MILLENIUM HX PORTABLE REVERSE OSMOSIS WATER PURIFICATION SYSTEM
K110578 · Mar Cor Purification, Inc. · FIP · May 20, 2011 · Gastroenterology, Urology
Device Facts
Record ID
K110578
Device Name
MILLENIUM HX PORTABLE REVERSE OSMOSIS WATER PURIFICATION SYSTEM
Applicant
Mar Cor Purification, Inc.
Product Code
FIP · Gastroenterology, Urology
Decision Date
May 20, 2011
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 876.5665
Device Class
Class 2
Indications for Use
The Mar Cor Purification Millenium HX Portable Reverse Osmosis Water Purification System is intended to be used as a dialysis accessory to produce water through reverse osmosis for use with hemodialysis equipment. The Millenium HX can be connected to hemodialysis equipment used in hospitals, clinics and in home environments, in conjunction with the appropriate pre and post treatment units, as a part of a water treatment system designed to meet current AAMI and Federal (U.S) standards.
Device Story
Portable reverse osmosis (RO) water purification system; produces purified water for hemodialysis equipment. Feed water enters unit; pump applies high hydrostatic pressure; RO membrane removes contaminants (excluding dissolved gases). System includes RO membrane, pump, water quality monitoring, programmable logic controller (PLC), and touch-screen interface. Features heat disinfection and chemical sanitization to maintain low microbiological levels. Used in hospitals, clinics, and home environments; operated by healthcare providers or patients. Provides audible/visual alarms; automatically diverts to drain if product water total dissolved solids (TDS) exceed set-point. Ensures water meets AAMI quality requirements for hemodialysis; requires pre/post-treatment (e.g., carbon adsorption) for chlorine/chloramine removal.
Clinical Evidence
Bench testing only. Performance data included system/RO membrane flow and water quality verification, heat disinfection process validation, chemical sanitization validation, material compatibility/biocompatibility, software validation, and electrical safety/EMC testing. Testing conducted per FDA guidance for water purification components and ANSI/AAMI RD62:2006.
Technological Characteristics
Portable RO system; utilizes high-pressure pump and RO membrane for purification. Includes programmable logic controller (PLC) with touch-screen interface. Features automated heat disinfection and chemical sanitization. Connectivity includes water quality monitoring with alarm outputs. Operates at minimum 0.26 gal/min (1.0 L/min).
Indications for Use
Indicated for use as a dialysis accessory to produce purified water via reverse osmosis for hemodialysis equipment in hospital, clinic, and home environments. Requires use with appropriate pre- and post-treatment units to meet AAMI/Federal standards.
Regulatory Classification
Identification
A water purification system for hemodialysis is a device that is intended for use with a hemodialysis system and that is intended to remove organic and inorganic substances and microbial contaminants from water used to dilute dialysate concentrate to form dialysate. This generic type of device may include a water softener, sediment filter, carbon filter, and water distillation system.
Special Controls
*Classification.* Class II (special controls). The device, when it is a water purification subsystem disinfectant, is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 876.9.
Mar Cor Millenium Portable RO System 750 (K964539)
Related Devices
K192398 — Millenium HX · Mar Cor Purification, Inc. · Oct 3, 2019
K200670 — EON Portable RO Water Purification System · Mar Cor Purification, Inc. · Apr 16, 2020
K171099 — EON Portable Reverse Osmosis Water Purification System · Cantel · Jan 5, 2018
K111740 — AMERIWATER MRO PORTABLE REVERSE OSMOSIS SYSTEM - SINGLE PATIENT, 1 TO 3 STATION · Ameriwater · Oct 5, 2011
K032004 — AQUABOSS ECORO DIA 50 · Lauer Ltd., Hong Kong · Oct 2, 2003
Submission Summary (Full Text)
{0}------------------------------------------------
K110578
Image /page/0/Picture/1 description: The image shows the logo for Mar Cor Purification, a Cantel Medical Company. The logo consists of a circular graphic on the left, with several curved lines inside. To the right of the graphic is the text "MAR COR" in large, bold letters, with "PURIFICATION" underneath in a smaller font. Below that is the text "A Cantel Medical Company" in an even smaller font.
