K023817 · Clio Designs, Inc. · MNW · Jun 25, 2003 · Cardiovascular
Device Facts
Record ID
K023817
Device Name
SMARTSHAPE BODY COMPOSITION ANALYZER, MODEL 4010
Applicant
Clio Designs, Inc.
Product Code
MNW · Cardiovascular
Decision Date
Jun 25, 2003
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 870.2770
Device Class
Class 2
Indications for Use
The SmartShape Body Composition Analyzer is a noninvasive bioimpedance analyzer intended for use in the estimation of percent body fat by body weight.
Device Story
Hand-held bioelectric impedance analyzer; measures body fat composition. User inputs weight, gender, age, and height; device transmits low-level electrical current through user to measure impedance. Device calculates percent body fat based on impedance data and user-provided physical parameters. Stores data for up to 10 individuals; includes clock and alarm functions. Used in home setting by individual users. Output provides body fat percentage to help users monitor body composition.
Clinical Evidence
Bench testing only. Performance testing evaluated functionality, reliability, repeatability, and reproducibility. Comparative testing against the predicate device demonstrated equivalent performance.
Technological Characteristics
Hand-held, programmable bioelectric impedance analyzer. Uses low-level electrical current transmission to measure impedance. Calculates body fat percentage using impedance data, weight, gender, age, and height. Battery-operated.
Indications for Use
Indicated for individuals seeking to estimate percent body fat by body weight via noninvasive bioimpedance analysis. Intended for personal home use.
Regulatory Classification
Identification
An impedance plethysmograph is a device used to estimate peripheral blood flow by measuring electrical impedance changes in a region of the body such as the arms and legs.
Special Controls
*Classification.* Class II (special controls). The device, when it is a body composition analyzer which is not intended to diagnose or treat any medical condition, is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 870.9.
Predicate Devices
Omron Healthcare, Inc. Body Fat Analyzer Model HBF-306 (K011652)
Related Devices
K043414 — HL 188 & HL 189 BODY FAT METER · Health & Life Co., Ltd. · Feb 9, 2005
K020738 — AVITA BF1 (VITALIO) BODY FAT ANALYZER · Avita Corporation · May 22, 2003
K052678 — AQ-830 SLIM BODY, BODY FAT MONITOR · Limax Electronics Co., Ltd. · Dec 27, 2005
K963575 — HBF-1000 OMRON BODY COMPOSITION ANALYZER · Omron Healthcare, Inc. · Jun 13, 1997
K030203 — LONG WELL LW-6DXX SERIES BODY FAT ANALYZER · Long Well Electronics Corp. · Apr 21, 2003
Submission Summary (Full Text)
{0}------------------------------------------------
# SUMMARY
This summary of 510K safety and effectiveness information is being submitted in accordance with the requirements of SMDA 1990 and 21 CFR §807.92.
# 1. APPLICANT INFORMATION
| a. Company Name: | Clio Designs Incorporated |
|---------------------|-------------------------------------------|
| b. Company Address: | 63 Herrick Rd.<br>Newton Centre, MA 02459 |
| c. Company Phone: | 617-969-7509 |
| d. Contact Person: | Bernard Fabricant<br>Vice President |
| e. Date of Summary: | 11/10/02 |
### 2. DEVICE IDENTIFICATION
| a. Trade/Proprietary Name: | SmartShape Body Composition<br>Analyzer 4010 |
|----------------------------|--------------------------------------------------|
| b. Classification Name: | Body Composition Analyzer<br>21 CFR 870.2770 MNW |
| c. Device Class: | Class II |
# 3. IDENTIFICATION OF PREDICATE DEVICE
| a. Company: | Omron Healthcare, Inc. |
|------------------|---------------------------------|
| b. Device: | Body Fat Analyzer Model HBF-306 |
| c. 510K No.: | K011652 |
| d. Date Cleared: | 11/07/01 |
### 4. DEVICE DESCRIPTION
{1}------------------------------------------------
The SmartShape Body Composition Analyzer is a hand-held device designed to determine body fat composition. The device offers the user the ability to measure their body fat level and store the data for analvsis. It is programmable for up to 10 separate individuals and uses bioelectric impedance technology for the estimation of percent body fat. The device also includes a clock and alarm function.
