Flexible Ureteroscope (U-Scope)(2.8/1.2) (OMI161-2F28-CH12-US); Flexible Ureteroscope (U-Scope)(2.5/1.2) (OMI161-2F25-CH12-US); Images Systems (Camera-Controlled Unit) (OMI01M12)

K242699 · Opcom Medical, Inc. · FGB · May 28, 2025 · Gastroenterology, Urology

Device Facts

Record IDK242699
Device NameFlexible Ureteroscope (U-Scope)(2.8/1.2) (OMI161-2F28-CH12-US); Flexible Ureteroscope (U-Scope)(2.5/1.2) (OMI161-2F25-CH12-US); Images Systems (Camera-Controlled Unit) (OMI01M12)
ApplicantOpcom Medical, Inc.
Product CodeFGB · Gastroenterology, Urology
Decision DateMay 28, 2025
DecisionSESE
Submission TypeTraditional
Regulation21 CFR 876.1500
Device ClassClass 2
AttributesTherapeutic

Intended Use

The Disposable Flexible Ureteroscope (U-Scope) is a sterile, single-use, flexible, digital video ureteroscope intended to be used to visualize organs, cavities, and canals in the urinary tract (urethra, bladder, ureter, calyces, and renal papillae) via trans-urethral or percutaneous access routes. It can also be used in conjunction with endoscopic instruments via its working channel to perform various diagnostic and therapeutic procedures in the urinary tract. Images Systems (Camera-Controlled Unit) is intended to provide power to and receive, process, display, and output recordings of images from compatible visualization devices, the intended medical indication will be defined by the connected visualization devices.

Device Story

System comprises single-use flexible digital video ureteroscope and reusable Camera-Controlled Unit (CCU). Ureteroscope features distal CMOS camera, LED illumination, and working channel for irrigation/instruments; connects to CCU for power and data. CCU includes video processor, touchscreen LCD, and I/O ports for image/video capture and external display. Used in clinical settings by physicians for urinary tract visualization and procedures. CCU processes and displays real-time video; enables image/video recording for clinical documentation. System facilitates minimally invasive diagnostic and therapeutic access to urinary anatomy.

Clinical Evidence

No clinical data. Substantial equivalence supported by bench testing, including electrical safety (IEC 60601-1), EMC (EN 60601-1-2), photobiological safety (IEC 62471), mechanical/optical performance (ISO 8600 series), biocompatibility (ISO 10993), and software verification.

Technological Characteristics

Single-use flexible ureteroscope with CMOS sensor, LED illumination, 270° bidirectional bending. Insertion tube OD 2.5/2.8mm, length 670mm. Working channel ID 1.2mm. Reusable CCU with 12.1" touchscreen, video processor, HDMI/USB connectivity. Sterilization: Ethylene Oxide. Compliance: IEC 60601-1, IEC 60601-1-2, IEC 60601-2-18.

Indications for Use

Indicated for visualization of the urinary tract (urethra, bladder, ureter, calyces, renal papillae) via trans-urethral or percutaneous access for diagnostic and therapeutic procedures in patients requiring endoscopic intervention.

Regulatory Classification

Identification

An endoscope and accessories is a device used to provide access, illumination, and allow observation or manipulation of body cavities, hollow organs, and canals. The device consists of various rigid or flexible instruments that are inserted into body spaces and may include an optical system for conveying an image to the user's eye and their accessories may assist in gaining access or increase the versatility and augment the capabilities of the devices. Examples of devices that are within this generic type of device include cleaning accessories for endoscopes, photographic accessories for endoscopes, nonpowered anoscopes, binolcular attachments for endoscopes, pocket battery boxes, flexible or rigid choledochoscopes, colonoscopes, diagnostic cystoscopes, cystourethroscopes, enteroscopes, esophagogastroduodenoscopes, rigid esophagoscopes, fiberoptic illuminators for endoscopes, incandescent endoscope lamps, biliary pancreatoscopes, proctoscopes, resectoscopes, nephroscopes, sigmoidoscopes, ureteroscopes, urethroscopes, endomagnetic retrievers, cytology brushes for endoscopes, and lubricating jelly for transurethral surgical instruments. This section does not apply to endoscopes that have specialized uses in other medical specialty areas and that are covered by classification regulations in other parts of the device classification regulations.