MAY 2 0 2011
Premarket Notification 510(k) Millenium HX 510(k) Summary OF 3
# Millenium HX Portable Reverse Osmosis Water Purification System
## 510(k) Summary of Safety and Effectiveness
Manufacturer:
Mar Cor Purification, A Cantel Medical Company
Address:
14550 28th Avenue North Minneapolis, MN 55447 USA (800) 633-3080
Official Contact: Brent Geiger, MS, RAC Senior RA Specialist
Trade Name: Millenium HX Common Name: Water Purification System Classification Name: Subsystem, water purification Product Code: FIP Device Class: II Classification Reg: 876.5665
Mar Cor Purification has supplied the following information to the US Food and Drug Administration to support substantial equivalence of the Millenium HX Portable Reverse Osmosis (RO) Water Purification System to other portable RO water purification systems currently cleared for sale in the U.S.
## 1. Device Description
The device is a portable water purification system that uses reverse osmosis to remove contaminants from water that is used to dilute dialysis concentrate to form dialysate for use in hemodialysis equipment. Feed water enters the unit and is directed through a pump into a RO membrane. The pump applies a high hydrostatic pressure that forces water from the concentrated (feed) side to the dilute (product) side of the RO membrane. As water flow across the membrane all types of water contaminants except dissolved gasses are removed and the purified product water is then supplied to hemodialysis equipment.
The Millenium HX is capable of generating purified water that meets AAMI water quality requirements for hemodialysis at a minimum of 0.26 US gallons/minute (1.0 liters/min). It must be used with appropriate pre and post treatment units, including at a minimum carbon adsorption media pretreatment in order to remove chlorine/chloramines. Additional pre and post treatment requirements may vary and are dependent on the quality of the local feed water supply and individual facility requirements.
> Mar Cor Purification Attachment 4 - Page 1 of 3
{1}------------------------------------------------
Image /page/1/Picture/0 description: The image shows the logo for Mar Cor Purification, a Cantel Medical Company. The logo consists of a circular graphic on the left, with a series of horizontal lines inside. To the right of the graphic is the text "MAR COR" in large, bold letters, with "PURIFICATION" underneath in a smaller font. Below that, it says "A Cantel Medical Company" in a smaller font.
K110578
Premarket Notification 510(k) Millenium HX 510(k) Summary ನ್ನ
The Millenium HX system is designed to maintain low microbiological levels in the flow pathway through regular heat disinfection and chemical sanitization. Notable components and features of the Millenium HX include:
- RO membrane .
- System pump .
- Water quality monitoring system .
- Operating panel and programmable logic controller (OPLC) .
- Heat disinfection and chemical sanitization capability .
- Audible and visual alarms .
- Automatic divert to drain mode upon start-up and anytime product water TDS is . above the quality set-point
- System control via a touch-screen user interface .
# 2. Intended Use
The Mar Cor Purification Millenium HX Portable Reverse Osmosis Water Purification System is intended to be used as a dialysis accessory to produce water through reverse osmosis for use with hemodialysis equipment.
The Millenium HX can be connected to hemodialysis equipment used in hospitals, clinics and in home environments, in conjunction with the appropriate pre and post treatment units, as a part of a water treatment system designed to meet current AAMI and Federal (U.S) standards.
# 3. Comparison to Other Devices in Commercial Distribution Within the United States
The Millenium HX is equivalent in function and indications to the Mar Cor (formerly Gambro) WRO 300H (K093608), the Mar Cor Semper Pure RO system (K003877) and the Mar Cor Millenium Portable RO System 750 (K964539). All of the products are portable reverse osmosis water purification systems with the same intended use and equivalent indications for use. Both the Millenium HX and the WRO 300H share the same capability for system heat disinfection.
# 4. Summary of Non-Clinical Performance Data
Mar Cor Purification has provided testing to show that the Millenium HX is safe and effective for its intended use based on the requirements listed in FDA's Guidance for the Content of Premarket Notifications for Water Purification Components and Systems for Hemodialysis (May 1997) and the FDA recognized consensus standard ANSI/AAMI RD62:2006 Water Treatment Equipment for Hemodialysis Applications. The following types of data were provided to FDA to support substantial equivalence to predicate devices and to demonstrate that the Millenium HX performs as intended.
