The device for flexible endoscopy and for compression of tissue in the gastrointestinal (GI) tract, for for treating gastrointestinal organ wall lesions, and for making of lesions. The Device is indicated for clip placement within the GI tract for the purpose of: - a .Endoscopic Marking - b. Hemostasis for: - Mucosal/Submucosal defects < 3cm - Bleeding Ulcers - Arteries <2 mm - Polyps < 1.5 cm in diameter - Diverticula in the Colon c. Closure of GI tract luminal perforations < 20mm that can be treated conservatively.
Device Story
Disposable, sterile, single-use endoscopic clipping device. Input: tissue within GI tract. Operation: negative pressure attracts tissue into barrel; handle operation releases nickel-titanium (NiTi) clip. Output: deployed clip for hemostasis, lesion closure, or marking. Used in clinical endoscopy settings by physicians. Output allows visualization of clip placement; aids in stopping bleeding, closing perforations, or marking sites. Benefits: minimally invasive tissue closure and hemostasis.
Clinical Evidence
No clinical data. Bench testing only. Performance verified via biocompatibility (ISO 10993-1), sterilization validation (ISO 11135), shelf-life/packaging integrity (ASTM F1980, ISO 11607), MR compatibility (ASTM F2503, F2052, F2119, F2182, F2213), and mechanical design verification (dimension, release performance, recovery, closure force, connection force, endoscope adaptability, deployment/persistence).
Technological Characteristics
Materials: Nickel-titanium (NiTi) alloy clip. Form factor: Circular clip (14.5mm, 17.0mm diameter); working lengths 1650mm-2700mm. Principle: Negative pressure tissue attraction and mechanical clip release. Sterilization: Ethylene Oxide (EO), SAL 10^-6. Connectivity: None (mechanical). Standards: ISO 10993-1 (biocompatibility), ISO 11135 (sterilization), ASTM F1980/ISO 11607 (packaging), ASTM F2503/F2052/F2119/F2182/F2213 (MR compatibility).
Indications for Use
Indicated for patients requiring endoscopic tissue compression in the GI tract for hemostasis, lesion treatment, or marking. Specific applications include mucosal/submucosal defects <3cm, bleeding ulcers, arteries <2mm, polyps <1.5cm, colonic diverticula, and GI luminal perforations <20mm treatable conservatively.
Regulatory Classification
Identification
A hemorrhoidal ligator is a device used to cut off the blood flow to hemorrhoidal tissue by means of a ligature or band placed around the hemorrhoid.
Special Controls
*Classification.* Class II (special controls). Except for a hemostatic metal clip intended for use in the gastrointestinal tract, the device is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 876.9.
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which is a blue square with the letters "FDA" in white. To the right of the blue square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue.
June 13, 2022
Micro-Tech (Nanjing) Co., Ltd. Sally He, RA Engineer No. 10 Gaoke Third Road. Naniing National Hi-Tech Industrial Development Zone Nanjing, Jiangsu 210032 CHINA
Re: K220157
> Trade/Device Name: Disposable Hemostatic Closure Clip Device Regulation Number: 21 CFR 876.4400 Regulation Name: Hemorrhoidal ligator Regulatory Class: Class II Product Code: PKL Dated: May 11, 2022 Received: May 13, 2022
Dear Sally He:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any
{1}------------------------------------------------
Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safetyreporting-combination-products); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-devices/medical-device-safety/medicaldevice-reporting-mdr-how-report-medical-device-problems.
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medical-device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-devices/device-advicecomprehensive-regulatory-assistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerelv.
Shanil P. Haugen, Ph.D. Assistant Director DHT3A: Division of Renal, Gastrointestinal, Obesity and Transplant Devices OHT3: Office of GastroRenal, ObGyn, General Hospital and Urology Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
# Indications for Use
510(k) Number (if known) K220157
Device Name
Disposable Hemostatic Closure Clip Device
Indications for Use (Describe)
The device for flexible endoscopy and for compression of tissue in the gastrointestinal (GI) tract, for for treating gastrointestinal organ wall lesions, and for making of lesions. The Device is indicated for clip placement within the GI tract for the purpose of:
- a .Endoscopic Marking
- b. Hemostasis for:
- Mucosal/Submucosal defects < 3cm
- Bleeding Ulcers
- Arteries <2 mm
- Polyps < 1.5 cm in diameter
- Diverticula in the Colon
c. Closure of GI tract luminal perforations < 20mm that can be treated conservatively.
