Rezum System

K180237 · Nxthera, Inc. · KNS · Feb 27, 2018 · Gastroenterology, Urology

Device Facts

Record IDK180237
Device NameRezum System
ApplicantNxthera, Inc.
Product CodeKNS · Gastroenterology, Urology
Decision DateFeb 27, 2018
DecisionSESE
Submission TypeSpecial
Regulation21 CFR 876.4300
Device ClassClass 2
AttributesTherapeutic

Intended Use

The Rezūm System is intended to relieve symptoms, obstructions, and reduce prostate tissue associated with BPH. It is indicated for men ≥ 50 years of age with a prostate volume ≥ 30cm3 and ≤ 80cm3. The Rezum System is also indicated for treatment of prostate with hyperplasia of the central zone and/or a median lobe.

Device Story

Rezūm System uses convective thermal energy to treat BPH. System comprises reusable RF generator and sterile, EtO-sterilized delivery device. Physician inserts delivery device into urethra; cystoscope provides visualization. Generator converts water to vapor; vapor delivered via needle into prostate tissue to cause thermal ablation. Generator also controls saline flush to cool urethra during procedure. Modifications include electrical needle delivery/retraction mechanism, updated vapor coil material, and relocated temperature monitoring. Device intended for clinical use to reduce prostate tissue, relieve symptoms, and obstructions.

Clinical Evidence

Bench testing only. Verification testing included dimensional, tensile/bond strength, functional, fluid resistance, calorimetry, corrosion resistance, packaging/distribution, sterility, shelf-life, software V&V, hardware tests, electrical safety (IEC 60601-1), and EMC (IEC 60601-1-2).

Technological Characteristics

System includes RF generator and delivery device. Vapor ablation via thermal energy. Features: electrical needle delivery/retraction, vapor coil, temperature monitoring, saline cooling. Electrical safety per IEC 60601-1; EMC per IEC 60601-1-2. Sterilization: EtO for delivery device. Software-controlled generator.

Indications for Use

Indicated for men ≥ 50 years old with BPH, prostate volume 30-80cm3, including hyperplasia of the central zone and/or median lobe.

Regulatory Classification

Identification

An endoscopic electrosurgical unit and accessories is a device used to perform electrosurgical procedures through an endoscope. This generic type of device includes the electrosurgical generator, patient plate, electric biopsy forceps, electrode, flexible snare, electrosurgical alarm system, electrosurgical power supply unit, electrical clamp, self-opening rigid snare, flexible suction coagulator electrode, patient return wristlet, contact jelly, adaptor to the cord for transurethral surgical instruments, the electric cord for transurethral surgical instruments, and the transurethral desiccator.

Predicate Devices

Related Devices

Submission Summary (Full Text)

