NXSTAGE PUREFLOW-B SOLUTION

K042045 · Nxstage Medical, Inc. · KPO · Aug 27, 2004 · Gastroenterology, Urology

Device Facts

Record IDK042045
Device NameNXSTAGE PUREFLOW-B SOLUTION
ApplicantNxstage Medical, Inc.
Product CodeKPO · Gastroenterology, Urology
Decision DateAug 27, 2004
DecisionSESE
Submission TypeSpecial
Regulation21 CFR 876.5820
Device ClassClass 2
AttributesTherapeutic

Indications for Use

NxStage PureFlow-B Solution is indicated for use with renal replacement therapy systems that utilize sterile premixed dialysate.

Device Story

NxStage PureFlow-B Solution is a non-pyrogenic dialysis solution provided in single-use flexible bags. It serves as a component for renal replacement therapy systems that require sterile premixed dialysate. The solution is used in clinical or home settings where such therapy is administered. It functions as the dialysate medium to facilitate solute removal and fluid balance during hemodialysis. Healthcare providers or patients (depending on the system) utilize the solution as part of the prescribed dialysis treatment to manage renal failure.

Clinical Evidence

No clinical data. Substantial equivalence is supported by design validation testing.

Technological Characteristics

Non-pyrogenic dialysis solution containing bicarbonate. Provided in single-use flexible bags. Designed for compatibility with renal replacement therapy systems utilizing sterile premixed dialysate.

Indications for Use

Indicated for use with renal replacement therapy systems requiring sterile premixed dialysate. Patient population includes individuals undergoing renal replacement therapy.

Regulatory Classification

Identification

A hemodialysis system and accessories is a device that is used as an artificial kidney system for the treatment of patients with renal failure or toxemic conditions and that consists of an extracorporeal blood system, a conventional dialyzer, a dialysate delivery system, and accessories. Blood from a patient flows through the tubing of the extracorporeal blood system and accessories to the blood compartment of the dialyzer, then returns through further tubing of the extracorporeal blood system to the patient. The dialyzer has two compartments that are separated by a semipermeable membrane. While the blood is in the blood compartment, undesirable substances in the blood pass through the semipermeable membrane into the dialysate in the dialysate compartment. The dialysate delivery system controls and monitors the dialysate circulating through the dialysate compartment of the dialyzer.(1) The extracorporeal blood system and accessories consists of tubing, pumps, pressure monitors, air foam or bubble detectors, and alarms to keep blood moving safely from the blood access device and accessories for hemodialysis (§ 876.5540) to the blood compartment of the dialyzer and back to the patient. (2) The conventional dialyzer allows a transfer of water and solutes between the blood and the dialysate through the semipermeable membrane. The semipermeable membrane of the conventional dialyzer has a sufficiently low permeability to water that an ultrafiltration controller is not required to prevent excessive loss of water from the patient's blood. This conventional dialyzer does not include hemodialyzers with the disposable inserts (Kiil type) (§ 876.5830) or dialyzers of high permeability (§ 876.5860). (3) The dialysate delivery system consists of mechanisms that monitor and control the temperature, conductivity, flow rate, and pressure of the dialysate and circulates dialysate through the dialysate compartment of the dialyzer. The dialysate delivery system includes the dialysate concentrate for hemodialysis (liquid or powder) and alarms to indicate abnormal dialysate conditions. This dialysate delivery system does not include the sorbent regenerated dialysate delivery system for hemodialysis (§ 876.5600), the dialysate delivery system of the peritoneal dialysis system and accessories (§ 876.5630), or the controlled dialysate delivery system of the high permeability hemodialysis system § 876.5860). (4) Remote accessories to the hemodialysis system include the unpowered dialysis chair without a scale, the powered dialysis chair without a scale, the dialyzer holder set, dialysis tie gun and ties, and hemodialysis start/stop tray.

Predicate Devices

Related Devices

Submission Summary (Full Text)

