Bicarby and Ci-Ca solutions are dialysis solutions for use in acute kidney replacement therapy for the correction of blood electrolytes and acid-base balance in an extracorporeal dialysis treatment.
Device Story
Sterile, single-use, 2-compartment dialysis solution bags; one compartment contains alkaline bicarbonate solution, other contains acidic electrolyte/glucose solution. Peel seam allows mixing before use to yield 5 L ready-to-use solution. Used in acute care settings for Kidney Replacement Therapy (KRT). Calcium-containing solutions used with heparin anticoagulation; calcium-free solutions used with regional citrate anticoagulation (RCA), requiring mandatory separate calcium infusion. Clinicians administer dialysate via extracorporeal circuit to correct blood electrolytes and acid-base balance. Benefits include maintenance of electrolyte homeostasis and anticoagulation support during KRT.
Clinical Evidence
No clinical studies were performed. Evidence consists of bench testing, including shipping/distribution integrity, connector mechanical testing (torque, breaking strength, leakage, pull-out force), and biocompatibility evaluation (cytotoxicity, sensitization, irritation, pyrogenicity, hemocompatibility, genotoxicity) per ISO 10993-1.
Technological Characteristics
Multi-layer gas barrier film bag; polyolefin/elastomer tubing; polypropylene/synthetic rubber injection port; polycarbonate/thermoplastic elastomer/silicone connectors. Steam sterilized. Single-use. Non-electrical. Ionic concentrations vary by formulation (Na, K, Mg, Ca, Cl, HCO3, Glucose).
Indications for Use
Indicated for use as a dialysis fluid for adult patients in acute care settings, requiring Kidney Replacement Therapies (KRT).
Regulatory Classification
Identification
A hemodialysis system and accessories is a device that is used as an artificial kidney system for the treatment of patients with renal failure or toxemic conditions and that consists of an extracorporeal blood system, a conventional dialyzer, a dialysate delivery system, and accessories. Blood from a patient flows through the tubing of the extracorporeal blood system and accessories to the blood compartment of the dialyzer, then returns through further tubing of the extracorporeal blood system to the patient. The dialyzer has two compartments that are separated by a semipermeable membrane. While the blood is in the blood compartment, undesirable substances in the blood pass through the semipermeable membrane into the dialysate in the dialysate compartment. The dialysate delivery system controls and monitors the dialysate circulating through the dialysate compartment of the dialyzer.(1) The extracorporeal blood system and accessories consists of tubing, pumps, pressure monitors, air foam or bubble detectors, and alarms to keep blood moving safely from the blood access device and accessories for hemodialysis (§ 876.5540) to the blood compartment of the dialyzer and back to the patient. (2) The conventional dialyzer allows a transfer of water and solutes between the blood and the dialysate through the semipermeable membrane. The semipermeable membrane of the conventional dialyzer has a sufficiently low permeability to water that an ultrafiltration controller is not required to prevent excessive loss of water from the patient's blood. This conventional dialyzer does not include hemodialyzers with the disposable inserts (Kiil type) (§ 876.5830) or dialyzers of high permeability (§ 876.5860). (3) The dialysate delivery system consists of mechanisms that monitor and control the temperature, conductivity, flow rate, and pressure of the dialysate and circulates dialysate through the dialysate compartment of the dialyzer. The dialysate delivery system includes the dialysate concentrate for hemodialysis (liquid or powder) and alarms to indicate abnormal dialysate conditions. This dialysate delivery system does not include the sorbent regenerated dialysate delivery system for hemodialysis (§ 876.5600), the dialysate delivery system of the peritoneal dialysis system and accessories (§ 876.5630), or the controlled dialysate delivery system of the high permeability hemodialysis system § 876.5860). (4) Remote accessories to the hemodialysis system include the unpowered dialysis chair without a scale, the powered dialysis chair without a scale, the dialyzer holder set, dialysis tie gun and ties, and hemodialysis start/stop tray.
