The 2-Way 100% Silicone ClearTract Catheter is intended for drainage of the urinary tract. Catheterization is accomplished by inserting the catheter through the urethra and into the bladder. Intended population is adults and pediatrics.
Device Story
The 2-Way 100% Silicone ClearTract Catheter is a single-use Foley catheter designed for urinary tract drainage. It features two lumens: one for balloon inflation/deflation and one for drainage. Constructed from medical-grade silicone, the device includes a zwitterionic polymer surface modification to reduce friction. It is inserted through the urethra into the bladder by a healthcare provider. The device connects to a standard urinary bag via a non-interconnectable connector. By facilitating continuous drainage, it assists in managing urinary retention or post-operative bladder care. No physical modifications were made to the device; this submission reflects updated labeling regarding the surface modification.
Clinical Evidence
No clinical data. Bench testing only. Friction testing performed per ASTM D1894-14 demonstrated a significant reduction in the coefficient of friction for the treated silicone surface compared to untreated silicone.
Technological Characteristics
Medical-grade silicone construction with zwitterionic polymer surface modification. 2-way Foley design with inflation and drainage lumens. Sizes: 14Fr/10cc, 16Fr/5cc, 18Fr/10cc. Sterilization: Ethylene Oxide gas. Performance standard: ASTM F623. Single-use, prescription-only.
Indications for Use
Indicated for drainage of the urinary tract in adult and pediatric patients via urethral insertion into the bladder.
Regulatory Classification
Identification
A urological catheter and accessories is a flexible tubular device that is inserted through the urethra and used to pass fluids to or from the urinary tract. This generic type of device includes radiopaque urological catheters, ureteral catheters, urethral catheters, coudé catheters, balloon retention type catheters, straight catheters, upper urinary tract catheters, double lumen female urethrographic catheters, disposable ureteral catheters, male urethrographic catheters, and urological catheter accessories including ureteral catheter stylets, ureteral catheter adapters, ureteral catheter holders, ureteral catheter stylets, ureteral catheterization trays, and the gastro-urological irrigation tray (for urological use).
K024010 — DOVER SILVER HYDROGEL COATED SILICONE FOLEY CATHETER · The Kendall Company Div. of Tyco Healthcare Group · Jun 11, 2003
Submission Summary (Full Text)
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). On the left, there is the Department of Health & Human Services logo. To the right of that is the FDA logo, which is a blue square with the letters "FDA" in white. To the right of the FDA logo is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue.
July 1, 2022
Silq Technologies Corporation Sigi Caron, MBA, RAC Vice President of Regulatory Affairs 323 Sunny Isles Blvd. Sunny Isles Beach, FL 33160
Re: K221625
> Trade/Device Name: 2-Way 100% Silicone ClearTract Catheter Regulation Number: 21 CFR§ 876.5130 Regulation Name: Urological catheter and accessories Regulatory Class: II Product Code: EZL Dated: June 3, 2022 Received: June 6, 2022
Dear Sigi Caron:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies.
{1}------------------------------------------------
You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatoryinformation/postmarketing-safety-reporting-combination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
For. Jessica K. Nguyen, Ph.D. Assistant Director DHT3B: Division of Reproductive, Gynecology and Urology Devices OHT3: Office of GastroRenal, ObGyn, General Hospital and Urology Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
| DEPARTMENT OF HEALTH AND HUMAN SERVICES |
|-----------------------------------------|
| Food and Drug Administration |
Indications for Use
510(k) Number (if known)
K221625
Device Name
2-Way 100% Silicone ClearTract Catheter
Indications for Use (Describe)
The 2-Way 100% Silicone ClearTract Catheter is intended for drainage of the winnary tract. Catheterization is accomplished by inserting the catheter through the wethra and into the bladder. Intended population is adults and pediatrics.
Type of Use (Select one or both, as applicable)
2 Prescription Use (Part 21 CFR 801 Subpart D)
_ Over-The-Counter Use (21 CFR 801 Subpart C)
Form Approved: OMB No. 0910-0120
Expiration Date: 06/30/2023
See PRA Statement below.
#### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW .*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
FORM FDA 3881 (6/20)
PSC Publishing Services (101)-46-6740
Silq Technologies Corporation Special 510(k) Submission
CONFIDENTIAL
{3}------------------------------------------------
## 1. SPECIAL 510(K) SUMMARY
This summary of 510(k) safety and effectiveness information is being submitted in accordance with the requirements of 21 CFR 807.92.