### 5. SUBSTANTIAL EQUIVALENCE
The SmartShape Body Composition Analyzer is substantially equivalent to the Omron Healthcare, Inc. Body Fat Analyzer Model HBF-306.
The fundamental technical characteristics and indications for use of the SmartShape Body Composition Analyzer are similar to those of the predicate device. Both devices use bioelectric impedance for the estimation of percent body fat. The body composition analyzer and the predicate devices are indicated for personal home use.
#### 6. INDICATIONS FOR USE
The SmartShape Body Composition Analyzer is a noninvasive bioimpedance analyzer intended for use in the estimation of percent body fat by body weight.
### 7. TECHNOLOGICAL CHARACTERISTICS
The SmartShape Body Composition Analyzer is a programmable hand-held device designed to determine body fat composition. The device utilizes the technology of bioelectric impedance to determine the transmission speed of a low level electrical current through the user. The device software utilizes the impedance data, patient weight, gender, age, and height data to calculate the patient percent body fat.
A comparison of the technological characteristics and performance testing of the SmartShape Body Composition Analyzer to those of the predicate device reveals that the devices are substantially equivalent.
### 8. PERFORMANCE DATA
Performance testing was conducted on the SmartShape Body Composition Analyzer for functionality, reliability, repeatability and reproducibility. All results were shown to be acceptable. In addition, coparison testing was performed using the SmartShape Body Composition Analyzer and the predicate device. Results of the testing showed that the SmartShape Body Composition Analyzer was equivalent in performance to the predicate
{2}------------------------------------------------
device. The SmartShape Body Composition Analyzer was shown to perform as intended.
# 9. 510(K) CHECKLIST
This notification contains all information required by 21 CFR 807.87. A completed copy of the Premarket Notification 510(k) Reviewer's Checklist is provided in this submission.
{3}------------------------------------------------
Image /page/3/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a circular seal with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES • USA" arranged around the perimeter. Inside the circle is a stylized image of an eagle's head, represented by three curved lines.
Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850
JUN 2 5 2003
Mr. Bernard Fabricant Vice President Clio Designs Incorporated 63 Herrick Road NEWTON CENTRE, MA 02459
Re: K023817
Trade/Device Name: SmartShape™ Body Composition Analyzer 4010 Regulation Number: 21 CFR §870.2770 Regulation Name: Impedance plethysmograph Regulatory Class: II Product Code: 74 MNW Dated: March 23, 2003 Received: March 27, 2003
Dear Mr. Fabricant:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug. and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (sections 531-542 of the Act); 21 CFR 1000-1050.
{4}------------------------------------------------
This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Office of Compliance at one of the following numbers, based on the regulation number at the top of the letter:
| 8xx. 1xxx | (301) 594-4591 |
|----------------------------------|----------------|
| 876.2xxx, 3xxx, 4xxx, 5xxx | (301) 594-4616 |
| 884.2xxx, 3xxx, 4xxx, 5xxx, 6xxx | (301) 594-4616 |
| 892.2xxx, 3xxx, 4xxx, 5xxx | (301) 594-4654 |
| Other | (301) 594-4692 |
Additionally, for questions on the promotion and advertising of your device, please contact the Office of Compliance at (301) 594-4639. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97) you may obtain. Other general information on your responsibilities under the Act may be obtained from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its Internet address http://www.fda.gov/cdrh/dsma/dsmamain.html.
Sincerely yours,
Nancy C. Brogdon
Nancy C. Brogdon Director, Division of Reproductive, Abdominal and Radiological Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{5}------------------------------------------------
# INDICATIONS FOR USE STATEMENT
| Applicant's Name: | Clio Designs Incorporated |
|-------------------|-------------------------------------------|
| Device Name: | SmartShape Body Composition Analyzer 4010 |
The SmartShape Body Composition Analyzer is a noninvasive bioimpedance analyzer intended for use in the estimation of percent body fat by body weight.
**Over-the-Counter Use**
Nancy C Hogdon
(Division Sign-Off)
Division of Reproductive, Abdominal, and Radiological Devic 510(k) Number.
Panel 1
/
Ready
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.