K190612 · Dornier Medtech America, Inc. · EZN · Dec 6, 2019 · Gastroenterology, Urology
Device Facts
Record ID
K190612
Device Name
EQUINOX Balloon Dilatation Catheter
Applicant
Dornier Medtech America, Inc.
Product Code
EZN · Gastroenterology, Urology
Decision Date
Dec 6, 2019
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 876.5470
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
The Dornier EQUINOX Balloon Dilatation Catheter is indicated for dilation of the urinary tract.
Device Story
Dornier EQUINOX Balloon Dilatation Catheter is a sterile, single-use, multi-lumen catheter with a distal dilatation balloon. Used by physicians to exert radial force to dilate narrow ureteral segments. Operated under x-ray fluoroscopy guidance; radiopaque material (BaSO4) and marker bands allow visualization. Physician connects an inflation device to the proximal luer lock to inflate the balloon. Device facilitates mechanical dilation of urinary tract strictures, potentially improving patient urinary flow or access.
Clinical Evidence
Bench testing only. Performance testing included sterility, packaging, biocompatibility (USP Class VI), balloon burst durability, balloon deflation time, balloon/scope compatibility, radiopacity, effective working length, catheter tip length, balloon diameter at rated burst pressure (RBP), balloon length at operating pressure (OP), and catheter shaft outer diameter.
Technological Characteristics
Multi-lumen catheter with distal dilatation balloon. Materials: medical grade nylon (USP Class VI), PET (balloon), BaSO4 (radiopaque). Dimensions: 5.8 Fr diameter, 80 cm length; balloon diameters 4-6mm, lengths 4-10cm. Features: luer lock connector, radiopaque marker bands. Sterilization: sterile, single-use. No software or electronic components.
Indications for Use
Indicated for dilation of the urinary tract in patients requiring ureteral segment dilation.
Regulatory Classification
Identification
A ureteral dilator is a device that consists of a specially shaped catheter or bougie and is used to dilate the ureter at the place where a stone has become lodged or to dilate a ureteral stricture.
Predicate Devices
Boston Scientific's Uromax Ultra Balloon Dilatation Catheter (K130804)
Submission Summary (Full Text)
{1}------------------------------------------------
Image /page/0/Picture/0 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which is a blue square with the letters "FDA" in white. To the right of the blue square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue.
December 6, 2019
Dornier MedTech America Inc John Hoffer Vice President Quality, Regulatory, Clinical 1155 Roberts Blvd, Suite 100 Kennesaw, GA 30144
Re: K190612 Trade/Device Name: EQUINOX Balloon Dilatation Catheter Regulation Number: 21 CFR 876.5470 Regulation Name: Ureteral dilator Regulatory Class: II Product Code: EZN, KOE Dated: October 22, 2019 Received: October 22, 2019
Dear John Hoffer:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part
{2}------------------------------------------------
801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4. Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
Jessica K. Nguyen, Ph.D. Acting Assistant Director DHT3B: Division of Reproductive, Gynecology and Urology Devices OHT3: Office of GastroRenal, ObGyn, General Hospital and Urology Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health
Enclosure
{3}------------------------------------------------
### DEPARTMENT OF HEALTH AND HUMAN SERVICES Food and Drug Administration Indications for Use
Form Approved: OMB No. 0910-0120 Expiration Date: January 31, 2017 See PRA Statement on last page
510(k) Number (if known)
K190612
Device Name
### Dornier EQUINOX Balloon Dilatation Catheter
Indications for Use (Describe)
The Dornier EQUINOX Balloon Dilatation Catheter is indicated for dilation of the urinary tract.
Type of Use (Select one or both, as applicable)
区 Prescription Use (Part 21 CFR 801 Subpart D) Subpart C)
□ Over-The-Counter Use (21 CFR 801
PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON A SEPARATE PAGE IF NEEDED.
FOR FDA USE ONLY
Concurrence of Center for Devices and Radiological Health (CDRH) (Signature)
This section applies only to requirements of the Paperwork Reduction Act of 1995. *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.* The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to: Department of Health and Human Services Food and Drug Administration
Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number.
