K240052 · Beijing L&Z Medical Technology Development Co., Ltd. · PIF · Jul 3, 2024 · Gastroenterology, Urology
Device Facts
Record ID
K240052
Device Name
Disposable Enteral Feeding Sets
Applicant
Beijing L&Z Medical Technology Development Co., Ltd.
Product Code
PIF · Gastroenterology, Urology
Decision Date
Jul 3, 2024
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 876.5980
Device Class
Class 2
Attributes
Therapeutic, Pediatric
Indications for Use
The Disposable Enteral Feeding Sets is intended to be used for enteral feeding nutrition into gastrointestinal tract by connecting with nasogastric tube by gravity. The devices may include a bag to contain the feeding nutrition and/or a spike to connect to a prefilled container. The device is used for children, adolescents, and adults.
Device Story
Disposable Enteral Feeding Sets facilitate gravity-fed enteral nutrition delivery. Components include a feeding bag for nutritional formula and/or a spike for connection to prefilled containers. The device connects to a patient's nasogastric tube. Operation is manual, relying on gravity to dispense liquid nutrients. Used in clinical or home settings by healthcare providers or patients/caregivers. The device provides a controlled flow of nutrition to support patients unable to ingest food orally, benefiting those requiring enteral support.
Clinical Evidence
Bench testing only. Performance testing conducted per ISO 20695:2020, ISO 8536-4:2022, ISO 8536-4:2020, ISO 80369-20:2015, and USP <71> Sterility Tests. Testing confirmed compliance with physical, chemical, and sterility specifications across multiple model variations.
Technological Characteristics
Gravity-fed enteral feeding set. Components include feeding bags and spikes for container connection. Materials evaluated for biocompatibility. Connectivity via standard enteral connectors (ENFit). Sterilization provided for sterile versions. Complies with ISO 20695, ISO 8536-4, and ISO 80369-20 standards.
Indications for Use
Indicated for enteral nutrition delivery into the gastrointestinal tract via nasogastric tube in children, adolescents, and adults.
Regulatory Classification
Identification
A gastrointestinal tube and accessories is a device that consists of flexible or semi-rigid tubing used for instilling fluids into, withdrawing fluids from, splinting, or suppressing bleeding of the alimentary tract. This device may incorporate an integral inflatable balloon for retention or hemostasis. This generic type of device includes the hemostatic bag, irrigation and aspiration catheter (gastric, colonic, etc.), rectal catheter, sterile infant gavage set, gastrointestinal string and tubes to locate internal bleeding, double lumen tube for intestinal decompression or intubation, feeding tube, gastroenterostomy tube, Levine tube, nasogastric tube, single lumen tube with mercury weight balloon for intestinal intubation or decompression, and gastro-urological irrigation tray (for gastrological use).
Special Controls
*Classification.* (1) Class II (special controls). The barium enema retention catheter and tip with or without a bag that is a gastrointestinal tube and accessory or a gastronomy tube holder accessory is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 876.9.(2) Class I (general controls) for the dissolvable nasogastric feed tube guide for the nasogastric tube. The class I device is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to § 876.9.
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which is a blue square with the letters "FDA" in white. To the right of the blue square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue.
July 3, 2024
Beijing L&Z Medical Technology Development Co., Ltd. Zhiling Li Product Registry Specialist No.8. M2-5 Block, Xinggu Industrial Developing Zone, Pinggu District Beijing, 101200 China
Re: K240052
Trade/Device Name: Disposable Enteral Feeding Sets Regulation Number: 21 CFR 876.5980 Regulation Name: Gastrointestinal Tube And Accessories Regulatory Class: Class II Product Code: PIF Dated: January 8, 2024 Received: June 5, 2024
Dear Zhiling Li:
We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device"
{1}------------------------------------------------
(https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download).
