Citrasate® Liquid Acid Concentrate is indicated for use in patients undergoing extracorporeal bicarbonate hemodialysis for acute and chronic renal failure. Citrasate® Liquid Acid Concentrate is intended to be used as one component in the preparation of dialysate in a 3-stream proportioning hemodialysis machine to a physician's prescription. NaturaLyte® Liquid Acid Concentrate is indicated for use in patients undergoing extracorporeal bicarbonate hemodialysis for acute and chronic renal failure. NaturaLyte® Liquid Acid Concentrate is intended to be used as one component in the preparation of dialysate in a 3-stream proportioning hemodialysis machine to a physician's prescription.
Device Story
Citrasate and NaturaLyte are liquid acid concentrates used as one component in 3-stream proportioning hemodialysis machines. The device is mixed with bicarbonate and water at a nominal ratio of 1:1.72:42.28 to generate dialysate. This dialysate is pumped through a dialyzer to create an osmotic gradient, facilitating solute exchange with patient blood. Used in clinical environments for acute and chronic renal failure treatment. The subject of this notification is a modified 1-gallon high-density polyethylene (HDPE) bottle and cap design, featuring a tamper-evident grooved ring pull tab and updated thread dimensions. The concentrates consist of electrolytes, dextrose, and an organic acid (citric acid for Citrasate; acetic acid for NaturaLyte).
Clinical Evidence
No clinical studies were performed. Evidence consists of bench testing, including stability verification, shipping verification per ASTM D4169-16, biocompatibility testing per ISO 10993-1:2018 (including cytotoxicity, sensitization, irritation, systemic toxicity, material-mediated pyrogenicity, and hemocompatibility), and a human factors assessment.
Technological Characteristics
Single-use, non-sterile liquid acid concentrates. Materials: High-density polyethylene (HDPE) bottles and caps. Components: Electrolytes (chloride salts of Na, Ca, Mg, K), dextrose, and organic acid (citric or acetic). Water quality meets ISO 13959. Proportioning ratio: 1:1.72:42.28 (45X system).
Indications for Use
Indicated for patients undergoing extracorporeal bicarbonate hemodialysis for acute and chronic renal failure.
Regulatory Classification
Identification
A hemodialysis system and accessories is a device that is used as an artificial kidney system for the treatment of patients with renal failure or toxemic conditions and that consists of an extracorporeal blood system, a conventional dialyzer, a dialysate delivery system, and accessories. Blood from a patient flows through the tubing of the extracorporeal blood system and accessories to the blood compartment of the dialyzer, then returns through further tubing of the extracorporeal blood system to the patient. The dialyzer has two compartments that are separated by a semipermeable membrane. While the blood is in the blood compartment, undesirable substances in the blood pass through the semipermeable membrane into the dialysate in the dialysate compartment. The dialysate delivery system controls and monitors the dialysate circulating through the dialysate compartment of the dialyzer.(1) The extracorporeal blood system and accessories consists of tubing, pumps, pressure monitors, air foam or bubble detectors, and alarms to keep blood moving safely from the blood access device and accessories for hemodialysis (§ 876.5540) to the blood compartment of the dialyzer and back to the patient. (2) The conventional dialyzer allows a transfer of water and solutes between the blood and the dialysate through the semipermeable membrane. The semipermeable membrane of the conventional dialyzer has a sufficiently low permeability to water that an ultrafiltration controller is not required to prevent excessive loss of water from the patient's blood. This conventional dialyzer does not include hemodialyzers with the disposable inserts (Kiil type) (§ 876.5830) or dialyzers of high permeability (§ 876.5860). (3) The dialysate delivery system consists of mechanisms that monitor and control the temperature, conductivity, flow rate, and pressure of the dialysate and circulates dialysate through the dialysate compartment of the dialyzer. The dialysate delivery system includes the dialysate concentrate for hemodialysis (liquid or powder) and alarms to indicate abnormal dialysate conditions. This dialysate delivery system does not include the sorbent regenerated dialysate delivery system for hemodialysis (§ 876.5600), the dialysate delivery system of the peritoneal dialysis system and accessories (§ 876.5630), or the controlled dialysate delivery system of the high permeability hemodialysis system § 876.5860). (4) Remote accessories to the hemodialysis system include the unpowered dialysis chair without a scale, the powered dialysis chair without a scale, the dialyzer holder set, dialysis tie gun and ties, and hemodialysis start/stop tray.
