SERIM BLOOD LEAK TEST STRIP

K990206 · Serim Research Corp. · FJD · Apr 21, 1999 · Gastroenterology, Urology

Device Facts

Record IDK990206
Device NameSERIM BLOOD LEAK TEST STRIP
ApplicantSerim Research Corp.
Product CodeFJD · Gastroenterology, Urology
Decision DateApr 21, 1999
DecisionSESE
Submission TypeTraditional
Regulation21 CFR 876.5820
Device ClassClass 2

Indications for Use

SERIM Blood Leak Test Strips provide a rapid and convenient means for testing spent dialysate for blood in hemodialysis clinics. The presence of blood in dialysate indicates a leak in the dialysis membrane.

Device Story

Device consists of reagent pad on polystyrene handle; used to detect occult blood in hemodialysis fluids. Operator immerses pad in dialysate sample; reacts for 60 seconds; compares color to bottle label blocks (Negative/Positive). Principle: peroxidase-like activity of hemoglobin catalyzes oxidation of chromogen to green color. Positive result indicates hemoglobin concentration ≥1.5 mg/dL (approx. 0.1 mL/min blood loss). Used by dialysis technicians to confirm photometric sensor alarms, reducing unnecessary session interruptions. Benefits: quick, convenient confirmation of membrane leaks; prevents premature termination of dialysis when false alarms occur.

Clinical Evidence

Bench testing and clinical comparison performed. 12 participants tested blood standards in blind studies (n=100 negative, n=100 positive) achieving 100% sensitivity and 100% specificity. Comparison study with 101 clinical dialysate samples showed equivalent performance to predicate Chemistrip occult blood strips; 33 spiked samples (1.5 mg/dL hemoglobin) correctly identified as positive by both methods.

Technological Characteristics

Reagent pad containing organic peroxide and chromogen attached to polystyrene handle. Sensing principle: peroxidase-like catalytic oxidation of chromogen by hemoglobin. Dimensions: 0.2 x 0.2 inch pad on 0.2 x 3.25 inch handle. Visual readout via colorimetric comparison. Non-electronic, manual test strip.

Indications for Use

Indicated for use in hemodialysis clinics to test spent dialysate for the presence of blood, which indicates a leak in the dialysis membrane.

Regulatory Classification

Identification

A hemodialysis system and accessories is a device that is used as an artificial kidney system for the treatment of patients with renal failure or toxemic conditions and that consists of an extracorporeal blood system, a conventional dialyzer, a dialysate delivery system, and accessories. Blood from a patient flows through the tubing of the extracorporeal blood system and accessories to the blood compartment of the dialyzer, then returns through further tubing of the extracorporeal blood system to the patient. The dialyzer has two compartments that are separated by a semipermeable membrane. While the blood is in the blood compartment, undesirable substances in the blood pass through the semipermeable membrane into the dialysate in the dialysate compartment. The dialysate delivery system controls and monitors the dialysate circulating through the dialysate compartment of the dialyzer.(1) The extracorporeal blood system and accessories consists of tubing, pumps, pressure monitors, air foam or bubble detectors, and alarms to keep blood moving safely from the blood access device and accessories for hemodialysis (§ 876.5540) to the blood compartment of the dialyzer and back to the patient. (2) The conventional dialyzer allows a transfer of water and solutes between the blood and the dialysate through the semipermeable membrane. The semipermeable membrane of the conventional dialyzer has a sufficiently low permeability to water that an ultrafiltration controller is not required to prevent excessive loss of water from the patient's blood. This conventional dialyzer does not include hemodialyzers with the disposable inserts (Kiil type) (§ 876.5830) or dialyzers of high permeability (§ 876.5860). (3) The dialysate delivery system consists of mechanisms that monitor and control the temperature, conductivity, flow rate, and pressure of the dialysate and circulates dialysate through the dialysate compartment of the dialyzer. The dialysate delivery system includes the dialysate concentrate for hemodialysis (liquid or powder) and alarms to indicate abnormal dialysate conditions. This dialysate delivery system does not include the sorbent regenerated dialysate delivery system for hemodialysis (§ 876.5600), the dialysate delivery system of the peritoneal dialysis system and accessories (§ 876.5630), or the controlled dialysate delivery system of the high permeability hemodialysis system § 876.5860). (4) Remote accessories to the hemodialysis system include the unpowered dialysis chair without a scale, the powered dialysis chair without a scale, the dialyzer holder set, dialysis tie gun and ties, and hemodialysis start/stop tray.

