The Flexima™ Ureteral Stent System and Flexima™ Ureteral Stent System Kit is intended to provide drainage from the ureteropelvic junction to the bladder and stenting of the ureter for all patients in whom it is desirable to place a drain which does not extend externally. The Flexima™ Biliary Catheter System / Flexima™ Biliary Catheter System Kit / Flexima™ Biliary Catheter System with Radiopaque Marker / Flexima™ Biliary Catheter System with Hydrophilic Dissolving Tip / Flexima™ Biliary Catheter System with Hydrophilic Dissolving Tip and Radiopaque Marker are intended to provide external and internal percutaneous drainage of the biliary system.
Device Story
Flexima Biliary Catheters and Ureteral Stents are passive drainage devices. Biliary catheters provide percutaneous biliary drainage; ureteral stents provide internal drainage from ureteropelvic junction to bladder. Devices are constructed of Tecoflex polyurethane-based resin with Glidex hydrophilic coating. Selected biliary models include 65% tungsten radiopaque markers or Temp-Tip dissolving distal tips to facilitate guidewire tracking. Devices are placed by physicians in clinical settings; indwelling time limited to 90 days. Output is physical drainage of fluids. This submission updates labeling with MRI safety information based on non-clinical testing (RF heating, displacement force, image artifacts).
Clinical Evidence
No clinical data. Bench testing only: RF heating (ASTM F2182), displacement force (ASTM F2052), and image artifact evaluation (ASTM F2119) performed to support MRI safety labeling.
Technological Characteristics
Materials: Tecoflex polyurethane-based resin, 65% tungsten (radiopaque markers). Coating: Glidex hydrophilic. Sterilization: 100% ethylene oxide (EO). Form factor: Catheters (8F-14F) and stents (6F-10F) with pigtail configurations. Connectivity: None. Software: None.
Indications for Use
Indicated for patients requiring drainage from the ureteropelvic junction to the bladder (ureteral stents) or external/internal percutaneous drainage of the biliary system (biliary catheters).
Regulatory Classification
Identification
A ureteral stent is a tube-like implanted device that is inserted into the ureter to provide ureteral rigidity and allow the passage of urine. The device may have finger-like protrusions or hooked ends to keep the tube in place. It is used in the treatment of ureteral injuries and ureteral obstruction.
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which is a blue square with the letters "FDA" in white. To the right of the blue square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue.
January 2, 2020
Boston Scientific Corporation Jois Zuniga-Gamboa Peripherical Interventional Regulatory Affairs Specialist Three Scimed Place Maple Grove, MN 55311
Re: K191446
Trade/Device Name: Flexima™ Biliary Catheter System Flexima™ Biliary Catheter System Kit Flexima™ Biliary Catheter System with Radiopaque Marker Flexima™ Biliary Catheter System with Hydrophilic Dissolving Tip FleximaTM Biliary Catheter System with Hydrophilic Dissolving Tip and Radiopaque Marker. Flexima™ Ureteral Stent System and Flexima™ Ureteral Stent System Kit Regulation Number: 21 CFR 876.4620 Regulation Name: Ureteral Stent Regulatory Class: Class II Product Code: FAD, FGE Dated: November 26, 2019 Received: December 4, 2019
Dear Jois Zuniga-Gamboa:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. However, you are responsible to determine that the medical devices you use as components in the kit have either been determined as substantially equivalent under the premarket notification process (Section 510(k) of the act), or were legally on the market prior to May 28, 1976, the
{1}------------------------------------------------
enactment date of the Medical Device Amendments. Please note: If you purchase your device components in bulk (i.e., unfinished) and further process (e.g., sterilize) you must submit a new 510(k) before including these components in your kit/tray. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn).
{2}------------------------------------------------
Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-devices/deviceadvice-comprehensive-regulatory-assistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely.
