Intended Use
The Mar Cor Purification Millenium HX Portable Reverse Osmosis Water Purification System is intended to be used as a dialysis accessory to produce water through reverse osmosis for use with hemodialysis equipment. The Millenium HX can be connected to hemodialysis equipment used in hospitals, clinics and in home environments, in conjunction with the appropriate pre and post treatment units, as a part of a water treatment system designed to meet current AAMI and Federal (U.S) standards.
Device Story
Portable reverse osmosis (RO) water purification system; processes feed water via pump-driven RO membrane to remove contaminants; produces purified water for hemodialysis equipment. Operated in hospitals, clinics, or home environments. Features include touch-screen interface, programmable logic controller (OPLC), water quality monitoring, and audible/visual alarms. System performs automatic divert-to-drain if product water TDS exceeds set-points. Supports heat disinfection and chemical sanitization to maintain low microbiological levels. Healthcare providers use the system as part of a water treatment train; output quality ensures safe dilution of dialysis concentrate. Benefits include consistent, AAMI-compliant water quality for hemodialysis therapy.
Clinical Evidence
Bench testing only. Performance data included system and RO membrane flow verification, product water quality testing across operating conditions, and software verification.
Technological Characteristics
Portable RO water purification system; utilizes RO membrane, system pump, and water quality monitoring. Features heat disinfection and chemical sanitization capabilities. Controlled via touch-screen interface and programmable logic controller (OPLC). Connectivity includes audible/visual alarms and automatic divert-to-drain functionality. Designed for integration with pre/post-treatment units (e.g., carbon adsorption).
Indications for Use
Indicated for use as a dialysis accessory to produce reverse osmosis water for hemodialysis equipment in hospitals, clinics, and home environments, requiring integration with pre and post-treatment units to meet AAMI/Federal water quality standards.
Regulatory Classification
Identification
A water purification system for hemodialysis is a device that is intended for use with a hemodialysis system and that is intended to remove organic and inorganic substances and microbial contaminants from water used to dilute dialysate concentrate to form dialysate. This generic type of device may include a water softener, sediment filter, carbon filter, and water distillation system.
Special Controls
*Classification.* Class II (special controls). The device, when it is a water purification subsystem disinfectant, is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 876.9.
Predicate Devices
- Millenium HX Portable Reverse Osmosis Water Purification System (K110578)
- WRO 300H (K093608)
Related Devices
- K200670 — EON Portable RO Water Purification System · Mar Cor Purification, Inc. · Apr 16, 2020
- K110578 — MILLENIUM HX PORTABLE REVERSE OSMOSIS WATER PURIFICATION SYSTEM · Mar Cor Purification, Inc. · May 20, 2011
- K151637 — UPT Series Medical RO Water System · Specialty Water Technologies · Sep 23, 2015
- K171099 — EON Portable Reverse Osmosis Water Purification System · Cantel · Jan 5, 2018
- K111740 — AMERIWATER MRO PORTABLE REVERSE OSMOSIS SYSTEM - SINGLE PATIENT, 1 TO 3 STATION · Ameriwater · Oct 5, 2011
Submission Summary (Full Text)
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). The logo consists of two parts: the Department of Health & Human Services logo on the left and the FDA logo on the right. The FDA logo is in blue and includes the letters "FDA" followed by the words "U.S. Food & Drug Administration".
October 3, 2019
Mar Cor Purification, Inc. % Mark Arnold Regulatory Affairs Manager, Life Sciences Cantel 9800 59th Ave. N Plymouth, MN 55442
Re: K192398
> Trade/Device Name: Millenium HX Regulation Number: 21 CFR 876.5665 Regulation Name: Water Purification System for Hemodialysis Regulatory Class: II Product Code: FIP Dated: August 27, 2019 Received: September 3, 2019
Dear Mark Arnold:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{1}------------------------------------------------
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely.
Carolyn Y. Neuland, Ph.D. Assistant Director DHT3A: Division of Renal, Gastrointestinal, Obesity and Transplant Devices OHT3: Office of GastroRenal, ObGyn, General Hospital and Urology Devices Office of Product Evaluation and Ouality Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
# Indications for Use
510(k) Number (if known) K192398
#### Device Name
Millenium HX Portable Reverse Osmosis Water Purification System
#### Indications for Use (Describe)
The Mar Cor Purification Millenium HX Portable Reverse Osmosis Water Purification System is intended to be used as a dialysis accessory to produce water through reverse osmosis for use with hemodialysis equipment.
The Millenium HX can be connected to hemodialysis equipment used in home environments, in conjunction with the appropriate pre and post treatment units, as a part of a water treatment system designed to meet current AAMI and Federal (U.S) standards.
Type of Use (Select one or both, as applicable)
| <span style="font-size:10px">☑</span> Prescription Use (Part 21 CFR 801 Subpart D) |
|------------------------------------------------------------------------------------|
| <span style="font-size:10px">☐</span> Over-The-Counter Use (21 CFR 801 Subpart C) |
#### CONTINUE ON A SEPARATE PAGE IF NEEDED.
{3}------------------------------------------------
Image /page/3/Picture/0 description: The image shows the logo for CANTEL. The logo consists of a blue and green globe on the left, followed by the word "CANTEL" in blue, sans-serif font. The globe is made up of curved stripes of blue and green, giving it a three-dimensional appearance. The word "CANTEL" is written in all capital letters and is positioned to the right of the globe.