Special Controls

*Classification* —(1)*Class II (special controls).* The device, when it is an endoscope disinfectant basin, which consists solely of a container that holds disinfectant and endoscopes and accessories; an endoscopic magnetic retriever intended for single use; sterile scissors for cystoscope intended for single use; a disposable, non-powered endoscopic grasping/cutting instrument intended for single use; a diagnostic incandescent light source; a fiberoptic photographic light source; a routine fiberoptic light source; an endoscopic sponge carrier; a xenon arc endoscope light source; an endoscope transformer; an LED light source; or a gastroenterology-urology endoscopic guidewire, is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 876.9.(2) Class I for the photographic accessories for endoscope, miscellaneous bulb adapter for endoscope, binocular attachment for endoscope, eyepiece attachment for prescription lens, teaching attachment, inflation bulb, measuring device for panendoscope, photographic equipment for physiologic function monitor, special lens instrument for endoscope, smoke removal tube, rechargeable battery box, pocket battery box, bite block for endoscope, and cleaning brush for endoscope. The devices subject to this paragraph (b)(2) are exempt from the premarket notification procedures in subpart E of part 807of this chapter, subject to the limitations in § 876.9.

Predicate Devices

Reference Devices

Related Devices

Submission Summary (Full Text)

{0} FDA U.S. FOOD &amp; DRUG ADMINISTRATION May 28, 2025 OPCOM Medical Inc. Chinglin Huang Regulatory Specialist 12F, No.6, JianKang Rd., Zhonghe Dist. New Taipei City, 23586 TAIWAN Re: K242699 Trade/Device Name: Flexible Ureteroscope (U-Scope)(2.8/1.2) (OMI161-2F28-CH12-US); Flexible Ureteroscope (U-Scope)(2.5/1.2) (OMI161-2F25-CH12-US); Images Systems (Camera-Controlled Unit) (OMI01M12) Regulation Number: 21 CFR 876.1500 Regulation Name: Endoscope and accessories Regulatory Class: II Product Code: FGB, FET Received: May 5, 2025 Dear Chinglin Huang: We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading. If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of {1} K242699 - Chinglin Huang Page 2 Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register. Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device" (https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download). Your device is also subject to, among other requirements, the Quality System (QS) regulation (21 CFR Part 820), which includes, but is not limited to, 21 CFR 820.30, Design controls; 21 CFR 820.90, Nonconforming product; and 21 CFR 820.100, Corrective and preventive action. Please note that regardless of whether a change requires premarket review, the QS regulation requires device manufacturers to review and approve changes to device design and production (21 CFR 820.30 and 21 CFR 820.70) and document changes and approvals in the device master record (21 CFR 820.181). Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reporting-combination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050. All medical devices, including Class I and unclassified devices and combination product device constituent parts are required to be in compliance with the final Unique Device Identification System rule ("UDI Rule"). The UDI Rule requires, among other things, that a device bear a unique device identifier (UDI) on its label and package (21 CFR 801.20(a)) unless an exception or alternative applies (21 CFR 801.20(b)) and that the dates on the device label be formatted in accordance with 21 CFR 801.18. The UDI Rule (21 CFR 830.300(a) and 830.320(b)) also requires that certain information be submitted to the Global Unique Device Identification Database (GUDID) (21 CFR Part 830 Subpart E). For additional information on these requirements, please see the UDI System webpage at https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance/unique-device-identification-system-udi-system. