K150786 · Nxthera, Inc. · KNS · Aug 27, 2015 · Gastroenterology, Urology
Device Facts
Record ID
K150786
Device Name
Rezum System
Applicant
Nxthera, Inc.
Product Code
KNS · Gastroenterology, Urology
Decision Date
Aug 27, 2015
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 876.4300
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
The Rezum System is intended to relieve symptoms, obstructions, and reduce prostate tissue associated with BPH. It is indicated for men ≥ 50 years of age with a prostate volume ≥ 30cm³ and ≤ 80cm³. The Rezūm System is also indicated for treatment of prostate with hyperplasia of the central zone and/or a median lobe.
Device Story
Rezūm System uses convective thermal energy to treat BPH; converts water to vapor outside body; delivers vapor via needle in sterile delivery device to prostate tissue; vapor ablates targeted tissue via thermal energy transfer. RF generator controls vapor delivery and saline flush for urethral cooling. Used in clinical setting by physicians. Output is ablated prostate tissue; reduces symptoms and obstructions. Benefits include relief of BPH symptoms and reduced prostate volume.
Clinical Evidence
Three studies: feasibility (n=15), pilot (n=50), and randomized placebo-controlled pivotal study (n=197). Pivotal study (136 treatment, 61 control) showed device safe (p<0.0001) and superior to control at 3 months (p<0.001). 75% responder rate at 6 months (p<0.0001) and 77% at 12 months (p<0.0001) based on IPSS improvement. No unanticipated adverse events reported.
Technological Characteristics
System includes reusable RF generator and sterile disposable delivery device. Uses convective water vapor for thermal ablation. Biocompatibility per ISO 10993-1. Sterilization via ethylene oxide (SAL 10⁻⁶).
Indications for Use
Indicated for men ≥ 50 years old with BPH, prostate volume 30-80cm³, including hyperplasia of the central zone and/or median lobe.
Regulatory Classification
Identification
An endoscopic electrosurgical unit and accessories is a device used to perform electrosurgical procedures through an endoscope. This generic type of device includes the electrosurgical generator, patient plate, electric biopsy forceps, electrode, flexible snare, electrosurgical alarm system, electrosurgical power supply unit, electrical clamp, self-opening rigid snare, flexible suction coagulator electrode, patient return wristlet, contact jelly, adaptor to the cord for transurethral surgical instruments, the electric cord for transurethral surgical instruments, and the transurethral desiccator.
{1}------------------------------------------------
Image /page/0/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a stylized image of an eagle with three human profiles incorporated into its design. The words "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" are arranged in a circular pattern around the eagle.
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002
September 12, 2016
NxThera, Inc. % Sew-Wah Tay. Ph.D. Regulatory Consultant Libra Medical, Inc. 8401 73rd Avenue North, Suite 63 Brooklyn Park, MN 55428
Re: K150786 Trade/Device Name: Rezūm System Regulation Number: 21 CFR§ 876.4300 Regulation Name: Endoscopic electrosurqical unit and accessories Regulatory Class: II Product Code: KNS Dated: August 21, 2015 Received: August 24, 2015
Dear Sew-Wah Tay:
This letter corrects our substantially equivalent letter of August 27, 2015.
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent for the indications for use stated in the enclosure to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{2}------------------------------------------------
Page 2 - Sew-Wah Tay, Ph.D.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies.
You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If vou desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled. "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerely,
For Division
# Douglas Silverstein -S 2016.09.12 16:11:58 -04'00'
Benjamin Fisher, Ph.D. Director Division of Reproductive, Gastro-Renal, and Urological Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{3}------------------------------------------------
| DEPARTMENT OF HEALTH AND HUMAN SERVICES<br>Food and Drug Administration | Form Approved: OMB No. 0910-0120<br>Expiration Date: January 31, 2017<br>See PRA Statement below. |
|-------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------|
|-------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------|
| 510(k) Number ( <i>if known</i> ) | K150786 |
|-----------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Device Name: | Rezūm System |
| Indications for Use ( <i>describe</i> ) | The Rezūm System is intended to relieve symptoms, obstructions, and reduce prostate tissue associated with BPH. It is indicated for men ≥ 50 years of age with a prostate volume ≥ 30cm³ and ≤ 80cm³. The Rezūm System is also indicated for treatment of prostate with hyperplasia of the central zone and/or a median lobe. |
| Type of Use ( <i>select one or both, as applicable</i> ) | |
|--------------------------------------------------------------------------------|--------------------------------------------------|
| <div> <b>X</b> Prescription Use (Part 21 CFR 801 Subpart D) </div> | Over-The-Counter Use (Part 21 CFR 801 Subpart C) |
CONTINUE ON A SEPARATE PAGE IF NEEDED
This section applies only to requirements of the Paperwork Reduction Act of 1995
*DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden to:
Department of Health and Human Services
Food and Drug Administration
Office of Chief Information Officer
Paperwork Reduction Act (PRA) Staff
PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{4}------------------------------------------------
# 510(K) SUMMARY (K150786) 1
#### ADMINISTRATIVE INFORMATION 1.1
Date of Summary Preparation: June 12 2015
# 1.2 CONTACT INFORMATION
| Primary Submission Contact | Sew-Wah Tay, PhD<br>Regulatory Consultant,<br>Libra Medical Inc. |
|----------------------------|------------------------------------------------------------------|
|----------------------------|------------------------------------------------------------------|
| Secondary Submission Contact | Julie Bodmer |
|------------------------------|------------------------|
| | Regulatory Consultant, |
| | Libra Medical Inc. |
#### 1.3 DEVICE INFORMATION
| Trade Name | Rezūm System |
|---------------------------|-------------------------------------------------|
| Common Name | Vapor BPH Ablation Device |
| Classification Name | Endoscopic electrosurgical unit and accessories |
| Classification Regulation | 876.4300 |
| Class | II |
| Panel | Gastroenterology/Urology |
| Product Code | KNS |
# 1.4 510(K) TYPE AND REASON FOR SUBMISSION
This 510(k) is a traditional 510(k) and is submitted to obtain marketing clearance for a new device - the Rezūm System.