| DATE PREPARED | June 16, 2022 |
|---------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| APPLICANT | Silq Technologies Corporation<br>323 Sunny Isles Blvd, 7th Floor<br>Sunny Isles, FL 33160 |
| OFFICIAL<br>CORRESPONDENT | Sigi Caron, MBA, RAC<br>Biologics and Medical Device Consulting Group (BioMDg)<br>377 Second Tee Dr, #5868<br>Incline Village, NV 89451<br>phone: (831) 346-6970<br>e-mail: sigi@bioMDg.com |
| TRADE NAME | 2-Way 100% Silicone ClearTract Catheter |
| COMMON NAME | Foley Catheter |
| DEVICE CLASSIFICATION | Name: Urological Catheter and Accessories<br>Regulation No: 21 CFR §876.5130<br>Product Code: EZL - Catheter, Retention Type, Balloon<br>Class: II |
| PREDICATE DEVICE | HDX 100% Silicone 2-Way Foley Catheter (K192034) |
#### DESCRIPTION OF THE DEVICE SUBJECT TO PREMARKET NOTIFICATION:
The 2-Way 100% Silicone ClearTract Catheter (previously known as the HDX 100% Silicone 2-Way Foley Catheter) is a standard single-use 2-way Foley catheter that is constructed of medical grade silicone with a surface modification. It incorporates two (2) lumens, one for inflation/deflation of the balloon and the other for drainage of the urinary tract. The drainage inlet is located distal to the catheter's balloon. The connection to the urinary bag is a standard non-interconnectable connector.
#### INDICATIONS FOR USE:
The indications for use remain the same. The 2-Way 100% Silicone ClearTract Catheter is intended for drainage of the urinary tract. Catheterization is accomplished by inserting the catheter through the urethra and into the bladder. Intended population is adults and pediatrics.
#### PREDICATE DEVICE
The renamed 2-Way 100% Silicone ClearTract Catheter has not been modified in any way. It is exactly the same as the HDX 100% Silicone 2-Way Catheter cleared under 510(k)
{4}------------------------------------------------
premarket notification K192034.
## SUMMARY OF MODIFICATIONS
This special 510(k) premarket notification only proposes an update to the product labeling. The following information has been added to the Instructions For Use:
Device description - "The 2-Way 100% Silicone ClearTract Catheter is a standard singleuse 2-way Foley catheter that is constructed of medical grade silicone with a surface modification. It incorporates two (2) lumens, one for inflation/deflation of the balloon and the other for drainage of the urinary tract."
Performance Characteristics - "The surface of the ClearTract Catheter has a surface modification that reduces surface friction."
No physical changes or modifications are proposed to the cleared device.
## PERFORMANCE STANDARDS:
No performance standards have been established by the Agency to date that apply to this device.
# SUMMARY OF NONCLINICAL TESTING:
The following bench testing was performed to support the additional information added to the product labeling:
- Friction testing based on ASTM D1894-14 which demonstrates a significant reduction in the coefficient of friction between the untreated and treated silicone surface.
# SUBSTANTIAL EQUIVALENCE - COMPARISON TO PREDICATE DEVICE
The 2-Way 100% Silicone ClearTract Catheter has not been modified in any way. It is exactly the same as the HDX 100% Silicone 2-Way Catheter cleared under 510(k) premarket notification K192034. Following is the SE Table.
{5}------------------------------------------------
| Element | Subject Device:<br>2-Way 100% Silicone<br>ClearTract Catheter | Predicate Device:<br>HDX 100% Silicone 2-Way<br>Foley Catheter | Comparison |
|----------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|------------|
| 510(k) Number | K221625 | K192034 | --- |
| Indications for use | ... intended for drainage of the<br>urinary tract. Catheterization<br>is accomplished by inserting<br>the catheter through the<br>urethra and into the bladder.<br>Intended population is adults<br>and pediatrics. | ... intended for drainage of the<br>urinary tract. Catheterization<br>is accomplished by inserting<br>the catheter through the<br>urethra and into the bladder.<br>Intended population is adults<br>and pediatrics. | Same |
| Design Features | | | |
| Type | 2-Way Foley Catheter with<br>inflation and drainage lumens | 2-Way Foley Catheter with<br>inflation and drainage lumens | Same |
| Size/Balloon | 14Fr/10cc<br>16Fr/5cc<br>18Fr/10cc | 14Fr/10cc<br>16Fr/5cc<br>18Fr/10cc | Same |
| Material of Construction | Silicone, Medical Grade | Silicone, Medical Grade | Same |
| Zwitterionic Polymer<br>Surface Modification | Yes | Yes | Same |
| Performance Standard | ASTM F623 | ASTM F623 | Same |
| Single Use? | Yes | Yes | Same |
| Prescription Use? | Yes | Yes | Same |
| Sterile? | Yes | Yes | Same |
| Sterilization Method | Ethylene Oxide Gas | Ethylene Oxide Gas | Same |
# CONCLUSION:
All changes were made in compliance with 21 CFR 820.30 for design controls. Based on a risk assessment, the proposed additions to the labeling do not change or add new risks to the CleartTract catheter. All performance bench testing was successfully completed and provided objective evidence on the performance characteristics of the ClearTract catheter. The ClearTract catheter is substantially equivalent to the predicate device.
Panel 1
/
Ready
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.