Type of Use (Select one or both, as applicable)X Prescription Use (Part 21 CFR 801 Subpart D)
Over-The-Counter Use (21 CFR 801 Subpart C)
### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
Image /page/3/Picture/1 description: The image shows a blue logo with the letters 'MT' in a stylized, blocky font. The letters are interconnected and have a three-dimensional appearance, with shading suggesting depth. A registered trademark symbol '®' is positioned in the upper right corner of the logo.
### 510K Summary
This 510(k) Summary of 510(k) safety and effectiveness information is being submitted in accordance with requirements of SMDA 1990 and Title 21, CFR Section 807.92. The assigned 510(k) Number: K220157
#### Date of Preparation: 2022-05-10 1.
#### 2. Sponsor Identification
Micro-Tech (Nanjing) Co., Ltd. No.10 Gaoke Third Road, Nanjing National Hi-Tech, Industrial Development Zone, Nanjing, Jiangsu Province, PRC Establishment Registration Number: 3004837686 Contact Person: Sally He Position: RA Engineer Tel: +86-25-58646395 Fax: +86-25-58350006 Email: RA.Micro-Tech@outlook.com
#### Identification of Proposed Device 3.
Product Name: Disposable Hemostatic Closure Clip Device Trade Name: / Common Name: Hemostasis Clip Regulatory Information Classification Name: Hemorrhoidal ligator Classification: 2 Product Code: PKL Regulation Number: 876.4400 Review Panel: Gastroenterology/Urology
{4}------------------------------------------------
K220157 Page 2 of 7
Image /page/4/Picture/1 description: The image shows a blue logo with the letters "MT" in a stylized, three-dimensional font. The letters are connected and appear to be made of solid blocks. A registered trademark symbol (®) is located in the upper right corner of the logo. The logo has a clean and modern design.
#### Identification of Predicate Device 4.
510(k) Number: K101428 Product Name: OTSC™ (over-the scope-clip) system set Manufacturer: Ovesco Endoscopy AG
#### 5. Indications for Use
The device for flexible endoscopy and for compression of tissue in the gastrointestinal (GI) tract, for hemostasis or for treating gastrointestinal organ wall lesions, and for making of lesions. The Device is indicated for clip placement within the GI tract for the purpose of:
- a .Endoscopic Marking
- b. Hemostasis for:
- Mucosal/Submucosal defects < 3cm
- Bleeding Ulcers
- Arteries < 2 mm
- Polyps < 1.5 cm in diameter
- Diverticula in the Colon
- c. Closure of GI tract luminal perforations < 20mm that can be treated conservatively
#### Device Description 6.
The Disposable Hemostatic Closure Clip Device is a sterile, single-use endoscopic clipping device, intended for flexible endoscopy and for compression of tissue in the gastrointestinal (GI) tract, for hemostasis or for treating gastrointestinal organ wall lesions, and for making of lesions. The device is indicated for clip placement within the GI tract for the purpose of.
- endoscopic marking a.
- hemostasis for: b.
- mucosal/submucosal defects < 3cm -
- bleeding ulcers
- arteries < 2 mm -
- polyps < 1.5 cm in diameter -
- diverticuli in the colon
- Closure of GI tract luminal perforations < 20mm that can be treated conservatively. C.
The Disposable Hemostatic Closure Clip Device consists of one pre-loaded clip and delivery device for single patient use only. The clip is made of nickel-titanium alloy with good superelasticity performance. The clip is pre-loaded in the barrel at the end of the delivery system through its deformation.
The proposed device has twenty-four (24) specifications that vary to the outer diameter of
{5}------------------------------------------------
Image /page/5/Picture/1 description: The image shows a blue logo with the letters "MT" in a stylized, three-dimensional font. The letters are bold and appear to be interconnected, creating a unified design. A registered trademark symbol is located in the upper right corner of the logo.
the clip: φ14.5 mm, φ17.0 mm; working length: 1650 mm, 1950 mm, 2350 mm, 2700 mm; and whether the outer tube is coated or not.
The proposed devices are EO sterilized to achieve the Sterility Assurance Level (SAL) of 10-6 and placed in a sterility maintenance package to ensure a shelf life of 3 years.