- System and RO Membrane Performance Flow and product water quality verification . over range of operating conditions
- Heat Disinfection Process Validation .
- . Chemical Sanitization Validation
- . Material Compatibility and Biocompatibility
Mar Cor Purification Attachment 4 - Page 2 of 3
{2}------------------------------------------------
K110578
Image /page/2/Picture/1 description: The image shows the logo for Mar Cor Purification, a Cantel Medical Company. The logo consists of a circular graphic on the left, with several horizontal lines inside. To the right of the graphic is the text "MAR COR" in large, bold letters, followed by "PURIFICATION" in smaller letters. Below that is the text "A Cantel Medical Company".
Premarket Notification 510(k) Millenium HX 510(k) Summary
PAGE 3 OF 3
- Chemical Sanitization and Cleaning Rinsing Verification .
- Software Validation .
- Electrical Safety and Electromagnetic Compatibility ●
- Risk Analysis .
### 5. Conclusion
Mar Cor Purification has provided appropriate premarket notification information in the form of a 510(k) to support the substantial equivalence of the Millenium HX to legally marketed predicate devices. The information and performance data provided indicates that the Millenium HX is safe and effective for its intended use when used in accordance with the device labeling.
{3}------------------------------------------------
Image /page/3/Picture/1 description: The image shows the logo for the Department of Health & Human Services USA. The logo is a circular seal with the words "DEPARTMENT OF HEALTH & HUMAN SERVICES USA" around the perimeter. Inside the circle is an image of an eagle with its wings spread. The eagle is facing to the left.
Food and Drug Administration 10903 New Hampshire Avenue Document Mail Center - WO66-G609 Silver Spring, MD 20993-0002
MAR COR® Purification, Inc. % Brent Geiger, M.S., RAC Senior RA Specialist == Minntech Corporation 14605 28" Avenue North MINNEAPOLIS MN 55447
MAY 2 0 2011
Re: K110578
Trade/Device Name: Millenium HX Regulation Number: 21 CFR §876.5665 Regulation Name: Water purification system for hemodialysis Regulatory Class: II Product Code: FIP Dated: February 28, 2011 Received: March 2, 2011
Dear Mr. Geiger:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related
{4}------------------------------------------------
#### Page 2
adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable; the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please go to http://www.fda.gov/AboutFDA/CentersOffices/CDRH/CDRHOffices/ucm115809.htm for the Center for Devices and Radiological Health's (CDRH's) Office of Compliance. Also, please note the regulation entitled. "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm.
Sincerely yours,
Herbert Lemmer MD
Herbert P. Lerner, M.D., Director (Acting) Division of Reproductive. Gastro-Renal and Urological Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{5}------------------------------------------------
Image /page/5/Picture/0 description: The image shows the logo for Mar Cor Purification, a Cantel Medical Company. The logo consists of a circular graphic on the left, with a series of horizontal lines inside. To the right of the graphic is the text "MAR COR" in large, bold letters, with "PURIFICATION" underneath in a smaller font. Below that, it says "A Cantel Medical Company".
Indications for Use
110579 510(k) number (if known):
Device Name: Millenium HX
Indications for Use:
The Mar Cor Purification Millenium HX Portable Reverse Osmosis Water Purification System is intended to be used as a dialysis accessory to produce water through reverse osmosio for yoten th hemodialysis equipment.
The Millenium HX can be connected to hemodialysis equipment used in hospitals, clinics and in home environments, in conjunction with the appropriate pre and post treatment units, as a part of a water treatment system designed to meet current AAMI and Federal (U.S) standards.
Prescription Use X (Part 21 CFR 801 Subpart D)
AND/OR
Over-the-Counter Use (Part 21 CFR 801 Subpart C)
(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
Doym In Whay
(Division Sign-Off)
Division of Reproductive, Gastro-Renal, and
Urological Devices
510(k) Number K110578
Mar Cor Puri
· Mar Cor Purification Attachment 3 - Page 1 of 1
Panel 1
/
Sort by
Ready
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.