GYRUS ACMI URO-EZDILATE URETERAL BALLOON DILATION CATHETER

K132181 · Olympus Surgical Technologies America · EZN · Oct 21, 2013 · Gastroenterology, Urology

Device Facts

Record IDK132181
Device NameGYRUS ACMI URO-EZDILATE URETERAL BALLOON DILATION CATHETER
ApplicantOlympus Surgical Technologies America
Product CodeEZN · Gastroenterology, Urology
Decision DateOct 21, 2013
DecisionSESE
Submission TypeTraditional
Regulation21 CFR 876.5470
Device ClassClass 2
AttributesTherapeutic

Indications for Use

The Gyrus ACMI Uro - EZDilate Ureteral Balloon Dilation Catheter is recommended for dilation of the urinary tract.

Device Story

Reinforced catheter with distal dilatation balloon; radiopaque tip and markers for visualization. Used for dilating urinary tract strictures. Guidewire-assisted insertion; balloon inflated to deliver radial force. Hydrophilic coating facilitates insertion, positioning, and removal. Operated by clinicians in clinical settings. Sold separately or in kits with inflation devices. Benefits include effective stricture dilation to restore urinary tract patency.

Clinical Evidence

Bench testing only. Testing included first article inspection, balloon burst, kink, durability (cycle), compliance, shape characteristics, insertion force, cystoscope compatibility, and deflation testing.

Technological Characteristics

Reinforced catheter; polycarbonate two-way hub; polyurethane strain relief; polyamide outer tube; Polyethylene Terephthalate (PET) balloon; polyamide inner catheter body. Hydrophilic coating. Sterilized via ethylene oxide. Complies with ISO 10993-5, ISO 10993-10, ANSI/AAMI/ISO 11607-1, ANSI/AAMI/ISO 11135-1, and ISO 14971.

Indications for Use

Indicated for dilation of the urinary tract in patients requiring stricture treatment.

Regulatory Classification

Identification

A ureteral dilator is a device that consists of a specially shaped catheter or bougie and is used to dilate the ureter at the place where a stone has become lodged or to dilate a ureteral stricture.

Predicate Devices

Related Devices

Submission Summary (Full Text)