Predicate Devices
Citrasate Liquid Acid Concentrate and NaturaLyte Liquid Acid Concentrate (K192017)
Submission Summary (Full Text)
{1}------------------------------------------------
Image /page/0/Picture/0 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). The logo consists of two parts: the Department of Health & Human Services logo on the left and the FDA text logo on the right. The FDA text logo is in blue and consists of the letters "FDA" stacked on top of the words "U.S. FOOD & DRUG ADMINISTRATION".
May 10, 2022
Fresenius Medical Care Renal Therapies Group, LLC Denise Oppermann Senior Director, Regulatory Affairs 920 Winter Street Waltham, MA 02451
Re: K212620
Trade/Device Name: Citrasate® Liquid Acid Concentrate, NaturaLyte® Liquid Acid Concentrate Regulation Number: 21 CFR 876.5820 Regulation Name: Hemodialysis system and accessories Regulatory Class: Class II Product Code: KPO Dated: August 17, 2021 Received: August 18, 2021
Dear Denise Oppermann:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for
{2}------------------------------------------------
devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely.
Gema Gonzalez, MS Acting Assistant Director DHT3A: Division of Renal, Gastrointestinal, Obesity and Transplant Devices OHT3: Office of GastroRenal, ObGyn, General Hospital and Urology Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health
Enclosure
{3}------------------------------------------------
# Indications for Use
510(k) Number (if known) K212620
Device Name Citrasate® Liquid Acid Concentrate NaturaLyte® Liquid Acid Concentrate
### Indications for Use (Describe)
Citrasate® Liquid Acid Concentrate is indicated for use in patients undergoing extracorporeal bicarbonate hemodialysis for acute and chronic renal failure. Citrasate® Liquid Acid Concentrate is intended to be used as one component in the preparation of dialysate in a 3-stream proportioning hemodialysis machine to a physician's prescription.
NaturaLyte® Liquid Acid Concentrate is indicated for use in patients undergoing extracorporeal bicarbonate hemodialysis for acute and chronic renal failure. NaturaLyte® Liquid Acid Concentrate is intended to be used as one component in the preparation of dialysate in a 3-stream proportioning hemodialysis machine to a physician's prescription.
| Type of Use (Select one or both, as applicable) |
|-------------------------------------------------|
|-------------------------------------------------|
X Prescription Use (Part 21 CFR 801 Subpart D)
Over-The-Counter Use (21 CFR 801 Subpart C)
### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{4}------------------------------------------------
Image /page/3/Picture/1 description: The image shows the logo for Fresenius Medical Care. On the left side of the logo is a blue symbol that looks like three chevrons stacked on top of each other. To the right of the symbol is the text "FRESENIUS MEDICAL CARE" in blue, with "FRESENIUS" on top and "MEDICAL CARE" on the bottom.
#### 5. 510(K) SUMMARY
This 510(k) Summary is in accordance with the requirements of the Safe Medical Device Act (SMDA) of 1990. The content of this 510(k) summary is provided in conformance with 21 CFR § 807.92.
#### 5.1. Submitter's Information
| Name: | Fresenius Medical Care Renal Therapies Group, LLC |
|-------------------|---------------------------------------------------|
| Address: | 920 Winter Street<br>Waltham, MA<br>02451-1457 |
| Phone: | (781) 996-9103 |
| Fax: | (781) 699-9635 |
| Contact Person: | Denise Oppermann, Senior Director |
| Preparation Date: | 17 August 2021 |
#### 5.2. Device Name
| Trade Name: | Citrasate® Liquid Acid Concentrate |
|--------------------|----------------------------------------------------------|
| | NaturaLyte® Liquid Acid Concentrate |
| Common Name: | Liquid Acid Concentrates |
| Regulation Name: | Hemodialysis system and accessories |
| Regulatory Class: | Class II per 21 CFR § 876.5820 |
| Product Code: | KPO |
| Product Code Name: | Dialysate Concentrate for Hemodialysis (Liquid or Powder |
| FDA Review Panel: | Gastroenterology/Urology |
#### 5.3. Legally Marketed Predicate Device
The legally marketed predicate devices are the Citrasate Liquid Acid Concentrate and NaturaLyte Liquid Acid Concentrate cleared under K192017.
These devices are not currently subject to a design-related recall.
#### 5.4. Device Description
#### 5.4.1. Device Identification- Citrasate
The Citrasate Liquid Acid Concentrate product line consists of three (3) formulations that differ in concentrations of potassium chloride. Each formulation is offered in two (2) presentations-a 1gallon high density polyethylene bottle and a 55-gallon high density polyethylene drum-for a total of six (6) product codes. Only the bottle presentation is the subject of this 510(k). Descriptions of the available Citrasate Liquid Acid Concentrate bottle formulations are provided in Table 1.
