HP M2376A DEVICE LINK SYSTEM

K984194 · Hewlett-Packard Co. · MQS · Apr 30, 1999 · Gastroenterology, Urology

Device Facts

Record IDK984194
Device NameHP M2376A DEVICE LINK SYSTEM
ApplicantHewlett-Packard Co.
Product CodeMQS · Gastroenterology, Urology
Decision DateApr 30, 1999
DecisionSESE
Submission TypeTraditional
Regulation21 CFR 876.5820
Device ClassClass 2

Indications for Use

The HP M2376A Device Link System is indicated for use in data collection and clinical information management, either connected directly or through networks, with the independent bedside devices that are listed below; - 1) IMED 1310 (PC-1) channel IV Pump - 2) In-line Diagnostic HR Crit-Line Monitor - 3) Newport Wave Ventilator

Device Story

HP M2376A DeviceLink System acts as a data gateway; receives digital data from external bedside devices (IV pumps, monitors, ventilators) via device-specific cables; converts data into HL7 format; transmits data to networked Clinical Information Systems. Device is not patient-connected; does not remotely control source devices; does not transmit waveforms, alarms, or images. Used in clinical environments to facilitate electronic data collection and information management for healthcare providers.

Clinical Evidence

Bench testing only; no clinical data provided.

Technological Characteristics

Data interface system; converts proprietary digital data from bedside devices to HL7 format for network transmission. Does not transmit waveforms, alarms, or images. Standalone hardware unit connecting to external bedside devices via specific cables.

Indications for Use

Indicated for clinical information management and data collection from specific bedside devices (IMED 1310 IV Pump, In-line Diagnostic HR Crit-Line Monitor, Newport Wave Ventilator) in a clinical setting.

Regulatory Classification

Identification

A hemodialysis system and accessories is a device that is used as an artificial kidney system for the treatment of patients with renal failure or toxemic conditions and that consists of an extracorporeal blood system, a conventional dialyzer, a dialysate delivery system, and accessories. Blood from a patient flows through the tubing of the extracorporeal blood system and accessories to the blood compartment of the dialyzer, then returns through further tubing of the extracorporeal blood system to the patient. The dialyzer has two compartments that are separated by a semipermeable membrane. While the blood is in the blood compartment, undesirable substances in the blood pass through the semipermeable membrane into the dialysate in the dialysate compartment. The dialysate delivery system controls and monitors the dialysate circulating through the dialysate compartment of the dialyzer.(1) The extracorporeal blood system and accessories consists of tubing, pumps, pressure monitors, air foam or bubble detectors, and alarms to keep blood moving safely from the blood access device and accessories for hemodialysis (§ 876.5540) to the blood compartment of the dialyzer and back to the patient. (2) The conventional dialyzer allows a transfer of water and solutes between the blood and the dialysate through the semipermeable membrane. The semipermeable membrane of the conventional dialyzer has a sufficiently low permeability to water that an ultrafiltration controller is not required to prevent excessive loss of water from the patient's blood. This conventional dialyzer does not include hemodialyzers with the disposable inserts (Kiil type) (§ 876.5830) or dialyzers of high permeability (§ 876.5860). (3) The dialysate delivery system consists of mechanisms that monitor and control the temperature, conductivity, flow rate, and pressure of the dialysate and circulates dialysate through the dialysate compartment of the dialyzer. The dialysate delivery system includes the dialysate concentrate for hemodialysis (liquid or powder) and alarms to indicate abnormal dialysate conditions. This dialysate delivery system does not include the sorbent regenerated dialysate delivery system for hemodialysis (§ 876.5600), the dialysate delivery system of the peritoneal dialysis system and accessories (§ 876.5630), or the controlled dialysate delivery system of the high permeability hemodialysis system § 876.5860). (4) Remote accessories to the hemodialysis system include the unpowered dialysis chair without a scale, the powered dialysis chair without a scale, the dialyzer holder set, dialysis tie gun and ties, and hemodialysis start/stop tray.

