K102712 · Imed Technology, Inc. · KNT · Dec 16, 2011 · Gastroenterology, Urology
Device Facts
Record ID
K102712
Device Name
MED GRAVITY SET, IMED PUMP SET
Applicant
Imed Technology, Inc.
Product Code
KNT · Gastroenterology, Urology
Decision Date
Dec 16, 2011
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 876.5980
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
The devices in the iMed Technology Enteral Feeding Set family are used to dispense liquid nutrients (feeding solutions) at a user controlled rate. These enteral feeding sets interface with the patient's feeding tube and may use gravity or an enteral feeding pump to dispense the feeding solution. The devices may include a bag to contain the feeding solution and/or a spike to connect to a pre-filled container. These devices are food containing. These devices are to be issued by prescription only.
Device Story
iMed Gravity Set and iMed Pump Set are enteral feeding delivery systems. Devices consist of tubing, connectors, and optional feeding solution bags or spikes for pre-filled containers. Function as interface between nutrient source and patient feeding tube. Gravity sets rely on hydrostatic pressure; pump sets interface with enteral feeding pumps to regulate flow rate. Used in clinical or home settings by patients or caregivers under prescription. Facilitates controlled delivery of liquid nutrition to patients unable to ingest food orally. No electronic or software components; purely mechanical fluid delivery accessories.
Clinical Evidence
Bench testing only.
Technological Characteristics
Mechanical enteral feeding delivery sets. Components include PVC or similar medical-grade tubing, drip chambers, flow control clamps, and connectors. Gravity-fed or pump-compatible designs. Non-sterile or sterile fluid path options. No energy source, software, or electronic components.
Indications for Use
Indicated for patients requiring enteral nutrition delivery via feeding tube; used to dispense liquid nutrients at a user-controlled rate via gravity or enteral feeding pump. Prescription use only.
Regulatory Classification
Identification
A gastrointestinal tube and accessories is a device that consists of flexible or semi-rigid tubing used for instilling fluids into, withdrawing fluids from, splinting, or suppressing bleeding of the alimentary tract. This device may incorporate an integral inflatable balloon for retention or hemostasis. This generic type of device includes the hemostatic bag, irrigation and aspiration catheter (gastric, colonic, etc.), rectal catheter, sterile infant gavage set, gastrointestinal string and tubes to locate internal bleeding, double lumen tube for intestinal decompression or intubation, feeding tube, gastroenterostomy tube, Levine tube, nasogastric tube, single lumen tube with mercury weight balloon for intestinal intubation or decompression, and gastro-urological irrigation tray (for gastrological use).
Special Controls
*Classification.* (1) Class II (special controls). The barium enema retention catheter and tip with or without a bag that is a gastrointestinal tube and accessory or a gastronomy tube holder accessory is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 876.9.(2) Class I (general controls) for the dissolvable nasogastric feed tube guide for the nasogastric tube. The class I device is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to § 876.9.
Related Devices
K132424 — METRIXCARE · The Metrix Company · Feb 12, 2014
K240052 — Disposable Enteral Feeding Sets · Beijing L&Z Medical Technology Development Co., Ltd. · Jul 3, 2024
K081725 — ENTERAL PUMP SET, MODEL 20-1010 · Viasys Medsystems · Nov 7, 2008
K082863 — DYNAREX ENTERAL DELIVERY PUMP SET WITH SPIKE (1000ML), DYNAREX ENTERAL FEEDING SETS FOR GRAVITY AND PUMP USE (1000ML) · Dynarex Corporation · Jan 9, 2009
K992719 — AMSURE ENTERAL DELIVERY PUMP SET WITH SPIKE · Amsino Intl., Inc. · Jan 12, 2000
Submission Summary (Full Text)
{0}------------------------------------------------
Image /page/0/Picture/1 description: The image shows the seal of the Department of Health & Human Services USA. The seal is circular and contains the department's name around the perimeter. In the center of the seal is an abstract image of an eagle.
Food and Drug Administration 10903 New Hampshire Avenue Document Mail Center - WO66-G609 Silver Spring, MD 20993-0002
Mr. Brandon Converse Director of Product Refinement iMed Technology, Inc. 17408 Tamaron Drive DALLAS TX 75287
DEC 1-6 2011
Re: K102712
Trade/Device Name: iMed Gravity Set, iMed Pump Set Regulation Number: 21 CFR\$ 876.5980 Regulation Name: Gastrointestinal tube and accessories Regulatory Class: II Product Code: KNT Dated: November 3, 2011 Received: December 5, 2011
Dear Mr. Converse:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related
{1}------------------------------------------------
## Page 2
adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please go to http://www.fda.gov/AboutFDA/CentersOffices/CDRHOffices/ucm115809.htm for the Center for Devices and Radiological Health's (CDRH's) Office of Compliance. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm.
Sincerely vours.
Hubert Lenoir MD
Herbert P. Lerner, M.D., Director (Acting) Division of Reproductive, Gastro-Renal and Urological Devices Office of Device Evaluation Center for Devices and Radiological Health
## Enclosure
{2}------------------------------------------------
Premarket Notification 510(k) Traditional
4.0
## INDICATIONS FOR USE STATMENT
The statement of Indications for Use for the Enteral Feeding Sets, in compliance with 21 CFR 807.87(e), is provided below:
Applicant: iMed Technology, Inc.
Device Name: iMed Gravity Set, iMed Pump Set
Indications For Use:
The devices in the iMed Technology Enteral Feeding Set family are used to dispense liquid nutrients (feeding solutions) at a user controlled rate. These enteral feeding sets interface with the patient's feeding tube and may use gravity or an enteral feeding pump to dispense the feeding solution. The devices may include a bag to contain the feeding solution and/or a spike to connect to a pre-filled container. These devices are food containing. These devices are to be issued by prescription only.
(PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
(Per 21 CFR 801.109)
(Division Sign-Off)
Division of Reproductive, Gastro-Renal, and
Urological Devices
510(k) Number K102712
For Prescription Use
(Per 21 CFR 801.109)
iMed Technology, Inc
Panel 1
/
Sort by
Ready
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.