{0}------------------------------------------------ Image /page/0/Picture/0 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). The logo consists of two parts: the Department of Health & Human Services logo on the left and the FDA logo on the right. The FDA logo features the letters "FDA" in a blue square, followed by the words "U.S. FOOD & DRUG ADMINISTRATION" in blue text. February 27, 2018 NxThera, Inc. Kathy Simpson VP Quality and Regulatory Affairs 7351 Kirkwood Land North, Suite 138 Maple Grove, MN 55369 Re: K180237 Trade/Device Name: Rezūm System Regulation Number: 21 CFR 876.4300 Regulation Name: Endoscopic Electrosurgical Unit and Accessories Regulatory Class: II Product Code: KNS Dated: January 25, 2018 Received: February 1, 2018 Dear Kathy Simpson: We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading. If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register. Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. {1}------------------------------------------------ You must comply with all the Act's requirements. including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance. For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/MedicalDevices/DeviceRegulationandGuidance/) and CDRH Learn (http://www.fda.gov/Training/CDRHLearn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (http://www.fda.gov/DICE) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100). Sincerely, # Benjamin R. Fisher -S Benjamin R. Fisher, Ph.D. Director Division of Reproductive, Gastro-Renal, and Urological Devices Office of Device Evaluation Center for Devices and Radiological Health Enclosure {2}------------------------------------------------ ## Indications for Use 510(k) Number (if known) K180237 Device Name Rezūm System Indications for Use (Describe) The Rezūm System is intended to relieve symptoms, obstructions, and reduce prostate tissue associated with BPH. It is indicated for men ≥ 50 years of age with a prostate volume ≥ 30cm3 and ≤ 80cm3. The Rezum System is also indicated for treatment of prostate with hyperplasia of the central zone and/or a median lobe. | Type of Use (Select one or both, as applicable) | | |------------------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------| | <div><span style="font-size:100%;">☒</span> Prescription Use (Part 21 CFR 801 Subpart D)</div> | <div><span style="font-size:100%;">☐</span> Over-The-Counter Use (21 CFR 801 Subpart C)</div> | CONTINUE ON A SEPARATE PAGE IF NEEDED. This section applies only to requirements of the Paperwork Reduction Act of 1995. ## *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.* The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to: > Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov "An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number." {3}------------------------------------------------ # 510(k) Summary #### l. SUBMITTER NxThera, Inc. 7351 Kirkwood Lane North Suite 138 Maple Grove, MN 55369 Phone: (763) 515-0404 Fax: (763) 515-2085 | Contact Person: | Kathy Simpson | |-----------------|------------------| | Date Prepared: | January 18, 2018 | #### ll. DEVICE | Name of Device: | Rezūm® System | |-----------------------|-----------------------------------------------------------------------| | Common or Usual Name: | Vapor Ablation System | | Classification Name: | Endoscopic electrosurgical unit and accessories<br>(21 CFR §876.4300) | | Regulatory Class: | II | | Product Code: | KNS | #### lll. PREDICATE DEVICE Rezūm System, K160417 #### IV. DEVICE DESCRIPTION The Rezum System consists of a reusable Generator and a sterile Delivery Device Kit consisting of one Delivery Device with cable and tubing, one syringe, one spike adaptor, and one 50mL sterile water vial. The Delivery Device is EtO sterilized. The Rezūm System converts water into vapor outside of the body with the vapor delivered to the prostate tissue via a needle within the sterile Delivery Device. The vapor ablates the targeted tissue within the prostate via thermal ablation as energy is transferred from the vapor to the prostate tissue. The amount of vapor delivered is controlled by a Radiofrequency (RF) Generator which also control the amount of saline flush used to cool the urethra during the treatment. {4}------------------------------------------------ Image /page/4/Picture/1 description: The image shows the word "NxTHERA." in a bold, sans-serif font. The letters are black, and there is a small, stylized graphic of a wave or swoosh in light blue underneath the "x" and "T" in the word. The background is white. #### INDICATIONS FOR USE V. The Rezūm System is intended to relieve symptoms, obstructions, and reduce prostate tissue associated with BPH. It is indicated for men ≥ 50 years of age with a prostate volume ≥ 30cm³ and ≤ 80cm³. The Rezūm System is also indicated for treatment of prostate with hyperplasia of the central zone and/or a median lobe. ### VI. COMPARISON OF TECHNOLOGICAL CHARACTERISTICS WITH THE PREDICATE DEVICE At a high level, the subject and predicate devices are based on the following technological elements that remain equivalent between the two devices: - . Cystoscope used within the Delivery Device to visualize the target treatment area - Device is inserted into the urethra to reach the targeted prostate tissue ● - Generation of water vapor via a specific caloric / energy requirement - Thermal ablation of the prostate tissue ● - Use of saline flush to cool the urethra during treatment . The following technological differences exist between the subject and predicate devices: - . Design enhancements of the user interface: - Reduced device handle envelope distance, improve handle angle o and improve fit and feel - Changed needle delivery and retraction mechanism from o mechanical to electrical - . Design changes to enhance manufacturability and reliability of the Delivery Device: - Change vapor coil material to a more efficient material o - Relocate temperature monitoring on vapor coil O - Group switching functions on one printed circuit board O - Cable changes including a plastic non-locking connector to O improve manufacturability - Replace prior adhesive with equivalent adhesive O - Removal of the device rotation capability O - . Design changes to accommodate device updates, manage component obsolescence, enhance manufacturability and reliability of the Generator: - Update circuit boards and software for new electrical needle o delivery and retraction mechanism, to manage obsolescence and improve manufacturability - о Update RF power supply to accommodate new vapor coil material {5}------------------------------------------------ Image /page/5/Picture/1 description: The image shows the word "NxTHERA." in a bold, sans-serif font. The "x" in "NxTHERA" is stylized with a swoosh-like design underneath it. The word is followed by a period. - Move ESD protection from Device to Generator o - Update wiring and connectors for manufacturability and servicing O #### VII. PERFORMANCE DATA The modifications made to the subject Rezūm System have been tested to ensure compliance to the initial device specifications. Based on the change assessment, the following design verification tests were repeated on the C2 Delivery Device. The test methods used were the same as those submitted for the predicate device: - Dimensional - Tensile / bond strength tests ● - Full Functional tests ● - Fluid resistance ● - Calorimetry tests ● - Corrosion resistance ● - Packaging and distribution testing - Sterility validation - Shelf-life aging ● - Software verification and validation ● - Hardware tests ● - Electrical safety testing (per IEC 60601-1) - Electromagnetic compatibility testing (per IEC 60601-1-2) ● The conclusion of the performance assessments demonstrate that the device continues to function as intended in a manner equivalent to the predicate device, and that there are no new issues of safety or effectiveness with the implementation of the above listed modifications. #### VIII. CONCLUSION Based on the test data and other characteristics of the subject device as compared to the predicate, the modified Rezūm System is substantially equivalent to its predicate Rezūm System.
Innolitics

Panel 1

/
Sort by
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...