# 1.5 PREDICATE DEVICE
The Rezūm System is substantially equivalent to the Medtronic Prostiva devices (K113380 and K142248).
#### 1.6 DEVICE DESCRIPTION
The reusable Rezūm Generator is provided with the following reusable components:
- Generator ●
- One Power Cord
The Rezūm Delivery Device Kit contains the following disposable components:
- 0 One sterile Delivery Device with cable and tubing
{5}------------------------------------------------
- One sterile Syringe
- One sterile Spike Adaptor ●
- One 50 ml Sterile Water Vial ●
Additional spike adaptor and syringe accessory is provided as an Accessory Pack.
# INDICATIONS FOR USE / INTENDED USE 1.7
The Rezum System is intended to relieve symptoms, obstructions, and reduce prostate tissue associated with BPH. It is indicated for men ≥ 50 years of age with a prostate volume ≥ 30cm3 and ≤ 80 cm³. The Rezūm System is also indicated for treatment of prostate with hyperplasia of the central zone and/or a median lobe.
# 1.8 TECHNOLOGICAL CHARACTERISTICS
The device converts water into vapor outside of the body and the vapor is delivered to the prostate tissue via a needle within the sterile Delivery Device. The vapor ablates the targeted tissue within the prostate via thermal ablation as energy is transferred from the vapor to the prostate tissue.
The amount of vapor delivered is controlled by an RF Generator which also controls the amount of saline flush used to cool the urethra.
# 1.9 PERFORMANCE DATA
The Rezūm System has been tested to meet the FDA 2010 BPH guidance where applicable for preclinical and clinical testing.
The Rezum System has been tested and meets all its physical and performance specifications on the bench including:
- Dimensions ●
- Tensile strength tests
- Full functional tests
- Calories tests
- Hardware tests ●
- Software verification and validation
- Packaging tests
- Distribution tests
In addition, the Delivery Device was tested for biocompatibility per ISO 10993-1 for short duration contact with tissue and mucosal membrane (<24 hours). The device is sterilized by ethylene oxide to an SAL 10° level. These performances are similar to that described by the predicate device. The testing showed that the device meets specifications before and after aging indicating that the device is as safe and effective as the predicate device.
{6}------------------------------------------------
The device has also been tested in 3 clinical studies to evaluate the safety and effectiveness of the Rezum device: 65 patients in the feasibility and pilot open label studies and in a 197 patient randomized placebo controlled study. All these studies showed that the device is safe and effective.
# 1.10 SUMMARY OF CLINICAL DATA
#### REZŪM FIM OPTIMIZATION STUDY 1.10.1
Fifteen subjects were enrolled in one site in this study to determine the optimum settings and vapor dosage for the Rezum System. Subjects were evaluated for outcomes, adverse events and prostate and lesion sizes at 1 week, 1 month, 3 months and 6 months.
Fifteen subjects were enrolled and treated in this study with 14 subjects having 6 months follow up data. All procedures were completed successfully and there were no unanticipated adverse event. The outcomes for the subjects were equivalent for the full range of energy tested although the higher energy group exhibited more acute AE. Site reported AE early in the study and was typical for thermal ablation of the prostate. Rapid IPSS improvement was observed at 1 month and continues to be improving through 6 months and 12 months.
### REZŪM I PILOT STUDY 1.10.2
This is an open label, single arm study involving 3 centers, two in Europe and one from Latin America. A total of 50 patients were enrolled in this study, 45 of whom were enrolled in the two European sites. Only one energy level was used in this study. The IPSS improvement is consistent with the results from Rezūm FIM Optimization Study and improvements in IPSS persist through the 2 year follow up. At month 3, 6, 12 and 24 months, more than 80% of the subjects treated were responders for all the time period. There were no unanticipated adverse events with most of the reported adverse event related to LUTS symptoms and occurred in the acute healing phase.
# REZŪM II PIVOTAL STUDY 1.10.3
This is a blinded, 2:1 randomized placebo control study. 197 subjects were enrolled in this study. 136 subjects were randomized to the treatment arm and 61 were assigned to the control arm. 83.6% of the control subjects crossed over to the treatment arm. The rest of the subjects either did not qualify at the time of the crossover or exited the study.
The primary objective is to establish the safety and effectiveness of the device. On an intention to treat basis, the results showed that the device is safe (p<0.0001). The effectiveness as measured by IPSS. At 3 months, the effectiveness of the Rezum System was superior to the control group (p<0.001). The long term effectiveness was measured at 6 and 12 months by comparing the IPSS to the subjects' baseline. At 6 months, 75% of the subjects were responders (p<0.0001) and at 12 months, 77% of the subjects were responders (p<0.0001).
There were no unanticipated adverse events.
{7}------------------------------------------------
# SUBSTANTIAL EQUIVALENCE 1.11
The Rezūm System is substantially equivalent to the Prostiva Device (K113380/K142248). It has the same intended use for thermal ablation of BPH tissue. The test and clinical data showed that the technological difference between the Rezūm and its predicate do not raise safety and efficacy issues. The Rezum II randomized placebo controlled clinical trial data showed that the device is effective in relieving the symptoms of BPH and does not raise new questions of safety.
# 1.12 CONCLUSION
Based on the test data and the same intended use, the Rezūm System is found to be substantially equivalent to its predicate.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.