POLYSCOPE FLEXIBLE ENDOSCOPY CATHETER; 10,000 PIXEL FIBER OPTIC; XENON LIGHT SOURCE, MODEL LS-200

K091962 · Lumenis, Inc. · ODF · Sep 30, 2009 · Gastroenterology, Urology

Device Facts

Record IDK091962
Device NamePOLYSCOPE FLEXIBLE ENDOSCOPY CATHETER; 10,000 PIXEL FIBER OPTIC; XENON LIGHT SOURCE, MODEL LS-200
ApplicantLumenis, Inc.
Product CodeODF · Gastroenterology, Urology
Decision DateSep 30, 2009
DecisionSESE
Submission TypeTraditional
Regulation21 CFR 876.1500
Device ClassClass 2
AttributesTherapeutic

Indications for Use

The PolyScope Flexible Endoscope is intended to provide direct visualization for diagnosis and treatment within the bladder, urethra, ureter, and kidney and during endoscopic procedures in the pancreaticobiliary system including the hepatic ducts. The PolyScope Xenon Light Source is intended to be used with an endoscope to provide illumination during endoscopic procedures.

Device Story

PolyScope system consists of disposable, flexible, steerable mini-endoscope and reusable optical components; paired with LS-200 Xenon light source. System provides white light illumination for endoscopic visualization. Used by clinicians in gastroenterology and urology settings for diagnostic and therapeutic procedures. Provides direct visualization of internal anatomy to assist physician in performing interventions. Benefits include flexible access to complex anatomical structures like hepatic/pancreatic ducts and renal system.

Clinical Evidence

Bench testing only. Safety, performance, and functional testing conducted to verify device performance against requirements.

Technological Characteristics

System comprises disposable, flexible, steerable mini-endoscope and modular, reusable optical components. Xenon light source (LS-200) provides white sunlight-spectrum illumination. Connectivity via standard endoscopic interfaces. No software-based algorithms or digital processing described.

Indications for Use

Indicated for diagnostic and therapeutic flexible endoscopic procedures in gastroenterology (biliary system, hepatic and pancreatic ducts) and urology (bladder, ureter, renal pelvis, kidney). Light source indicated for illumination during video, fiber, micro-endoscopy, rigid optics, and forehead illumination.

Regulatory Classification

Identification

An endoscope and accessories is a device used to provide access, illumination, and allow observation or manipulation of body cavities, hollow organs, and canals. The device consists of various rigid or flexible instruments that are inserted into body spaces and may include an optical system for conveying an image to the user's eye and their accessories may assist in gaining access or increase the versatility and augment the capabilities of the devices. Examples of devices that are within this generic type of device include cleaning accessories for endoscopes, photographic accessories for endoscopes, nonpowered anoscopes, binolcular attachments for endoscopes, pocket battery boxes, flexible or rigid choledochoscopes, colonoscopes, diagnostic cystoscopes, cystourethroscopes, enteroscopes, esophagogastroduodenoscopes, rigid esophagoscopes, fiberoptic illuminators for endoscopes, incandescent endoscope lamps, biliary pancreatoscopes, proctoscopes, resectoscopes, nephroscopes, sigmoidoscopes, ureteroscopes, urethroscopes, endomagnetic retrievers, cytology brushes for endoscopes, and lubricating jelly for transurethral surgical instruments. This section does not apply to endoscopes that have specialized uses in other medical specialty areas and that are covered by classification regulations in other parts of the device classification regulations.

Special Controls

*Classification* —(1)*Class II (special controls).* The device, when it is an endoscope disinfectant basin, which consists solely of a container that holds disinfectant and endoscopes and accessories; an endoscopic magnetic retriever intended for single use; sterile scissors for cystoscope intended for single use; a disposable, non-powered endoscopic grasping/cutting instrument intended for single use; a diagnostic incandescent light source; a fiberoptic photographic light source; a routine fiberoptic light source; an endoscopic sponge carrier; a xenon arc endoscope light source; an endoscope transformer; an LED light source; or a gastroenterology-urology endoscopic guidewire, is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 876.9.(2) Class I for the photographic accessories for endoscope, miscellaneous bulb adapter for endoscope, binocular attachment for endoscope, eyepiece attachment for prescription lens, teaching attachment, inflation bulb, measuring device for panendoscope, photographic equipment for physiologic function monitor, special lens instrument for endoscope, smoke removal tube, rechargeable battery box, pocket battery box, bite block for endoscope, and cleaning brush for endoscope. The devices subject to this paragraph (b)(2) are exempt from the premarket notification procedures in subpart E of part 807of this chapter, subject to the limitations in § 876.9.

Predicate Devices

Related Devices

Submission Summary (Full Text)

