ALTRATOUCH 1000 HEMODIALYSIS DELIVERY SYSTEM
Device Facts
| Record ID | K954987 |
|---|---|
| Device Name | ALTRATOUCH 1000 HEMODIALYSIS DELIVERY SYSTEM |
| Applicant | Althin Medical AB an Affiliate of Baxter Intl |
| Product Code | KDI · Gastroenterology, Urology |
| Decision Date | Aug 8, 1996 |
| Decision | SESE |
| Submission Type | Traditional |
| Regulation | 21 CFR 876.5860 |
| Device Class | Class 2 |
| Attributes | Therapeutic |
Indications for Use
To provide hemodialysis treatment to patients with acute or chronic renal failure.
Device Story
Hemodialysis delivery system; provides dialysate at prescribed temperature/ionic concentration; monitors machine, dialysate, and blood circuit functions. Operator interface via interactive touch screen; treatment parameters displayed on LCD monitor. Used in clinical settings by trained personnel. Device modifies predicate by replacing CRT with LCD, reducing dialysate flow rate by 20%, and adjusting water temperature requirements. Safety systems identical to predicate.
Clinical Evidence
Bench testing only; no clinical data provided.
Technological Characteristics
Hemodialysis delivery system; LCD display; interactive touch screen interface. Components largely identical to predicate. Software controls display color palette. Safety monitoring systems for machine, dialysate, and blood circuits.
Indications for Use
Indicated for patients with acute or chronic renal failure requiring hemodialysis treatment.
Regulatory Classification
Identification
A high permeability hemodialysis system is a device intended for use as an artificial kidney system for the treatment of patients with renal failure, fluid overload, or toxemic conditions by performing such therapies as hemodialysis, hemofiltration, hemoconcentration, and hemodiafiltration. Using a hemodialyzer with a semipermeable membrane that is more permeable to water than the semipermeable membrane of the conventional hemodialysis system (§ 876.5820), the high permeability hemodialysis system removes toxins or excess fluid from the patient's blood using the principles of convection (via a high ultrafiltration rate) and/or diffusion (via a concentration gradient in dialysate). During treatment, blood is circulated from the patient through the hemodialyzer's blood compartment, while the dialysate solution flows countercurrent through the dialysate compartment. In this process, toxins and/or fluid are transferred across the membrane from the blood to the dialysate compartment. The hemodialysis delivery machine controls and monitors the parameters related to this processing, including the rate at which blood and dialysate are pumped through the system, and the rate at which fluid is removed from the patient. The high permeability hemodialysis system consists of the following devices:(1) The hemodialyzer consists of a semipermeable membrane with an in vitro ultrafiltration coefficient (K uf ) greater than 8 milliliters per hour per conventional millimeter of mercury, as measured with bovine or expired human blood, and is used with either an automated ultrafiltration controller or anther method of ultrafiltration control to prevent fluid imbalance.(2) The hemodialysis delivery machine is similar to the extracorporeal blood system and dialysate delivery system of the hemodialysis system and accessories (§ 876.5820), with the addition of an ultrafiltration controller and mechanisms that monitor and/or control such parameters as fluid balance, dialysate composition, and patient treatment parameters (e.g., blood pressure, hematocrit, urea, etc.). (3) The high permeability hemodialysis system accessories include, but are not limited to, tubing lines and various treatment related monitors (e.g., dialysate pH, blood pressure, hematocrit, and blood recirculation monitors).
Special Controls
*Classification.* Class II. The special controls for this device are FDA's:(1) “Use of International Standard ISO 10993 ‘Biological Evaluation of Medical Device—Part I: Evaluation and Testing,’ ” (2) “Guidance for the Content of 510(k)s for Conventional and High Permeability Hemodialyzers,” (3) “Guidance for Industry and CDRH Reviewers on the Content of Premarket Notifications for Hemodialysis Delivery Systems,” (4) “Guidance for the Content of Premarket Notifications for Water Purification Components and Systems for Hemodialysis,” and (5) “Guidance for Hemodialyzer Reuse Labeling.”
Predicate Devices
- Drake Willock® System 1000® Single Patient Hemodialysis Delivery System (K910215)