{0}------------------------------------------------ Ko42045 R275/ of. AUG 27 2004 # Section 7 510(K) Summary of Safety Effectiveness This 510(k) summary of safety and effectiveness information is being submitted in accordance with the requirements of the Safe Medical Device Act (SMDA) of 1990. The contents of this 510(k) summary have been provided in conformance with 21 CFR 807.92. | Date: | July 29, 2004 | |-------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------| | Common/Usual Name: | Premixed Dialysate for Hemodialysis | | Trade/Proprietary Name: | NxStage PureFlow-B Solution | | Classification Name: | Hemodialysis systems and accessories<br>(21 CFR 876.5820) | | Device Classification: | Class II | | Product Code: | 78 KPO - Dialysate Concentrate for Hemodialysis<br>(Liquid or Powder) | | Device Panel: | Gastroenterology-Urology (GU)/Gastro-Renal<br>(GRDB) | | 510(k) Sponsor &<br>Owner/Operator: | NxStage Medical, Inc<br>439 South Union St, Suite 501<br>Lawrence, MA 01843<br>Owner/Operator No. 9045797<br>Establishment Registration #3003464075 | | Contact Person: | Norma LeMay<br>Manager, Regulatory Affairs | #### Device Description: The NxStage PureFlow Solution with bicarbonate (NxStage PureFlow-B Solution) is a non-pyrogenic dialysis solution provided in single use flexible bags. The PureFlow-B Solution is intended for use with renal replacement therapy systems that utilize sterile premixed dialysate. Special 510(k) Premarket Notification Device Modification NxStage Medical, Inc. NxStage PureFlow-B Solution {1}------------------------------------------------ ## Substantial Equivalence: This submission is a Special 510(k) Device Modification as described in FDA's guidance document entitled "The New 510(k) Paradigm - Alternate Approaches to Demonstrating Substantial Equivalence in Pre-Market Notifications." In support of this 510(k), NxStage has provided certification of compliance to 21 CFR 820.30 Design Control requirements. Design validation testing has been performed to ensure that the modified device meets design specifications. The modified NxStage PureFlow Solution has been compared to the baseline as cleared in K033386 and found to be substantially equivalent. ### Conclusion: Based on the device indications for use, comparison of descriptive and technological characteristics, and design control certification, the modified NxStage PureFlow Solution has been shown to meet the minimum requirements that are considered acceptable for its intended use. {2}------------------------------------------------ Image /page/2/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo features a stylized depiction of an eagle or bird-like figure with three curved lines forming its body and wings. The logo is surrounded by text that reads "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" in a circular arrangement. The text is in all caps and appears to be in a sans-serif font. Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850 AUG 27 2004 Ms. Norma LeMay Manager, Regulatory Affairs NxStage Medical, Inc. 439 South Union Street 5th Floor LAWRENCE MA 01843 Re: K042045 KU42045 Trade/Device Name: NxStage PureFlow Solution with Bicarbonate (PureFlow-B) Regulation Number: 21 CFR §876.5820 Regulation Name: Hemodialysis system and accessories Regulatory Class: II Product Code: 78 KPO Dated: July 29, 2004 Received: July 30, 2004 Dear Ms. LeMay: We have reviewed your Section 510(k) premarket notification of intent to market the device w o forrowed your and have determined the device is substantially equivalent (for the indications forchenood above and harsure) to legally marketed predicate devices marketed in interstate for associated in the May 28, 1976, the enactment date of the Medical Device Amendments, or to conimores that have been reclassified in accordance with the provisions of the Federal Food, Drug, de Hoos that have been that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The r ou may, dicrerere, mains of the Act include requirements for annual registration, listing of general controls provision practice, labeling, and prohibitions against misbranding and adulteration. If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it your device is easonivals. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may round in further announcements concerning your device in the Federal Register. Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act that I Drines interves and regulations administered by other Federal agencies. You must or any Federal statuter and and ing, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (sections 531-542 of the Act); 21 CFR 1000-1050. {3}------------------------------------------------ This letter will allow you to begin marketing your device as described in your Section 510(k) The same and the same of the first of Substantial equivalence of your device to This letter will allow you to begin marketing your actives of your device of your device to a legally premarket notification The FDA indines of substantial equivalence of y premarket notification - The FDA linding of substantial or your device and thus, permits your device to proceed to the market. If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please s and the many of the many of the fellouing numbers, hased on the regulat If you desire specific advice for your device on our laboling toge. contact the Office of Compliance at one of the following numbers, based on the regulation in number at the top of the letter: | 8xx. Ixxx | (301) 594-4591 | |----------------------------------|----------------| | 876.2xxx, 3xxx, 4xxx, 5xxx | (301) 594-4616 | | 884.2xxx, 3xxx, 4xxx, 5xxx, 6xxx | (301) 594-4616 | | 892.2xxx, 3xxx, 4xxx, 5xxx | (301) 594-4654 | | Other | (301) 594-4692 | Additionally, for questions on the promotion and advectising of your device, please contact the Additionally, for questions on the production and the regulation entitled, "Maintanding Office of Compliance at (301) 594-4639 - Association may obtain - Other oceecal Office of Compliance at (301) 394-4057. Also, posts and obtain. Other general by reference to premarket notification (210, 10, 10, 10, 10, 2011, 10:11 the Division of Small information on your responsibilities under the Act may be obtained from the Div information on your responsibilities and its its tall-free number (800) -638-2041 or Manufacturers, International and Consumer Assistance at its toll-free maman huml Manufacturers, International and Collisumer Assistantoo at le 18.1.2017 (301) 443-6597 or at its Internet address http://www.fda.gov/cdrh/dsma/dsmamain.html. Sincerely yours, Nancy C. Brogdon Nancy C. Brogdon Director, Division of Reproductive, Abdominal and Radiological Devices Office of Device Evaluation Center for Devices and Radiological Health Enclosure {4}------------------------------------------------ # Indications for Use 510(k) Number (if known): NxStage PureFlow-B Solution Device Name: Indications for Use: NxStage PureFlow-B Solution is indicated for use with renal replacement therapy systems that utilize sterile premixed dialysate. Prescription Use _____________________________________________________________________________________________________________________________________________________________ × (Part 21 CFR 801 Subpart D) AND/OR Over-The-Counter Use_ (21 CFR 807 Subpart C) (PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED) Concurrence of CDRH, Office of Device Evaluation (ODE) Daniel R. Larson (Division Sign-Off) Division of Reproductive, A and Radiological Devices 51()(k) Number _ Page 1 of 1 .
Innolitics

Panel 1

/
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...