K201771 · Anrei Medical (Hangzhou) Co., Ltd. · PKL · Oct 19, 2020 · Gastroenterology, Urology
Device Facts
Record ID
K201771
Device Name
Single Use Rotatable and Repositionable Hemoclip
Applicant
Anrei Medical (Hangzhou) Co., Ltd.
Product Code
PKL · Gastroenterology, Urology
Decision Date
Oct 19, 2020
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 876.4400
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
Single Use Rotatable and Repositionable Hemoclip is indicated for clip placement within the Gastrointestinal (GI) tract for the purpose of: 1) Endoscopic marking 2) Hemostasis for: Mucosal/sub-mucosal defects < 3cm, Bleeding ulcers. Arteries < 2mm. Polyps < 1.5cm in diameter, Diverticula in the colon, 3) As a supplementary method, closure of GI tract luminal perforations < 20mm that can be treated conservatively
Device Story
Sterile, single-use endoscopic clipping device; consists of delivery system and clip. Used by physicians in clinical settings during endoscopic procedures. Device allows clip rotation and repositioning; jaws open/close up to five times prior to final deployment. Physician operates delivery system to place clip at lesion site. Provides mechanical closure of tissue defects, aiding hemostasis or marking. Benefits include reduced bleeding risk and non-surgical closure of small perforations.
Clinical Evidence
No clinical data. Bench testing only. Performance verified via mechanical integrity, clamping strength, tensile strength, release force, rotation, and repositionability tests. Biocompatibility verified per ISO 10993 (cytotoxicity, sensitization, irritation, systemic toxicity) and pyrogen testing (USP <151>).
Technological Characteristics
Single-use endoscopic clip with delivery system. Working lengths: 1650mm, 1950mm, 2300mm, 2700mm. Minimum working channel: 2.8mm. Open widths: 9mm, 11mm, 13mm, 16mm. Sterilization: Ethylene oxide (SAL 10^-6). Biocompatibility: ISO 10993 compliant. Packaging: ASTM F1929-15 (dye penetration), ASTM F88/F88M-15 (seal strength).
Indications for Use
Indicated for patients requiring endoscopic clip placement in the GI tract for marking, hemostasis (mucosal/sub-mucosal defects <3cm, bleeding ulcers, arteries <2mm, polyps <1.5cm, colonic diverticula), or closure of GI luminal perforations <20mm treatable conservatively.
Regulatory Classification
Identification
A hemorrhoidal ligator is a device used to cut off the blood flow to hemorrhoidal tissue by means of a ligature or band placed around the hemorrhoid.
Special Controls
*Classification.* Class II (special controls). Except for a hemostatic metal clip intended for use in the gastrointestinal tract, the device is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 876.9.
{1}------------------------------------------------
Image /page/0/Picture/0 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). The logo consists of two parts: the Department of Health & Human Services seal on the left and the FDA acronym along with the full name of the agency on the right. The FDA part of the logo is in blue, with the acronym in a square and the full name written out to the right of the square.
October 19, 2020
Anrei Medical (Hangzhou) Co., Ltd % Diana Hong General Manager Mid-Link Consulting Co., Ltd P.O. Box 120-119 Shanghai. 200120 CHINA
Re: K201771
Trade/Device Name: Single Use Rotatable and Repositionable Hemoclip Regulation Number: 21 CFR 876.4400 Regulation Name: Hemorrhoidal Ligator Regulatory Class: II Product Code: PKL Dated: September 16, 2020 Received: September 17, 2020
Dear Diana Hong:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal
{2}------------------------------------------------
statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
For
Shani P. Haugen, Ph.D. Assistant Director DHT3A: Division of Renal, Gastrointestinal, Obesity and Transplant Devices OHT3: Office of GastroRenal, ObGyn, General Hospital and Urology Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health
Enclosure
{3}------------------------------------------------
## Indications for Use
510(k) Number (if known) K201771
Device Name
Single Use Rotatable and Repositionable Hemoclip
Indications for Use (Describe) Single Use Rotatable and Repositionable Hemocip is indicated for clip placement within the Gastrointestinal (GI) tract for the purpose of: 1) Endoscopic marking 2) Hemostasis for: Mucosal/sub-mucosal defects < 3cm, Bleeding ulcers. Arteries < 2mm. Polyps < 1.5cm in diameter, Diverticula in the colon,
3) As a supplementary method, closure of GI tract luminal perforations < 20mm that can be treated conservatively
|X Prescription Use (Part 21 CFR 801 Subpart D)
Over-The-Counter Use (21 CFR 801 Subpart C)
### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
Form Approved: OMB No. 0910-0120 Expiration Date: 06/30/2020 See PRA Statement below.