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-devices/medical-device-safety/medical-device-reporting-mdr-how-report-medical-device-problems. For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the {2} K242699 - Chinglin Huang Page 3 Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100). Sincerely, Mark J. Antonino -S Mark J. Antonino, M.S. Assistant Director DHT3B: Division of Reproductive, Gynecology, and Urology Devices OHT3: Office of Gastrorenal, ObGyn, General Hospital, and Urology Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health Enclosure {3} DEPARTMENT OF HEALTH AND HUMAN SERVICES Food and Drug Administration Indications for Use Form Approved: OMB No. 0910-0120 Expiration Date: 07/31/2026 See PRA Statement below. Submission Number (if known) K242699 Device Name Flexible Ureteroscope (U-Scope)(2.8/1.2) (OMI161-2F28-CH12-US); Flexible Ureteroscope (U-Scope)(2.5/1.2) (OMI161-2F25-CH12-US); Images Systems (Camera-Controlled Unit) (OMI01M12) Indications for Use (Describe) The Disposable Flexible Ureteroscope (U-Scope) is a sterile, single-use, flexible, digital video ureteroscope intended to be used to visualize organs, cavities, and canals in the urinary tract (urethra, bladder, ureter, calyces, and renal papillae) via trans-urethral or percutaneous access routes. It can also be used in conjunction with endoscopic instruments via its working channel to perform various diagnostic and therapeutic procedures in the urinary tract. Images Systems (Camera-Controlled Unit) is intended to provide power to and receive, process, display, and output recordings of images from compatible visualization devices, the intended medical indication will be defined by the connected visualization devices. Type of Use (Select one or both, as applicable) ☑ Prescription Use (Part 21 CFR 801 Subpart D) ☐ Over-The-Counter Use (21 CFR 801 Subpart C) ## CONTINUE ON A SEPARATE PAGE IF NEEDED. This section applies only to requirements of the Paperwork Reduction Act of 1995. *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.* The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to: Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov “An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number.” {4} K242699 Page 1 of 5 # 510(k) Summary ## 1. Submitter OPCOM Medical, Inc. 12F, No.6, JianKang Rd., Zhonghe Dist., New Taipei City 23586, Taiwan (R.O.C.) ## 2. Contact Person ChingLin Huang, Regulatory specialist Phone: +886 2 2221 7201 #1904 Email: hazel.huang@opcommedical.com ## 3. Date Prepared May 27, 2025 ## 4. Trade Name Flexible Ureteroscope (U-Scope) (2.8/1.2) (OMI161-2F28-CH12-US); Flexible Ureteroscope (U-Scope) (2.5/1.2) (OMI161-2F25-CH12-US); Images Systems (Camera-Controlled Unit) (OMI01M12) ## 5. Common Name Endoscope and accessories ## 6. Classification Name Ureteroscope And Accessories, Flexible/Rigid ## 7. Regulation Number 876.1500 ## 8. Product Code FGB, FET ## 9. Predicate Device K231878: Uretero1™ Ureteroscope System, STERIS Corporation ## 10. Reference Device K221158: Single-Use Video Flexible Ureterorenoscope System, Guangzhou Red Pine Medical Instrument Co., Ltd ## 11. Device Description The Flexible Ureteroscope (U-Scope) is a sterile, single-use, flexible, digital video ureteroscope. It contains a miniature CMOS camera, light-emitting diodes (LED) illumination module at the tip and an EEPROM in the handle to store use time. The tip of the U-Scope has a bending portion with $270^{\circ} \pm 15^{\circ}$ bidirectional bend angle in the bending section. The U-Scope connects to the I through a separate electrical connector for sending image data and receiving LED power. The U-Scope has an inner working channel no less than $1.20\mathrm{mm}$ for the infusion fluid and instrument passage with two access ports. The insertion tube distal tip OD is available in two size configurations: Flexible Ureteroscope (2.8/1.2) and Flexible Ureteroscope (2.5/1.2). Apart from the size, the endoscopes share a similar design and working length of the ureteroscope is $670~\mathrm{mm}$. The Luer Port working {5} K242699 Page 2 of 5 lumen ID is 1.35mm and irrigation lumen ID is 1.3mm that is used for irrigation connection and accessory device access. The reusable Images System (Camera-Controlled Unit) contains most electronics, including a power on/off button, touch screen, video processor, LCD, power management electronics and microcontrollers. The system includes the necessary hardware, software, and firmware to drive the endoscope CMOS camera and light-emitting diodes (LED), adjust live view images, capture images, save images to an external source, capture videos, save videos to an external source, manage