COBE C HEMODIALYSIS DELIVERY SYSTEM , COBE CARTRIDGE BLOOD TUBING SET
Device Facts
| Record ID | K982760 |
|---|---|
| Device Name | COBE C HEMODIALYSIS DELIVERY SYSTEM , COBE CARTRIDGE BLOOD TUBING SET |
| Applicant | Gambro Healthcare |
| Product Code | KDI · Gastroenterology, Urology |
| Decision Date | Nov 4, 1998 |
| Decision | SESE |
| Submission Type | Traditional |
| Regulation | 21 CFR 876.5860 |
| Device Class | Class 2 |
| Attributes | Therapeutic |
Indications for Use
The COBE C Hemodialysis Delivery System is indicated for patients in acute or chronic renal failure and when the physician prescribes hemodialysis or ultrafiltration. The COBE® Cx System may be used with both high permeability and low permeability (conventional) dialyzers. The COBE® Cartridge Set is intended for use with the COBE Centrysystem 3 Hemodialysis Delivery System, or the COBE C* Hemodialysis Delivery System only.
Device Story
COBE Cx Hemodialysis Delivery System and COBE Cartridge Blood Tubing Set facilitate hemodialysis and ultrafiltration for patients with renal failure. System manages extracorporeal blood circuit; controls fluid removal; monitors dialysis parameters. Used in clinical settings by trained healthcare professionals. Physician prescribes therapy; system executes delivery via integrated tubing set. Benefits include effective solute clearance and fluid management for renal patients.
Clinical Evidence
No clinical data provided; substantial equivalence based on device description and intended use.
Technological Characteristics
Hemodialysis delivery system with integrated cartridge blood tubing set. Supports high and low permeability dialyzers. Operates as a fluid management and extracorporeal circuit control system.
Indications for Use
Indicated for patients with acute or chronic renal failure requiring hemodialysis or ultrafiltration. Compatible with high and low permeability dialyzers.
Regulatory Classification
Identification
A high permeability hemodialysis system is a device intended for use as an artificial kidney system for the treatment of patients with renal failure, fluid overload, or toxemic conditions by performing such therapies as hemodialysis, hemofiltration, hemoconcentration, and hemodiafiltration. Using a hemodialyzer with a semipermeable membrane that is more permeable to water than the semipermeable membrane of the conventional hemodialysis system (§ 876.5820), the high permeability hemodialysis system removes toxins or excess fluid from the patient's blood using the principles of convection (via a high ultrafiltration rate) and/or diffusion (via a concentration gradient in dialysate). During treatment, blood is circulated from the patient through the hemodialyzer's blood compartment, while the dialysate solution flows countercurrent through the dialysate compartment. In this process, toxins and/or fluid are transferred across the membrane from the blood to the dialysate compartment. The hemodialysis delivery machine controls and monitors the parameters related to this processing, including the rate at which blood and dialysate are pumped through the system, and the rate at which fluid is removed from the patient. The high permeability hemodialysis system consists of the following devices:(1) The hemodialyzer consists of a semipermeable membrane with an in vitro ultrafiltration coefficient (K uf ) greater than 8 milliliters per hour per conventional millimeter of mercury, as measured with bovine or expired human blood, and is used with either an automated ultrafiltration controller or anther method of ultrafiltration control to prevent fluid imbalance.(2) The hemodialysis delivery machine is similar to the extracorporeal blood system and dialysate delivery system of the hemodialysis system and accessories (§ 876.5820), with the addition of an ultrafiltration controller and mechanisms that monitor and/or control such parameters as fluid balance, dialysate composition, and patient treatment parameters (e.g., blood pressure, hematocrit, urea, etc.). (3) The high permeability hemodialysis system accessories include, but are not limited to, tubing lines and various treatment related monitors (e.g., dialysate pH, blood pressure, hematocrit, and blood recirculation monitors).
Special Controls
*Classification.* Class II. The special controls for this device are FDA's:(1) “Use of International Standard ISO 10993 ‘Biological Evaluation of Medical Device—Part I: Evaluation and Testing,’ ” (2) “Guidance for the Content of 510(k)s for Conventional and High Permeability Hemodialyzers,” (3) “Guidance for Industry and CDRH Reviewers on the Content of Premarket Notifications for Hemodialysis Delivery Systems,” (4) “Guidance for the Content of Premarket Notifications for Water Purification Components and Systems for Hemodialysis,” and (5) “Guidance for Hemodialyzer Reuse Labeling.”
Reference Devices
- COBE Centrysystem 3 Hemodialysis Delivery System
Related Devices
- K113023 — STREAMLINE AIRLESS SYSTEM SET WITH PRE-ATTACHED DIALYZER · Nxstage Medical, Inc. · Dec 21, 2012
- K030099 — BAXTER ARENA HEMODIALYSIS DELIVERY SYSTEM, MODEL 1571278000 · Baxter Healthcare Corp · Jun 17, 2003
- K041424 — NXSTAGE SYSTEM ONE · Nxstage Medical, Inc. · Jul 2, 2004
- K030470 — NXSTAGE SYSTEM ONE · Nxstage Medical, Inc. · Jul 7, 2003
- K040696 — MODIFICATION TO NXSTAGE SYSTEM ONE · Nxstage Medical, Inc. · May 26, 2004