K052393 — B. BRAUN MODIFIED BICARBONATE DIALYSATE · B.Braun Medical, Inc. · Nov 9, 2005
Submission Summary (Full Text)
{0}
FDA U.S. FOOD & DRUG ADMINISTRATION
April 4, 2025
Fresenius Medical Care Renal Therapies Group, LLC
Timothy Groves
Regulatory Affairs Senior Lead
920 Winter Street
Waltham, MA 02451
Re: K243786
Trade/Device Name: Bicarby Dialysate RFP-402 (RFP-402-G); Bicarby Dialysate RFP-400 (RFP-400-G); Bicarby Dialysate RFP-407 (RFP-407-G); Bicarby Dialysate RFP-401 (RFP-401-G); Bicarby Dialysate RFP-404 (RFP-404-G); Bicarby Dialysate RFP-456 (RFP-456-G); Ci-Ca Dialysate 2K (RFP-457-G); Ci-Ca Dialysate 4K (RFP-458-G)
Regulation Number: 21 CFR § 876.5820
Regulation Name: Hemodialysis system and accessories
Regulatory Class: II
Product Code: KPO
Dated: December 9, 2024
Received: March 6, 2025
Dear Timothy Groves:
We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
{1}
K243786 - Timothy Groves
Page 2
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device" (https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download).
Your device is also subject to, among other requirements, the Quality System (QS) regulation (21 CFR Part 820), which includes, but is not limited to, 21 CFR 820.30, Design controls; 21 CFR 820.90, Nonconforming product; and 21 CFR 820.100, Corrective and preventive action. Please note that regardless of whether a change requires premarket review, the QS regulation requires device manufacturers to review and approve changes to device design and production (21 CFR 820.30 and 21 CFR 820.70) and document changes and approvals in the device master record (21 CFR 820.181).
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reporting-combination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050.
All medical devices, including Class I and unclassified devices and combination product device constituent parts are required to be in compliance with the final Unique Device Identification System rule ("UDI Rule"). The UDI Rule requires, among other things, that a device bear a unique device identifier (UDI) on its label and package (21 CFR 801.20(a)) unless an exception or alternative applies (21 CFR 801.20(b)) and that the dates on the device label be formatted in accordance with 21 CFR 801.18. The UDI Rule (21 CFR 830.300(a) and 830.320(b)) also requires that certain information be submitted to the Global Unique Device Identification Database (GUDID) (21 CFR Part 830 Subpart E). For additional information on these requirements, please see the UDI System webpage at https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance/unique-device-identification-system-udi-system.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-devices/medical-device-safety/medical-device-reporting-mdr-how-report-medical-device-problems.
{2}
K243786 - Timothy Groves
Page 3
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
Maura Rooney -S
Maura Rooney
Assistant Director
DHT3A: Division of Renal, Gastrointestinal, Obesity and Transplant Devices
OHT3: Office of GastroRenal, ObGyn, General Hospital and Urology Devices
Office of Product Evaluation and Quality
Center for Devices and Radiological Health
Enclosure
{3}
DEPARTMENT OF HEALTH AND HUMAN SERVICES
Food and Drug Administration
Indications for Use
Form Approved: OMB No. 0910-0120
Expiration Date: 07/31/2026
See PRA Statement below.
Submission Number (if known)
K243786
Device Name
Bicarby Dialysate RFP-402 (RFP-402-G);
Bicarby Dialysate RFP-400 (RFP-400-G);
Bicarby Dialysate RFP-407 (RFP-407-G);
Bicarby Dialysate RFP-401 (RFP-401-G);
Bicarby Dialysate RFP-404 (RFP-404-G);
Bicarby Dialysate RFP-456 (RFP-456-G);
Ci-Ca Dialysate 2K (RFP-457-G);
Ci-Ca Dialysate 4K (RFP-458-G)
Indications for Use (Describe)
Bicarby™ Dialysate solutions are indicated for use as a dialysis fluid for adult patients in acute care settings, requiring Kidney Replacement Therapies (KRT).
Ci-Ca Dialysate solutions are indicated for use as a dialysis fluid for adult patients in acute care settings, requiring Kidney Replacement Therapies (KRT).