{4}------------------------------------------------
## Exhibit #8 510(k) Summary
This 510(k) Summary is being submitted in accordance with requirements of Title 21, CFR Section 807.92.
The assigned 510(k) Number: K201771
- 1. Date of Preparation: 09/07/2020
- 2. Sponsor Identification
## Anrei Medical (Hangzhou) Co., Ltd
No.3 Ave. 8, Hangzhou Economic Development Area, Hangzhou, People's Republic of China
Establishment Registration Number: 3006621103
Contact Person: Huibing Yang Position: RA Director Tel: +86-571-86913333 Fax: +86-571-87603502 Email: huibing.yang@aanrui.com.cn
- 3. Designated Submission Correspondent
Ms. Diana Hong (Primary Contact Person) Ms. Ying Xu (Alternative Contact Person)
Mid-Link Consulting Co., Ltd P.O. Box 120-119, Shanghai, 200120, China
Tel: +86-21-22815850, Fax: 360-925-3199 Email: info@mid-link.net
{5}------------------------------------------------
#### Identification of Proposed Device 4.
510(k) Summary
Trade Name: Single Use Rotatable and Repositionable Hemoclip Common Name: Hemorrhoidal ligator
### Regulatory Information
Classification Name: Hemorrhoidal ligator Classification: II; Product Code: PKL; Regulation Number: 21CFR 876.4400 Review Panel: Gastroenterology/Urology;
### Indication for Use Statement:
Single Use Rotatable and Repositionable Hemoclip is indicated for clip placement within the Gastrointestinal (GI) tract for the purpose of:
- 1. Endoscopic marking,
- 2. Hemostasis for: Mucosal/sub-mucosal defects <3cm,
Bleeding ulcers,
Arteries < 2mm,
Polyps < 1.5cm in diameter,
Diverticula in the colon,
3. As a supplementary method, closure of GI tract luminal perforations < 20mm that can be treated conservatively.
### Device Description
The proposed device is a sterile, single-use endoscopic clipping device. It is consisted of two main components: the delivery system and the clip. The delivery system is available in three different working length. The clip is deployed from the delivery system during use. The hemoclip jaws can be opened and closed up to five times prior to deployment, aiding in repositioning of the clip at the lesion site.
#### ર. Identification of Predicate Devices
Predicate Device 1 510(k) Number: K161463 Product Name: SureClip(TM) Repositionable Hemostasis Clip
{6}------------------------------------------------
Predicate Device 2 510(k) Number: K152001 Product Name: Sterile Repositionable Hemostasis Clipping Device
#### 6. Non-Clinical Test Conclusion
Non clinical tests were conducted to verify that the proposed device met all design specifications as was Substantially Equivalent (SE) to the predicate device. The test results demonstrated that the proposed device complies with the following standards:
- ISO 10993-5:2009 Biological evaluation of medical devices- Part 5: Test for in vitro cytotoxicity >
- > ISO 10993-7:2008 Biological evaluation of medical devices- Part 7: Ethylene Oxide Sterilization Residuals
- > ISO 10993-10:2010 Biological evaluation of medical devices- Part 10: Test for irritation and delayed-type hypersensitivity
- > ISO 10993-11: 2017 Biological evaluation of medical devices - Part 11: Tests for systemic toxicity
- > ASTM F1929-15 Standard test method for detecting seal leaks in porous medical packaging by dye penetration.
- > ASTM F88/F88M-15 standard method for seal strength of flexible barrier materials
- > USP 42 NF 37 <151> Pyrogen Test
Dimension test was performed on the proposed device and the test result demonstrated that the device can meet its design specification requirement.
