K983559 · Icu Medical, Inc. · FJK · Dec 24, 1998 · Gastroenterology, Urology
Device Facts
Record ID
K983559
Device Name
BLOOD TUBING ACCESSORY WITH CLAVE CONNECTOR
Applicant
Icu Medical, Inc.
Product Code
FJK · Gastroenterology, Urology
Decision Date
Dec 24, 1998
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 876.5820
Device Class
Class 2
Attributes
Expedited Review
Indications for Use
The 'Blood Tubing Accessory with CLAVE Connector' is intended for use in hemodialysis and other chronic applications where blood sampling is required. Luer-locking adaptors allow secure, in-line placement between the blood tubing set and the catheter or venipuncture device. Use of a CLAVE connector allows the clinician needle-free blood withdrawal.
Device Story
Blood tubing accessory; single-use; sterile; non-pyrogenic. Device functions as in-line adapter between blood tubing sets and catheters/venipuncture devices. Features CLAVE connector; enables needle-free blood withdrawal. Operated by clinicians in hemodialysis or chronic care settings. Design utilizes luer-locking adaptors; straight-line tubing minimizes turbulence. Benefits patient by providing needle-free access; reduces risk of needle-stick injuries; maintains sterile barrier during repeated blood sampling.
Clinical Evidence
Bench testing only. Hemolysis study conducted by independent laboratory showed no significant hemolysis. Microbial challenge study demonstrated maintenance of sterile barrier for 6 days (144 hours) with 24 activations per day, indicating no increased infection risk under simulated clinical conditions.
Indicated for patients undergoing hemodialysis or other chronic applications requiring blood sampling. Intended for clinician use for needle-free blood withdrawal.
Regulatory Classification
Identification
A hemodialysis system and accessories is a device that is used as an artificial kidney system for the treatment of patients with renal failure or toxemic conditions and that consists of an extracorporeal blood system, a conventional dialyzer, a dialysate delivery system, and accessories. Blood from a patient flows through the tubing of the extracorporeal blood system and accessories to the blood compartment of the dialyzer, then returns through further tubing of the extracorporeal blood system to the patient. The dialyzer has two compartments that are separated by a semipermeable membrane. While the blood is in the blood compartment, undesirable substances in the blood pass through the semipermeable membrane into the dialysate in the dialysate compartment. The dialysate delivery system controls and monitors the dialysate circulating through the dialysate compartment of the dialyzer.(1) The extracorporeal blood system and accessories consists of tubing, pumps, pressure monitors, air foam or bubble detectors, and alarms to keep blood moving safely from the blood access device and accessories for hemodialysis (§ 876.5540) to the blood compartment of the dialyzer and back to the patient. (2) The conventional dialyzer allows a transfer of water and solutes between the blood and the dialysate through the semipermeable membrane. The semipermeable membrane of the conventional dialyzer has a sufficiently low permeability to water that an ultrafiltration controller is not required to prevent excessive loss of water from the patient's blood. This conventional dialyzer does not include hemodialyzers with the disposable inserts (Kiil type) (§ 876.5830) or dialyzers of high permeability (§ 876.5860). (3) The dialysate delivery system consists of mechanisms that monitor and control the temperature, conductivity, flow rate, and pressure of the dialysate and circulates dialysate through the dialysate compartment of the dialyzer. The dialysate delivery system includes the dialysate concentrate for hemodialysis (liquid or powder) and alarms to indicate abnormal dialysate conditions. This dialysate delivery system does not include the sorbent regenerated dialysate delivery system for hemodialysis (§ 876.5600), the dialysate delivery system of the peritoneal dialysis system and accessories (§ 876.5630), or the controlled dialysate delivery system of the high permeability hemodialysis system § 876.5860). (4) Remote accessories to the hemodialysis system include the unpowered dialysis chair without a scale, the powered dialysis chair without a scale, the dialyzer holder set, dialysis tie gun and ties, and hemodialysis start/stop tray.
