GAMBRO POLYFLUX 11S, 14S, 17S & 21S HEMODIALYZERS/ HEMOFILTERS

K982414 · Gambro Healthcare · KDI · Mar 26, 1999 · Gastroenterology, Urology

Device Facts

Record IDK982414
Device NameGAMBRO POLYFLUX 11S, 14S, 17S & 21S HEMODIALYZERS/ HEMOFILTERS
ApplicantGambro Healthcare
Product CodeKDI · Gastroenterology, Urology
Decision DateMar 26, 1999
DecisionSESE
Submission TypeTraditional
Regulation21 CFR 876.5860
Device ClassClass 2
AttributesTherapeutic

Indications for Use

The Gambro Polyflux 11S, 14S, 17S, and 21S hemodialyzers / filters can be used whenever hemodialysis or hemofiltration is indicated. These can be used for acute or chronic hemodialysis or hemofiltration. In these therapies, monitoring of patient vital signs, the therapy delivery system, heparin administration and clotting times should be performed under the supervision / direction of a physician.

Device Story

Hemodialyzers/hemofilters for acute/chronic renal failure, intoxications, and fluid overload. Device uses polyarylethersulfone (PES) hollow fibers housed in polycarbonate. Blood enters inlet port, traverses inside of hollow fibers; dialysate flows counter-current in compartment. Solutes/plasma water removed via diffusion and convection across membrane driven by hydrostatic/transmembrane pressure. Used in clinical settings under physician supervision. Output is filtered blood returned to patient; dialysate/filtrate exits via outlet port. Benefits include removal of uremic toxins and waste products.

Clinical Evidence

No clinical data. Bench testing only. In vitro testing evaluated blood/dialysate side priming volumes, flow resistance, ultrafiltration coefficient, and urea/vitamin B12 clearances.

Technological Characteristics

Hollow fiber hemodialyzer/hemofilter. Materials: polyarylethersulfone (PES) membrane, polycarbonate housing, polyurethane potting. Fiber ID: 215 microns; wall thickness: 50 microns. Surface areas: 1.1, 1.4, 1.7, 2.1 m2. Operation: diffusion and convection via hydrostatic/transmembrane pressure. Sterilization method not specified.

Indications for Use

Indicated for patients requiring acute or chronic hemodialysis or hemofiltration, including treatment of renal failure, certain intoxications, and acute fluid overload.

Regulatory Classification

Identification

A high permeability hemodialysis system is a device intended for use as an artificial kidney system for the treatment of patients with renal failure, fluid overload, or toxemic conditions by performing such therapies as hemodialysis, hemofiltration, hemoconcentration, and hemodiafiltration. Using a hemodialyzer with a semipermeable membrane that is more permeable to water than the semipermeable membrane of the conventional hemodialysis system (§ 876.5820), the high permeability hemodialysis system removes toxins or excess fluid from the patient's blood using the principles of convection (via a high ultrafiltration rate) and/or diffusion (via a concentration gradient in dialysate). During treatment, blood is circulated from the patient through the hemodialyzer's blood compartment, while the dialysate solution flows countercurrent through the dialysate compartment. In this process, toxins and/or fluid are transferred across the membrane from the blood to the dialysate compartment. The hemodialysis delivery machine controls and monitors the parameters related to this processing, including the rate at which blood and dialysate are pumped through the system, and the rate at which fluid is removed from the patient. The high permeability hemodialysis system consists of the following devices:(1) The hemodialyzer consists of a semipermeable membrane with an in vitro ultrafiltration coefficient (K uf ) greater than 8 milliliters per hour per conventional millimeter of mercury, as measured with bovine or expired human blood, and is used with either an automated ultrafiltration controller or anther method of ultrafiltration control to prevent fluid imbalance.(2) The hemodialysis delivery machine is similar to the extracorporeal blood system and dialysate delivery system of the hemodialysis system and accessories (§ 876.5820), with the addition of an ultrafiltration controller and mechanisms that monitor and/or control such parameters as fluid balance, dialysate composition, and patient treatment parameters (e.g., blood pressure, hematocrit, urea, etc.). (3) The high permeability hemodialysis system accessories include, but are not limited to, tubing lines and various treatment related monitors (e.g., dialysate pH, blood pressure, hematocrit, and blood recirculation monitors).

Special Controls

*Classification.* Class II. The special controls for this device are FDA's:(1) “Use of International Standard ISO 10993 ‘Biological Evaluation of Medical Device—Part I: Evaluation and Testing,’ ” (2) “Guidance for the Content of 510(k)s for Conventional and High Permeability Hemodialyzers,” (3) “Guidance for Industry and CDRH Reviewers on the Content of Premarket Notifications for Hemodialysis Delivery Systems,” (4) “Guidance for the Content of Premarket Notifications for Water Purification Components and Systems for Hemodialysis,” and (5) “Guidance for Hemodialyzer Reuse Labeling.”

