QAG · Endoscopic Electrosurgical Clip Cutting System

Obstetrics/Gynecology · 21 CFR 876.4310 · Class 2

Overview

Product CodeQAG
Device NameEndoscopic Electrosurgical Clip Cutting System
Regulation21 CFR 876.4310
Device ClassClass 2
Review PanelObstetrics/Gynecology

Identification

An endoscopic electrosurgical clip cutting system is a prescription device that applies electrical energy to fragment metallic clips, which are devices placed in the digestive tract to close gastrointestinal perforations, hemorrhages, or perform resection. The system includes instruments that are then used to remove the fragmented clips from the digestive tract.

Classification Rationale

Class II (special controls). The special controls for this device are:

Special Controls

In combination with the general controls of the FD&C Act, the endoscopic electrosurgical clip cutting system is subject to the following special controls:

*Classification.* Class II (special controls). The special controls for this device are:(1) Non-clinical performance testing must demonstrate that the device performs as intended under anticipated conditions of use. The following performance characteristics must be tested: (i) Performance bench testing to evaluate the functionality (including stress, compatibility, usability, and reliability) of the device during use; (ii) Electrical and thermal safety testing; and (iii) Electromagnetic compatibility testing. (2) Animal testing must evaluate tissue damage, including thermal effects, during the clip removal procedure. This testing must also evaluate usability and effectiveness of the device. (3) The patient-contacting components of the device must be demonstrated to be biocompatible. (4) Performance data must demonstrate the sterility of the device components intended to be provided sterile. (5) Performance data must support shelf life by demonstrating continued sterility of the device (or the sterile components), package integrity, and device functionality over the labeled shelf life. (6) Labeling of the device must include: (i) Instructions for use, and (ii) A shelf life for single use components.

Recent Cleared Devices (1 of 1)

RecordDevice NameApplicantDecision DateDecision
DEN160014remOVE SystemOvesco Endoscopy AGDec 22, 2017DENG

Top Applicants

Innolitics

Panel 1

/
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...