K163025 — DORA Disposable A.V. Fistula Needle Sets · Bain Medical Equipment (Guangzhou) Co., Ltd. · Jul 20, 2017
K970855 — CLAVE CONNECTOR · Icu Medical, Inc. · Jun 24, 1997
K102994 — K-SHIELD ARTERIAL VENOUS FISTULA SET WITH ANTINEEDLE STICK PROTECTOR · Kawasumi Laboratories, Inc. · Nov 3, 2010
K103165 — BLOOD SAMPLING SET · Acacia, Inc. · Nov 10, 2010
K241586 — PIVO Pro Needle-free Blood Collection Device · Becton Dickinson Infusion Therapy Systems, Inc. · Feb 27, 2025
Submission Summary (Full Text)
{0}------------------------------------------------
### ICCI Medical Inc.
Blood Tubing Accessory with CLAVE Connector
R983559
Page 1 of 2
510(k) Notification
Page 39 of 40
DEC 2 4 1998
# 10.0 510(k) SUMMARY OF SAFETY AND EFFECTIVENESS
The following section is included as required by the Safe Medical Device Act (SMDA) of 1990.
Name: Address: ICU Medical, Inc. 951 Calle Amanecer San Clemente, CA 92673
Contact Person: Phone Number: Fax Number:
Sheryl Sáenz (949)366-2183 (949)366-8368
This summary of 510(k) safety and effectiveness information is being submitted in accordance with the requirements of SMDA 1990 and 21 CFR 807.92.
#### Applicant Information:
| Date Prepared: | October 7, 1998 |
|----------------|----------------------------------------------|
| Name: | ICU Medical Inc. |
| Address: | 951 Calle Amanecer<br>San Clemente, CA 92673 |
| Contact Person: | Sheryl Sáenz |
|-----------------|----------------|
| Phone Number: | (949) 366-2183 |
| Fax Number: | (949) 366-8368 |
#### Device Information:
| Trade Name: | Blood Tubing Accessory with CLAVE Connector |
|----------------------|---------------------------------------------|
| Common Name: | Blood Access Accessory |
| Classification Name: | Blood Access Device / Accessories |
#### Equivalent Device:
MEDISYSTEMS ReadySet® Hemodialysis Blood Tubing Set, needle injection sites
#### Device Description:
The blood tubing accessory with CLAVE Connector is a single use, sterile, non-pyrogenic device intended for use as an adjunct to blood tubing sets or venipuncture devices for needle-free withdrawal of blood.
{1}------------------------------------------------
510(k) Notification Page 40 of 40
Blood Tubing Accessory with CLAVE Connector
#### Intended Use:
The 'Blood Tubing Accessory with CLAVE Connector' is intended for use in hemodialysis and other chronic applications where blood sampling is required. Luer-locking adaptors allow secure, in-line placement between the blood tubing set and the catheter or venipuncture device. Use of a CLAVE connector allows the clinician needle-free blood withdrawal.
#### Comparison To Predicate Device:
| Characteristic | ICU MEDICAL'S BLOOD TUBING<br>ACCESSORY WITH CLAVE®<br>CONNECTOR | Injection sites included on MEDISYSTEMS<br>READYSET HEMODIALYSIS BLOOD<br>TUBING SET |
|---------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------------------|
| Product<br>Labeling | Sterile Use, Non-Pyrogenic, Sterile fluid<br>pathway in unopened, undamaged package.<br>Directions for use on labeling. | Sterile Use, Directions for use on labeling |
| Intended Use | Needle-free port to access any vein, or artery. | Needle injection sites on arterial and venous tubing<br>sides for blood aspiration from tubing. |
| Design | One-piece connector activated by luer<br>connection to allow blood withdrawal. Tubing<br>and connectors are straight-line to minimize<br>turbulence. | Pierceable septum secured in durable housing<br>which is bonded to blood tubing set. |
| Materials | Y-Connector-Polyester<br>Male Luer: Polypropylene<br>Female Luer: Polyester<br>Blood Tubing Extension: Polyvinylchloride<br><br>CLAVE Connector: Polycarbonate,<br>Polyester, Silicone, Fluorosilicone,<br>Polypropylene, Polyethylene<br><br>Packaging - Medical packaging grade fiber-<br>free peelable paper lidding and pouching<br>material. | Septum -Thermoplastic elastomer<br>Housing - Polycarbonate<br>Blood Tubing - Polyvinylchloride |
Conclusions of nonclinical tests:
- . Hemolysis study was conducted by independent laboratory and found that no significant hemolysis occurs with the use of the CLAVE Connector.
