K233213 · Nxstage Medical, Inc. · KPO · Jan 18, 2024 · Gastroenterology, Urology
Device Facts
Record ID
K233213
Device Name
NxStage PureFlow Solution
Applicant
Nxstage Medical, Inc.
Product Code
KPO · Gastroenterology, Urology
Decision Date
Jan 18, 2024
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 876.5820
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
NxStage PureFlow Solution is intended for use with renal replacement therapy systems that utilize sterile premixed dialysate. The solutions are not designed for use with intermittent proportioning hemodialysis systems. The NxStage offering of premixed dialysate solutions allows the physician to prescribe different electrolyte compositions that meet the specific needs of individual patients.
Device Story
NxStage PureFlow Solution is a family of sterile, non-pyrogenic, fully constituted dialysate solutions provided in 5L single-use flexible bags. Used as dialysate during hemodialysis therapy in renal replacement systems; not for use with proportioning systems or direct infusion. Solutions contain electrolytes (sodium, potassium, calcium, magnesium chlorides), sodium lactate, dextrose, and water. Used in dialysis clinics, healthcare facilities, and home/community settings. Physician prescribes specific electrolyte compositions based on patient needs. Device provides necessary chemical environment for hemodialysis, facilitating solute removal and fluid balance. No electronic or software components.
Clinical Evidence
No clinical data. Bench testing only, including analytical testing per USP/EP, physical container testing, container closure integrity (ISO 11607-1), temperature excursion, shelf-life (ICH Q1A), and ship testing (ASTM D4169-22). Biocompatibility evaluated per ISO 10993-1.
Technological Characteristics
Sterile, non-pyrogenic, single-chamber flexible bags (5L). Materials: Polypropylene, SEBS, SIS, Polycarbonate. Sterilization: Moist heat per ISO 17665-1:2006/(R)2013 (SAL 10^-6). No electronic features, no software, no energy source.
Indications for Use
Indicated for use with NxStage renal replacement therapy systems that utilize sterile premixed dialysate during hemodialysis for chronic care patients.
Regulatory Classification
Identification
A hemodialysis system and accessories is a device that is used as an artificial kidney system for the treatment of patients with renal failure or toxemic conditions and that consists of an extracorporeal blood system, a conventional dialyzer, a dialysate delivery system, and accessories. Blood from a patient flows through the tubing of the extracorporeal blood system and accessories to the blood compartment of the dialyzer, then returns through further tubing of the extracorporeal blood system to the patient. The dialyzer has two compartments that are separated by a semipermeable membrane. While the blood is in the blood compartment, undesirable substances in the blood pass through the semipermeable membrane into the dialysate in the dialysate compartment. The dialysate delivery system controls and monitors the dialysate circulating through the dialysate compartment of the dialyzer.(1) The extracorporeal blood system and accessories consists of tubing, pumps, pressure monitors, air foam or bubble detectors, and alarms to keep blood moving safely from the blood access device and accessories for hemodialysis (§ 876.5540) to the blood compartment of the dialyzer and back to the patient. (2) The conventional dialyzer allows a transfer of water and solutes between the blood and the dialysate through the semipermeable membrane. The semipermeable membrane of the conventional dialyzer has a sufficiently low permeability to water that an ultrafiltration controller is not required to prevent excessive loss of water from the patient's blood. This conventional dialyzer does not include hemodialyzers with the disposable inserts (Kiil type) (§ 876.5830) or dialyzers of high permeability (§ 876.5860). (3) The dialysate delivery system consists of mechanisms that monitor and control the temperature, conductivity, flow rate, and pressure of the dialysate and circulates dialysate through the dialysate compartment of the dialyzer. The dialysate delivery system includes the dialysate concentrate for hemodialysis (liquid or powder) and alarms to indicate abnormal dialysate conditions. This dialysate delivery system does not include the sorbent regenerated dialysate delivery system for hemodialysis (§ 876.5600), the dialysate delivery system of the peritoneal dialysis system and accessories (§ 876.5630), or the controlled dialysate delivery system of the high permeability hemodialysis system § 876.5860). (4) Remote accessories to the hemodialysis system include the unpowered dialysis chair without a scale, the powered dialysis chair without a scale, the dialyzer holder set, dialysis tie gun and ties, and hemodialysis start/stop tray.