Performance test was performed on both the proposed device and predicate device and the test result demonstrated that there was no significant difference between them, the test include following items
Mechanical Integrity of Clip Assembly Clamping Strength Testing Tensile Strength Testing Release Force Testing Rotation Testing Repositionability
- 7. Clinical Test Conclusion
No clinical study is included in this submission.
{7}------------------------------------------------
#### 8. Substantially Equivalent (SE) Comparison
| Item | Proposed Device | Predicate Device 1 K161463 | Predicate Device 2 K152001 |
|-------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Product Code | PKL | PKL | PKL |
| Regulation Number | 21CFR 876.4400 | 21CFR 876.4400 | 21CFR 876.4400 |
| Indication for Use | Single Use Rotatable and<br>Repositionable Hemoclip is indicated<br>for clip placement within the<br>Gastrointestinal (GI) tract for the<br>purpose of:<br>1. Endoscopic marking,<br>2. Hemostasis for:<br>Mucosal/sub-mucosal defects <3cm,<br>Bleeding ulcers,<br>Arteries < 2mm,<br>Polyps < 1.5cm in diameter,<br>Diverticula in the colon,<br>3. As a supplementary method,<br>closure of GI tract luminal perforations<br>< 20mm that can be treated<br>conservatively. | The SureClip(TM) Repositionable<br>Hemostasis Clip is indicated for<br>endoscopic clip placement within the<br>gastrointestinal tract for the purpose of:<br>(1) endoscopic marking,<br>(2) hemostasis for<br>(a) mucosal / sub-mucosal defects <<br>3cm,<br>(b) bleeding ulcers,<br>(c) polyps < 1.5cm in diameter,<br>(d) diverticula in the colon,<br>(3) as a supplementary method, closure<br>of GI tract luminal perforations <20mm<br>that can be treated conservatively | The Sterile Repositionable Hemostasis<br>Clipping Device is indicated for<br>endoscopic clip placement within the<br>gastrointestinal tract for the purpose of:<br>(1) endoscopic marking,<br>(2) hemostasis for<br>(a) mucosal / sub-mucosal defects <<br>3cm,<br>(b) bleeding ulcers,<br>(c) polyps < 1.5cm in diameter,<br>(d) diverticula in the colon,<br>(3) as a supplementary method, closure<br>of GI tract luminal perforations <20mm<br>that can be treated conservatively. |
| Configuration | Delivery system and jaw | Delivery system and jaw | Delivery system and jaw |
| Rotation function | rotatable | rotatable | rotatable |
| Open width | 9mm, 11mm, 13mm and 16mm | 8mm, 11mm and 16mm | 11mm |
| Minimal working channel | 2.8mm | 2.8mm | 2.8mm |
# Table 1 Comparison of Technology Characteristics
{8}------------------------------------------------
| Working length | 1650mm, 1950mm, 2300mm and<br>2700mm | 1650mm, 1950mm and 2350mm | 1650mm, 1950mm, 2350mm and<br>2700mm |
|-----------------------------|--------------------------------------|---------------------------------|--------------------------------------|
| Single Use | Single Use | Single Use | Single Use |
| Labeling | Comply with 21 CFR Part 801 | Comply with 21 CFR Part 801 | Comply with 21 CFR Part 801 |
| Biocompatibility | | | |
| Cytotoxicity | No cytotoxicity | Comply with ISO 10993 standards | Comply with ISO 10993 standards |
| Skin Sensitization | No skin sensitization | | |
| Irritation | No irritation | | |
| Acute Systemic Toxicity | No acute toxicity | | |
| Sub-acute Systemic Toxicity | No sub-acute toxicity | | |
| Pyrogen | No pyrogen | | |
| Sterilization | | | |
| Method | Ethylene oxide | Ethylene oxide | Ethylene oxide |
| SAL | 10-6 | 10-6 | 10-6 |
| Endotoxin Limit | 20EU | 20EU | 20EU |
| Shelf life | 3 years | 3 years | 3 years |
{9}------------------------------------------------
- 9. Substantially Equivalent (SE) Conclusion
The nonclinical tests demonstrate that the device is as safe, as effective, and performs as well as the legally marketed device predicate device.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.