SLIMMANAGER BODY COMPOSITION ANALYZER, MODELS N40 AND SM-300
K073227 · Aiia Communication, Inc. · MNW · Mar 20, 2008 · Cardiovascular
Device Facts
Record ID
K073227
Device Name
SLIMMANAGER BODY COMPOSITION ANALYZER, MODELS N40 AND SM-300
Applicant
Aiia Communication, Inc.
Product Code
MNW · Cardiovascular
Decision Date
Mar 20, 2008
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 870.2770
Device Class
Class 2
Indications for Use
For Measurement in healthy subjects of: Estimated : ECW, ICW, TBW, ECW/TBW, Body Fat, Body Lean +Dry Lean Metabolic Rates, Segmental Lean Mass. Model N40 adds: Skeletal Muscle Mass; Actual: Weight, BMI, Impedance Values
Device Story
SLIMMANAGER (Models N40, SM-300) is a body composition analyzer using Bioelectrical Impedance Analysis (BIA). Device sends weak electric current through body; detects impedance response based on water content in muscle/cellular fluids versus fat/bone tissues. Transforms impedance data into estimates of lean tissue, fat, and water mass. N40 features integrated Linux system with touch monitor; SM-300 interfaces with external PC. Used by individuals for self-monitoring; measurements saved locally or transmitted via Internet to personal web page for progress tracking without expert intervention. Output provides users with body composition metrics to monitor health changes over time.
Clinical Evidence
Bench and clinical testing performed. Results indicate device is as safe and effective as predicate devices. No specific clinical trial metrics (e.g., sensitivity, specificity) provided in summary.
Technological Characteristics
BIA technology; 8-electrode tactile configuration. N40: Linux-based with touch monitor. SM-300: PC-interfaced. Power: AC100-120/200-240V. Connectivity: Internet transmission for web-based data storage. Measuring scope: 1-200kg (N40), 1-150kg (SM-300).
Indications for Use
Indicated for measurement of body composition parameters in healthy subjects. Parameters include ECW, ICW, TBW, ECW/TBW, body fat, body lean, dry lean, metabolic rates, segmental lean mass, and (for N40) skeletal muscle mass, weight, BMI, and impedance values.
Regulatory Classification
Identification
An impedance plethysmograph is a device used to estimate peripheral blood flow by measuring electrical impedance changes in a region of the body such as the arms and legs.
Special Controls
*Classification.* Class II (special controls). The device, when it is a body composition analyzer which is not intended to diagnose or treat any medical condition, is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 870.9.
AIIA communication Inc. SLIMMANAGER(SM-X, SM-E) (K051589)
Submission Summary (Full Text)
{0}------------------------------------------------
PAGE 1 of 2
MAR 2 0 2008
## 510(k) Summary K073227 AIIA Communication Inc. Geumwon building, #714-1, Bokjung-dong, Sujung-gu, Sungnam-si, Gyunggi-do. Korea Tel : +82-31-751-3697 Fax : +82-31-751-2648 Homepage : http://www.aiia.co.kr March 10, 2007 Contact: Ho Lee, R&D Manager
- 1. Identification of the Device: Proprietary-Trade Name: SLIMMANAGER, Body composition analyzer (Models: N40, SM-300). Classification Names: 74 MNW ANALYZER, BODY COMPOSITION Common/Usual Name: Body composition analyzer
- Equivalent legally marketed devices Biospace InBody 230 K062603; AIIA communication ni Inc. SLIMMANAGER(SM-X, SM-E) K051589
- 3. Indications for Use (intended use) (Models SM-300 and N40) For Measurement in healthy subjects of: Estimated : ECW, ICW, TBW, ECW/TBW, Body Fat, Body Lean +Dry Lean Metabolic Rates, Segmental Lean Mass. Model N40 adds: Skeletal Muscle Mass; Actual: Weight, BMI, Impedance Values
- 4. Description of the Device: SLIMMANAGER is a body composition analyzer and uses Bioelectrical Impedance Analysis technology to determine body composition parameters. Using a principle that an electric current has a low impedance when it flows in muscle or cellular fluids containing lots of water while it has a high impedance such as in fat and bone tissues slowing and stopping an applied current, BIA sends an extremely weak electric current through the body and detects the relative response after it flows our body's various substances, and then it can determines lean tissue, fat, water mass. All measurements made by SLIMMANAGER can be recorded & saved in the website real-time by the Internet transmission as well in the device, and in his own page people could check their changing progress even a minute change without an expert's advice at anywhere, at anytime. N-40 is designed by Linux system with touch monitor, while SM-300 should be interfaced with customer's PC.
- 5. Safety and Effectiveness, comparison to predicate device. The results of bench and clinical testing indicates that the new device is as safe and effective as the predicate devices.
{1}------------------------------------------------
Image /page/1/Picture/0 description: The image shows the text "K073227 PAGE 2 OF 2". The text appears to be handwritten. The numbers are slightly distorted, but still legible.