{4}------------------------------------------------
# 510(k) SUMMARY Dornier EQUINOX Balloon Dilatation Catheter
### Submitter's Name, Address, Telephone Number, Contact Person and Date Prepared
Dornier MedTech America, Inc. 1155 Roberts Blvd., Suite 00 Kennesaw, GA 30144
Date Prepared: October 18, 2019
Contact Person: John Hoffer
Phone: 770-514- 6163
#### Name of Device and Name/Address of Sponsor
Dornier EQUINOX Balloon Dilatation Catheter
Dornier MedTech America, Inc. 1155 Roberts Blvd. Kennesaw, GA 30144
Common or Usual Name
Ureteral Balloon Dilatation Catheter
#### Classification Name
According to 21 C.F.R. § 876.5470, the Food and Drug Administration ("FDA" or "the agency") has classified Ureteral Dilators as Class II devices (performance standards). The Product Code for these Dilators is EZN.
#### Predicate Device
Boston Scientific's Uromax Ultra Balloon Dilatation Catheter (K130804)
#### Purpose of the 510(k) Notice
The purpose of this submission is to obtain marketing clearance for the Dornier EQUINOX Balloon Dilatation Catheter product line.
#### Intended Use/Indications for Use
The Dornier EQUINOX Balloon Dilatation Catheter is indicated for dilation of the urinary tract.
{5}------------------------------------------------
## Device Description
The Dornier EQUINOX Balloon Dilatation Catheter is a multiple lumen catheter with a dilatation balloon mounted at the distal tip. Dilatation balloon catheters are used to exert radial force to dilate narrow ureteral segments.
The Dornier EQUINOX Balloon Dilatation Catheters are sterile, single-use devices. The EQUINOX Balloon Dilatation Catheters are available in 5.8 French (Fr) diameter, with a catheter length of 80 centimeters (cm). There will be three (3) balloon diameters, (4mm, 5mm and 6mm) and two (2) balloon lengths (4cm and 10cm)
The Dornier EQUINOX Balloon Dilatation Catheters are constructed of a medical grade nylon material. This material has been USP Class VI tested. The base material is compounded with BaSO4 to render the Dornier EQUINOX Balloon Dilatation Catheter radiopaque under x-ray fluoroscopy. The catheter also has radio-opaque marker bands on the proximal and distal ends of the balloon for visualization during use. All colorants used are compliant with FDA standards.
# Technological Characteristics
The Dornier EQUINOX Balloon Dilatation Catheters have the same basic technological characteristics as the predicate device, Boston Scientific's Uromax Ultra Balloon Dilatation Catheter. They both are a multiple lumen catheter with a dilatation balloon mounted at the distal tip. They are both manufactured using PET for the inflation balloon which is the primary functional component of the devices. At the proximal end, they both have leur lock connectors that will attach to the inflation device to inflate the balloon during use. As noted above, both devices are compounded with Barium sulfate and radiopaque marking bands to allow visualization under fluoroscopy.
# Performance Data
The Dornier EQUINOX Balloon Dilatation Catheter was subjected to the following tests to assure design and performance under the specified testing parameters:
- Sterilitv Packaging Biocompatibility Balloon Burst Durability Test Balloon Deflation Time Balloon/Scope Compatibility Radiopacity Effective Working Length Catheter Tip Length Balloon Diameter @ RBP Balloon Length @ OP Catheter Shaft OD
{6}------------------------------------------------
All testing was found to be acceptable and substantially equivalent to those of the predicate device.
## Substantial Equivalence
A comparison of design characteristics has been performed and demonstrates that the proposed Dornier EQUINOX Balloon Dilatation Catheter is substantially equivalent to the predicate device in terms of intended use, technological characteristics, type of materials and performance characteristics. Therefore, the proposed Dornier EQUINOX Balloon Dilatation Catheter is as safe, as effective, and performs as well as the predicate device.
## Conclusion
Based on the data and information comparing the Dornier EQUINOX Balloon Dilatation Catheter and the predicate device we conclude they are substantially equivalent as they have the same intended use, basic design, principle of operation, technology, materials, and performance to the predicate. Any minor differences between the subject and predicate devices do not raise any concerns regarding the overall safety or effectiveness. Thus, the Dornier EQUINOX Balloon Dilatation Catheter is substantially equivalent to its predicate device.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.