K080838 · Alveolus, Inc. · ESW · Apr 24, 2008 · General, Plastic Surgery
Device Facts
Record ID
K080838
Device Name
ALVEOLUS ALIMAXX-E ESOPHAGEAL STENT SYSTEM
Applicant
Alveolus, Inc.
Product Code
ESW · General, Plastic Surgery
Decision Date
Apr 24, 2008
Decision
SESE
Submission Type
Special
Regulation
21 CFR 878.3610
Device Class
Class 2
Attributes
Therapeutic
Intended Use
The Alveolus ALIMAXX-E™ Esophageal Stent System is intended for maintaining esophageal lumen patency in esophageal strictures caused by intrinsic and / or extrinsic malignant tumors and for occlusion of esophageal fistulae.
Device Story
The ALIMAXX-E™ Esophageal Stent System is a self-expanding, completely covered Nitinol stent used to maintain esophageal lumen patency. It is deployed via a delivery system to treat strictures resulting from malignant tumors or to occlude esophageal fistulae. The device is intended for single-patient use. It is utilized by clinicians in a medical setting to provide mechanical support to the esophageal wall, thereby preventing obstruction and improving patient quality of life by facilitating swallowing and reducing symptoms associated with malignant esophageal disease.
Clinical Evidence
Bench testing only. Biocompatibility and physical performance testing were conducted in accordance with FDA guidance for esophageal and tracheal prostheses.
Technological Characteristics
Self-expanding, completely covered Nitinol stent; delivery system; available in various diameters and lengths; single-patient-use; non-sterile.
Indications for Use
Indicated for patients with esophageal strictures caused by intrinsic or extrinsic malignant tumors and for patients requiring occlusion of esophageal fistulae.
Regulatory Classification
Identification
An esophageal prosthesis is a rigid, flexible, or expandable tubular device made of a plastic, metal, or polymeric material that is intended to be implanted to restore the structure and/or function of the esophagus. The metal esophageal prosthesis may be uncovered or covered with a polymeric material. This device may also include a device delivery system.
Special Controls
The special control for this device is FDA's “Guidance for the Content of Premarket Notification Submissions for Esophageal and Tracheal Prostheses.”
Predicate Devices
Alveolus ALIMAXX-E™ Esophageal Stent System (K051621)
Related Devices
K041648 — NITI-S STENT & INTRODUCER, MODEL EOXXXX · Taewoong Medical Co., Ltd. · Apr 28, 2005
K233939 — Ultraflex Esophageal NG Stent System; WallFlex Esophageal Stent System · Boston Scientific · Apr 29, 2024
K111611 — ENDOMAXX FULLY COVERED ESOPHAGEAL STENT · Merit Medical Systems, Inc. · Jul 29, 2011
K113551 — ESOPHAGEAL TTS STENT · Taewoong Medical Co., Ltd. · Feb 14, 2012
K060239 — ALEOLUS/ENOVENTIONS,ALIMAXX-E DV ESOPHAGEAL STENT SYSTEM · Alveolus, Inc. · Mar 22, 2006
Submission Summary (Full Text)
{0}------------------------------------------------
# 14 510(k) SUMMARY
### 510(k) SUMMARY OF SAFETY AND EFFECTIVENESS (Per 21 CFR 807.92)
APR 2 4 2008
## General Company Information
| Name: | Alveolus, Inc. |
|----------|--------------------------|
| Contact: | Tony Alexander |
| | EVP Corporate Compliance |
Address:
9013 Perimeter Woods, Suite A Charlotte, NC 28216
| Telephone: | (704) 926 - 4864 |
|------------|------------------|
| Fax: | (704) 926 - 4838 |
#### Date Prepared: March 24, 2008
#### General Device Information
| Product Name: | ALIMAXX-E™ Esophageal Stent System |
|-----------------|--------------------------------------------------------------------------|
| Classification: | "Esophageal Prosthesis", Product code: ESW<br>21 CFR 878.3610 - Class II |
#### Predicate Device
Alveolus ALIMAXX-E™ Esophageal Stent System [510(k) Number K051621]
#### Description
The Alveolus ALIMAXX-E™ Esophageal Stent System is designed to maintain lumen patency in esophageal strictures caused by intrinsic and / or extrinsic malignant tumors. The stent is also indicated for occlusion of esophageal fistulae.
