PCU · Pancreatic Stent, Covered, Metallic, Removable

Gastroenterology, Urology · 21 CFR 876.5015 · Class 2

Overview

Product CodePCU
Device NamePancreatic Stent, Covered, Metallic, Removable
Regulation21 CFR 876.5015
Device ClassClass 2
Review PanelGastroenterology, Urology
ImplantYes

Identification

A pancreatic drainage stent is a prescription device that consists of a self-expanding, covered, metallic stent, intended for placement to facilitate transmural endoscopic drainage of pancreatic pseudocysts. This stent is intended to be removed upon confirmation of pseudocyst resolution. This device may also include a delivery system.

Classification Rationale

Class II (special controls). The special controls for this device are:

Special Controls

In combination with the general controls of the FD&C Act, the AXIOS Stent and Delivery System is subject to the following special controls:

*Classification.* Class II (special controls). The special controls for this device are:(1) The device and elements of the delivery device that may contact the patient must be demonstrated to be biocompatible. (2) Performance data must demonstrate the sterility of patient-contacting components of the device. (3) Performance data must support the shelf life of the device by demonstrating continued sterility, package integrity, and device functionality over the requested shelf life. (4) Non-clinical testing data must demonstrate that the stent and delivery system perform as intended under anticipated conditions of use. The following performance characteristics must be tested: (i) Deployment testing of the stent and delivery system must be conducted under simulated use conditions. (ii) Removal force testing must be conducted. The removal force testing must demonstrate that the stent can be safely removed, and that the stent will remain in place when subjected to forces encountered during use. (iii) Expansion force testing must be conducted. The expansion force must demonstrate that the forces exerted by the stent will not damage the tissue surrounding the stent. (iv) Compression force testing must be conducted. The compression force must demonstrate that the stent will withstand the forces encountered during use. (v) Dimensional verification testing must be conducted. (vi) Tensile testing of joints and materials must be conducted. The minimum acceptance criteria must be adequate for its intended use. (vii) Fatigue testing must be conducted. Material strength must demonstrate that the stent will withstand forces encountered during use. (viii) Corrosion testing must be conducted. Corrosion resistance must demonstrate that the stent will withstand conditions encountered during use. (5) Non-clinical testing must evaluate the compatibility of the stent in a magnetic resonance (MR) environment. (6) Well-documented clinical experience must demonstrate safe and effective use, and capture any adverse events observed during clinical use. (7) Labeling must include the following: (i) Appropriate instructions, warnings, cautions, limitations, and information related to the safe use of the device, including deployment of the device, maintenance of the drainage lumen, and removal of the device. (ii) A warning that the safety and patency of the stent has not been established beyond the duration of the documented clinical experience. (iii) Specific instructions and the qualifications and clinical training needed for the safe use of the device, including deployment of the device, maintenance of the drainage lumen, and removal of the device. (iv) Information on the patient population for which the device has been demonstrated to be effective. (v) A detailed summary of the clinical experience pertinent to use of the device. (vi) A detailed summary of the device technical parameters. (vii) A detailed summary of the device- and procedure-related complications pertinent to use of the device. (viii) An expiration date/shelf life.

Recent Cleared Devices (11 of 11)

RecordDevice NameApplicantDecision DateDecision
K252648Niti-S SPAXUS StentTaewoong Medical Co., Ltd.Mar 13, 2026SESE
K220112AXIOS Stent and Electrocautery-Enhanced Delivery SystemBoston Scientific CorporationSep 7, 2022SESE
K203132AXIOS Stent and Electrocautery Enhanced Delivery SystemBoston Scientific CorporationApr 15, 2021SESE
K192043AXIOS Stent and Electrocautery Enhanced Delivery SystemBoston Scientific CorporationAug 26, 2019SESE
K181905AXIOS Stent and Delivery System, AXIOS Stent and Electrocautery Enhanced Delivery SystemBoston Scientific CorporationMar 25, 2019SESE
K163272AXIOS Stent and Electrocautery Enhanced Delivery SystemBoston Scientific CorporationFeb 18, 2017SESE
K153088AXIOS Stent with Delivery System, AXIOS Stent with Electro-cautery Enhanced Delivery SystemBoston Scientific CorporationDec 11, 2015SESE
K152572AXIOS Stent and Delivery SystemBoston Scientific CorporationOct 6, 2015SESE
K150692AXIOS Stent with Electrocautery Enhanced Delivery SystemBoston Scientific CorpAug 5, 2015SESE
K140561AXIOS STENT AND DELIVERY SYSTEM (WITH 10MM X 10MM STENT) AXIOS STENT AND DELIVERY SYSTEM (WITH 15MM X 10MM STENT)Xlumena, Inc.Apr 23, 2014SESE
DEN130007AXIOS STENT AND DELIVERY SYSTEMXlumena, Inc.Dec 18, 2013DENG

Top Applicants

Innolitics

Panel 1

/
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...