Jessica K. Nguyen, Ph.D. Acting Assistant Director DHT3B: Division of Reproductive, Gynecology and Urology Devices OHT3: Office of GastroRenal, ObGyn, General Hospital and Urology Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health
Enclosure
{3}------------------------------------------------
# Indications for Use
510(k) Number (if known) K191446
#### Device Name
Flexima™ Ureteral Stent System and Flexima™ Ureteral Stent System Kit. Flexima™ Biliary Catheter System Kit / Flexima™ Biliary Catheter System with Radiopaque Marker / Flexima™ Biliary Catheter System with Hydrophilic Dissolving Tip / Flexima™ Biliary Catheter System with Hydrophilic Dissolving Tip and Radiopaque Marker.
Indications for Use (Describe)
The Flexima™ Ureteral Stent System and Flexima™ Ureteral Stent System Kit is intended to provide drainage from the ureteropelvic junction to the bladder and stenting of the ureter for all patients in whom it is desirable to place a drain which does not extend externally.
The Flexima™ Biliary Catheter System / Flexima™ Biliary Catheter System Kit / Flexima™ Biliary Catheter System with Radiopaque Marker / Flexima™ Biliary Catheter System with Hydrophilic Dissolving Tip / Flexima™ Biliary Catheter System with Hydrophilic Dissolving Tip and Radiopaque Marker are intended to provide external and internal percutaneous drainage of the biliary system.
| Type of Use (Select one or both, as applicable) | |
|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| <span style="font-family: Arial, sans-serif;"> <span style="font-size: 11pt;"> <span style="color: black;"> <span style="font-weight: normal;"> <span style="font-style: normal;"> <span style="text-decoration: none;"> <span style="vertical-align: baseline;"> <span style="white-space: pre-wrap;"> <input checked="" type="checkbox"/> Prescription Use (Part 21 CFR 801 Subpart D) </span> </span> </span> </span> </span> </span> </span> </span> | <span style="font-family: Arial, sans-serif;"> <span style="font-size: 11pt;"> <span style="color: black;"> <span style="font-weight: normal;"> <span style="font-style: normal;"> <span style="text-decoration: none;"> <span style="vertical-align: baseline;"> <span style="white-space: pre-wrap;"> <input type="checkbox"/> Over-The-Counter Use (21 CFR 801 Subpart C) </span> </span> </span> </span> </span> </span> </span> </span> |
#### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{4}------------------------------------------------
# 510(k) Summary
510(k) Summary Complying with 21 CFR 807.92
#### I. SUBMITTER INFORMATION
Submitter name: Boston Scientific Corporation
Submitter address: Three Scimed Place Maple Grove, MN 55311-1566 USA
Telephone: 763-494-2211 Fax: 763-494-2222 e-mail: JoisDayana.ZunigaGamboa@bsci.com
Contact person name: Jois Zuniga G. Date Prepared: December 11, 2019
#### II. DEVICE INFORMATION
Table 1, 2 and 3 as follows summarizes the relevant device information for the subject devices.