{0}------------------------------------------------ SEP 90 2009 ## 510(k) SUMMARY OF SAFETY AND EFFECTIVENESS INFORMATION This 510(k) summary of safety and effectiveness information is being submitted in accordance with the requirements of SMDA 1990. #### SUBMITTER INFORMATION: A. Submitter's Name: Address: Lumenis, Inc. · 5302 Betsy Ross Drive Santa Clara, CA 95054 500 0 6 033 Contact: Mike Aymami Phone: 408-764-3000 Fax: 408-764-3999 . Date of Preparation: June 22, 2009 #### B. DEVICE NAME: | B.1 – PolyScope Flexible Endoscope | | |------------------------------------|------------------------------------------------------------------------------------------------| | Trade Name(s): | PolyScope Flexible Endoscope | | Common/Usual Name: | Flexible Endoscope for diagnosis and treatment in<br>gastroenterology and urology applications | | Classification Names: | KOG, Endoscope and accessories<br>FGB. Ureteroscope and accessories, flexible/rigid | | CFR Reference: | 21 CFR 876.1500, Gastroenterology-Urology Devices | Endoscope and accessories ### B.2 PolyScope Xenon Light Source Information | Device Name: | Endoscope and/or accessories | |-----------------------|----------------------------------------------------| | Trade Name(s): | PolyScope Xenon Light Source model LS-200 | | Common/Usual name: | Xenon Light Source for endoscopy applications | | Classification Names: | GCT, Endoscope and accessories | | CFR Reference: | 21 CFR 876.1500, Gastroenterology-Urology Devices, | | | Endoscope and accessories | | Classification Panel: | Gastroenterology/Urology | ## PREDICATE DEVICE: C. C.1 - PolyScope Flexible Endoscope - The Boston Scientific SpyGlass direct visualization probe (K050403). 1. - 2. The Olympus VISERA Uretero-Reno Videoscope, Olympus XURF type V -(K072957) ## C.2 - PolyScope Xenon Light Source The Olympus EVIS EXERA Xenon Light Source CLV-160A (K051645) {1}------------------------------------------------ ### DEVICE DESCRIPTION: Rog1962 pg 2 of 2 The PolyScope Flexible Endoscope is comprised from two main components: a disposable, flexible, steerable mini-endoscope and a flexible, modular, reusable optical disposable, Holters, etc. to the catheter to allow the visualization of the desired area. The PolyScope LS 200 XENON-Endoscopic light source provides white examination light (sunlight spectrum) for all endoscopic applications including: - During video endoscopy. . - During fiber endoscopy. � - During micro-endoscopy. ◆ - During endoscopy using rigid optics. . - During usage of forehead illumination. ● #### E. INTENDED USE: The PolyScope Flexible Endoscope is intended to provide direct visualization for diagnosis and treatment within the bladder, urethra, ureter, and kidney and during endoscopic procedures in the pancreaticobiliary system including the hepatic ducts. The PolyScope Xenon Light Source is intended to be used with an endoscope to provide illumination during endoscopic procedures. # F. # TECHNOLOGICAL CHARACTERISTICS SUMMARY & SUBSTANTIAL EQUIVALENCE STATEMENT: The subject device, the PolyScope Flexible Endoscope, has the same intended use, general design and same fundamental scientific technology as the predicate devices (K050403, K072957). Also, the PolyScope Xenon Light Source has the same intended (11050-103) 110727 110 same fundamental scientific technology as the predicate device (K051645). The PolyScope Flexible Endoscope and the PolyScope Xenon Light Source use a The I orgoode I londolo equivalent to their predicates (K050403, K072957 and K051645 technology submantary og new hazards or ne question regarding safety and effectiveness introduced by the PolyScope Flexible Endoscope and the PolyScope Xenon Light Source as compared with the predicate devices. #### PERFORMANCE DATA SUMMARY: (". The appropriate testing including safety, performance and functional testing to determine The uppropriate coning after PolyScope Flexible Endoscope and the PolyScope Xenon Light Source has been conducted. та извързанит автория и подрежния D. {2}------------------------------------------------ ## DEPARTMENT OF HEALTH & HUMAN SERVICES Public Health Service SEP 30 2009 Food and Drug Administration 10903 New Hampshire Avenue Document Control. Room - WO66-G609 Silver Spring, MD 20993-0002 Mr. Mike Aymami Global Director Regulatory and Quality Systems Compliance Lumenis, Inc. 5302 Betsy Ross Drive SANTA CLARA CA 95054 Re: K091962 Trade/Device Name: PolyScope Flexible Endoscope Regulation Number: 21 CFR 876.1500 Regulation Name: Endoscope and accessories Regulatory Class: II Product Code: ODF Dated: June 29, 2009 Received: July 1, 2009 Dear Mr. Aymami: We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. · If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register. Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical {3}------------------------------------------------ Page 2 - device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050. If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please go to http://www.fda.gov/AboutFDA/CentersOffices/CDRH/CDRHOffices/ucm115809.htm for the Center for Devices and Radiological Health's (CDRH's) Office of Compliance. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance. You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.html Sincerely yours, Janine M. Morris Janine M. Morris Acting Director, Division of Reproductive, Abdominal, and Radiological Devices Office of Device Evaluation Center for Devices and Radiological Health Enclosure {4}------------------------------------------------ #### Indications for Use Statement 1.3 510(k) Number (if known): K 0 9 19 62 Device Name: PolyScope Flexible Endoscope PolyScope Xenon Light Source Indications for Use: The PolyScope Flexible Endoscope is indicated in flexible endoscopic procedures for diagnostic and therapeutic applications where flexible endoscopes are standard of care including: - Gastroenterology, including procedure of the Biliary system involving the hepatic 피 and pancreatic ducts. - Urology, including procedures involving the bladder, ureter, renal pelvis and kidney. 1 The PolyScope Xenon-Endoscopic light source LS 200 provides white examination light (sunlight spectrum) for all endoscopic applications including: - During video endoscopy. 1 - During fiber endoscopy. ■ - During micro-endoscopy. . - During endoscopy using rigid optics. - During usage of forehead illumination. . X Prescription Use (Part 21 CFR 801 Subpart D) AND/OR Over-The-Counter Use (21 CFR 801 Subpart C) (PLEASE DO NOT WRITE BELOW THIS LINE -CONTINUE ON ANOTHER PAGE IF NEEDED) concurrence of cdrh, office of device evaluation (ode) Holmen (Division Sign-Off) (Division Sign-On) Division of Reproductive, Abdominal, and Radiological Devices 510(k) Number 6
Innolitics

Panel 1

/
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...