NXSTAGE DIALYSATE PREPARATION MODULE, MODEL NX0269

K043436 · Nxstage Medical, Inc. · FKR · Mar 17, 2005 · Gastroenterology, Urology

Device Facts

Record IDK043436
Device NameNXSTAGE DIALYSATE PREPARATION MODULE, MODEL NX0269
ApplicantNxstage Medical, Inc.
Product CodeFKR · Gastroenterology, Urology
Decision DateMar 17, 2005
DecisionSESE
Submission TypeTraditional
Regulation21 CFR 876.5820
Device ClassClass 2
AttributesTherapeutic

Indications for Use

The NxStage Dialysate Preparation Module is an optional accessory to the NxStage System One that prepares dialysate for use during hemodialysis, as prescribed by the physician.

Device Story

The NxStage Dialysate Preparation Module is an accessory to the NxStage System One hemodialysis system. It functions by mixing purified water (AAMI standard) with AAMI-compliant concentrate to produce dialysate meeting ANSI/AAMI RD52:2004 and RD62:2001 standards. The device incorporates safety features including conductivity measurement, an automatic safety clamp, a mix-to-use timer, and ultrafilters/I.V. sterilization filters to ensure dialysate purity. It is intended for use in hemodialysis therapies as prescribed by a physician. The module automates the preparation process, ensuring the dialysate meets required purity and composition specifications before use in the dialysis circuit, thereby supporting safe and effective hemodialysis treatment.

Clinical Evidence

No clinical data. Performance was demonstrated through bench testing and simulated use testing to characterize the module's performance and verify safety features against predicate device standards.

Technological Characteristics

Proportioning subsystem for hemodialysis. Mixes purified water and concentrate to produce AAMI-compliant dialysate. Features: conductivity measurement, automatic safety clamp, mix-to-use timer, ultrafilters, and I.V. sterilization filters. Designed for use with the NxStage System One.

Indications for Use

Indicated for use as an optional accessory to the NxStage System One to prepare dialysate for hemodialysis patients as prescribed by a physician. No known contraindications.

Regulatory Classification

Identification

A hemodialysis system and accessories is a device that is used as an artificial kidney system for the treatment of patients with renal failure or toxemic conditions and that consists of an extracorporeal blood system, a conventional dialyzer, a dialysate delivery system, and accessories. Blood from a patient flows through the tubing of the extracorporeal blood system and accessories to the blood compartment of the dialyzer, then returns through further tubing of the extracorporeal blood system to the patient. The dialyzer has two compartments that are separated by a semipermeable membrane. While the blood is in the blood compartment, undesirable substances in the blood pass through the semipermeable membrane into the dialysate in the dialysate compartment. The dialysate delivery system controls and monitors the dialysate circulating through the dialysate compartment of the dialyzer.(1) The extracorporeal blood system and accessories consists of tubing, pumps, pressure monitors, air foam or bubble detectors, and alarms to keep blood moving safely from the blood access device and accessories for hemodialysis (§ 876.5540) to the blood compartment of the dialyzer and back to the patient. (2) The conventional dialyzer allows a transfer of water and solutes between the blood and the dialysate through the semipermeable membrane. The semipermeable membrane of the conventional dialyzer has a sufficiently low permeability to water that an ultrafiltration controller is not required to prevent excessive loss of water from the patient's blood. This conventional dialyzer does not include hemodialyzers with the disposable inserts (Kiil type) (§ 876.5830) or dialyzers of high permeability (§ 876.5860). (3) The dialysate delivery system consists of mechanisms that monitor and control the temperature, conductivity, flow rate, and pressure of the dialysate and circulates dialysate through the dialysate compartment of the dialyzer. The dialysate delivery system includes the dialysate concentrate for hemodialysis (liquid or powder) and alarms to indicate abnormal dialysate conditions. This dialysate delivery system does not include the sorbent regenerated dialysate delivery system for hemodialysis (§ 876.5600), the dialysate delivery system of the peritoneal dialysis system and accessories (§ 876.5630), or the controlled dialysate delivery system of the high permeability hemodialysis system § 876.5860). (4) Remote accessories to the hemodialysis system include the unpowered dialysis chair without a scale, the powered dialysis chair without a scale, the dialyzer holder set, dialysis tie gun and ties, and hemodialysis start/stop tray.