Your device is also subject to, among other requirements, the Quality System (QS) regulation (21 CFR Part 820), which includes, but is not limited to, 21 CFR 820.30. Design controls; 21 CFR 820.90. Nonconforming product; and 21 CFR 820.100, Corrective and preventive action. Please note that regardless of whether a change requires premarket review. the OS regulation requires device manufacturers to review and approve changes to device design and production (21 CFR 820.30 and 21 CFR 820.70) and document changes and approvals in the device master record (21 CFR 820.181).
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
> Sincerely. Stephanie Cole -S
for Anthony C. Lee, PhD, MBA Assistant Director DHT3A: Division of Renal, Gastrointestinal, Obesity and Transplant Devices OHT3: Office of Gastrorenal, ObGyn, General Hospital, and Urology Devices
{2}------------------------------------------------
Office of Product Evaluation and Quality Center for Devices and Radiological Health
{3}------------------------------------------------
### DEPARTMENT OF HEALTH AND HUMAN SERVICES Food and Drug Administration
### Indications for Use
Form Approved: OMB No. 0910-0120 Expiration Date: 07/31/2026 See PRA Statement below.
Submission Number (if known) K240052
Device Name
Disposable Enteral Feeding Sets
Indications for Use (Describe)
The Disposable Enteral Feeding Sets is intended to be used for enteral feeding nutrition into gastrointestinal tract by connecting with nasogastric tube by gravity. The devices may include a bag to contain the feeding nutrition and/or a spike to connect to a prefilled container. The device is used for children, adolescents, and adults.
Type of Use (Select one or both, as applicable)
X Prescription Use (Part 21 CFR 801 Subpart D)
Over-The-Counter Use (21 CFR 801 Subpart C)
#### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{4}------------------------------------------------
K240052
| 510(k) #: K240052 | 510(k) Summary | K240052Page 1 of 2Prepared on: 2024-07-01 |
|-------------------|----------------|-------------------------------------------|
|-------------------|----------------|-------------------------------------------|
| Contact Details | 21 CFR 807.92(a)(1) |
|-----------------------------|-----------------------------------------------------------------------------------------------------------------------|
| Applicant Name | Beijing L&Z Medical Technology Development Co., Ltd. |
| Applicant Address | No.8, M2-5 Block, Xinggu Industrial Developing Zone, Pinggu District 101200 BEIJING, P.R. CHINA. Beijing 101200 China |
| Applicant Contact Telephone | 15104955040 |
| Applicant Contact | Ms. Zhiling Li |
| Applicant Contact Email | lizl@lingze.com |
| Device Name | 21 CFR 807.92(a)(2) |
|---------------------|---------------------------------------------------------|
| Device Trade Name | Disposable Enteral Feeding Sets |
| Common Name | Gastrointestinal tube and accessories |
| Classification Name | Gastrointestinal Tubes With Enteral Specific Connectors |
| Regulation Number | 21 CFR 876.5980 |
| Product Code(s) | PIF |
| Legally Marketed Predicate Devices | 21 CFR 807.92(a)(3) |
|------------------------------------|---------------------|
|------------------------------------|---------------------|
| Predicate # | Predicate Trade Name (Primary Predicate is listed first) | Product Code |
|-------------|--------------------------------------------------------------------|--------------|
| K142539 | Enteralite Infinity Enteral Pump Delivery Set, 1200 ML Enteral Fee | PIF |
| Device Description Summary | 21 CFR 807.92(a)(4) |
|----------------------------|---------------------|
|----------------------------|---------------------|
The proposed device, Disposable Enteral Feeding Sets, is intended to be used for enteral feeding nutrition into gastrointestinal tract by connecting with nasogastric tube by gravity. The devices may include a bag to contain the feeding nutrition and/or a spike to connect toconnecting with nasogastric tube by gravity. The devices may include a bag to contain the feeding nutrition and/or a spike to connect to a prefilled container. The device is used for children, adolescents and adults.
The proposed device is provided sterile or non-sterile and single use.