| Device Trade Name | Model Number |
|--------------------------------------|--------------|
| Flexima™ Biliary Catheter System | M001271540 |
| | M001271550 |
| | M001271560 |
| | M001271570 |
| | M001271580 |
| | M001271590 |
| | M001271600 |
| | M001271610 |
| | M001271620 |
| Flexima™ Biliary Catheter System Kit | M001271630 |
| | M001271640 |
| | M001271650 |
| | M001271660 |
#### Table 1. Flexima™ Biliary Catheters Name of Devices
{5}------------------------------------------------
| Device Trade Name | Model Number |
|-------------------------------------------------------------------------------------------|--------------|
| | M001272600 |
| | M001272610 |
| | M001272620 |
| Flexima™ Biliary Catheter System with Radiopaque Marker | M001272630 |
| | M001272640 |
| | M001272650 |
| | M001272660 |
| | M001272670 |
| | M001281560 |
| Flexima™ Biliary Catheter System with Hydrophilic Dissolving Tip | M001281570 |
| | M001281580 |
| | M001281590 |
| | M001282610 |
| Flexima™ Biliary Catheter System with Hydrophilic Dissolving Tip and<br>Radiopaque Marker | M001282600 |
# Table 2, Elexima™ Ureteral Stents Name of Devices
| Table 2. Flexima TM Ureteral Stents Name of Devices<br>Device Trade Name | Model Number |
|--------------------------------------------------------------------------------------|--------------|
| FleximaTM Ureteral Stent System | M001274000 |
| | M001274010 |
| | M001274020 |
| | M001274030 |
| | M001274040 |
| | M001274050 |
| | M001274060 |
| | M001274070 |
| | M001274080 |
| | M001274090 |
| M001274100 | |
| M001274110 | |
| M001274120 | |
| M001274130 | |
| FleximaTM Ureteral Stent System Kit | M001274140 |
| | M001274150 |
| | M001274160 |
| | M001274170 |
| | M001274180 |
| | M001274190 |
{6}------------------------------------------------
| Common or<br>Usual Name | Classification<br>Number | Classification<br>Name | Product<br>Code | Product<br>Code Name | Regulatory<br>Class |
|-------------------------------|--------------------------|----------------------------------------|-----------------|-------------------------------------------------------------|---------------------|
| Biliary Drainage<br>Catheters | 21 CFR Part<br>876.5010 | Biliary catheter<br>and<br>accessories | FGE | stents, drains<br>and dilators<br>for the biliary<br>ducts. | II |
| Ureteral Stent<br>Systems | 21 CFR Part<br>876.4620 | Ureteral stent | FAD | stent, ureteral | II |
#### III. PREDICATE DEVICE IDENTIFICATION
#### Name of Predicate Device
Flexima Biliary Catheter K023870 (Predicate for Flexima Biliary Catheters)
Flexima Drainage Catheters and Stents K944290 (Predicate for Flexima Ureteral Stents)
Predicate devices referenced above have not been subject to a design-related recall.
#### IV. DEVICE DESCRIPTION
The Flexima Biliary Catheters are constructed of Flexima, a Tecoflex Polyeurethane-based resin, and selected subfamilies also have a radiopaque marker (RO) made from 65%Tungsten. The radiopaque marker is located approximately 5mm proximal to the most proximal drainage hole to aid in proper placement of the catheter. Additionally, designated Flexima Biliary Catheters are available with a biocompatible, dissolving distal tip (Temp-Tip™). This dissolvable tip facilities tracking over a guidewire for percutaneous placement and dissolves within 24 hours of placement to provide a larger drainage lumen. These biliary catheters provide external and internal percutaneous drainage of the biliary system. Where long-term use is indicated, it is recommended that indwelling time not exceed 90 days.
The Flexima Ureteral Stent is also constructed of Tecoflex Polyeurethane-based resin. The pigtails on either end of the stent, are in opposite directions so that the proximal pigtail can form in the renal pelvis, while the distal pigtail forms in the bladder, to provide drainage from the ureteropelvic junction to the bladder and stenting of the ureter for all patients in whom it is desirable to place a drain which does not extend externally. The double pigtail stent prevents upward and downward movement. Also, where longterm use is indicated, this stent is recommended not to exceed 90 days of indwelling time.
The Flexima Biliary Catheters and Flexima Ureteral Stents are provided sterile, using 100% ethylene oxide (EO) gas sterilization, and are intended for hospital and single use only. These devices utilize Glidex™ Hydrophillic Coating for the reduction of surface friction during placement.
{7}------------------------------------------------
Table 4 and Table 5 below provide information on the differences between the models, in relation to the outside diameter and length dimensions of the Flexima Biliary Catheters and Flexima Ureteral Stents respectively.