Type of Use (Select one or both, as applicable)
☑ Prescription Use (Part 21 CFR 801 Subpart D)
☐ Over-The-Counter Use (21 CFR 801 Subpart C)
## CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
*DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
Department of Health and Human Services
Food and Drug Administration
Office of Chief Information Officer
Paperwork Reduction Act (PRA) Staff
PRAStaff@fda.hhs.gov
“An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number.”
{4}
FRESENIUS MEDICAL CARE
510(k) Summary Bicarby and Ci-Ca Solutions Traditional 510(k)
K243786
Page 1 of 6
# 1. 510(K) SUMMARY
This 510(k) Summary is in accordance with the requirements of the Safe Medical Device Act (SMDA) of 1990. The content of this 510(k) summary is provided in conformance with 21 CFR § 807.92.
## 1.1. Submitter's Information
Name: Fresenius Medical Care Renal Therapies Group, LLC
Address: 920 Winter Street
Waltham, MA 02451-1457
Phone: (781) 460-1087
Fax: (781) 699-9635
Contact Person: Timothy Groves, Senior Lead
Preparation Date: 06 December 2024
## 1.2. Device Name
Trade Names: Bicarby™ Dialysate and Ci-Ca Dialysate
Common Name: Dialysis Solutions for Kidney Replacement Therapy (KRT)
Regulation Name: Hemodialysis System and Accessories
Regulatory Class: Class II per 21 CFR § 876.5820
Product Code: KPO
Product Code Name: Dialysate Concentrate for Hemodialysis (Liquid or Powder)
FDA Review Panel: Gastroenterology/Urology
## 1.3. Legally Marketed Predicate Device
The legally marketed predicate device for the proposed Bicarby and Ci-Ca solutions is the pureFLOW 400 series dialysis solutions (K233950). The pureFLOW 500 series dialysis solutions are being used as a secondary predicate device (K233159). Neither of these devices are currently subject to design-related recalls.
The B. Braun Modified Duosol Bicarbonate Dialysate (K052393) and Baxter PrismaSATE solutions (K162887) are used as reference devices.
## 1.4. Device Description
### 1.4.1. Device Identification
#### 1.4.1.1. Bicarby and Ci-Ca Solutions
The Bicarby and Ci-Ca solutions are single-use, sterile, ready-to-use dialysis solutions consisting of sodium chloride, potassium chloride, magnesium chloride, calcium chloride (select solutions), sodium bicarbonate, and glucose. The calcium-containing solutions are available in five (5)
Confidential & Proprietary
Page 1
{5}
FRESENIUS MEDICAL CARE
510(k) Summary Bicarby and Ci-Ca Solutions Traditional 510(k)
K243786
Page 2 of 6
formulations (Table 1), while the calcium-free solutions are available in three (3) formulations (Table 2).
Table 1: Calcium-Containing Solutions
| Part Number | Description |
| --- | --- |
| RFP-402-G | Bicarby™ Dialysate RFP-402 |
| RFP-400-G | Bicarby™ Dialysate RFP-400 |
| RFP-407-G | Bicarby™ Dialysate RFP-407 |
| RFP-401-G | Bicarby™ Dialysate RFP-401 |
| RFP-404-G | Bicarby™ Dialysate RFP-404 |
Table 2: Calcium-Free Solutions
| Part Number | Description |
| --- | --- |
| RFP-456-G | Bicarby™ Dialysate RFP-456 |
| RFP-457-G | Ci-Ca Dialysate 2K |
| RFP-458-G | Ci-Ca Dialysate 4K |
## 1.4.2. Device Characteristics
The Bicarby and Ci-Ca solutions are single-use, steam sterilized dialysis solutions for use in Kidney Replacement Therapy (KRT).
## 1.4.3. Environment of Use
The Bicarby and Ci-Ca solutions are used in acute care environments where KRT is performed.