{5}------------------------------------------------
Image /page/4/Picture/1 description: The image shows the Fresenius Medical Care logo. On the left is a blue symbol that consists of three chevrons stacked on top of each other, pointing downwards. To the right of the symbol is the text "FRESENIUS MEDICAL CARE", with "FRESENIUS" on the top line and "MEDICAL CARE" on the bottom line. The text is also in blue.
| Part Number | Part Number Description |
|-------------|--------------------------------------------------------------|
| 08-1251-CA | Citrasate Liquid Acid – 1.0 K, 2.5 Ca, 1.0 Mg – 1-gal bottle |
| 08-2251-CA | Citrasate Liquid Acid – 2.0 K, 2.5 Ca, 1.0 Mg – 1-gal bottle |
| 08-3251-CA | Citrasate Liquid Acid – 3.0 K, 2.5 Ca, 1.0 Mg – 1-gal bottle |
#### Table 1: Citrasate Liquid Acid Concentrate Products (1-Gallon Bottles)
#### 5.4.2. Device Identification- NaturaLyte
The NaturaLyte Liquid Acid Concentrate product line consists of 16 formulations that differ in concentrations of potassium chloride and calcium chloride. All formulations are offered in a 1-gallon high density polyethylene bottle and six (6) formulations are offered in a 55-gallon high density polyethylene drum, for a total of 22 product codes. Only the bottle presentation is the subject of this 510(k). Descriptions of the available NaturaLyte Liquid Acid Concentrate bottle formulations are provided in Table 2.
| Part Number | Part Number Description |
|-------------|----------------------------------------------------------------|
| 08-0231-4 | NaturaLyte Liquid Acid - 0.0 K, 2.25 Ca, 1.0 Mg - 1-gal bottle |
| 08-1001-0 | NaturaLyte Liquid Acid - 1.0 K, 0.0 Ca, 1.0 Mg - 1- gal bottle |
| 08-1201-8 | NaturaLyte Liquid Acid - 1.0 K, 2.0 Ca, 1.0 Mg - 1- gal bottle |
| 08-1231-3 | NaturaLyte Liquid Acid - 1.0 K, 2.25 Ca, 1.0 Mg - 1-gal bottle |
| 08-1251-1 | NaturaLyte Liquid Acid - 1.0 K, 2.5 Ca, 1.0 Mg - 1-gal bottle |
| 08-1301-4 | NaturaLyte Liquid Acid - 1.0 K, 3.0 Ca, 1.0 Mg - 1-gal bottle |
| 08-2201-5 | NaturaLyte Liquid Acid - 2.0 K, 2.0 Ca, 1.0 Mg - 1-gal bottle |
| 08-2231-2 | NaturaLyte Liquid Acid - 2.0 K, 2.25 Ca, 1.0 Mg - 1-gal bottle |
| 08-2251-0 | NaturaLyte Liquid Acid - 2.0 K, 2.5 Ca, 1.0 Mg - 1-gal bottle |
| 08-2301-3 | NaturaLyte Liquid Acid - 2.0 K, 3.0 Ca, 1.0 Mg - 1-gal bottle |
| 08-2351-8 | NaturaLyte Liquid Acid - 2.0 K, 3.5 Ca, 1.0 Mg - 1-gal bottle |
| 08-3201-4 | NaturaLyte Liquid Acid - 3.0 K, 2.0 Ca, 1.0 Mg - 1-gal bottle |
| 08-3231-1 | NaturaLyte Liquid Acid - 3.0 K, 2.25 Ca, 1.0 Mg - 1-gal bottle |
| 08-3251-9 | NaturaLyte Liquid Acid - 3.0 K, 2.5 Ca, 1.0 Mg - 1-gal bottle |
| 08-3301-2 | NaturaLyte Liquid Acid - 3.0 K, 3.0 Ca, 1.0 Mg - 1-gal bottle |
| 08-4231-0 | NaturaLyte Liquid Acid - 4.0 K, 2.25 Ca, 1.0 Mg - 1-gal bottle |
Table 2: NaturaLyte Liquid Acid Concentrate Products (1-Gallon Bottles)
{6}------------------------------------------------
Image /page/5/Picture/1 description: The image shows the logo for Fresenius Medical Care. On the left is a blue symbol that looks like three chevrons stacked on top of each other, pointing downwards. To the right of the symbol is the text "FRESENIUS MEDICAL CARE" in blue, with "FRESENIUS" on top and "MEDICAL CARE" on the bottom.