Predicate Devices

Reference Devices

Related Devices

Submission Summary (Full Text)

{0}------------------------------------------------ 4/2i/99 Image /page/0/Picture/1 description: The image shows the text "K990206 Page 1 of 3". The text appears to be handwritten in black ink on a white background. The text is likely part of a document or report, possibly indicating a page number or identification code. Image /page/0/Picture/2 description: The image shows the logo for Serim Research Corp. The logo features the word "Serim" in a stylized, bold font, with the letters connected. Above the "im" in Serim is a shape that looks like a flame or a stylized drop. Below the word "Serim" is the text "RESEARCH CORP" in a smaller, less stylized font. P.O. Box 4002, Elkhart, IN 46514-0002 • (219) 264-3440 • FAX (219) 266-6222 Image /page/0/Picture/4 description: The image shows a document with the text "FDA/CDRH/ODE/DMG" at the top right corner. Below this, the date "Jan 21 4 23 PM '99" is printed diagonally. The word "RECEIVED" is stamped vertically on the left side of the image, indicating the document's processing status. # 510(k) SUMMARY Serina [M Blood Leat. Test Strips ### Submitted by: Robert J. Carrico Serim Research Corporation P.O. Box 4002 Elkhart IN 46514 Phone: (219) 264-3440 Fax: (219) 266-6222 Contact person: Robert J. Carrico Date prepared: January 19, 1999 Robert J. Zinnick No 9. 300 #### Device Name: - SERIM™ Blood Leak Test Strips Trade name: Common name: Occult blood test strip Classitication name: Dialysis Blood Leak Test #### Legally Marketed Equivalent Device: This product is similar in design, composition and function to the occult blood test pad on Chemistrip 10 urine test strip manufactured by Boehringer Mannheim Corporation which was the subject of Premarket Notification #K896454 The product is similar in utility to the cyanomethemoglobin assay. Supportive data for substantial equivalence to the latter was obtained. # Description of the SERIMIN Blood Leak Test Strip Sk.5-8 {1}------------------------------------------------ K990206 Page 2 of 3 2 of 36 Page Serim Research Corporation 510(k) Summary - SERIM Blood Leak Test Strip The SERIM Blood Leak Test Strip consists of a 0.2 x 0.2 inch reagent pad attached to one end of a white 0.2 x 3.25 inch polystyrene handle. The reagent pad is immersed into a sample, removed immediately and allowed to react for 60 seconds. Then the reagent pad color is compared to two color blocks on the bottle label. One color block marked Negative is yellow with a few green speckles. If a reagent pad has a similar vellow color the dialysate does not contain significant blood. The second color block is green with small yellow speckles and is marked Positive. A reagent pad with green color equal to or darker than this color block indicates the presence of a significant blood leak. The Association for the Advancement of Medical Instrumentation(AAMI) standards recommend that blood loss be limited to 0.35 mL/minute or less. This calculates to 5.5 mg/dL hemoglobin at 25% hematocrit. (AAMI Standards and Recommended Practices, Vol. 3, 1998, p. 57) SERIM Blood Leak Test Strips are designed to give a positive readings with 1.5 mg/dL hemoglobin in acid/bicarbonate buffer, pH 7.4. Thus, a positive test strip reading corresponds to approximately 0.1 mL/minute blood loss. ## Intended Use: Hemodialyzers are equipped with photometric sensors in the dialysate stream to detect blood leaks at the dialysis membranes. The sensors measure red color in the dialysate due to hemoglobin in red blood cells. Sensors are calibrated to sound an alarm when blood loss reaches 0.35 mL/minute. A confirmed blood leak requires termination of the dialysis session and restarting the dialysis with a new dialyzer. Occasionally false alarms occur possibly due to gas bubbles in the sensor. Since interrupting a dialysis session is stressful for the patient, technicians prefer to confirm an alarm before taking further steps. SERIM Blood Leak Test Strips provide a quick and convenient means to test for occult blood in hemodialysis fluids. ## Technological Comparison to Predicate Device: SERIM Blood Leak Test Strips and Chemstrip occult blood test both utilize the peroxidase-like activity of hemoglobin to detect whole blood in fluids. Both tests have an organic peroxide and a chromogen dried in test pads. When the pads are exposed to test samples the organic peroxide and chromogen react extremely slowly unless hemoglobin from red blood cells is present. Hemoglobin catalyzes the oxidation of the chromogen to produce green color. The rate of color formation is dependent on the concentration of hemoglobin in the test sample. In some applications the level of occult blood in samples is estimated from the amount of color produced. However, a positive/negative answer is preferable in the hemodialysis application. SERIM Blood Leak Test Strips have adequate sensitivity to detect hemoglobin at the required levels as discussed above. Two color blocks