## 1.4.4. Brief Written Description of the Device
Bicarby and Ci-Ca solutions are sterile dialysates for use in KRT. The solutions are each provided in a 2-compartment bag. One compartment contains 4.75 L of a slightly alkaline bicarbonate solution and the other compartment contains 0.25 L of an acidic electrolyte, glucose solution. The 2 solutions are mixed before use by opening the peel seam between the compartments, yielding 5 L of a ready-to-use sterile solution. The dialysate bags are made of a gas barrier foil, a material manufactured without PVC, latex, or plasticizer. The calcium-containing solutions are designed for modalities using heparin anticoagulation, while the calcium-free solutions are intended for modalities using regional citrate anticoagulation (RCA). The calcium-free solutions sustain the anticoagulatory effect of citrate in the filter. Therefore, a separate infusion of calcium is mandatory when using these solutions.
## 1.4.5. Materials of Use
The Bicarby and Ci-Ca solutions are classified as externally communicating, blood path indirect, prolonged contact (>24 hours to 30 days) duration, Class II (Category B) devices in accordance with FDA guidance document Use of International Standard ISO 10993-1, Biological evaluation
Confidential & Proprietary
Page 2
{6}
FRESENIUS MEDICAL CARE
510(k) Summary Bicarby and Ci-Ca Solutions Traditional 510(k)
K243786
Page 3 of 6
of medical devices – Part 1: Evaluation and testing within a risk management process (08 September 2023).
The calcium-containing and calcium-free solution container closure systems are constructed from identical materials, with the exception of colorants used in the HF-connectors. Each bag is equipped with an HF-connector, a Luer-lock connector, and an injection port. The HF-connectors of the calcium-containing and calcium-free container closure systems contain blue and yellow colorants, respectively. The bags are composed of the materials listed in Table 3.
Table 3: Bicarby and Ci-Ca Dialysate Bag Materials
| Component | Material |
| --- | --- |
| Solution Bag | Multi-layer Gas Barrier Film |
| Connective Tubing | Polyolefin/Elastomer |
| Injection Port | Polypropylene |
| | Synthetic Rubber |
| Luer-Lock Connector with Cap | Polycarbonate |
| | Thermoplastic Elastomer |
| HF Connector with Cap | Polycarbonate |
| | Silicone |
| Overwrap | Polyolefin/Elastomer |
## 1.4.6. Key Performance Specifications/Characteristics
The key performance specifications of the solutions are outlined in Table 4.
Table 4: Key Performance Specifications
| Chemical Component | Ionic Contribution (mEq/L, mixed) | Ionic Contribution (mEq/L, mixed) |
| --- | --- | --- |
| | Calcium-containing | Calcium-free |
| Sodium (Na⁺) | 140 | 133 or 140 |
| Potassium (K⁺) | 0, 2, 3, or 4 | 2 or 4 |
| Magnesium (Mg²⁺) | 1.0 or 1.5 | 1.5 |
| Calcium (Ca²⁺) | 2.5 or 3.0 | 0 |
| Chloride (Cl⁻) | 109, 111, 112, or 113 | 116.5, 118.5, or 120.5 |
| Bicarbonate (HCO₃⁻) | 35 | 20 or 25 |
| Glucose (g/L) | 1 | 1 |
Confidential & Proprietary
Page 3
{7}
FRESENIUS MEDICAL CARE
510(k) Summary Bicarby and Ci-Ca Solutions Traditional 510(k)
K243786
Page 4 of 6
# 1.5. Intended Use
Bicarby and Ci-Ca solutions are dialysis solutions for use in acute kidney replacement therapy for the correction of blood electrolytes and acid-base balance in an extracorporeal dialysis treatment.
# 1.6. Indications for Use
## Bicarby Dialysate
Bicarby™ Dialysate solutions are indicated for use as a dialysis fluid for adult patients in acute care settings, requiring Kidney Replacement Therapies (KRT).
## Ci-Ca Dialysate
Ci-Ca Dialysate solutions are indicated for use as a dialysis fluid for adult patients in acute care settings, requiring Kidney Replacement Therapies (KRT).