- 6. Substantial Equivalence Chart, SLIMMANAGER, Body composition analyzer (Models: N40, SM-300).
| Company | Biospace | AIIA communication<br>Inc. | AIIA communication<br>Inc. | AIIA<br>communication Inc. | AIIA<br>communication Inc. |
|--------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------|----------------------------|
| Model | InBody 230 | SLIMMANAGER(SM-X,<br>SM-E) | SLIMMANAGER(N40) | SLIMMANAGER(SM-<br>300) | |
| 510(k)<br>no. | K062603 | K051589 | New | New | |
| Method | Bioelectrical<br>Impedance<br>Analysis | Bioelectrical Impedance<br>Analysis | Bioelectrical<br>Impedance Analysis | Bioelectrical<br>Impedance Analysis | |
| Electrode | 8 electrodes | 8 electrodes | 8 electrodes | 8 electrodes | |
| Type | tactile | tactile | tactile | tactile | |
| Intended<br>Use | Estimated :<br>ECW, ICW, TBW<br>ECW/TBW<br>Body Fat<br>Body Lean +Dry Lean<br>Metabolic Rates<br>Segmental Lean Mass<br>Skeletal Muscle Mass<br>Actual :<br>Weight<br>BMI<br>Impedance Values | Estimated :<br>ECW, ICW, TBW<br>ECW/TBW<br>Body Fat<br>Body Lean +Dry Lean<br>Metabolic Rates<br>Segmental Lean Mass<br>Actual :<br>Weight<br>BMI<br>Impedance Values | Estimated :<br>ECW, ICW, TBW<br>ECW/TBW<br>Body Fat<br>Body Lean +Dry Lean<br>Metabolic Rates<br>Segmental Lean Mass<br>Skeletal Muscle Mass<br>Actual :<br>Weight<br>BMI<br>Impedance Values | Estimated :<br>ECW, ICW, TBW<br>ECW/TBW<br>Body Fat<br>Body Lean +Dry Lean<br>Metabolic Rates<br>Segmental Lean Mass<br>Actual :<br>Weight<br>BMI<br>Impedance Values | |
| Frequency | 20, 100KHz | 5, 50, 250, 500KHz | 500Hz, 50, 500KHz | 50KHz | |
| Measuring<br>scope | 10 ~ 250 kg | 1 ~ 250 kg | 1 ~ 200 kg | 1 ~ 150 kg | |
| Position | Upright | Upright | Upright | Upright | |
| Display | 240x320 STN LCD | 1024x768<br>Color TFT<br>Touch<br>Screen LCD | 640x480 Color TFT<br>Touch Screen LCD | PC interface | |
| Power | AC100~120/200~240V | AC100~120/200~240V | AC100~120/200~240V | AC100~120/200~240V | |
## 7. Conclusion
After analyzing both bench and clinical testing data, it is the conclusion of AIIA Communication Inc.that the SLIMMANAGER, Body composition analyzer (Models: N40, SM-300). as safe and effective as the predicate devices, has few technological differences, and has no new indications for use, thus rendering it substantially equivalent to the predicate devices.
{2}------------------------------------------------
## DEPARTMENT OF HEALTH & HUMAN SERVICES
Image /page/2/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a stylized eagle with three stripes extending from its back, representing the department's mission to protect the health of all Americans. The text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" is arranged in a circular pattern around the eagle.
MAR 2 0 2008
Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850
AIIA Communication Inc. c/o Daniel Kamm, P.E. Kamm & Associates P.O. Box 7007 DEERFIELD IL 60015
K073227 Re:
> Trade/Device Name: SLIMMANAGER, Body composition analyzer (Models N40, SM-300) Regulation Number: 21 CFR §870.2770. Regulation Name: Impedance plethysmograph Regulatory Class: II Product Code: MNW Dated: March 10, 2008 Received: March 13, 2008
Dear Mr. Kamm:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration. Iisting of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{3}------------------------------------------------
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally. marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Center for Devices and Radiological Health's (CDRH's) Office of Compliance at one of the following numbers, based on the regulation number at the top of this letter.
| 21 CFR 876.xxxx | (Gastroenterology/Renal/Urology) | 240-276-0115 |
|-----------------|----------------------------------|--------------|
| 21 CFR 884.xxxx | (Obstetrics/Gynecology) | 240-276-0115 |
| 21 CFR 892.xxxx | (Radiology) | 240-276-0120 |
| Other | | 240-276-0100 |
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding postmarket surveillance, please contact CDRH's Office of Surveillance and Biometric's (OSB's) Division of Postmarket Surveillance at 240-276-3474. For questions regarding the reporting of device adverse events (Medical Device Reporting (MDR)), please contact the Division of Surveillance Systems at 240-276-3464. You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (240) 276-3150 or at its Internet address http://www.fda.gov/cdrb/industry/support/index.html.
Sincerely vours.
Nancy C Brogdon
Nancy C. Brogdon Director, Division of Reproductive. Abdominal, and Radiological Devices Office of Device Eyaluation Center for Devices and Radiological Health
Enclosure
{4}------------------------------------------------
## Indications for Use
510(k) Number (if known): K073227
Device Name: SLIMMANAGER, Body composition analyzer (Models N40, SM-300)
Indications For Use: For Measurement in healthy subjects of:
Estimated : ECW, ICW, TBW, ECW/TBW, Body Fat, Body Lean +Dry Lean Metabolic Rates, Segmental Lean Mass. Model N40 adds: Skeletal Muscle Mass Actual : Weight, BMI, Impedance Values
Prescription Use (Part 21 CFR 801 Subpart D) AND/OR
Over-The-Counter Use X (21 CFR 807 Subpart C)
(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
Divided Sine Corp.
(Division Sign-Off) Division of Reproductive, Abdominal, and Radiological Devices 510(k) Number
Page 1 of 1
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.