Alveolus ALIMAXX-E Device Modification: Premarket Notification Page 67 of 73
{1}------------------------------------------------
K080838
Page 2 of 2
The Alveolus ALIMAXX-E™ Esophageal Stent System is comprised of two components: the completely covered, self-expanding Nitinol stent and the delivery system. The ALIMAXX-E™ Esophageal Stent System is available in several diameters and lengths.
The single-patient-use components of the ALIMAXX-E™ Esophageal Stent System are provided non-sterile.
#### Intended Use (Indications)
The Alveolus ALIMAXX-E™ Esophageal Stent System is intended for maintaining esophageal lumen patency in esophageal strictures caused by intrinsic and / or extrinsic malignant tumors. The stent is also indicated for occlusion of esophageal fistulae.
#### Substantial Equivalence
This Notice supports the position that the modified Alveolus ALIMAXX-E™ Esophageal Stent is substantially equivalent to the Alveolus ALIMAXX-E™ Esophageal Stent System [510(k) Number K051621].
Comparisons of the modified Alveolus ALIMAXX-E™ Esophageal Stent System and the ALIMAXX-E™ Stent System predicate device (K051621) show that technological characteristics such as materials, biocompatibility, performances properties, sterilization and packaging are substantially equivalent to the currently marketed predicate device. Physical test results were performed and i biocompatibility test results included tests recommended in FDA "Guidance for the Content of Premarket Notifications for Esophageal and Tracheal Prostheses" (issued on April 28, 1998).
#### Conclusions
Alveolus believes that the information provided establishes that similar legally marketed devices have been used for the same clinical applications as the subject device. The materials from which the Alveolus device is fabricated have an established history of use in clinical applications, and the devices produced by Alveolus have been tested in accordance with applicable FDA guidelines.
{2}------------------------------------------------
Image /page/2/Picture/0 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo features a stylized depiction of a human figure, represented by three curved lines, overlaid on a circle. The text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" is arranged around the upper portion of the circle.
Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850
# APR 2 4 2008
Mr. Tony Alexander Chief Legal Officer Alveolus®, Inc. 9013-A Perimeter Woods Drive CHARLOTTE NC 28216
K080838 Re:
> Trade/Device Name: ALIMAXX-E" Esophagea1 Stent System Regulation Number: 21 CFR §878.3610 Regulation Name: Esophageal prosthesis Regulatory Class: II Product Code: ESW Dated: March 24, 2008 Received: March 25, 2008
Dear Mr. Alexander:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug. and Cosmetic act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations. Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{3}------------------------------------------------
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally. marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Center for Devices and Radiological Health's (CDRH's) Office of Compliance at one of the following numbers, based on the regulation number at the top of this letter,
| 21 CFR 876.xxxx | (Gastroenterology/Renal/Urology) | 240-276-0115 |
|-----------------|----------------------------------|--------------|
| 21 CFR 884.xxxx | (Obstetrics/Gynecology) | 240-276-0115 |
| 21 CFR 892.xxxx | (Radiology) | 240-276-0120 |
| Other | | 240-276-0100 |
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding postmarket surveillance, please contact CDRH's Office of Surveillance and Biometric's (OSB's) Division of Postmarket Surveillance at 240-276-3474. For questions regarding the reporting of device adverse events (Medical Device Reporting (MDR)), please contact the Division of Surveillance Systems at 240-276-3464. You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers. International and Consumer Assistance at its toll-free number (800) 638-2041 or (240) 276-3150 or at its Internet address http://www.fda.gov/cdrh/industry/support/index.html.
Sincerely vours.
Nancy C. Brogdon
Nancy C. Brogdon Director, Division of Reproductive, Abdominal, and Radiological Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{4}------------------------------------------------
#### INTENDED USE STATEMENT 1
### There are no changes in the Intended Use.
The Intended Use for the Subject Device is the same as the Intended Use for the ALIMAXX-E™ predicate system (K051621).
Device Name: Alveolus, ALIMAXX-E™ Esophageal Stent System
Indications for Use:
The Alveolus ALIMAXX-E™ Esophageal Stent System is intended for maintaining esophageal lumen patency in esophageal strictures caused by intrinsic and / or extrinsic malignant tumors and for occlusion of esophageal fistulae.
Prescription Use (Part 21 CFR 801 Subpart D)
AND/OR
Over-The-Counter Use (21 CFR 807 Subpart C)
(PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
Huld Leun
(Division Sign-Off)
Division of Reproductive, Abdominal, and Radiological Devices
510(k) Number K080838
Panel 1
/
Sort by
Ready
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.