Advanix™ Pancreatic Stent and NaviFlex™ Rapid Exchange (RX) Pancreatic Delivery System and Pushers

K251291 · Boston Scientific · FGE · Jul 3, 2025 · Gastroenterology, Urology

Device Facts

Record IDK251291
Device NameAdvanix™ Pancreatic Stent and NaviFlex™ Rapid Exchange (RX) Pancreatic Delivery System and Pushers
ApplicantBoston Scientific
Product CodeFGE · Gastroenterology, Urology
Decision DateJul 3, 2025
DecisionSESE
Submission TypeTraditional
Regulation21 CFR 876.5010
Device ClassClass 2
AttributesTherapeutic

Indications for Use

The Advanix™ Pancreatic Stent and NaviFlex™ Rapid Exchange (RX) Pancreatic Delivery System and Pushers are intended for delivery of the stent to the pancreatic duct (PD): • Used to drain pancreatic ducts

Device Story

The Advanix™ Pancreatic Stent and NaviFlex™ RX Pancreatic Delivery System and Pushers consist of a plastic stent and a delivery catheter/pusher system. The device is used by physicians in an endoscopic setting to facilitate drainage of the pancreatic duct. The stent is delivered to the target site within the pancreatic duct using the RX delivery system. Stents are available in straight or single pigtail configurations, featuring various tip designs (rounded or tapered), barbs, and lateral drainage holes to assist in placement and patency. Markers (endoscopic or fluoroscopic) are integrated to assist with depth visualization during placement. The device provides a mechanical pathway for fluid drainage, relieving obstruction or supporting ductal patency. The system is intended for prescription use.

Clinical Evidence

Bench testing only. Testing included performance bench testing (corrosion, tensile, lumen patency, side drainage hole, integrity, length, outer diameter, pigtail shape/diameter, marker location, pusher deployment/trackability forces) and comprehensive biocompatibility testing (cytotoxicity, sensitization, intracutaneous reactivity, acute systemic toxicity, pyrogenicity, genotoxicity, implantation, and chemical characterization). Results demonstrate the device is safe and effective.

Technological Characteristics

Plastic pancreatic stent and delivery system. Features include straight or single pigtail shapes, trailing/leading barbs, lateral drainage holes, and endoscopic/fluoroscopic markers. Connectivity: None. Energy source: None (mechanical). Sterilization: Not specified. Software: None.

Indications for Use

Indicated for patients requiring drainage of the pancreatic duct.

Regulatory Classification

Identification

A biliary catheter and accessories is a tubular flexible device used for temporary or prolonged drainage of the biliary tract, for splinting of the bile duct during healing, or for preventing stricture of the bile duct. This generic type of device may include a bile collecting bag that is attached to the biliary catheter by a connector and fastened to the patient with a strap.

Special Controls

*Classification.* Class II (special controls). The device, when it is a bile collecting bag or a surgical biliary catheter that does not include a balloon component, is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 876.9.

Predicate Devices

Related Devices

Submission Summary (Full Text)