- . Microbial challenge study showed that the CLAVE Connector maintained a sterile barrier for six days or 144 hours while administering 24 repeat activations per day. These results indicate that the CLAVE Connector, when using a standard disinfection protocol, did not increase infection rates under worst case clinical simulation
simulation.
Sheryl Sáenz
Sheryl Sáenz
Regulatory Affairs Specialist ICU Medical, Inc.
{2}------------------------------------------------
Image /page/2/Picture/0 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a circular seal with the words "DEPARTMENT OF HEALTH & HUMAN SERVICES USA" around the perimeter. Inside the circle is an abstract symbol that resembles an eagle or bird-like figure.
Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850
Re: K983559
Blood Tubing Accessory with CLAVE® Connector Dated: October 7, 1998 Received: October 13, 1998 Regulatory Class: II 21 CFR 876.5540/Procode: 78 FJK
Dear Ms. Saenz:
Ms. Sheryl Saenz
ICU Medical, Inc.
951 Calle Amanecer
Regulatory Affairs Specialist
San Clemente, CA 92673
DEC 24 1998
We have reviewed your Section 510(k) notification of intent to market the device referenced above and we have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into either class III (Premarket Approval), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 895. A substantially equivalent determination assumes compliance with the Current Good Manufacturing Practice requirements, as set forth in the Quality System Regulation (OS) for Medical Devices: General regulation (21 CFR Part 820) and that, through periodic QS inspections, the Food and Drug Administration (FDA) will verify such assumptions. Failure to comply with the GMP regulation may result in regulatory action. In addition, FDA may publish further announcements concerning your device in the Federal Register. Please note: this response to your premarket notification submission does not affect any obligation you might have under sections 542 of the Act for devices under the Electronic Product Radiation Control provisions, or other Federal laws or regulations.
This letter will allow you to begin marketing your device as described in your 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801 and additionally 809.10 for in vitro diagnostic devices), please contact the Office of Compliance at (301) 594-4613. Additionally, for questions on the promotion and advertising of your device, please contact the Office of Compliance at (301) 594-4639. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). Other general information on your responsibilities under the Act may be obtained from the Division of Small Manufacturers Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its Internet address "http://www.fda.gov/odrb/dsmaldsmamain.html".
Sincerely your
Lillian Yin, Ph.D.
Director, Division of Reproductive, Abdominal, Ear, Nose and Throat and Radiological Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{3}------------------------------------------------
# Device Name: Blood Tubing Accessory with CLAVE Connector
# INDICATIONS FOR USE
The 'Blood Tubing Accessory with CLAVE Connector' is intended for use in hemodialysis and other chronic applications where blood sampling is required. Luer-locking adaptors allow secure, in-line placement between the blood tubing set and the catheter or venipuncture device. Use of a CLAVE Connector allows the clinician needle-free blood withdrawal.
## (PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
| Prescription Use<br>(Per 21 CFR 801.109) | OR | Over-the-Counter-Use |
|------------------------------------------|----|----------------------|
|------------------------------------------|----|----------------------|
:
| (Division Sign-Off) | |
|-----------------------------------------------------------------------|---------|
| Division of Reproductive, Abdominal, ENT,<br>and Radiological Devices | |
| 510(k) Number | K983559 |
. 1
: 上 i
Concurrence of CDRH, Office of Device Evaluation (ODE)
Panel 1
/
Ready
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.