{0}------------------------------------------------ Gyrus ACMI Uro-EZDilate Ureteral Balloon Dilation Catheter Gyrus ACMI, Inc. Traditional 510(k) Notification July 12, 2013 ### 510(k) Summary of Safety and Effectiveness Gyrus ACMI, Inc. Gyrus ACMI Uro - EZDilate Ureteral Balloon Dilation Catheter #### General Information Contract Manufacturer: Future Matrix Interventional, Inc. 1605 Enterprise Street Athens, TX 75751 Phone: 903-677-9166 Establishment Registration Number: Establishment Registration Number: 510(k) Submitter: Contact Person: Date Prepared: Device Description Classification Name: Gyrus ACMI, Inc. **OCT** 2 1 2013 136 Turnpike Rd. Southborough, MA 01772-2104 3003790304 1646831 Neil Kelly Regulatory Affairs Specialist 508-804-2690 Neil.kelly@olympus-osta.com July 12, 2013 Dilator, Urethral Dilator, Catheter, Ureteral 21 CFR 876.5470, 876.5520 EZN. KOE Class II Gastroenterology/Urology Gyrus ACMI Uro – EZDilate Ureteral Balloon Dilation Catheter Balloon Dilation Catheter Predicate Devices Generic/Common Name: Trade Name: Boston Scientific UroMax Ultra K130804 {1}------------------------------------------------ Gyrus ACMI Uro - EZDilate Ureteral Balloon Dilation Catheter Gyrus ACMI, Inc. Traditional 510(k) Notification July 12, 2013 #### Product Description The Gyrus ACMI Uro - EZDilate Ureteral Balloon Dilation Catheter is a reinforced catheter attached to a distal dilatation balloon. It has a radiopaque tip and two radiopaque markers positioned inside the balloon that define the working length. The balloon catheter can be used to dilate strictures of the urinary tract. #### Technological Characteristics The Gyrus ACMI Uro - EZDilate Ureteral Balloon Dilation Catheter is a reinforced catheter attached to a distal dilatation balloon. Upon inflation, a radial force is delivered over the length of the balloon. A hydrophilic coating applied to the balloon increases ease of insertion, positioning of the balloon within the ureter, and device removal. The dilation catheter(s) is sold separately, or as part of a kit containing an inflation device. The inflation device is also available for individual sale. #### Material A polycarbonate two-way hub at the proximal end leads into a polyurethane strain relief and into polyamide outer tube. The Polyethylene Terephthalate balloon joins the . outer catheter body and sits over a non-patient contacting loaded polyamide inner catheter body. A guidewire is positioned within the patient and the proposed device is then fed over the guidewire and into position. #### Intended Uses The Gyrus ACMI Uro - EZDilate Ureteral Balloon Dilation Catheter is recommended for dilation of the urinary tract. #### Compliance to Voluntary Standards The design of the proposed device complies with the following standards: ISO 10993-5. 2009 ISO 10993-10, 2010 ANSI/AAMI/ISO 11607-1, 2006 ANSI/AAMI/ISO 11135-1, 2007 ISO 14971, 2007 {2}------------------------------------------------ Gyrus ACMI Uro-EZDilate Ureteral Balloon Dilation Catheter Gyrus ACMI, Inc. Traditional 510(k) Notification July 12, 2013 #### Summary of Sterilization and Shelf Life Discussion The Gyrus ACMI Uro - EZDilate Ureteral Balloon Dilation Catheter is delivered in a sterile state and is intended for single patient use only. The sterilization method used is ethylene oxide and has a shelf.life of one(1) year. #### Summary of Performance Testing The following performance tests were conducted: - First Article Inspection . - . Balloon Burst Testing - Balloon Kink Testing . - Durability Testing (cycle testing) . - Compliance Testing . - Balloon Shape Characteristics Testing . - . Balloon Insertion Force Testing - Balloon Cystoscope Compatibility Testing D - Balloon Deflation Testing . #### Substantial Equivalence The proposed Gyrus ACMI Uro - EZDilate Ureteral Balloon Dilation Catheter has the same intended use, design, and scientific technology as the Predicate Boston Scientific UroMax Ultra Balloon Dilation Catheter (K130804). Both devices are of similar design and there were no new issues of safety or effectiveness with the proposed device. #### Conclusion: In summary, the Gyrus ACMI Uro - EZDilate Ureteral Balloon Dilation Catheter is substantially equivalent to the predicate devices and presents no new questions of safety or efficacy. {3}------------------------------------------------ Image /page/3/Picture/0 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a circular seal with the words "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" arranged around the perimeter. Inside the circle is a stylized image of an eagle with its wings spread, symbolizing the department's mission to protect the health of all Americans. Food and Drug Administration 10903 New Hampshire Avenue Document Cantrol Center - WO66-G609 Silver Spring MD 20993-0002 October 21, 2013 Olympus Surgical Technologics America Gyrus ACMI % Neil Kelly Regulatory Affairs Specialist 136 Turnpike Road Southborough, MA 01772 Re: K132181 Trade/Device Name: Gyrus ACM1 Uro – EZDilate Ureteral Balloon Dilation Catheter Regulation Number: 21 CFR§ 876.5470 Regulation Name: Ureteral dilator Regulatory Class: II Product Code: EZN, KOE Dated: September 25, 2013 Received: September 26, 2013 Dear Neil Kelly, We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have deternined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976. the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The rou may the organisions of the Act include requirements for annual registration, listing of general controls providering practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading. If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register. Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complics with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. {4}------------------------------------------------ Page 2 - Neil Kelly You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050. If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please go to http://www.fda.gov/AboutFDA/CentersOffices/CDRH/CDRHOffices/ucm115809.htm for the Center for Devices and Radiological Health's (CDRH's) Office of Compliance. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance. You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Sincerely yours, # Herbert P. Lerner -S for Benjamin R. Fisher, Ph.D. Director Division of Reproductive, Gastro-Renal, and Urological Devices Office of Device Evaluation Center for Devices and Radiological Health Enclosure {5}------------------------------------------------ Gyros ACMI URO - Ureteral EZDitate Ureteral Balloon Dilation Catheter Gyrus ACMI, Inc. Traditional 510(k) Notification July 12, 2013 ## Indications for Use Device Name: Gyrus ACMI Uro - EZDilate Ureteral Balloon Dilation Catheter 510(k) Number: TBD K132181 Indications for use: The Gyrus ACMI Uro - EZDilate Ureteral Balloon Dilation Catheter is recommended for dilation of the urinary tract. Prescription Use: ____________________________________________________________________________________________________________________________________________________________ OR Over-the-Counter Use: _ (Per 21 CFR 801.109) (PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NEEDED) Concurrence of CDRH, Office of Device Evaluation (ODE) # Herbert P. Lerner -S 2013.10.21 15:40:37 -04'00'
Innolitics

Panel 1

/
Sort by
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...