{0} FDA U.S. FOOD & DRUG ADMINISTRATION July 3, 2025 Boston Scientific Yanine Garcia-Quezada Principle Regulatory Affairs Specialist 300 Boston Scientific Way Marlborough, Massachusetts 01752 Re: K251291 Trade/Device Name: Advanix™ Pancreatic Stent and NaviFlex™ Rapid Exchange (RX) Pancreatic Delivery System and Pushers Regulation Number: 21 CFR 876.5010 Regulation Name: Biliary Catheter and Accessories Regulatory Class: Class II Product Code: FGE Dated: April 25, 2025 Received: April 25, 2025 Dear Yanine Garcia-Quezada: We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading. If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register. Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device" U.S. Food & Drug Administration 10903 New Hampshire Avenue Silver Spring, MD 20993 www.fda.gov {1} K251291 - Yanine Garcia-Quezada Page 2 (https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download). Your device is also subject to, among other requirements, the Quality System (QS) regulation (21 CFR Part 820), which includes, but is not limited to, 21 CFR 820.30, Design controls; 21 CFR 820.90, Nonconforming product; and 21 CFR 820.100, Corrective and preventive action. Please note that regardless of whether a change requires premarket review, the QS regulation requires device manufacturers to review and approve changes to device design and production (21 CFR 820.30 and 21 CFR 820.70) and document changes and approvals in the device master record (21 CFR 820.181). Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reporting-combination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050. All medical devices, including Class I and unclassified devices and combination product device constituent parts are required to be in compliance with the final Unique Device Identification System rule ("UDI Rule"). The UDI Rule requires, among other things, that a device bear a unique device identifier (UDI) on its label and package (21 CFR 801.20(a)) unless an exception or alternative applies (21 CFR 801.20(b)) and that the dates on the device label be formatted in accordance with 21 CFR 801.18. The UDI Rule (21 CFR 830.300(a) and 830.320(b)) also requires that certain information be submitted to the Global Unique Device Identification Database (GUDID) (21 CFR Part 830 Subpart E). For additional information on these requirements, please see the UDI System webpage at https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance/unique-device-identification-system-udi-system. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-devices/medical-device-safety/medical-device-reporting-mdr-how-report-medical-device-problems. {2} K251291 - Yanine Garcia-Quezada Page 3 For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100). Sincerely, ANTHONY LEE -S Anthony C. Lee, Ph.D., M.B.A. Assistant Director DHT3A: Division of Renal, Gastrointestinal, Obesity, and Transplant Devices OHT3: Office of Gastrorenal, ObGyn, General Hospital, and Urology Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health Enclosure {3} Advanix™ Pancreatic Stent and NaviFlex™ Rapid Exchange (RX) Pancreatic Delivery System and Page 8 of 37 | Indications for Use | | | | --- | --- | --- | | Please type in the marketing application/submission number, if it is known. This textbox will be left blank for original applications/submissions. | K251291 | ? | | Please provide the device trade name(s). | | ? | | Advanix™ Pancreatic Stent and NaviFlex™ Rapid Exchange (RX) Pancreatic Delivery System and Pushers | | | | Please provide your Indications for Use below. | | ? | | The Advanix™ Pancreatic Stent and NaviFlex™ Rapid Exchange (RX) Pancreatic Delivery System and Pushers are intended for delivery of the stent to the pancreatic duct (PD): • Used to drain pancreatic ducts | | | | Please select the types of uses (select one or both, as applicable). | ☑ Prescription Use (Part 21 CFR 801 Subpart D) ☐ Over-The-Counter Use (21 CFR 801 Subpart C) | ? | {4} | 510(k) #: | 510(k) Summary | Prepared on: 2025-06-26 | | --- | --- | --- | | Contact Details | | 21 CFR 807.92(a)(1) | | Applicant Name | Boston Scientific | | | Applicant Address | 300 Boston Scientific Way Marlborough MA 01752 United States | | | Applicant Contact Telephone | 508-382-0240 | | | Applicant Contact | Yanine Garcia-Quezada | | | Applicant Contact Email | yanine.garcia-quezada@bsci.com | | | Device Name | | 21 CFR 807.92(a)(2) | | Device Trade Name | Advanix™ Pancreatic Stent and NaviFlex™ Rapid Exchange (RX) Pancreatic Delivery System and Pushers | | | Common Name | Biliary catheter and accessories | | | Classification Name | Stents, Drains And Dilators For The Biliary Ducts | | | Regulation Number | 21 CFR 876.5010 | | | Product Code(s) | FGE | | | Legally Marketed Predicate Devices | | 21 CFR 807.92(a)(3) | | Predicate # | Predicate Trade Name (Primary Predicate is listed first) | Product Code | | K223592 | Advanix™ Pancreatic Stent and NaviFlex™ Rapid Exchange (RX) Pancreatic Delivery System and Pusher | FGE | | Device Description Summary | | 21 CFR 807.92(a)(4) | | The Advanix™ Pancreatic Stent and NaviFlex™ RX Pancreatic Delivery System and Pushers is a plastic pancreatic stent designed for the delivery of the stent to the pancreatic duct and used to drain pancreatic ducts. The pancreatic stents are provided in straight or single pigtail shape. The straight shape stents have trailing barbs and/or leading barbs depending on application, a rounded or tapered leading end tip to facilitate access through the papilla, and a rounded trailing end to abut the push catheter portion of the delivery system or stent pusher. The single pigtail stent may or may not have leading end barbs depending on application. Some stents have lateral drainage holes in the pigtails, a tapered or rounded leading end tip, and a rounded trailing end. All stents have either an endoscopic marker, fluoroscopic marker, or both on the trailing or leading end of the stent to assist with depth of placement in the pancreatic duct. The location and presence of an endoscopic or fluoroscopic marker is dependent on the length and diameter size of the stent. Some codes have side port holes in the body of the stent. | | | | Intended Use/Indications for Use | | 21 CFR 807.92(a)(5) | | The Advanix™ Pancreatic Stent and NaviFlex™ Rapid Exchange (RX) Pancreatic Delivery System and Pushers are intended for delivery of the stent to the pancreatic duct (PD): • Used to drain pancreatic ducts | | | | Indications for Use Comparison | | 21 CFR 807.92(a)(5) | | The indications for use of the proposed Advanix™ Pancreatic Stent and NaviFlex™ Rapid Exchange (RX) Pancreatic Delivery System and Pushers device are identical to the predicate device. | | | {5} Technological Comparison 21 CFR 807.92(a)(6) The proposed Advanix™ Pancreatic Stent and NaviFlex™ Rapid Exchange (RX) Pancreatic Delivery System and Pushers is identical to the predicate's design, intended use, fundamental technology, operating principle, manufacturing process and all materials with the exception of the stent material for the 4F and 5F sizes listed in this submission. Non-Clinical and/or Clinical Tests Summary & Conclusions 21 CFR 807.92(b) This 510(k) notification includes testing for the following: - Performance Bench Testing - Stent Corrosion Testing - Stent Tensile Testing - Lumen Patency Testing - Side Drainage Hole Testing - Stent Integrity Testing - Stent Length Testing - Stent Outer Diameter Testing - Pigtail Stent Shape and Diameter Testing - Stent Endoscopic Marker Location Testing - Pusher Deployment Force Testing - Pusher Trackability Force Testing - Biocompatibility Testing: - Cytotoxicity Testing - ISO Guinea Pig Maximization Sensitization Testing - ISO Intracutaneous Reactivity Study in Rabbits - ISO Acute Systemic Toxicity Study in Mice - USP Rabbit Pyrogen Study, Materials Mediated - Genotoxicity: Bacterial Reverse Mutation Study - Genotoxicity: Mouse Lymphoma Assay - Implantation & Systemic Toxicity - Chemical Characterization The results of the testing listed above demonstrate that the proposed Advanix Pancreatic Stent and NaviFlex RX Pancreatic Delivery System and Pushers is considered safe and effective for its intended use. The data also demonstrates its substantially equivalent to the predicate device (K223592).
Innolitics

Panel 1

/
Sort by
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...