requirements | Image /page/1/Figure/11 description: The image appears to be a high-contrast, black-and-white image with a grainy or speckled texture. The distribution of black and white pixels is uneven, with clusters of black dots scattered across the white background. There are no discernible objects or patterns, and the image lacks any clear structure or recognizable content. The overall impression is that of random noise or a heavily degraded image. {2}------------------------------------------------ | | regarding aseptic processing | |-----------------------------|----------------------------------------------------------------------------------------------| | Demonstrate Chemical purity | Incoming Raw Material testing using<br>USP, ACS, FCC, GIBCO, or Cell<br>Culture requirements | # Stability/Shelf-Life - - Life Technologies Corporation performs shelf life testing for StemPro® MSC SFM Medium using retained product stored at 2° - 8° C. In addition, a minimum of one new production lot of StemPro® MSC SFM Medium is tested each year to verify that the product continues to meet the established shelf life. Based on analysis of product performance over time, Life Technologies Corporation has established a shelf life of twelve months for the StemPro® MSC SFM Medium formulation. Stability testing involves the assessment of these functional aspects of media, including demonstrating: (1) that the pH continues to meet specifications; and (2) the media is not cytotoxic and supports the growth of mammalian cells. The pH is tested to demonstrate that the media is not chemically altered during its storage. In assessing cytoxicity, Life Technologies Corporation demonstrates that the media functions in supporting the growth of mammalian cells and that the media does not become toxic to mammalian cells during storage. In addition, results from the studies indicate the container/closure system provides protection from microbial.contamination. ### D. Conclusion # ... . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . StemPro® MSC SFM and Knockout™ SR Medium are used for human ex vivo ticayo and coll culture processing and have the seme principles of tissue and cell culture processing applications and have the same principles of operation, technological characteristics, efficacy (generic cellular growth and maintenance) and safety (consistency in chemical content and formulation, biocompatibility with cells, and purity). Their efficacy in supporting the survival, growth, development, and maintenance of human cells or tissue culture systems has been well established in scientific publications included in this submission. StemPro® MSC SFM and Knockout™ SR Medium are manufacturer in accordance with QSR requirements and are labeled as aseptically processed. Thus, StemPro® MSC SFM is substantially equivalent to the legally marketed device intended for the human ex vivo tissue and cell culture processing applications. K10 3302 {3}------------------------------------------------ Image /page/3/Picture/0 description: The image shows the seal of the Department of Health & Human Services (HHS) of the United States. The seal features a stylized eagle with three lines representing its wings, and the text "DEPARTMENT OF HEALTH & HUMAN SERVICES · USA" encircles the eagle. The text is arranged in a circular fashion around the eagle, with the department's name at the bottom and "USA" at the top. # DEPARTMENT OF HEALTH & HUMAN SERVICES Food and Drug Administration 10903 New Hampshire Avenue Document Mail Center - WO66-G609 Silver Spring, MD 20993-0002 Kelli L. Tanzella, Ph.D. Sr. Manager, Americas Regulatory Affairs Life Technologies" Inc. 3175 Staley Road GRAND ISLAND NY 14072 FEB 18 201 Re: K103302 Trade/Device Name: StemPro® MSC SFM Medium – StemPro® MSC SFM Basal Medium and StemPro® MSC SFM Supplement for ex-vivo Tissue and Cell Culture Regulation Number: 21 CFR §876.5885 Regulation Name: Tissue culture media for human ex vivo tissue and cell culture processing applications. Regulatory Class: II Product Code: NDS Dated: February 17, 2011 Received: February 18, 2011 Dear Dr. Tanzella: We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading. If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register. Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements; including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related {4}------------------------------------------------ adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050. If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please go to http://www.fda.gov/AboutFDA/CentersOffices/CDRH/CDRHOffices/ucm115809.htm for the Center for Devices and Radiological Health's (CDRH's) Office of Compliance. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance. You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm. Sincerely yours, Hubert Humphrey Herbert P. Lerner, M.D., Director (Acting) Division of Reproductive, Gastro-Renal and Urological Devices Office of Device Evaluation Center for Devices and Radiological Health Enclosure {5}------------------------------------------------ KIO3302 510(k) Number (if known): __ Device Name: StemPro® MSC SFM Medium # Indications for Use: StemPro® MSC SFM Medium is a liquid tissue culture medium products intended for human ex vivo tissue and cell culture processing applications. (PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHRE PAGE IF NEEDED) Concurrence of CDRH, Office of Evaluation (ODE) . Prescription Use (Per 21 CFR 801.109) OR Roger M. Whang stro-Renal, and Over-The -Counter Use (Optional Format 1-2-96)
Innolitics

Panel 1

/
Sort by
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...