# 1.7. Comparison of Technological Characteristics with the Predicate Device
The proposed and predicate solutions share the same technological characteristics:
- Chemical composition
- Packaging configuration
- Shelf-life
- Sterilization method
- Single use
The following technological characteristics of the proposed solutions are substantially equivalent to those of the predicate pureFLOW solutions (K233950) and secondary predicate pureFLOW solutions (K233159):
- Indications for Use and Principle of Operation – Continuous Renal Replacement Therapy (CRRT) was updated to Kidney Replacement Therapy (KRT) as these solutions can be used in other modalities in acute care settings
- Intended Use – A high-level description was added to the Intended Use to meet both U.S. and EU labeling requirements. The proposed solutions are still intended to be used as a dialysis solution/dialysate for treatment of patients with acute renal failure
- Ionic Contributions (mixed) – New formulations contain different ionic contributions (mixed). Reference devices support the ionic contributions of the new formulations.
- Packaging materials – A new white hot stamp (UDI) material was introduced. Results of biocompatibility evaluations demonstrate that the proposed bag materials are biologically safe for their intended use
These differences do not raise any different questions of safety and effectiveness.
Confidential & Proprietary
Page 4
{8}
FRESENIUS MEDICAL CARE
510(k) Summary Bicarby and Ci-Ca Solutions Traditional 510(k)
K243786
Page 5 of 6
# 1.8. Performance Data
The container closure system of the proposed Bicarby and Ci-Ca solutions is identical to the predicate pureFLOW (K233950) and secondary predicate pureFLOW (K233159) solutions, with the exception of the UDI stamp material. Testing conducted to support the determination of substantial equivalence is summarized in Table 5.
Table 5: Performance Testing Summary
| Test Conducted | Test Method Description |
| --- | --- |
| Shipping and Distribution | Demonstrate the integrity and robustness of the bag system packaging within the distribution environment |
| HF-Connector | • Evaluate torque for removing the protective cap from the female connector and the torque of male and female connector
• Evaluate breaking strength of the cone of the HF-connector
• Evaluate the leakage and tightness of the connectors
• Evaluate the pull-out force of the connectors and injection ports |
# 1.8.1. Biocompatibility Testing
Biocompatibility testing was conducted to evaluate the material changes in accordance with ISO 10993-1:2018 and FDA guidance document Use of International Standard ISO 10993-1, "Biological evaluation of medical devices – Part 1: Evaluation and testing within a risk management process" (08 September 2023). The following endpoints were assessed to support the biological safety of the Bicarby and Ci-Ca solutions container closure system:
- Chemical Characterization
- Cytotoxicity
- Sensitization
- Irritation
- Material Mediated Pyrogenicity
- Hemocompatibility
- Genotoxicity
A toxicological risk assessment was also performed.
# 1.8.2. Human Factors Validation Testing
The Bicarby and Ci-Ca solutions were found to be safe and effective for their intended users, uses, and use environments.
# 1.8.3. Electrical Safety and Electromagnetic Compatibility (EMC)
Not applicable. The Bicarby and Ci-Ca solutions are not electrical mechanical devices.
Confidential & Proprietary
Page 5
{9}
FRESENIUS MEDICAL CARE
510(k) Summary Bicarby and Ci-Ca Solutions Traditional 510(k)
K243786 Page 6 of 6
## 1.8.4. Software Verification and Validation Testing
Not applicable. The Bicarby and Ci-Ca solutions do not contain software.
## 1.8.5. Animal Studies
No animal studies were performed.
## 1.8.6. Clinical Studies
No clinical studies were performed.
## 1.9. Conclusion
The intended use, design, principle of operation, and performance specifications of the Bicarby and Ci-Ca solutions are substantially equivalent to those of the predicate and secondary predicate devices. Differences between the Bicarby and Ci-Ca solutions and the predicates do not raise new concerns with regard to safety or efficacy. FMCRTG concludes that within the meaning of the Medical Device Amendments Act of 1976, Bicarby and Ci-Ca solutions are safe and effective for their intended use.
Confidential & Proprietary
Page 6
Panel 1
/
Sort by
Ready
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.