#### 5.4.3. Device Characteristics
Citrasate Liquid Acid Concentrate and NaturaLyte Liquid Acid Concentrate products are single-use, non-sterile devices intended for use in hemodialysis therapy for acute and chronic renal failure. Both products consist of electrolytes (chloride salts of sodium, calcium, magnesium, and potassium), dextrose, and an organic acid source (citric acid in Citrasate and acetic acid in NaturaLyte). Citrasate also contains sodium acetate as a secondary pH adjuster.
All formulation components are mixed in a single solution with dialysis quality water and provided to the customer ready for use.
#### 5.4.4. Environment of Use
Citrasate Liquid Acid Concentrate and NaturaLyte Liquid Acid Concentrate are used in environments where acute and chronic hemodialysis is performed.
#### 5.4.5. Brief Written Description of the Device
Citrasate Liquid Acid Concentrate and NaturaLyte Liquid Acid Concentrate are each intended to be used as one (1) component in the preparation of dialysate in a 3-stream proportioning hemodialysis machine according to a physician's prescription. Both concentrates are formulated for use in 45X proportioning systems which proportion a nominal ratio of 1 : 1.72 : 42.28 (acid : bicarbonate : water) to generate dialysate. The dialysate is intended to be pumped through a dialyzer, creating an osmotic gradient across the dialyzer membrane to exchange solutes with blood during hemodialysis.
The proposed Citrasate Liquid Acid Concentrate and NaturaLyte Liquid Acid Concentrate 1-gallon bottles have a modified cap design with alternate raw materials. The proposed bottle cap is composed of high density polyethylene (HDPE). The tamper-evident cap has a grooved ring pull tab intended to provide a thread lock. Due to the difference in cap design, the bottle threads are modified to match the threads and dimensions of the proposed cap. Dimensional changes are limited to the threads and bottle termination area only and do not affect the main bottle dimensions. There is no change in wall thickness or material used to manufacture the Citrasate Liquid Acid Concentrate and NaturaLyte Liquid Acid Concentrate 1-gallon bottles.
#### 5.4.6. Materials of Use
Citrasate Liquid Acid Concentrate and NaturaLyte Liquid Acid Concentrate are classified as externally communicating, blood path indirect, prolonged contact (>24 hours to 30 days) duration, Class II (Category B) devices in accordance with FDA guidance document Use of International Standard ISO 10993-1, "Biological evaluation of medical devices-Part 1: Evaluation and testing within a risk management process" (04 September 2020). Packaging for the Citrasate Liquid Acid Concentrate and NaturaLyte Liquid Acid Concentrate bottles consists of the following materials:
- 1-gallon bottle - high density polyethylene
- Bottle cap high density polyethylene ●
{7}------------------------------------------------
Image /page/6/Picture/1 description: The image contains the logo for Fresenius Medical Care. On the left side of the logo is a blue symbol that consists of three downward-pointing chevrons stacked on top of each other. To the right of the symbol is the text "FRESENIUS MEDICAL CARE" in blue, with "FRESENIUS" on the top line and "MEDICAL CARE" on the bottom line.
Citrasate Liquid Acid Concentrate and NaturaLyte Liquid Acid Concentrate are composed of USP grade raw chemicals. Purified water meeting the requirements of ISO 13959 is used in the manufacture of Citrasate Liquid Acid Concentrate and NaturaLyte Liquid Acid Concentrate.
#### 5.4.7. Key Performance Specifications/Characteristics
Citrasate Liquid Acid Concentrate and NaturaLyte Liquid Acid Concentrate are each used as a component of the dialysate for hemodialysis treatments. Both concentrates are intended to be used in 45X proportioning systems which proportion a nominal ratio of 1 : 1.72 : 42.28 (acid : bicarbonate : water) to generate dialysate. The water used for the final dialysate meets ISO 13959 or ANSI/AAMI RD62 requirements.
#### 5.5. Intended Use
Citrasate Liquid Acid Concentrate and NaturaLyte Liquid Acid Concentrate are each intended for use in hemodialysis therapy for acute and chronic renal failure.
#### 5.6. Indications for Use
Citrasate® Liquid Acid Concentrate is indicated for use in patients undergoing extracorporeal bicarbonate hemodialysis for acute and chronic renal failure. Citrasate® Liquid Acid Concentrate is intended to be used as one component in the preparation of dialysate in a 3-stream proportioning hemodialysis machine according to a physician's prescription.