October 2, 2019
## Millenium HX Portable Reverse Osmosis Water Purification System
Section 07 - 510(k) Summary
| Manufacturer: | Mar Cor Purification, Inc. |
|----------------------|------------------------------------------------------------------|
| Address: | 14550 28th Avenue North<br>Minneapolis, MN 55447<br>844-348-5636 |
| Official Contact: | Mark Arnold<br>Regulatory Affairs Manager, Cantel |
| Trade Name: | Millenium HX |
| Common Name: | Water Purification System |
| Classification Name: | Water Purification System for Hemodialysis |
| Product Code: | FIP |
| Device Class: | II |
| Regulation No: | 21 CFR 876 5665 |
Cantel has supplied the following information to the US Food and Drug Administration to support substantial equivalence of the Millenium HX Portable Reverse Osmosis Water Purification System to its predicates - Millenium HX Portable Reverse Osmosis Water Purification System, 510(k) K110578 and WRO 300H, 510(k) K093608 - both currently cleared for sale in the U.S.
#### 1. Device Description
The device is a portable water purification system that uses reverse osmosis to remove contaminants from water that is used to dilute dialysis concentrate to form dialysate for use in hemodialysis equipment. Feed water enters the unit and is directed through a pump into a RO membrane. The pump applies a high hydrostatic pressure that forces water from the concentrated (feed) side to the dilute (product) side of the RO membrane. As water flow across the membrane all types of water contaminants except dissolved gasses are removed and the purified product water is then supplied to hemodialysis equipment.
The Millenium HX is capable of generating purified water that meets AAMI water quality requirements for hemodialysis at a minimum of 1.4 liters/min. It must be used with appropriate pre and post treatment units, including at a minimum carbon adsorption media pretreatment in order
CONFIDENTIAL
{4}------------------------------------------------
Image /page/4/Picture/1 description: The image shows the Cantel logo. The logo features a blue and green globe-like design on the left, followed by the word "CANTEL" in large, bold, blue letters. The globe design consists of curved segments in varying shades of blue and green, creating a layered effect.
to remove chlorine/chloramines. Additional pre and post treatment requirements may vary and are dependent on the quality of the local feed water supply and individual facility requirements.
The Millenium HX system is designed to maintain low microbiological levels in the flow pathway through regular heat disinfection and chemical sanitization. Notable components and features of the Millenium HX include:
- RO membrane .
- System pump .
- Water quality monitoring system .
- Operating panel and programmable logic controller (OPLC) .
- Heat disinfection and chemical sanitization capability .
- Audible and visual alarms .
- Automatic divert to drain mode upon start-up and anytime product water TDS is above . the quality set-point
- System control via a touch-screen user interface .
#### 2. Intended Use
The Mar Cor Purification Millenium HX Portable Reverse Osmosis Water Purification System is intended to be used as a dialysis accessory to produce water through reverse osmosis for use with hemodialysis equipment.
The Millenium HX can be connected to hemodialysis equipment used in hospitals, clinics and in home environments, in conjunction with the appropriate pre and post treatment units, as a part of a water treatment system designed to meet current AAMI and Federal (U.S) standards.
## 3. Comparison to Other Devices in Commercial Distribution Within the United States
The Millenium HX is equivalent in intended use, functions and fundamental technology to its predicate devices, the Mar Cor Millenium HX Portable Reverse Osmosis Water Purification System (K110578) and Mar Cor (formerly Gambro) WRO 300H (K093608). All of the products are portable reverse osmosis water purification systems with the same intended use and equivalent indications for use.
#### Similarities between Subject and Predicate Devices
The Millenium HX is equivalent in function, intended use, scientific technology, principle of operation, performance characteristics and components its predicates. Both the subject and predicate devices utilize chemical sanitization and heat disinfection of the RO loop and machine.
CONFIDENTIAL
Page 29 of 268
{5}------------------------------------------------
Image /page/5/Picture/0 description: The image shows the logo for Cantel Medical. The logo consists of a blue and green globe on the left, followed by the word "CANTEL" in large, blue, sans-serif letters. The globe is made up of several curved segments of blue and green.
### Differences between Subject and Predicate Devices
The subject device is identical in build to the Millenium HX Portable Reverse Osmosis Water Purification System (K110578), except it uses the membrane from the WRO 300H (K093608), and has several minor component changes to adjust for dimensional fit.
### 4. Summary of Non-Clinical Performance Data
Cantel has conducted the following testing to demonstrate that the use of the WRO 300H membrane in the Millenium HX is safe and effective for its intended use:
- System and RO Membrane Performance - Flow and product water quality verification over range of operating conditions
- Software Verification
### 5. Conclusion
Based on the intended use, fundamental technology and performance data, the subject device is substantially equivalent to and is as safe and as effective as the legally marketed predicate devices.
CONFIDENTIAL
Page 30 of 268