Predicate Devices

Related Devices

Submission Summary (Full Text)

{0}------------------------------------------------ | 6/99 | | K98241<br>Page<br>510K Notification<br>Gambro Polyflux 11S, 14S, 17S and 21S<br>July 7th, 1998 | |-----------------------|------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------| | 510K(k) SUMMARY | | | | SUBMITTER: | Gambro Healthcare<br>1185 Oak Street<br>Lakewood, CO 80215<br>(303) 231-4436 | | | DATE PREPARED: | June 1st, 1998 | | | DEVICE NAME: | Gambro Polyflux 11S, 14S, 17S, & 21S<br>Hemodialyzers/filters | | | CLASSIFICATION NAMES: | High Permeability Hemodialyzer / Hemofilter | | | PREDICATE DEVICE: | Cobe Centrysystem 14 PES Hemodialyzer | | | Device Description: | | | 3/26 # Gambro Polyflux 11S, 14S, 17S, and 21S, Hemodialyzers/Filters The Gambro Polyflux 11S, 14S, & 21S Hemodialyzers/Hemofiters are identical in construction in function to other hemodialyzers/filters currently marketed in the United States. These devices are intended for use in hemodialysis and hemofiltration for the treatment of acute and chronic renal failure and for certain types of intoxications. They may also be used in cases of acute fluid overload for the removal of plasma water. The membrane used in this device is polyarylether sulfone (polyethersulfone) (PES) which is identical to the membrane utilized in the Cobe Centrysystem 14 PES hemodialyzer which has been previously approved for marketing in the United States under a 510K Notification (K95-5592). A copy of this approval letter is included in Part D. of Section XIV of this Notification. Blood enters a blood inlet port where it is distributed to polyarylethersulfone hollow fibers. Each hollow fiber has an inner diameter of approximately 215 microns (wet hollow fiber internal diameter) and a wall thickness of 50 microns. The number of polyarylethersulfone hollow fibers in each hemodialyzer / hemofilter are 8000 for the 11S. 10000 for the 14S, 10000 for the 17S, and 12500 for the 21S. These dialyzers have effective membrane lengths of either 210mm (11S & 14S) or 250 mm (17S & 21S). The effective membrane surface areas are 1.1 m2 for the 11S, 1.4 m2 for the 14S, 1.7 m2 for the 17S, and 2.1 m2 for the 21S. The housing of this hemodialyzer is also made of polycarbonate. The fibers used in this device are identical in design and materials to the previously approved Cobe Centrysystem 14 PES (K95 5592) (please refer to Parts A., B. C. and D. of Section XIV. of this Notification for information on the Cobe Centrysystem 14 PES Hemodialyzer). The patient's blood traverses the inside of the hollow fibers and exits the device via a blood exit port. By means of a hydrostatic pressure or transmembrane pressure which is created by a combination of positive and negative pressures across the polyarylethersulfone membrane, # 000083 {1}------------------------------------------------ Page 2 cf 3 Gambro Polyflux 11S, 14S, 17S and 21S July 7th , 1998 K982414 plasma water along with certain lower molecular weight solutes of plasma water pass through the membrane and into the dialysate or filtrate compartment of the device. Removal of uremic toxins and waste products are removed from the patient's blood in this device by means of both diffusion and convection through polyarylethersulfone membrane and into the counter current flowing dialysis solution during hemodialysis. The dialysate exits the devices via a dialysate outlet port. #### Predicate Devices: The Gambro Polyflux 11S, 14S, 17S, and 21S hemodialyzers/filters are substantially equivalent in construction, design, intended use, function and materials to other high 11S 14C 17S and 215 1 115, 14S, 17S, and 21S hemodialyzers/filter are substantially equivalent in function, design, composition, materials, and operation, to the Cobe Centrysystem 14 PES hemodialyzer (K95 5592) which is currently in commercial distribution in the United States. #### Intended Use: The Gambro Polyflux 11S, 14S, 17S, and 21S hemodialyzers can be used whenever hemodialysis and hemofiltration is indicated. This can be used for acute or chronic hemodialysis and hemofiltration. In these therapies, monitoring of patient vital signs, the therapy delivery system, heparin administration, and clotting times should be performed under the direction of a physician. This indication statement is essentially the same as the indication statement for the predicate device. ## Technological Characteristics: Comparing the proposed device to the predicate device, some similarities and differences are noted in the design employed to accomplish the same intended use. Both the proposed and predicate devices use the same polyarylethersulfone, hollow fiber membrane. Both the proposed and predicate devices use polycerbonate for the housing and header material and polyurethane for the membrane potting material. The predicate device is different from the proposed device in that it uses a different membrane surface areas and has different blood side priming volumes and performance parameters (i.e. clear and " ### Summary of Non-Clinical Tests: In vitro testing was performed on