#### 7. Comparison of Technological Characteristics
The Disposable Hemostatic Closure Clip Device incorporates substantially equivalent device materials, design, configuration, packaging fundamental technology, sterilization process and intended use as those featured in the Predicate Device OTSC™ (Over-The-Scope-Clip) System Set (K101428).
| ITEM | Proposed Device<br>Disposable Hemostatic<br>Closure Clip Device | Predicate Device (K101428)<br>OTSCTM<br>(Over-The-Scope-Clip) System<br>Set | Remark |
|--------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------|
| Product Code | PKL | PKL | Same |
| Regulation No. | 876.4400 Hemorrhoidal<br>ligator | 876.4400 Hemorrhoidal ligator | Same |
| Class | II | II | Same |
| Intended use | The device for flexible<br>endoscopy and for<br>compression of tissue in the<br>gastrointestinal (GI) tract, for<br>hemostasis or for treating<br>gastrointestinal organ wall<br>lesions, and for making of<br>lesions. The Device is<br>indicated for clip placement<br>within the GI tract for the<br>purpose of:<br>a .Endoscopic Marking<br>b. Hemostasis for:<br>- Mucosal/Submucosal defects<br>< 3cm<br>- Bleeding Ulcers<br>- Arteries < 2 mm<br>- Polyps < 1.5 cm in diameter<br>- Diverticula in the Colon<br>c. Closure of GI tract luminal | The OTSC system set is<br>indicated for use in flexible<br>endoscopy and for the<br>compression of tissue in the<br>gastrointestinal tract, for<br>haemostasis or for treating<br>lesions of the wall of<br>gastrointestinal organs.<br>Marking of lesions.<br>The OTSC clip is indicated for<br>clip placement within the<br>gastrointestinal (GI) tract for<br>the purpose of:<br>- Endoscopic marking -<br>Hemostasis for:<br>- Mucosal/submucosal defects<br>< 3 cm<br>- Bleeding ulcers<br>- Arteries < 2 mm<br>- Polyps < 1.5 cm in diameter | Same |
| ITEM | Proposed Device<br>Disposable Hemostatic<br>Closure Clip Device<br>perforations < 20mm that can<br>be treated conservatively | Predicate Device (K101428)<br>OTSCTM<br>(Over-The-Scope-Clip) System<br>Set<br>- Diverticula in the colon<br>- Closure of GI tract luminal<br>perforations < 20 mm that can<br>be treated conservatively | Remark |
| Single Use | Yes | Yes | Same |
| Supplied in<br>Sterile | Yes | Yes | Same |
| Configuration | Clip, delivery system | Clip, Delivery System | Same |
| Clip (Material) | Niti | Niti | Same |
| Clip Width (mm) | φ14.5, φ17 | 9, 10, 11 | The clip of<br>proposed<br>device is<br>circular in<br>shape, while<br>the clip of<br>predicate<br>device is<br>rectangular<br>in shape |
| Scope<br>Compatibility<br>(mm) | φ 9.5---φ14 | φ 9.5---φ14 | Same |
| Working Length<br>(mm) | 1650, 1950, 2350, 2700 | 1650, 2200 | Similar |
| Principles of<br>Operation | Use negative pressure to<br>attract tissue, operate handle<br>and release clip | Use negative pressure to attract<br>tissue, operate handle and<br>release clip | Same |
| Biocompatibility | Comply with ISO10993-1 | Comply with ISO10993-1 | Same |
| Packaging | Single-use EO sterilized<br>pouch with one device per<br>pouch | Single-use EO sterilized pouch<br>with one device per pouch | Same |
| MRI | Comply with ASTM F 2503,<br>ASTM F 2052, ASTM F2119,<br>ASTM F2182, ASTM F2213 | Comply with ASTM F 2503,<br>ASTM F 2052, ASTM F2119,<br>ASTM F2182, ASTM F2213 | Same |
| Sterilization | EO Sterilized, SAL: 10-6 | EO Sterilized, SAL: 10-6 | Same |
| Shelf Life | 3 years | 3 years | Same |
| ITEM | Proposed Device<br>Disposable Hemostatic<br>Closure Clip Device | Predicate Device (K101428)<br>OTSCTM<br>(Over-The-Scope-Clip) System<br>Set | Remark |
| Labeling | Conforms to 21 CFR part 801 | Conforms to 21 CFR part 801 | Same |
Comparison to Predicate Devices:
{6}------------------------------------------------
Image /page/6/Picture/1 description: The image shows a blue logo with the letters "MT" in a stylized, three-dimensional font. The letters are bold and appear to be interconnected, creating a sense of unity and strength. A small registered trademark symbol is visible in the upper right corner of the logo.
{7}------------------------------------------------
Image /page/7/Picture/1 description: The image shows a blue logo with the letters 'MT' in a stylized, three-dimensional font. The letters are connected and appear to be angled slightly, giving them a dynamic look. A registered trademark symbol '®' is positioned in the upper right corner of the logo.