Predicate Devices

Related Devices

Submission Summary (Full Text)

{0}------------------------------------------------ 4/30/99 ## 5.0 510(k) Summary ## 510(k) Summary This summary of 510(k) safety and effectiveness information is submitted in accordance with the requirements of the Safety and onestical 1990 and 21 C.F.R. §.807.92. The name of this device is the Hewlett Packard M2376A DeviceLink System. The common name is the HP DeviceLink. Classification names are as follows: | REGULATION | CLASSIFICATION NAME | PANEL | PROCODE | |------------|----------------------------------------------------------|----------------|---------| | NUMBER | | | | | 868.5895 | Continuous ventilator | Anesthesiology | 73 CBK | | 876.5820 | System, hemodialysis, access<br>recirculation monitoring | Gastro-urology | 78 MQS | | 880.5725 | Pump, infusion | Gnr'l Hospital | 80 FRN | The above device is substantially equivalent to the HP M1032A VueLink Interface Plug-in Module marketed pursuant to K923682. The Hewlett Packard M2376A DeviceLink System receives digital data frontuced by external devices through device specific cables, converts that data into the HL7 format and transmits that information to any networked Clinical Information System. When connected to a bedside device, the HP M2376A DeviceLink System is intended for electronic data collection and clinical Information management. DeviceLink is neither patient connected, nor does it remotely control the attached source device. The technological characteristics are the same or similar to those found with the predicate device, with the exception that neither waveforms, alarms, nor image data are transmitted in the DeviceLink System. {1}------------------------------------------------ Image /page/1/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a stylized eagle with three lines forming its body and wings. The eagle is facing right. The text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" is arranged in a circular fashion around the eagle. Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850 APR 3 0 1999 Mr. Mike Hudon Requlatory Affairs Engineer Medical Products Group Hewlett-Packard GmbH Herrenberger Strasse 110-140 71034 Boeblingen GERMANY Re: K984194 Hewlett-Packard Model M2376A Device Link System Regulatory Class: II (two) Product Code: MQS Dated: March 11, 1999 Received: March 12, 1999 Dear Mr. Hudon: We have reviewed your Section 510(k) notification of intent to market the device referenced above and we have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. If your device is classified (see above) into either class II (Special Controls) or class III (Premarket Approval), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Requlations, Title 21, Parts 800 to 895. A substantially equivalent determination assumes compliance with the Current Good Manufacturing Practice requirements, as set forth in the Quality System Regulation (QS) for Medical Devices : General regulation (21 CFR Part 820) and that, through periodic QS inspections, the Food and Drug Administration (FDA) will verify such assumptions. Failure to comply with the GMP regulation may result in regulatory action. In addition, FDA may publish further announcements {2}------------------------------------------------ concerning your device in the Federal Register. Please note: this response to your premarket notification submission does not affect any obligation you might have under sections 531 through 542 of the Act for devices under the Electronic Product Radiation Control provisions, or other Federal laws or regulations. This letter will allow you to begin marketing your device as described in your 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market. If you desire specific advice for your device on our labeling regulation (21 CFR Part 801 and additionally 809.10 for in vitro diagnostic devices), please contact the Office of Compliance at (301) 594-4648. Additionally, for questions on the promotion and advertising of your device, please contact the Office of Compliance at (301) 594-4639. Also, please note the regulation entitled, 2Misbranding by reference to premarket notification² (21CFR 807.97). Other general information on your responsibilities under the Act may be obtained from the Division of Small Manufacturers Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its internet address "http://www.fda.gov/cdrh/dsma/dsmamain.html". Sincerely yours, Thomas J. Callahson Thomas J. Callahan, Ph.D. Director Division of Cardiovascular, Respiratory, and Neurological Devices Office of Device Evaluation Center for Devices and Radiological Health Enclosure {3}------------------------------------------------ Page ______________________________________________________________________________________________________________________________________________________________________________ 510(k) Number (if known): ____________________________________________________________________________________________________________________________________________________ Hewlett-Packard M2376A Devicel .ink System Ocvice Name: Indications for Use: The HP M2376A Device Link System is indicated for use in data collection and clinical information management, either connected directly or through networks, with the independent bedside devices that are listed below; - 1) IMED 1310 (PC-1) channel IV Pump - 2) In-line Diagnostic HR Crit-Line Monitor - 3) Newport Wave Ventilator ## (PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDIED) Concurrence of CDRH, Office of Device Evaluation (ODE) Q. K. Stillwell 510(k) Number K192444 Prescription Use (Per 21 CFR 801.109) OR Over-The-Counter Use ---------- (Optional Format 1-2-96)
Innolitics

Panel 1

/
Sort by
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...