K022913 — NXSTAGE PREMIXED DIALYSATE · Nxstage Medical, Inc. · Oct 21, 2002
K053286 — MODIFICATION TO: NXSTAGE PUREFLOW-B SOLUTION · Nxstage Medical, Inc. · Dec 22, 2005
K042045 — NXSTAGE PUREFLOW-B SOLUTION · Nxstage Medical, Inc. · Aug 27, 2004
K033386 — NXSTAGE PUREFLOW SOLUTION · Nxstage Medical, Inc. · Nov 26, 2003
K013655 — NXSTAGE DIALYSATE CONCENTRATE, MODEL RFC-100 · Nxstage Medical, Inc. · Feb 4, 2002
Submission Summary (Full Text)
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). The logo consists of two parts: the Department of Health & Human Services logo on the left and the FDA acronym with the full name of the agency on the right. The FDA part of the logo is in blue, with the acronym in a square and the full name written out to the right of the square.
January 18, 2024
NxStage Medical, Inc. Caitlin Kalda Senior Lead, Regulatory Affairs North America 350 Merrimack Street Lawrence, Massachusetts 01843
Re: K233213
Trade/Device Name: NxStage PureFlow Solution Regulation Number: 21 CFR 876.5820 Regulation Name: Hemodialysis system and accessories Regulatory Class: Class II Product Code: KPO Dated: December 19, 2023 Received: December 19, 2023
Dear Caitlin Kalda:
We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device"
{1}------------------------------------------------
(https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download).
Your device is also subject to, among other requirements, the Quality System (QS) regulation (21 CFR Part 820), which includes, but is not limited to, 21 CFR 820.30. Design controls; 21 CFR 820.90. Nonconforming product; and 21 CFR 820.100, Corrective and preventive action. Please note that regardless of whether a change requires premarket review. the OS regulation requires device manufacturers to review and approve changes to device design and production (21 CFR 820.30 and 21 CFR 820.70) and document changes and approvals in the device master record (21 CFR 820.181).
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic.
{2}------------------------------------------------
See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
# Gema Gonzalez -S
Maura Rooney, MS Assistant Director DHT3A: Division of Renal, Gastrointestinal, Obesity and Transplant Devices OHT3: Office of Gastrorenal, ObGyn, General Hospital, and Urology Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health
Enclosure
{3}------------------------------------------------
# Indications for Use
510(k) Number (if known) K233213
Device Name NxStage PureFlow Solution
Indications for Use (Describe)
NxStage PureFlow Solution is indicated for use with NxStage renal replacement therapy systems that utilize sterile premixed dialysate during hemodialysis for chronic care patients.
Type of Use (Select one or both, as applicable)
| <span style="font-family: Arial, sans-serif;"> <input checked="true" type="checkbox"/> Prescription Use (Part 21 CFR 801 Subpart D)</span> |
|--------------------------------------------------------------------------------------------------------------------------------------------|
| <span style="font-family: Arial, sans-serif;"> <input type="checkbox"/> Over-The-Counter Use (21 CFR 801 Subpart C)</span> |
## CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{4}------------------------------------------------
Image /page/4/Picture/2 description: The image shows the Fresenius Medical Care logo. The logo consists of a blue chevron-like symbol on the left and the text "FRESENIUS MEDICAL CARE" on the right. The text is also in blue, with "FRESENIUS" on the top line and "MEDICAL CARE" on the bottom line.
NxStage PureFlow Solution Traditional 510(k)
#### 510(K) SUMMARY 5.
This 510(k) Summary is in accordance with the requirements of the Safe Medical Device Act (SMDA) of 1990. The content of this 510(k) summary is provided in conformance with 21 CFR § 807.92.
#### 5.1. Submitter's Information
| Name: | NxStage Medical, Inc. |
|-------------------|-----------------------------------------------|
| Address: | 350 Merrimack Street<br>Lawrence MA 01843 USA |
| Phone: | (617) 831-3469 |
| Fax: | (781) 699-9635 |
| Contact Person: | Caitlin Kalda, Senior Lead |
| Preparation Date: | 28 September 2023 |
#### 5.2. Device Name
| Trade Name: | NxStage PureFlow Solution |
|--------------------|--------------------------------------------------------------|
| Common Name: | Premixed Dialysate |
| Regulation Name: | Hemodialysis Systems and Accessories |
| Regulatory Class: | Class II per 21 CFR § 876.5820 |
| Product Code: | 78 KPO |
| Product Code Name: | Dialysate Concentrate for Hemodialysis<br>(Liquid of Powder) |
| FDA Review Panel: | Gastroenterology/Urology |
#### 5.3. Legally Marketed Predicate Device
The legally marketed predicate for the proposed device is the NxStage PureFlow Solution, as cleared through K033386 (November 26, 2003). This predicate has not been subject to a design-related recall.