saved image and files, and manage saved video files. The reusable Images System (Camera-Controlled Unit) has connectors for attaching and detaching the endoscope, external monitor, external memory, and DC power supply. The LCD has a touchscreen function for user interaction with the GUI to control the CCU functions. The display unit has a VESA mount incorporated into the rear of the enclosure for attachment to a cart or any existing customer mounting bracket that adheres to the same standard mounting pattern. Both the reusable Images System (Camera-Controlled Unit) and the single-use Flexible Ureteroscope together complete the "Ureteroscope System" enabling a Flexible Ureteroscopy procedure to take place. ## 12. Indications for Use The Disposable Flexible Ureteroscope (U-Scope) is a sterile, single-use, flexible, digital video ureteroscope intended to be used to visualize organs, cavities, and canals in the urinary tract (urethra, bladder, ureter, calyces, and renal papillae) via trans-urethral or percutaneous access routes. It can also be used in conjunction with endoscopic instruments via its working channel to perform various diagnostic and therapeutic procedures in the urinary tract. Images Systems (Camera-Controlled Unit) is intended to provide power to and receive, process, display, and output recordings of images from compatible visualization devices, the intended medical indication will be defined by the connected visualization devices. ## 13. Indications for Use Comparison The subject device indications for use are similar to the predicate device. Both devices are intended for use for visualization in the urethra, bladder, ureter, calyces, and renal papillae via trans-urethral or percutaneous access routes to perform diagnostic and therapeutic procedures. Compared to the predicate device, the subject device indications include a separate description of the image processor function. This difference does not raise any different questions of safety or effectiveness. ## 14. Technological Comparison | Table 1 Comparison with Predicate Devices | | | | --- | --- | --- | | Characteristic | Flexible Ureteroscope System | Uretero1™ Ureteroscope System | | Manufacturer | OPCOM Medical. Inc. | Steris Corporation | | Product Code | FGB, FET | FGB | | Regulation Number | 21 CFR 876.1500 | 21 CFR 876.1500 | {6} K242699 Page 3 of 5 | Indications for Use | The Disposable Flexible Ureteroscope (U-Scope) is a sterile, single-use, flexible, digital video ureteroscope intended to be used to visualize organs, cavities, and canals in the urinary tract (urethra, bladder, ureter, calyces, and renal papillae) via trans-urethral or percutaneous access routes. It can also be used in conjunction with endoscopic instruments via its working channel to perform various diagnostic and therapeutic procedures in the urinary tract. Images Systems (Camera-Controlled Unit) is intended to provide power to and receive, process, display, and output recordings of images from compatible visualization devices, the intended medical indication will be defined by the connected visualization devices. | The single use Uretero1™ Ureteroscope System is intended to be used to visualize organs, cavities and canals in the urinary tract (urethra, bladder, ureter, calyces and renal papillae) via transurethral or percutaneous access routes. It can also be used in conjunction with endoscopic accessories to perform various diagnostic and therapeutic procedures in the urinary tract. | | --- | --- | --- | | Image sensor resolution | 400*400 | 400*400 | | Tube size | OD: 2.5 /2.8mm Length:670mm | OD: 2.98mm Length:680mm | | Working channel | ID: 1.2mm | ID: 1.2mm | | Depth of field | 5-50 mm | 2-50 mm | | Field of View | Nominal: 120°± 3°(diagonal) Measured: 110°± 3°(diagonal) | 120° | | Bending angle | Up / Down 270°±15° | Up 275° / Down 275° | | Working hours | 6 hours | 4 hours | | Illumination Source | 5mm > 2600 lux 20mm > 700 lux 25mm > 500 lux | None disclose | | Sterilization Method | Ethylene Oxide | Ethylene Oxide | | Waterproof | U-SCOPE: IPX7 | None disclose | | Operating conditions | 10°C to 40°C 30%RH - 85%RH 50 KPa - 106 Kpa | 17°C to 25°C 20%RH - 75%RH | | Transport and Storage conditions | <70°C, < 95%RH 50 KPa - 106 Kpa | -20°C to 60°C 15%RH - 90%RH | {7} K242699 Page 4 of 5 | LCD Size | 12.1" | 