Urological catheter is intended for drainary tract. Catheterization is accomplished by inserting the catheter through the urethra and into the bladder. Intended population is adults and pediatrics.
Device Story
Single-use, 2-way Foley catheter; constructed of medical-grade silicone. Features two lumens: one for balloon inflation/deflation, one for urinary drainage. Distal balloon maintains catheter position in urinary tract; proximal connector uses standard slip/taper luer for syringe inflation. Drainage inlet located distal to balloon; connects to standard urinary bag. Used in clinical settings for urinary drainage. Healthcare provider inserts catheter through urethra into bladder; inflates balloon to secure; deflates for removal. Benefits patient by facilitating urine drainage.
Clinical Evidence
Bench testing only. Evaluations included biocompatibility, sterilization, packaging integrity, transportation, and performance/functionality testing in accordance with FDA guidance and ASTM F623 standards. All tests yielded successful results.
Indicated for drainage of the urinary tract in adult and pediatric populations via urethral insertion.
Regulatory Classification
Identification
A urological catheter and accessories is a flexible tubular device that is inserted through the urethra and used to pass fluids to or from the urinary tract. This generic type of device includes radiopaque urological catheters, ureteral catheters, urethral catheters, coudé catheters, balloon retention type catheters, straight catheters, upper urinary tract catheters, double lumen female urethrographic catheters, disposable ureteral catheters, male urethrographic catheters, and urological catheter accessories including ureteral catheter stylets, ureteral catheter adapters, ureteral catheter holders, ureteral catheter stylets, ureteral catheterization trays, and the gastro-urological irrigation tray (for urological use).
{1}------------------------------------------------
Image /page/0/Picture/0 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). The logo consists of two parts: the Department of Health and Human Services seal on the left and the FDA acronym along with the full name of the agency on the right. The FDA part of the logo is in blue, with the acronym in a solid blue square and the agency's name in a lighter shade of blue.
April 21, 2020
Pathway, LLC Aaron Rogers Director of Regulatory and Quality 8779 Cottonwood Avenue, Suite 105 Santee, CA 92071
Re: K192034 Trade/Device Name: HDX 100% Silicone 2-way Foley Catheter, 14Fr/10cc HDX 100% Silicone 2-way Foley Catheter, 16Fr/5cc HDX 100% Silicone 2-way Foley Catheter, 18Fr/10cc Regulation Number: 21 CFR 876.5130 Regulation Name: Urological catheter and accessories Regulatory Class: II Product Code: EZL Dated: March 19, 2020 Received: March 23, 2020
Dear Aaron Rogers:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling
{2}------------------------------------------------
(21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safetyreporting-combination-products); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-devices/medical-device-safety/medical-devicereporting-mdr-how-report-medical-device-problems.
For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
Purva Pandya Acting Assistant Director DHT3B: Division of Reproductive, Gynecology and Urology Devices OHT3: Office of GastroRenal, ObGyn, General Hospital and Urology Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health
Enclosure
{3}------------------------------------------------
# Indications for Use
510(k) Number (if known) K192034
Device Name
HYDROPHILIX 100% Silicone 2-Way Foley Catheter, 14Fr/10cc HYDROPHILIX 100% Silicone 2-Way Foley Catheter, 16Fr/5cc HYDROPHILIX 100% Silicone 2-Way Foley Catheter, 18Fr/10cc
Indications for Use (Describe)
Urological catheter is intended for drainary tract. Catheterization is accomplished by inserting the catheter through the urethra and into the bladder. Intended population is adults and pediatrics.
Type of Use (Select one or both, as applicable)
X Prescription Use (Part 21 CFR 801 Subpart D)
| Over-The-Counter Use (21 CFR 801 Subpart C)
#### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{4}------------------------------------------------
K192034 page 1 of 3
Image /page/3/Picture/1 description: The image contains the logo for Hydrophilix. On the left is a blue hexagon shape with a circular design in the middle. To the right of the shape is the company name, HYDROPHILIX, in blue font.
Image /page/3/Picture/2 description: The image shows the word "pathway" in lowercase letters, with a stylized logo to the right of the word. The logo consists of two curved shapes, one above the other, forming a diamond-like shape. The text and logo are in a gray color. The image appears to be a logo or branding element.