| Product Trade Name | Model Number | Dimensions<br>(OD/Length) | | | | | | | | | | | | | | | | | | | | | | | | | | | | |
|-------------------------------------------------------------------------------------------|---------------------------------------------------------|---------------------------|------------|----------|------------|----------|------------|----------|------------|---------|------------|----------|------------|----------|------------|----------|---------------------------------------------------------------------|------------|---------|------------|----------|------------|----------|------------|----------|-------------------------------------------------------------------------------------------|------------|---------|------------|----------|
| Flexima™ Biliary Catheter System | M001271540 | 8F/35cm | | | | | | | | | | | | | | | | | | | | | | | | | | | | |
| | M001271550 | 10F/35cm | | | | | | | | | | | | | | | | | | | | | | | | | | | | |
| | M001271560 | 8F/35cm | | | | | | | | | | | | | | | | | | | | | | | | | | | | |
| | M001271570 | 10F/35cm | | | | | | | | | | | | | | | | | | | | | | | | | | | | |
| | M001271580 | 12F/35cm | | | | | | | | | | | | | | | | | | | | | | | | | | | | |
| | M001271590 | 14F/35cm | | | | | | | | | | | | | | | | | | | | | | | | | | | | |
| | M001271600 | 10F/35cm | | | | | | | | | | | | | | | | | | | | | | | | | | | | |
| | M001271610 | 12F/35cm | | | | | | | | | | | | | | | | | | | | | | | | | | | | |
| | M001271620 | 14F/35cm | | | | | | | | | | | | | | | | | | | | | | | | | | | | |
| Flexima™ Biliary Catheter System Kit | M001271630 | 8F/35cm | | | | | | | | | | | | | | | | | | | | | | | | | | | | |
| | M001271640 | 10F/35cm | | | | | | | | | | | | | | | | | | | | | | | | | | | | |
| | M001271650 | 8F/35cm | | | | | | | | | | | | | | | | | | | | | | | | | | | | |
| | M001271660 | 10F/35cm | | | | | | | | | | | | | | | | | | | | | | | | | | | | |
| Flexima™ Biliary Catheter System with Radiopaque Marker | M001272600 | 8F/35cm | M001272610 | 10F/35cm | M001272620 | 12F/35cm | M001272630 | 14F/35cm | M001272640 | 8F/35cm | M001272650 | 10F/35cm | M001272660 | 10F/35cm | M001272670 | 12F/35cm | Flexima™ Biliary Catheter System with Hydrophilic Dissolving<br>Tip | M001281560 | 8F/35cm | M001281570 | 10F/35cm | M001281580 | 12F/35cm | M001281590 | 14F/35cm | Flexima™ Biliary Catheter System with Hydrophilic Dissolving<br>Tip and Radiopaque Marker | M001282600 | 8F/35cm | M001282610 | 10F/35cm |
| | Flexima™ Biliary Catheter System with Radiopaque Marker | M001272600 | 8F/35cm | | | | | | | | | | | | | | | | | | | | | | | | | | | |
| | | M001272610 | 10F/35cm | | | | | | | | | | | | | | | | | | | | | | | | | | | |
| | | M001272620 | 12F/35cm | | | | | | | | | | | | | | | | | | | | | | | | | | | |
| | | M001272630 | 14F/35cm | | | | | | | | | | | | | | | | | | | | | | | | | | | |
| | | M001272640 | 8F/35cm | | | | | | | | | | | | | | | | | | | | | | | | | | | |
| | | M001272650 | 10F/35cm | | | | | | | | | | | | | | | | | | | | | | | | | | | |
| | | M001272660 | 10F/35cm | | | | | | | | | | | | | | | | | | | | | | | | | | | |
| M001272670 | | 12F/35cm | | | | | | | | | | | | | | | | | | | | | | | | | | | | |
| Flexima™ Biliary Catheter System with Hydrophilic Dissolving<br>Tip | M001281560 | 8F/35cm | | | | | | | | | | | | | | | | | | | | | | | | | | | | |
| | M001281570 | 10F/35cm | | | | | | | | | | | | | | | | | | | | | | | | | | | | |
| | M001281580 | 12F/35cm | | | | | | | | | | | | | | | | | | | | | | | | | | | | |
| | M001281590 | 14F/35cm | | | | | | | | | | | | | | | | | | | | | | | | | | | | |
| Flexima™ Biliary Catheter System with Hydrophilic Dissolving<br>Tip and Radiopaque Marker | M001282600 | 8F/35cm | | | | | | | | | | | | | | | | | | | | | | | | | | | | |
| | M001282610 | 10F/35cm | | | | | | | | | | | | | | | | | | | | | | | | | | | | |