# Intended Use/Indications for Use
The Disposable Enteral Feeding Sets is intended to be used for enteral feeding nutrition into gastrointestinal tract by connecting with nasogastric tube by gravity. The devices may include a bag to contain the feeding nutrition and/or a spike to a prefilled container. The device is used for children, adolescents and adults.
# Indications for Use Comparison
Indication for Use:
Proposed Device: The Disposable Enteral Feeding Sets is intended to be used for enteral feeding nutrition into gastrointestinal tract by connecting with nasgastric tube by gravity. The devices may include a bag to contain the feeding nutrition and/or a spike to connect to a prefilled container. The device is used for children, adolescents and adults.
Predicate Device: The devices in this product family are used to dispense liquid nutrients (feeding solution) at a preprogrammed pump or user controlled rate. These enteral feeding sets interface with the patient's feeding tube and may use gravity or an enteral feeding pump to dispense feeding solution. The devices may include a bag to contain the feeding solution and/or a spike to connect to a prefilled container.
# 21 CFR 807.92(a)(5)
21 CFR 807.92(a)(5)
{5}------------------------------------------------
## Different - Indication for Use
The predicate device can be used for gravity feeding. Therefore, the description of the indication for use of the proposed device differs from that of the predicate device. The proposed device only uses gravity to dispense feeding solution. The indications for use of the proposed device is covered by that of the predicate device. Therefore, this difference will not raise new question on the safety and effectiveness of the proposed device.
# Technological Comparison
## Different - Main Configuration
The configuration of the proposed device is the same as that of the predicate device (1200 mL Enteral Feeding Delivery Set), only the naming of the components is different. Therefore, this difference will not raise new question on the new safety and effectiveness of the proposed device.
### Different - Bag capacity
The Bag capacity of the proposed device is different from that of the proposed device offers more options. However, the bag is just used to store the nutritional formula. In addition, the performed on the proposed device and the test result of bag part showed that the proposed device met the pre-determined acceptance criteria. Therefore, this difference on bag capacity will not raise new question on the new safety and effectiveness of the proposed device.
### Different - Spike types
The Soike types of the proposed device is different from that of the proposed device offers more options. However, the spike is just used to connect to a prefilled container. In addition, the performance test has been performed on the proposed device and the test result of spike part showed that the performance of the pro-determined acceptance criteria. Therefore, this difference on bag capacity will not raise new question on the new safety and effectiveness of the proposed device.
### Different - Patient-contact material
The patient-contact material of the proposed device is different from that of the biocompatibility test has been performed on the proposed device and the test results show that the materials of proposed device will not have the on the patient. Therefore, this difference will not raise new question on the new safety and effectiveness of the proposed device.
#### Non-Clinical and/or Clinical Tests Summary & Conclusions 21 CFR 807.92(b)
The performance test was conducted on the proposed device to evaluate the proposed device according to Final product inspection procedure, ISO 20695:2020 Annex C and Annex D, ISO 8536-4:2022 Annex B, ISO 8536-4:2020 Annex B.4,1SO 80369-20:2015 and USP <71> Sterility Tests. The specification BEEGA1 22211, BEEGA1 32211, BEEGA1 4221, BEEGA1 52211, BEEGA1 62211, BEEGB1 53211, BEEGB1 13011 have already covered the variation of all specifications of Disposable Enteral Feeding Sets. Therefore, the specification BEGA1 22211, BEEGA1 32211, BEEGA1 42211, BEEGA1 52211, BEEGA1 62211, BEEGB1 53211, BEGB1 13011 were selected to conduct the performance test and the test result showed that the aged proposed device met the acceptance criteria. Copies of the test report are provided in as follow:
Tab #24 Chemical Performance Test Report Tab #25 Physical Performance Test Report Tab #26 Sterility Test Exhibit #29 ENfit Physical Performance Test Report
## 21 CFR 807.92(a)(6)
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.