Cytology Brush
K172663 · Micro-Tech (Nanjing) Co., Ltd. · FDX · Mar 5, 2018 · Gastroenterology, Urology
Device Facts
| Record ID | K172663 |
| Device Name | Cytology Brush |
| Applicant | Micro-Tech (Nanjing) Co., Ltd. |
| Product Code | FDX · Gastroenterology, Urology |
| Decision Date | Mar 5, 2018 |
| Decision | SESE |
| Submission Type | Traditional |
| Regulation | 21 CFR 876.1500 |
| Device Class | Class 2 |
Intended Use
The Cytology Brush is used to collect cells from the bronchi and upper and lower gastrointestinal tracts.
Device Story
Cytology Brush is a manual endoscopic accessory used to collect cells from bronchi and gastrointestinal tracts. Device consists of a brush head, outer sheath, and handle. Operator (physician) inserts device through an endoscope; moves finger ring on handle to extend/retract brush head at target site to collect cellular samples. Device is used in clinical settings during endoscopic procedures. Collected cells are used for diagnostic evaluation. Benefits include minimally invasive tissue sampling for pathology.
Clinical Evidence
No clinical data. Bench testing only, including dimension testing and tensile strength testing, performed to demonstrate substantial equivalence.
Technological Characteristics
Manual endoscopic brush. Components: brush head, outer sheath, handle. Sterilization: Ethylene Oxide (ISO 11135:2014) to SAL 10^-6. Biocompatibility: ISO 10993-1:2009/(R)2013. EO residuals: ISO 10993-7:2008. Seven specifications varying by brush diameter, sheath diameter, working length, and finger ring color.
Indications for Use
Indicated for the collection of cells from the bronchi and upper and lower gastrointestinal tracts in patients requiring diagnostic cytology.
Regulatory Classification
Identification
An endoscope and accessories is a device used to provide access, illumination, and allow observation or manipulation of body cavities, hollow organs, and canals. The device consists of various rigid or flexible instruments that are inserted into body spaces and may include an optical system for conveying an image to the user's eye and their accessories may assist in gaining access or increase the versatility and augment the capabilities of the devices. Examples of devices that are within this generic type of device include cleaning accessories for endoscopes, photographic accessories for endoscopes, nonpowered anoscopes, binolcular attachments for endoscopes, pocket battery boxes, flexible or rigid choledochoscopes, colonoscopes, diagnostic cystoscopes, cystourethroscopes, enteroscopes, esophagogastroduodenoscopes, rigid esophagoscopes, fiberoptic illuminators for endoscopes, incandescent endoscope lamps, biliary pancreatoscopes, proctoscopes, resectoscopes, nephroscopes, sigmoidoscopes, ureteroscopes, urethroscopes, endomagnetic retrievers, cytology brushes for endoscopes, and lubricating jelly for transurethral surgical instruments. This section does not apply to endoscopes that have specialized uses in other medical specialty areas and that are covered by classification regulations in other parts of the device classification regulations.
Special Controls
*Classification* —(1)*Class II (special controls).* The device, when it is an endoscope disinfectant basin, which consists solely of a container that holds disinfectant and endoscopes and accessories; an endoscopic magnetic retriever intended for single use; sterile scissors for cystoscope intended for single use; a disposable, non-powered endoscopic grasping/cutting instrument intended for single use; a diagnostic incandescent light source; a fiberoptic photographic light source; a routine fiberoptic light source; an endoscopic sponge carrier; a xenon arc endoscope light source; an endoscope transformer; an LED light source; or a gastroenterology-urology endoscopic guidewire, is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 876.9.(2) Class I for the photographic accessories for endoscope, miscellaneous bulb adapter for endoscope, binocular attachment for endoscope, eyepiece attachment for prescription lens, teaching attachment, inflation bulb, measuring device for panendoscope, photographic equipment for physiologic function monitor, special lens instrument for endoscope, smoke removal tube, rechargeable battery box, pocket battery box, bite block for endoscope, and cleaning brush for endoscope. The devices subject to this paragraph (b)(2) are exempt from the premarket notification procedures in subpart E of part 807of this chapter, subject to the limitations in § 876.9.
Predicate Devices
- Wilson-Cook Cytology Brush (K896318)
Related Devices
- K103437 — US ENDOSCOPY CYTOLOGY BRUSH · United States Endoscopy Group, Inc. · Dec 23, 2010
- K190293 — Single Use Cytology Brush BC-205D · Olympus Medical Systems Corp. · Aug 2, 2019
- K021998 — HORIZONS CYTOLOGY BRUSH · Horizons Intl. Corp. · Jul 12, 2002
- K220063 — Single Use Cytology Brush · Zhejiang Chuangxiang Medical Technology Co., Ltd. · Jul 28, 2022
- K171607 — Bronchi and Gastrointestinal Cytology Brush · Wilson-Cook Medical Inc./Cook Endoscopy · Jul 31, 2017
Submission Summary (Full Text)
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health and Human Services logo. To the right of that is the FDA logo, which is a blue square with the letters "FDA" in white. To the right of the square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue.
March 5, 2018
Micro-Tech (Nanjing) CO., Ltd. Becky Li Quality and Regulatory Affairs Director NO. 10 Gaoke Third Road Nanjing, Jiangsu 210032 China
Re: K172663 Trade/Device Name: Cytology Brush Regulation Number: 21 CFR& 876.1500 Regulation Name: Endoscope and accessories Regulatory Class: II Product Code: FDX Dated: January 31, 2018 Received: February 2, 2018
Dear Becky Li:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{1}------------------------------------------------
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies.