21.5" | | --- | --- | --- | | Output Resolution | HDMI 1920 x 1080 30 fps Touch screen LVDS 1280 x 800 30fps | Full HD 1920(H) x 1080(V) pixels, 30Hz Touch screen | | VESA | 75 x 75 mm | 100 x 100 mm | | Electrical Power | 12V/2A DC input 0.8A-1.2A AC 100-240V 50/60 Hz | 12VDC, 5A AC 110V - 230V , 50 - 60 Hz | | I/O | Right side: sensor and power connector Bottom side: USB, HDMI, DC Jack | USB, HDMI, power connector | | Regulation Compliance | IEC 60601-1 IEC 60601-1-2 IEC 60601-2-18 | IEC 60601-1 IEC 60601-1-2 IEC 60601-2-18 | **Note:** Flexible Ureteroscope differs from its predicate device in some technical specifications, including insertion portion, Image sensor resolution, Depth of field and working hours. These differences do raise different questions of safety and performance. Flexible Ureteroscope has passes the performance testing. Image system (Camera-Controlled Unit) differs from its predicate device in some technical specifications, including LCD Size. These differences do raise different questions of safety and performance. Image system (Camera-Controlled Unit) has passed the performance testing, **15. Intended Use** The Disposable Flexible Ureteroscope (U-Scope) is a sterile, single-use, flexible, digital video ureteroscope intended to be used to visualize organs, cavities, and canals in the urinary tract (urethra, bladder, ureter, calyces, and renal papillae) via trans-urethral or percutaneous access routes. It can also be used in conjunction with endoscopic instruments via its working channel to perform various diagnostic and therapeutic procedures in the urinary tract. **16. Non-clinical Testing in Support of Substantial Equivalence Determination** Non-clinical tests were conducted to verify that the proposed device meets all design specifications and is as safe and effective as the predicate device. The reference device was used to support the optical performance characteristics of the subject device. **Electrical Safety and Electromagnetic Compatibility** The test results demonstrated that the proposed device complies with the following standards: IEC 60601-1:2005, IEC 60601-1:2005/AMD:2012, IEC 60601-1:2005/AMD2:2020 Medical electrical equipment – Part 1: General requirements for basic safety and essential performance. EN 60601-1-2:2014/AMD1:2020, Medical electrical equipment - Part 1-2: General requirements for basic safety and essential performance - Collateral standard: Electromagnetic compatibility - Requirements and tests. IEC 60601-2-18:2009, Medical electrical equipment - Part 2-18: Particular requirements for the basic safety and essential performance of endoscopic equipment. **Photobiological Safety** Photobiological safety of the subject device was assessed according to IEC 62471:2006. {8} K242699 Page 5 of 5 ## Mechanical and Optical Performance - Bending performance - Working channel performance - Flow rate - Tensile and torsional strength - Field of view (ISO 8600-3:2019 Method B) - Direction of view (ISO 8600-3:2019) - Resolution (ISO 8600-5:2020 Characteristic C) - Noise and dynamic range - Geometric distortion - Image intensity uniformity ## Software Software Verification Test was performed to verify the software functions against its intended use. ## Biocompatibility Biocompatibility has been assessed according to: - In Vitro Cytotoxicity tested according to ISO 10993-5:2009 - Skin Sensitization tested according to ISO 10993-10:2021 - Acute Systemic Toxicity tested according to ISO 10993-11:2017 - Penile Irritation Study tested according to ISO 10993-23:2021 - Vaginal Irritation Study tested according to ISO 10993-23:2021 - Pyrogen Study tested according to USP46:2023 ## Sterilization and shelf life - Sterilization validation according to ISO 11135:2014 - Environmental conditioning (ASTM D4332-22) and simulated shipping distribution according to ASTM D4169-22 - Package integrity and device performance testing to support a three-year shelf-life for the Flexible Ureteroscope ## 17. Conclusion Based on the intended use, technological characteristics and non-clinical performance data, the subject device is as safe and effective and substantially equivalent to the legally marketed predicate device.
Innolitics

Panel 1

/
Sort by
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from the tree

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...