Predicate Devices

Related Devices

Submission Summary (Full Text)

{0}------------------------------------------------ #### NxStage Medical, Inc. NxStage Dialysate Prep Module MAR 1 7 2005 510(k) Premarket Notification K043436 Page 1 of 2 # Section VIII: 510(k) Summary of Safety & Effectiveness This 510(k) Summary of Safety and Effectiveness is being submitted in accordance with the requirements of SMDA 1990. #### A. Submitter's Information: | Name: | NxStage Medical, Inc. | |----------------------|---------------------------------------------------------| | Address: | 439 South Union Street, Suite 501<br>Lawrence, MA 01843 | | Phone: | (978) 687-4700 | | Fax: | (978) 687-4800 | | Contact Person: | Norma LeMay<br>Manager, Regulatory Affairs | | Date of Preparation: | December 10, 2004 | #### B. Device Name: | Trade Name: | NxStage Dialysate Preparation Module | |----------------------|--------------------------------------------------------------------------------------------------------------| | Common/Usual Name: | Subsystem, proportioning | | Classification Name: | Hemodialysis System & Accessories (Class I<br>medical device under 21 CFR 875.5820, (Product<br>Code 78 FKR) | ### C. Substantial Equivalence/Predicate Devices: The NxStage Dialysate Prep Module is substantially equivalent to the following legally marketed predicate devices previously cleared by FDA: - Aksys PHD System, K010131 (cleared on 03/26/02) . - Fresenius 2008K Hemodialysis System, K994267 (cleared on 03/16/00) . - Gambro Bicarbonate Module BCM-10, K810127 (cleared on 02/25/81) . {1}------------------------------------------------ # NxStage Medical, Inc. NxStage Dialysate Prep Module 510(k) Premarket Notification K043436 Page 2 of 2 # Section VIII: 510(k) Summary of Safety & Effectiveness # D. Device Description/Indications for Use: Device Booonproniment and optional accessory to the NxStage System The NAGtage Dialyone in that mixes with concentrate to produce One that pro treats crequirements of the ANSI/AAMI Standards, RD52:2004 and The dialysate is to be used with the NxStage System One during RD62:2001. hemodialysis therapies. ### Indications for use: The NxStage Dialysate Preparation Module is an optional accessory to the NxStage rne NAGtago Dialysate for use during hemodialysis, as prescribed by the physician. There are no known contraindications with the use of the NxStage Dialysate Preparation Module. ## E. Technological Characteristics: The proposed Prep Module has the same technological characteristics and is similar in design and configuration as compared to the predicate devices. It uses purified AAMI water that mixes with AAMI concentrate to produce AAMI dialysate. The proposed device is designed and assembled with components commonly found in the predicate devices. Importantly, the NxStage Prep Module incorporates the proper safety features, such as conductivity measurement, automatic safety clamp, mix to use timer, and ultrafilters and I.V. sterilization filters to insure dialysate purity. ## F. Summary of Non-Clinical Test/Performance Testing NxStage believes that the information and data provided in this submission clearly describes the proposed device and demonstrates that the device is adequately designed for the labeled indication for use. As there is no technological differences as compared to the predicate devices, performance testing was deemed appropriate and was conducted to characterize the performance of the proposed Prep Module to provide a basis of comparison to the predicate devices. The Prep Module utilizes FDA cleared devices with minor modifications, and additional components which are simple to operate using the Verification and Validation testing, which includes provided device labeling. simulated use testing, sufficiently demonstrates that the Prep Module can be used safely as intended. All testing performed is summarized in Section V of this 510(k) submission. Results of the testing presented in this 510(k) have demonstrated that the NxStage Prep Module is substantially equivalent to the predicate devices and is suitable for the labeled indication for use. {2}------------------------------------------------ Image /page/2/Picture/1 description: The image shows a circular seal or logo. The seal features an abstract image of a bird in flight, composed of three curved lines. Encircling the bird is text that reads "DEPARTMENT OF HEALTH & HUMAN SERVICES U.S.A." Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850 MAR 1 7 2005 Ms. Norma LeMay Manager, Regulatory Affairs NxStage® Medical, Inc. 439 South Union Street, 5th Floor LAWRENCE MA 01843 Re: K043436 K043430 Trade/Device Name: NxStage® Dialysate Preparation Module Regulation Number: 21 CFR §876.5820 Regulation Name: Hemodialysis system and accessories Regulatory Class: II Product Code: 78 KPO and FKR Dated: March 3, 2005 Received: March 4, 2005 Dear Ms. LeMay: We have reviewed your Section 510(k) premarket notification of intent to market the device referenced we have review of your better of substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to the cholosate) to regary mantowa of the Medical Device Amendments, or to devices that have been ready 20, 1776, the chance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that recrassition in accercation in premarket approval application (PMA). You may, therefore, market the do not require upproval er a ve visions of the Act. The general controls provisions of the Act de fice, subjoct to the general registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. If your device is classified (see above) into either class II (Special Controls) or class III (Premarket It your devise is elective (t to such additional controls. Existing major regulations affecting your Applorary, the lia) 00 bely works of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register. Please be advised that FDA's issuance of a substantial equivalence determination does not mean that I reast of advised that 127 i our device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the I carral statues and regulations association and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set forth in the quality systems (QS) (21 CFR Part 801); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050. {3}------------------------------------------------ This letter will allow you to begin marketing your device as described in your Section 510(k) This letter will allow you to begin marketing your actrice of your device of your device to a legally premarket notification. The FDA finding of substantial equipand this pe premarket notification. The FDA Imaling of substantial equivalent of your device to proceed to the market. If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please If you desire specific advice for your device on our labering regarantists. contact the Office of Compliance at one of the following numbers, based on the regulation number a the top of this letter: | 21 CFR 876.xxxx | (Gastroenterology/Renal/Urology) | 240-276-0115 | |-----------------|----------------------------------|--------------| | 21 CFR 884.xxxx | (Obstetrics/Gynecology) | 240-276-0115 | | 21 CFR 892.xxxx | (Radiology) | 240-276-0120 | | Other | | 240-276-0100 | Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Also, please note the regulation entitled, "onsonmants" of the sunder the Act from the 807.97). You may obtain other gelleral intornation on your respeared its toll-free number (800) Division of Small Manufacturers, International and Consumer Assistance at its Division of Small 1143-6597 or at its Internet address 056-2041 or (2019/01/2019/11/industry/support/index.html. Sincerely yours, Nancy C. Brogdon Nancy C. Brogdon Director, Division of Reproductive, Abdominal, and Radiological Devices Office of Device Evaluation Center for Devices and Radiological Health Enclosure {4}------------------------------------------------ # Indications for Use | 510(k) Number (if known): | K043436 | |---------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------| | Device Name: | NxStage Dialysate Preparation Module | | Indications for Use: | The NxStage Dialysate Preparation Module is an optional<br>accessory to the NxStage System One that prepares<br>dialysate for use during hemodialysis, as prescribed by the<br>physician. | Prescription Use _____________________________________________________________________________________________________________________________________________________________ Use (Part 21 CFR 801 Subpart D) Over-The-Counter AND/OR (21 CFR 807 Subpart C) (PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED) # Concurrence of CDRH, Office of Device Evaluation (ODE) Nancy C. Brogdon (Division Sign-Off) Division of Reproductive, Abdominal, and Radiological Devices 510(k) Number _ Page 1 of 1
Innolitics

Panel 1

/
Sort by
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...