#### 5.4. Device Description
#### Device Identification: 5.4.1.
The NxStage PureFlow Solution is a family of premixed dialysate solution formulations which are sterile, non-pyrogenic, fully constituted dialysate solutions provided in 5L single use flexible bags.
| Catalog Number | Description |
|----------------|------------------------|
| RFP-204-A* | 5000mL, 40 Lactate, 1K |
| RFP-204-B* | |
| RFP-204-C* | |
#### Table 1: Device Identification
{5}------------------------------------------------
Image /page/5/Picture/1 description: The image contains the logo for Fresenius Medical Care. The logo consists of a blue symbol on the left and the text "FRESENIUS MEDICAL CARE" on the right. The symbol is made up of three downward-pointing chevrons stacked on top of each other. The text is in a bold, sans-serif font, with "FRESENIUS" on the top line and "MEDICAL CARE" on the bottom line.
| RFP-205-A | 5000mL, 35 Lactate, 3K |
|-----------|------------------------|
| RFP-205-B | |
| RFP-205-C | |
| RFP-207-A | 5000mL, 45 Lactate, 1K |
| RFP-207-B | |
| RFP-207-C | |
| RFP-209-A | 5000mL, 45 Lactate, 2K |
| RFP-209-B | |
| RFP-209-C | |
| RFP-211-A | 5000mL, 40 Lactate, 2K |
| RFP-211-B | |
| RFP-211-C | |
| RFP-220-A | 5000mL, 40 Lactate, 2K |
| RFP-220-B | |
| RFP-220-C | |
| RFP-221-A | 5000mL, 40 Lactate, 3K |
| RFP-221-B | |
| RFP-221-C | |
| RFP-222-A | 5000mL, 45 Lactate, 3K |
| RFP-222-B | |
| RFP-222-C | |
| RFP-223-A | 5000mL, 35 Lactate, 2K |
| RFP-223-B | |
| RFP-223-C | |
| RFP-224-A | 5000mL, 35 Lactate, 4K |
| RFP-224-B | |
| RFP-224-C | |
#### Table 1: Device Identification
* suffix of A, B, or C indicates country or region specific labeling content and language translations for the same dialysate formulation.
#### Device Characteristics 5.4.2.
The NxStage PureFlow Solution is a sterile, nonpyrogenic, single-use device provided in a single chamber flexible bag. The device is sterilized using moist steam sterilization and SAL=10° using a validated sterilization cycle per ISO 17665-1:2006/(R)2013, Sterilization of health care products – Moist heat – Part 1: Requirements for the development, validation and routine control of a sterilization process for medical devices.
{6}------------------------------------------------
Image /page/6/Picture/0 description: The image shows the Fresenius Medical Care logo. The logo consists of a blue symbol on the left and the text "FRESENIUS MEDICAL CARE" on the right, also in blue. The symbol is made up of three downward-pointing chevrons stacked on top of each other. The text is in a bold, sans-serif font, with "FRESENIUS" on the top line and "MEDICAL CARE" on the bottom line.
#### 5.4.3. Environment of Use
The intended environments of use are:
- Dialysis Clinics ●
- Health Care Facilities
- Home
- Community at Large (e.g. travel/hotel) ●
#### Brief Written Description of the Device 5.4.4.
The NxStage PureFlow Solution is a product family of fully constituted, sterile, non-pyrogenic dialysate solutions that are provided in single use, single chamber 5L flexible bags. The solutions are intended for use with renal replacement therapy systems that utilize sterile premixed dialysate.
A range of formulations is offered which allows the physician to prescribe the electrolyte compositions that meet the specific needs of individual patients. The solutions are not designed for use with proportioning hemodialysis systems and are not intended for direct infusion into the patient.
| 5.4.5. | Materials of Use |
|--------|------------------|
|--------|------------------|
#### Table 2: Materials
| Where Used | Materials |
|---------------------|------------------------------------------|
| Dialysate Fluid | Sodium chloride |
| | Potassium chloride dihydrate |
| | Calcium chloride dihydrate |
| | Magnesium chloride hexahydrate |
| | Sodium lactate |
| | Dextrose monohydrate |
| | Hydrochloric acid |
| | Water for injection |
| Container Materials | Polypropylene (PP) |
| | Styrene ethylene butylene styrene (SEBS) |
| | Styrene-isoprene-rubber (SIS) |
| | Polycarbonate (PC) |
{7}------------------------------------------------
Image /page/7/Picture/0 description: The image contains the logo for Fresenius Medical Care. On the left is a blue symbol that looks like three chevrons stacked on top of each other. To the right of the symbol is the text "FRESENIUS MEDICAL CARE" in blue, with "FRESENIUS" on top of "MEDICAL CARE".