You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/MedicalDevices/DeviceRegulationandGuidance/) and CDRH Learn (http://www.fda.gov/Training/CDRHLearn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (http://www.fda.gov/DICE) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely.
Charles Viviano -S
Benjamin R. Fisher, Ph.D. For Director Division of Reproductive, Gastro-Renal, and Urological Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
DEPARTMENT OF HEALTH AND HUMAN SERVICES Food and Drug Administration
# Indications for Use
510(k) Number (if known)
K172663
Device Name
Cytology Brush
Indications for Use (Describe)
The Cytology Brush is used to collect cells from the bronchi and upper and lower gastrointestinal tracts.
Type of Use (Select one or both, as applicable)
X Prescription Use (Part 21 CFR 801 Subpart D)
_ Over-The-Counter Use (21 CFR 801 Subpart C)
#### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Druq Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
Form Approved: OMB No. 0910-0120 Expiration Date: January 31, 2017 See PRA Statement below.
{3}------------------------------------------------
Image /page/3/Picture/1 description: The image shows the logo for Micro-Tech. The logo consists of a blue, three-dimensional block letter "M" and "T" on the left side. To the right of the letters is the text "MICRO-TECH" in blue, with a line underneath it. Below the line are two Chinese characters, also in blue.
Tab 2
510K Summary
# 510K Summary
This 510(k) Summary of 510(k) safety and effectiveness information is being submitted in accordance with requirements of SMDA 1990 and Title 21, CFR Section 807.92.
The assigned 510(k) Number: __K172663
# 1. Date of Preparation: 2018-01-02
# 2. Sponsor Identification
# Micro-Tech (Nanjing) Co., Ltd.
No.10 Gaoke Third Road, Nanjing National Hi-Tech, Industrial Development
Zone, Nanjing, Jiangsu Province, PRC
Establishment Registration Number: 3004837686
Contact Person: Becky Li
Position: Quality and Regulatory Affairs Director
Tel: +86-25-58646378
Fax: +86-25-58350006
Email: In@micro-tech.com.cn
## 3. Identification of Proposed Device
Trade Name: Cytoloqy Brush
Common Name: Endoscopic Cytology Brush
## Regulatory Information
Classification Name: Endoscopes and Accessories
Classification: 2
Product Code: FDX
Regulation Number: 876.1500
Review Panel: Gastroenterology/Urology
Intended Use Statement:
{4}------------------------------------------------
Image /page/4/Picture/1 description: The image contains the logo for MICRO-TECH. The logo consists of a stylized blue "MT" and the words "MICRO-TECH" in blue. Below the words "MICRO-TECH" are two Chinese characters, also in blue. A blue line underlines the word "MICRO-TECH".
510K Summary
The Cytology Brush is used to collect cells from the bronchi and upper and lower gastrointestinal tracts.
#### 4. Identification of Predicate Device
510(k) Number: K896318 Product Name: Wilson-Cook Cytology Brush Manufacturer: Wilson-Cook Medical, Inc.
#### 5. Indications for Use
The Cytology Brush is used to collect cells from the bronchi and upper and lower gastrointestinal tracts.
#### 6. Device Description
The main component of the proposed device is Brush Head, Outer Sheath and Handle. The main operation is move the Finger Ring back and forth to achieve the movement of Brush Head, then the Brush Head can collect cells from the target site.
The proposed device has seven (7) specifications, the main differences of these specifications are Diameter of Brush Head, Diameter of Outer Sheath and Working Length, Color of Finger Ring.
The proposed devices are EO sterilized to achieve the Sterility Assurance Level (SAL) of 10° and placed in a sterility maintenance package to ensure a shelf life of 2 years.
# 7. Comparison of Technological Characteristics
The Cytology Brush of Micro-Tech (Nanjing) CO., Ltd. incorporates substantially equivalent materials, design, configuration, packaging fundamental technology, manufacturing processes, sterilization process and intended use as those featured in the predicate devices (K896318).
{5}------------------------------------------------
Image /page/5/Picture/1 description: The image contains the logo for MICRO-TECH. The logo consists of the letters "MT" in a stylized, three-dimensional font, with the letters appearing to be constructed from blue blocks. To the right of the "MT" is the text "MICRO-TECH" in a simple, sans-serif font, with a thin line underneath. Below the line are two Chinese characters.
#### 8. Performance Data
The proposed device Cytology Brush meets the requirements of AAMI ANSI ISO 10993-1 :2009/(R)2013 "Biological Evaluation of Medical Devices – Part 1: Evaluation and Testing Within A Risk Management Process", ISO 11135:2014 "Sterilization of Health Care products - Ethylene Oxide - Requirements for Development, Validation, and Routine Control of Sterilization processes for Medical Devices", and ISO 10993-7 Second Edition 2008-10-15 "Biological Evaluation Of Medical Devices - Part 7: Ethylene Oxide Sterilization Residuals [Including: Technical Corrigendum 1 (2009)]".
The device specific guidance document was consulted in preparing this premarket submission, <Guidance for the Content of Premarket Notifications for Biopsy Devices Used in Gastroenterology and Urology>. Performance testing such as Dimension Testing, Tensile Strength Testing and etc, were performed.
The testing performed demonstrated that the proposed device meets the same performance requirements and is Substantially Equivalent (SE) to the currently cleared predicate device (K896318).
## 9. Clinical Test Conclusion
Clinical Test is not applicable for the proposed device. No clinical study is included in this submission.
#### 10. Substantially Equivalent (SE) Conclusion
Based on the indications for use, technological characteristics, and safety and performance testing, the Cytology Brush has been shown to be appropriate for its intended use and is considered to be substantially equivalent to the currently cleared predicate devices (K896318).