NaturaLyte® Liquid Acid Concentrate is indicated for use in patients undergoing extracorporeal bicarbonate hemodialysis for acute and chronic renal failure. NaturaLyte® Liquid Acid Concentrate is intended to be used as one component in the preparation of dialysate in a 3-stream proportioning hemodialysis machine according to a physician's prescription.
#### 5.7. Comparison of Technological Characteristics with the Predicate Device
The Citrasate Liquid Acid Concentrate and NaturaLyte Liquid Acid Concentrate are substantially equivalent to the predicate devices (K192017) with regard to the following technological characteristics:
- Intended use
- Design
- Principle of operation
- Materials of construction ●
- . Performance specifications
#### 5.8. Performance Data
Performance testing was conducted for the Citrasate Liquid Acid Concentrate and NaturaLyte Liquid Acid Concentrate products. Results of performance testing support substantial equivalence, safety, and efficacy of the Citrasate Liquid Acid Concentrate and NaturaLyte Liquid Acid Concentrate.
{8}------------------------------------------------
Image /page/7/Picture/1 description: The image shows the logo for Fresenius Medical Care. The logo consists of a blue symbol on the left and the text "FRESENIUS MEDICAL CARE" on the right. The symbol is made up of three downward-pointing chevrons, stacked on top of each other. The text is in a bold, sans-serif font, with "FRESENIUS" on the top line and "MEDICAL CARE" on the bottom line.
#### Stability Design Verification 5.8.1.
Stability evaluations were conducted for Citrasate Liquid Acid Concentrate and NaturaLyte Liquid Acid Concentrate to support the 24-month (2-year) shelf life. Stability is monitored as part of routine production testing.
#### 5.8.2. Shipping Verification
Shipping and distribution verification testing was performed for the Citrasate Liquid Acid Concentrate and NaturaLyte Liquid Acid Concentrate products in accordance with ASTM D4169-16, Standard Practice for Performance Testing of Shipping Containers and Systems. Results support that the products' packaging is able to withstand the distribution environment.
#### 5.8.3. Biocompatibility Testing
Biocompatibility testing was conducted in accordance with ISO 10993-1:2018 and FDA guidance document Use of International Standard ISO 10993-1, "Biological evaluation of medical devices – Part 1: Evaluation and testing within a risk management process" (04 September 2020). The following endpoints were assessed to support the biological safety of the Citrasate Liquid Acid Concentrate and NaturaLyte Liquid Acid Concentrate bottles:
- Chemical characterization ●
- Cytotoxicity ●
- Sensitization ●
- Irritation
- Systemic toxicity ●
- Material mediated pyrogenicity ●
- Hemocompatibility ●
A toxicological risk assessment was also performed.
#### Human Factors Validation Testing 5.8.4.
A Human Factors assessment was conducted for the Citrasate Liquid Acid Concentrate and NaturaLyte Liquid Acid Concentrate to demonstrate their safe and effective use in accordance with FDA guidance document Applying Human Factors and Usability Engineering to Medical Devices (03 February 2016).
#### 5.8.5. Electrical Safety and Electromagnetic Compatibility (EMC)
Not applicable. Citrasate Liquid Acid Concentrate and NaturaLyte Liquid Acid Concentrate are not electrical mechanical devices.
{9}------------------------------------------------
Image /page/8/Picture/1 description: The image contains the logo for Fresenius Medical Care. The logo consists of a blue symbol on the left and the text "FRESENIUS MEDICAL CARE" on the right, also in blue. The symbol is made up of three downward-pointing chevrons stacked on top of each other.
#### Software Verification and Validation Testing 5.8.6.
Not applicable. Citrasate Liquid Acid Concentrate and NaturaLyte Liquid Acid Concentrate do not contain software.
#### 5.8.7. Animal Studies
No animal studies were performed for Citrasate Liquid Acid Concentrate or NaturaLyte Liquid Acid Concentrate.
#### 5.8.8. Clinical Studies
No clinical studies were performed for Citrasate Liquid Acid Concentrate or NaturaLyte Liquid Acid Concentrate.
#### 5.9. Conclusion
The intended use, design, principle of operation, and materials of construction of the Citrasate Liquid Acid Concentrate and NaturaLyte Liquid Acid Concentrate are substantially equivalent to those of the predicate devices. Differences between the Citrasate Liquid Acid Concentrate and NaturaLyte Liquid Acid Concentrate and the predicates do not raise new concerns with regard to safety or efficacy. FMCRTG concludes that within the meaning of the Medical Device Amendments Act of 1976, Citrasate Liquid Acid Concentrate and NaturaLyte Liquid Acid Concentrate are safe and effective for their intended use.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.