# 510(K) SUMMARY - K192034
## Submitter Information
| Company Name: | Hydrophilix, Inc. |
|------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Company Address: | 12100 Wilshire Blvd<br>Suite 800<br>Los Angeles, CA 90025 |
| Contact Person: | Aaron Rogers<br>Pathway, LLC.<br>Director of Regulatory & Quality<br>(619) 415-0103<br>arogers@pathwaynpi.com |
| Date: | April 21, 2020 |
| Device Identification | |
| Device Trade Name: | HYDROPHILIX 100% Silicone 2-Way Foley<br>Catheter, 14Fr/10cc<br>HYDROPHILIX 100% Silicone 2-Way Foley<br>Catheter, 16Fr/5cc<br>HYDROPHILIX 100% Silicone 2-Way Foley<br>Catheter, 18Fr/10cc |
| Common Name: | Foley Catheter |
| Classification Name: | Urological catheter and accessories |
| Classification Number: | 876.5130 |
| Regulatory Class: | Class II |
| Product Code(s): | EZL |
| Product Code Name: | Catheter, Retention Type, Balloon |
| Advisory Panel: | Gastroenterology/Urology |
# Identification of Predicate Device
| Predicate Device | | | | |
|--------------------------------|------------------------------------------------|--------------|---------------|------------------|
| Device Name | Regulation No. | Product Code | 510(K) Number | Clearance Date |
| Bardex Lubi-Sil Foley Catheter | 876.5130 - Urological catheter and accessories | EZL | K984084 | February 1, 1999 |
# Device Description
The Subject Device is a single-use, 2-way Foley catheter that is constructed of medical grade silicone. It incorporates two (2) lumens, one for inflation/deflation of the other for drainage of the urinary tract.
{5}------------------------------------------------
The balloon, at the distal end of the catheter, is inflated to the labeled fill volume and remains inflated via the incorporated two-way valve in the connector at the proximal end of the inflation lumen. The connector engages with slip/taper luer connection that is standard on syringes. With the balloon inflated, the catheter maintains position in the urinary tract for the duration of the clinical application. Prior to removal, the balloon is deflated via the 2-way valve connection.
The drainage inlet is located distal to the catheter's balloon. The urinary bag is a standard non-interconnectable connector.
The Subiect Device is packaged inside a perforated plastic sleeve that is inserted into a sterile barrier pouch. Ten (10) units are packaged into a carton, which is the saleable unit. The Subject Device is terminally sterilized via Ethylene Oxide (EO).
The Subject Device will be initially offered in 14. 16. 18Fr sizes. The entire length of the Subject Device is disposable. The Subject Device is not sold as a set, there are no accessories or components included.
#### Indications for Use
Urological catheter intended for drainage of the urinary tract. Catheterization is accomplished by inserting the catheter through the urethra and into the bladder. Intended population is adults and pediatrics.
| Comparison Table - Subject and Predicate Devices | | |
|--------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Comparison<br>Feature | Subject Device | Predicate Device |
| Indications for Use | Urological catheter intended for drainage of<br>the urinary tract. Catheterization is<br>accomplished through the urinary tract.<br>Intended population is adults and pediatrics. | Used in the drainage and/or collection<br>and/or measurement of urine. Generally,<br>drainage is accomplished by inserting the<br>catheter through the urethra and into the<br>bladder. However, drainage is sometimes<br>accomplished by suprapubic or other<br>placement of the catheter, such as a<br>nephrostomy tract. |
| Type | 2-way with inflation and drainage lumens | 2-way with inflation and drainage lumens |
| Size/Balloon Size | 14Fr/10cc | 8, 10Fr/3cc Balloon |
| | 16Fr/5cc | 12-24Fr, even sizes/5cc Balloon |
| | 18Fr/10cc | 16-24Fr, even sizes/30cc Balloon |
| Material | Silicone, Medical Grade | Silicone, High Grade |
| Performance<br>Standard | ASTM F623 | ASTM F623 |
| Single-Use? | Yes | Yes |
| Prescription Use? | Yes | Yes |
| Sterile? | Yes | Yes |
## Comparison of Technological Characteristics with Predicate Device
## Summary of Evaluations Performed
The test program was performed in accordance to FDA guidance and recognized performance states, which includes the following:
- Biocompatibility
- Sterilization
- Packaging Integrity (i.e., Sterile Barrier) ●
{6}------------------------------------------------
K192034 page 3 of 3
- Transportation .
- Performance/Functionality .
Successful results were achieved with all evaluations conducted.
# Conclusion
The Subject Device has demonstrated it is substantially equivalent to the commercially available predicate.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.