the Gambro Polyflux 11S, 14S, 17S, and 21S hemodialyzers/hemofilter to determine the following: blood side priming volume, dialysate side priming volume, dialysate and blood flow resistance, ultrafitration coefficient, urea, that the proposed douise and vitamin B12, clearances. The results of these tests confirmed that the proposed device is substantially equivalent to the proposed device for these parameters. ## Clinical Test Results: {2}------------------------------------------------ K 982414 Page 3 of 3 510K Notification Gambro Polyflux 11S, 14S, 17S and 21S July 7th , 1998 Clinical testing was not performed # Conclusions: Testing performed on the Gambro Polyflux 11S, 14S, 17S, and 21S indicates that they are safe, effective, and perform as well as the predicate device, when used in accordance the instructions for use. {3}------------------------------------------------ Image /page/3/Picture/1 description: The image contains the text "Public Health Service". The text is written in a bold, sans-serif font. The words are arranged horizontally, with "Public" and "Health" on the first line and "Service" on the second line. Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850 MAR 2 6 1999 Jeffrey R. Shideman, Ph.D. Gambro Healthcare 7307 Glouchester Drive Edina, MN 55435 Re: K982414 Gambro Polyflux 11S, 14S, 17S, and 21S Hemodialyzers/Hemofilters Dated: January 31, 1999 Received: February 3, 1999 Regulatory Class: III 21 CFR 876.5860/Procode: 78 KDI Dear Dr. Shideman: We have reviewed your Section 510(k) notification of intent to market the device referenced above and we have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Medical Detroo Annoneneric Act (Act). You may, therefore, market the device, subject to the general controls roder Pool, Drag, and e octors of the Act include requirements for annual registration, listing of provisions of the Not - 110 g practice, labeling, and prohibitions against misbranding and adulteration. If your device is classified (see above) into either class III (Special Controls) or class III (Premarket Approval), it may ir your do to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 895. A substantially equivalent determination assumes compliance with the Current Good Manufacturing Practice requirements, as set forth in the Quality System Regulation (QS) for the Ourient Ooo manationing the CFR Part 820) and that, through periodic QS inspections, the Food and Medial Donolar of Ochoral royalerify such assumptions. Failure to comply with the GMP regulation may result in Drug Nammondon (r Dri) mill rom) butlish further announcements concerning your device in the Federal regarder. Please note: this response to your premarket notification submission does not affect any obligation you might have under sections 531 through 542 of the Act for devices under the Electronic Product Radiation Control provisions, or other Federal laws or regulations. This letter will allow you to begin marketing your device as described in your 510(k) premarket notification. The FDA This reter will and you to begin marked predicate device results in a classification for your device and thus, permits your device to proceed to the market. If you desire specific advice for your device on our labeling regulation (21 CFR Part 801 and additionally 809.10 for in if you desire tipeonic devices), please contact the Office of Compliance at (301) 594-4613. Additionally, for questions on the promotion and advertising of your device, please contact the Office of Compliance at (301) 594-4639. Also, the promote the regulation entitled, "Misbranding by reference to premarket notification"(21 CFR 807.97). Other please note rogalation on your responsibilities under the Act may be obtained from the Division of Small Manufacturers Assistance at its toll-free number (800) 638-2041 or (301) 443-6597, or at its internet address "http://www.fda.gov/cdrh/dsma/dsmamain.html". Sincerety vours, APT Daniel G. Schultz, M.D. Acting Director, Division of Reproductive, Abdominal, Ear, Nose and Throat, and Radiological Devices Office of Device Evaluation Center for Devices and Radiological Health Enclosure {4}------------------------------------------------ rage of 510 (k) NUMBER (IF KNOWN) : K982414 DEVICE NAME: Gambro Polyflux 11S, 14S, 17S, and 21S Hemodialyzers/ Filters INDICATIONS FOR USE: The Gambro Polyflux 11S, 14S, 17S, and 21S hemodialyzers / filters can be used whenever hemodialysis or hemofiltration is indicated. These can be used for acute or chronic hemodialysis or hemofiltration. In these therapies, monitoring of patient vital signs, the therapy delivery system, heparin administration and clotting times should be performed under the supervision / direction of a physician. (PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED. ) Concurrence of CDRH, Office of Device Evaluation (ODE) Prescription Use OR Over-The-Counter-Use (Per 21 CFR 801.109) (Optional Format 1-2-96) (Division Sign-Off) Division of Reproductive, Abdominal, ENT, and Radiological De 510(k) Number
Innolitics

Panel 1

/
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...