The proposed device is similar in design to Predicate Device OTSC™ (Over-The-Scope-Clip) System Set (K101428), the dimensions of proposed device are different to the predicate device. All comparative non-clinical performance testings have been tested and have met the requirements of substantial equivalence to the predicate device.
Therefore, the difference between proposed device and predicated device is considered not to affect the substantially equivalency between the proposed and predicate devices concerning the safety and effectiveness.
#### Performance Data 8.
The biocompatibility evaluation for the Disposable Hemostatic Closure Clip Device was conducted in accordance with FDA Guidance, Use of International Standard ISO-10993-1, 'Biological Evaluation of Medical Devices Part 1: Evaluation and Testing within a risk management process" issued on September 4, 2020. The following tests were conducted for the clip of Disposable Hemostatic Closure Clip Device:
- Cytotoxicity Study 1)
- ISO Guinea Pig Maximization Sensitization Test 2)
- 3) ISO Intracutaneous Study in Rabbits
- 4) Acute Systemic Toxicity
- 5) Material Mediated Pyrogenicity in Rabbits
- Hemolysis 6)
- 7) ISO Muscle Implantation Study, 1 Weeks
- ISO Muscle Implantation Study, 4 Weeks 8)
- 9) ISO Muscle Implantation Study, 13 Weeks
The following tests were conducted for the delivery system of Disposable Hemostatic Closure Clip Device:
- Cytotoxicity Study 1)
- 2) ISO Guinea Pig Maximization Sensitization Test
- ISO Intracutaneous Study in Rabbits 3)
- ISO Acute Systemic Toxicity Study 4)
- ર) USP Rabbit Pyrogen Study
Sterilization validation of Disposable Hemostatic Closure Clip Device was carried out in
{8}------------------------------------------------
Image /page/8/Picture/1 description: The image shows a blue logo with the letters 'M' and 'T' stacked on top of each other. The 'M' is on the left and the 'T' is on the right, and they are both in a bold, sans-serif font. The logo has a 3D effect, with the letters appearing to be slightly raised. There is a registered trademark symbol in the upper right corner of the image.
accordance with ISO 11135 Second edition 2014-07-15 Sterilization of health-care products -Ethylene oxide - Requirements for the development, validation and routine control of a sterilization process for medical devices [Including: Amendment 1 (2018)]
Shelf-life testing and packaging integrity testing of Disposable Hemostatic Closure Clip Device was conducted based on an accelerated aging test in accordance with ASTM F1980-16 Standard Guide for Accelerated Aging of Sterile Barrier Systems for Medical Devices and ISO 11607-1:2019: Packaging for terminally sterilized medical devices - Part 1: Requirements for materials, sterile barrier systems and packaging systems and ISO 11607-2:2019: Packaging for terminally sterilized medical devices - Part 2: Validation requirements for forming, sealing and assembly processes. Three-year accelerated aging test have been performed to demonstrate the stability during the shelf life.
MR compatibility was evaluated in accordance with ASTM F 2503. ASTM F 2052. ASTM F2119, ASTM F2182, ASTM F2213 and FDA guidance "Establishing Safety and Compatibility of Passive Implants in the Magnetic Resonance (MR) Environment," issued on May 20, 2021.
During design verification, the following bench tests were performed on Disposable Hemostatic Closure Clip Device:
- A Dimension
- A Release Performance
- > Clip Recovery Performance
- > Closure Force of Clip
- > Connection Force between Cable and Handle
- > Connection Force between Barrel and Soft Boot
- > Endoscope Adaptability
- > Deployment and Clip Persistence
The results of all the performance testings demonstrated that the proposed device met the predetermined acceptance criteria and is substantial equivalence to the predicate device.
#### Animal Test Conclusion 9.
No animal study is included in this submission.
### 10. Clinical Test Conclusion
No clinical study is included in this submission.
{9}------------------------------------------------
Image /page/9/Picture/1 description: The image shows a blue logo with the letters "MT" in a stylized, three-dimensional font. The letters are connected and appear to be leaning slightly to the right. A circled "R" symbol, indicating a registered trademark, is positioned in the upper right corner of the logo.
## 11. Substantially Equivalent (SE) Conclusion
Based on the indications for use, technological characteristics, and safety and performance testing, the Disposable Hemostatic Closure Clip Device has been shown to be appropriate for its intended use and is considered to be substantially equivalent to the currently cleared predicate device OTSC™ (Over-The-Scope-Clip) System Set (K101428).
Panel 1
/
Sort by
Ready
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.