## Table 4. Flexima Biliary Catheters Dimensions
{8}------------------------------------------------
| Product Trade Name | Model Number | Dimensions<br>(OD/Length) |
|------------------------------------|--------------|---------------------------|
| Flexima™ Ureteral Stent System | M001274000 | 8F/20cm |
| | M001274010 | 8F/22cm |
| | M001274020 | 8F/24cm |
| | M001274030 | 8F/26cm |
| | M001274040 | 8F/28cm |
| | M001274050 | 10F/20cm |
| | M001274060 | 10F/22cm |
| | M001274070 | 10F/24cm |
| | M001274080 | 10F/26cm |
| | M001274090 | 10F/28cm |
| Flexima™ Ureteral Stent System Kit | M001274100 | 6F/20cm |
| | M001274110 | 6F/22cm |
| | M001274120 | 6F/24cm |
| | M001274130 | 6F/26cm |
| | M001274140 | 6F/28cm |
| | M001274150 | 8F/20cm |
| | M001274160 | 8F/22cm |
| | M001274170 | 8F/24cm |
| | M001274180 | 8F/26cm |
| | M001274190 | 8F/28cm |
### Table 5. Flexima Ureteral Stents Dimensions
#### V.INDICATIONS FOR USE
Flexima Biliary Catheters
To provide external and internal percutaneous drainage of the biliary system.
Flexima Ureteral Stents
To provide drainage from the ureteropelvic junction to the bladder and stenting of the ureter for all patients in whom it is desirable to place a drain which does not extend externally.
Predicate and subject device Intended use and Indications for Use are the same.
{9}------------------------------------------------
#### VI. COMPARISON OF TECHNOLOGICAL CHARACTERISTICS
There are no differences in the technological characteristics between the predicate and subject devices. The Flexima Biliary Catheters and Flexima Ureteral Stents maintain the same fundamental scientific technology, indications for use and operating principles as the predicate devices - K023870 and K944290 respectively.
The purpose of this 510(k) submission is to receive clearance for the addition of MRI Magnetic Resonance information to the product directions for use (DFU).
#### VII. SUMMARY OF NON-CLINICAL PERFORMANCE TESTING
In order to support the addition of Magnetic Resonance Safety Information to the Directions for Use the following non-clinical performance testing was performed:
- . Radio Frequency Heating
- Force Measurement
- . Image Artifact
FDA guidance Establishing Safety and Compatibility of Passive Implants in the Magnetic Resonance (MR) Environment. Dec. 2014 and the following standards were used to guide the generation of the nonclinical data:
| Document Name | Document<br>Number | Document<br>Version |
|-----------------------------------------------------------------------------------------------------------------------|--------------------|---------------------|
| Measurement of Radio Frequency Induced<br>Heating Near Passive Implants During Magnetic<br>Resonance Imaging | ASTM F2182 | -11a |
| Measurement of Magnetically Induced<br>Displacement Force on Medical Devices in the<br>Magnetic Resonance Environment | ASTM F2052 | - 14 |
| Evaluation of MR Image Artifacts from Passive<br>Implants | ASTM F2119 | - 07 |
#### VIII. CONCLUSION
Based on the intended use, technological characteristics, and non-clinical performance data provided, the Biliary Drainage Catheters and the Ureteral Stent Systems are substantially equivalent to the predicate devices K944290 and K023870 respectively. The Directions for Use update with magnetic resonance safety information for the subject devices does not raise new questions of safety or effectiveness, and the subject devices are as safe and effective as the predicate devices.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Report an issue or suggest a feature
Spot a broken citation, a wrong extraction, a missing document, or something you wish this app did? Tell us — no account needed.
or drag and drop files here
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.