Duration and type of contact:
- Per ISO 10993-1 NxStage PureFlow Solution is categorized as: o
- External communicating device —
- -Blood path, indirect
- । Long term (>30 d)
#### Key Performance Specifications/Characteristic 5.4.6.
The NxStage PureFlow Solution is used as dialysate during hemodialysis therapy.
#### 5.5. Intended Use
The NxStage PureFlow Solution is intended for use with renal replacement therapy systems that utilize sterile premixed dialysate. The solutions are not designed for use with intermittent proportioning hemodialysis systems. The NxStage offering of premixed dialysate solutions allows the physician to prescribe different electrolyte compositions that meet the specific needs of individual patients.
#### Indications for Use 5.6.
NxStage PureFlow Solution is indicated for use with NxStage renal replacement therapy systems that utilize sterile premixed dialysate during hemodialysis for chronic care patients.
#### 5.7. Comparison of Technological Characteristics with the Predicate Device
The proposed device has the same intended use, technological characteristics, and principle of operation as the predicate. The proposed device is similar in container design and configuration as compared to the predicate device and has similar performance specifications.
#### 5.8. Performance Data
Performance testing has been conducted to support the determination of substantial equivalence with regards to design, safety, and efficacy. The testing is summarized in Table 3.
{8}------------------------------------------------
Image /page/8/Picture/0 description: The image contains the logo for Fresenius Medical Care. On the left side of the logo is a blue symbol that consists of three downward-pointing chevrons stacked on top of each other. To the right of the symbol is the text "FRESENIUS MEDICAL CARE" in a bold, blue sans-serif font. The text is arranged in two lines, with "FRESENIUS" on the top line and "MEDICAL CARE" on the bottom line.
| Test Conducted | Test Method Description |
|-------------------------------------|-----------------------------------------------------------------|
| Analytical testing | Per device specifications and applicable sections of USP and EP |
| Physical testing of container (bag) | Per device specifications |
| Container closure integrity testing | Per ISO 11607-1 |
| Temperature excursion testing | Per device specifications |
| Stability/Shelf-life testing | Per ICH Q1A to support 24-month shelf life |
| Ship testing | Per ASTM D4169-22 |
#### Table 3: Performance Testing Summary
#### Biocompatibility Testing 5.8.1.
The NxStage PureFlow Solution has been evaluated for biological safety in accordance with ISO 10993 -1 and the FDA guidance document Use of International Standard ISO 10993-1, "Biological evaluation of medical devices - Part 1 : Evaluation and testing within a risk management process", published Sept 4, 2020. The following testing was performed:
- Chemical characterization
- Toxicological Risk Assessment ●
- Cytotoxicity ●
- Sensitization ●
- Irritation
- Pyrogenicity ●
- Hemocompatibility ●
- Systemic toxicity
#### Human Factors Validation Testing 5.8.2.
The NxStage PurceFlow Solution is found to be safe and effective for its intended user, use, and use environments in accordance with the FDA guidance document "Applying Human Factors and Usability Engineering to Medical Devices (Feb 3, 2016) and applicable sections of AAMI/ANSI/EC 62366-1.
#### 5.8.3. Electrical Safety and Electromagnetic Compatibility (EMC)
Not Applicable - The device subject of this submission does not have any electronic features
#### Software Verification and Validation Testing 5.8.4.
Not Applicable - The device subject of this submission does not have any software features.
{9}------------------------------------------------
Image /page/9/Picture/0 description: The image contains the logo for Fresenius Medical Care. On the left side of the logo is a blue symbol that consists of three downward-pointing chevrons stacked on top of each other. To the right of the symbol is the text "FRESENIUS" in a bold, blue font, with the words "MEDICAL CARE" stacked below it in the same font.
#### Animal Studies 5.8.5.
Not Applicable – The device subject of this submission did not use any animal studies
#### 5.8.6. Clinical Studies
Not Applicable - The device subject of this submission did not require clinical studies
#### 5.9. Conclusion
NxStage believes that the information and data provided in this submission clearly describes the proposed device and demonstrates that the device is adequately designed for the labeled indications for use. Verification and validation was conducted to characterize the proposed device and the predetermined acceptance criteria were met. Results of this testing have documented that the proposed NxStage PureFlow Solution is suitable for the labeled indications for use.
The proposed modifications to the NxStage PureFlow Solution are substantially equivalent to the predicate device with regards to design, intended use, principle of operation and materials of construction. Differences between the proposed device and the predicate device do not introduce any new concerns regarding the safety of efficacy of the device. NxStage Medical, Inc. concludes that within the meaning of the Medical Device Amendments Act of 1976, the NxStage PureFlow Solution product family is safe and effective for its intended use.
Panel 1
/
Sort by
Ready
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.