used in the test provide a clear distinction between Positive and Negative readings. If a reading between the two blocks is obtained a very small leak may be {2}------------------------------------------------ K990206 Payg 3 of 3 Serim Research Corporation 510(k) Summary - SERIM Blood Leak Test Strip 3 of 36 Page present and dialysis can be continued because the leak can seal. ## Statement of Substantial Equivalence SERIM Blood Leak Test Strips from two production lots were used by 12 participants to test blood standards in blind studies. One hundred negative strip readings (100% specificity) were obtained with dialysis buffer without blood. One hundred positive strip readings (100% sensitivity) were obtained with dialysis buffer containing non-hemolyzed blood at 1.5 mg/dL hemoglobin, determined by the cyanmethemoglobin method. SERIM Blood Leak Test Strips and Chemstrip occult blood strips (the test for which equivalence is claimed) were used to test 101 dialysates collected at a dialysis clinic. Negative readings for blood leaks were obtained with both tests. Non-hemolyzed blood was added to thirty-three of the dialysate samples to give 1.5 mg/dL hemoglobin. The spiked samples all gave positive blood leak readings with both types of strips. {3}------------------------------------------------ Image /page/3/Picture/1 description: The image is a black and white seal for the Department of Health & Human Services - USA. The seal is circular with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" around the perimeter. In the center of the seal is a stylized image of an eagle. Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850 APR 2 1 1999 Robert J. Carrico, Ph.D. Staff Scientist Serim Research Corporation P.O. Box 4002 Elkhart, IN 46514-0002 Re: K990206 Serim™ Blood Leak Test Strips Dated: January 19, 1999 Received: January 21, 1999 Regulatory Class: II 21 CFR 876.5820/Procode: 78 FJD Dear Dr. Carrico: We have reviewed vour Section 510(k) notification of intent to market the device referenced above and we have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. If your device is classified (see above) into either class II (Special Controls) or class III (Premarket Approval), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 895. A substantially equivalent determination assumes compliance with the Current Good Manufacturing Practice requirements, as set forth in the Quality System Regulation (QS) for Medical Devices: General regulation (21 CFR Part 820) and that, through periodic QS inspections, the Food and Drug Administration (FDA) will verify such assumptions. Failure to comply with the GMP regulation may result in regulatory action. In addition, FDA may publish further announcements concerning your device in the Federal Register. Please note: this response to your premarket notification submission does not affect any obligation you might have under sections 531 through 542 of the Act for devices under the Electronic Product Radiation Control provisions, or other Federal laws or regulations. This letter will allow you to begin marketing your device as described in your 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market. If you desire specific advice for your device on our labeling regulation (21 CFR Part 801 and additionally 809.10 for in vitro diagnostic devices), please contact the Office of Compliance at (301) 594-4613. Additionally, for questions on the promotion and advertising of your device, please contact the Office of Compliance at (301) 594-4639. Also, please note the requlation entitled. "Misbranding by reference to premarket notification"(21 CFR 807.97). Other general information on your responsibilities under the Act may be obtained from the Division of Small Manufacturers Assistance at its toll-free number (800) 638-2041 or (301) 443-6597, or at its internet address "http://www.fda.gov/cdrh/dsma/dsmamain.html". Sincerely yours, CAPT Daniel G. Schultz, M.D. Acting Director, Division of Reproductive, Abdominal, Ear, Nose and Throat, and Radiological Devices Office of Device Evaluation Center for Devices and Radiological Health Enclosure {4}------------------------------------------------ 510(k) Number (if known): K 99030(2 Oevice Name:_Serim_Blood_Leak_Test_Strips Indications For Use: SERIM Blood Leak Test Strips provide a rapid and convenient means for testing spent dialysate for blood in hemodialysis The presence of blood in dialysate indicates a clinics. leak in the dialysis membrane. (PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED) Concurrence of CDRH, Office of Device Evaluation (ODE) Prescription Use V (Per 21 CFR 801.109) ેમ Over-The-Counter Use__________________________________________________________________________________________________________________________________________________________ (Optional Format 1-2-96) Simid h. Siggm (Division Sign-Off) Division of Reproductive, and Radiological 510(k) Number
Innolitics

Panel 1

/
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...