EnlightenVue Microendoscopy System

K192300 · Enlightenvue, Inc. · GCJ · Mar 20, 2020 · Gastroenterology, Urology

Device Facts

Record IDK192300
Device NameEnlightenVue Microendoscopy System
ApplicantEnlightenvue, Inc.
Product CodeGCJ · Gastroenterology, Urology
Decision DateMar 20, 2020
DecisionSESE
Submission TypeTraditional
Regulation21 CFR 876.1500
Device ClassClass 2

Indications for Use

The enlightenVue Microendoscopy System is intended for visualization of body cavities, hollow organs, and canals. The surgiVue microendoscope is designed to be introduced through natural body cavities or surgical incisions through introducers, trocars, needles, sheaths, or other devices with lumens having inside diameters larger than the outer diameter of the microendoscope.

Device Story

System comprises surgiVue microendoscope and surgiTrac laser light source; provides internal anatomy imaging for diagnostic/interventional procedures. Laser light source focuses white light onto fiber; interfaces with disposable microendoscope and PC for image illumination/capture. Used in clinical settings by physicians. Output displayed on PC for visualization; assists clinical decision-making by enabling direct anatomical inspection. Benefits include minimally invasive access via natural orifices or existing surgical ports. Features expandable tip, two accessory channels, integrated camera, and light source.

Clinical Evidence

Bench testing only. Includes biocompatibility (cytotoxicity, irritation, sensitization), accelerated/real-time aging, shelf-life verification, mechanical/dimensional verification, laser safety testing, and functional verification of PEMS, FPGA, video system, and camera specifications. No clinical data presented.

Technological Characteristics

Disposable microendoscope; materials include Polypropylene, Polyimide, Pebax (7233, 5533, 3533, 4033), Nylon, Polysulfone, Polyester, PMMA, Fluorinated Polymer, Stainless Steel, Polycarbonate, Polyolefin, ABS, PVDF. Features 120° FOV, 90° direction of view. Class 3R laser light source. Sterilized via EtO. Includes two accessory channels (max 1.0mm diameter). PC-based video system connectivity.

Indications for Use

Indicated for visualization of body cavities, hollow organs, and canals in patients requiring endoscopic examination via natural orifices or surgical incisions.

Regulatory Classification

Identification

An endoscope and accessories is a device used to provide access, illumination, and allow observation or manipulation of body cavities, hollow organs, and canals. The device consists of various rigid or flexible instruments that are inserted into body spaces and may include an optical system for conveying an image to the user's eye and their accessories may assist in gaining access or increase the versatility and augment the capabilities of the devices. Examples of devices that are within this generic type of device include cleaning accessories for endoscopes, photographic accessories for endoscopes, nonpowered anoscopes, binolcular attachments for endoscopes, pocket battery boxes, flexible or rigid choledochoscopes, colonoscopes, diagnostic cystoscopes, cystourethroscopes, enteroscopes, esophagogastroduodenoscopes, rigid esophagoscopes, fiberoptic illuminators for endoscopes, incandescent endoscope lamps, biliary pancreatoscopes, proctoscopes, resectoscopes, nephroscopes, sigmoidoscopes, ureteroscopes, urethroscopes, endomagnetic retrievers, cytology brushes for endoscopes, and lubricating jelly for transurethral surgical instruments. This section does not apply to endoscopes that have specialized uses in other medical specialty areas and that are covered by classification regulations in other parts of the device classification regulations.

Special Controls

*Classification* —(1)*Class II (special controls).* The device, when it is an endoscope disinfectant basin, which consists solely of a container that holds disinfectant and endoscopes and accessories; an endoscopic magnetic retriever intended for single use; sterile scissors for cystoscope intended for single use; a disposable, non-powered endoscopic grasping/cutting instrument intended for single use; a diagnostic incandescent light source; a fiberoptic photographic light source; a routine fiberoptic light source; an endoscopic sponge carrier; a xenon arc endoscope light source; an endoscope transformer; an LED light source; or a gastroenterology-urology endoscopic guidewire, is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 876.9.(2) Class I for the photographic accessories for endoscope, miscellaneous bulb adapter for endoscope, binocular attachment for endoscope, eyepiece attachment for prescription lens, teaching attachment, inflation bulb, measuring device for panendoscope, photographic equipment for physiologic function monitor, special lens instrument for endoscope, smoke removal tube, rechargeable battery box, pocket battery box, bite block for endoscope, and cleaning brush for endoscope. The devices subject to this paragraph (b)(2) are exempt from the premarket notification procedures in subpart E of part 807of this chapter, subject to the limitations in § 876.9.

In combination with the general controls of the FD&C Act, the integrated operating table-electromechanical surgical system is subject to the following special controls: 1. (1) Premarket clinical performance testing, or a combination of premarket clinical performance testing and postmarket surveillance (in accordance with special control (2)), must include the following: 1. (i) Objective performance measures (e.g., rate and number of conversions to other surgical modalities, rate of device related adverse events (including tissue injury, hematoma, and increased blood loss), and their severity, cause, and outcomes) must be reported with relevant descriptive comparator performance measures. 2. (ii) The data must demonstrate the performance of the device for providing accurate and precise control of attached surgical instruments in range of clinical conditions relevant to the device's intended use. 3. (iii) The test dataset must include data collected from a patient population representative of the intended patient population under anticipated conditions of use. 2. (2) Data obtained from postmarket surveillance must demonstrate, in consideration of the premarket data obtained in accordance with special control (1), that the device performs in accordance with special control (1), unless FDA determines, based on the totality of the premarket data, that data from postmarket surveillance is not required to demonstrate that the device performs as intended. Such postmarket surveillance must be conducted per a protocol determined appropriate by FDA to demonstrate that the device performs as intended (in consideration of the premarket data obtained in accordance with special control (1)), and must include initiation, enrollment, and reporting requirements to ensure timely periodic updates to FDA on post-market surveillance progress and outcomes. 3. (3) Animal performance testing must evaluate the extent of port site trauma due to repositioning of table during surgical procedures when utilizing robotic minimally invasive and laparoscopic approaches 4. (4) The device manufacturer must develop, and update as necessary, a device-specific use training program that ensures proper device setup/use/shutdown, accurate control of instruments to perform the intended surgical procedures, troubleshooting and handling during unexpected events or emergencies, and safe practices to mitigate use error. 5. (5) The device manufacturer may only distribute the device to facilities that implement and maintain the device-specific use training program and ensure that users of the device have completed the device-specific use training program. (6) Human factors assessment must demonstrate that the user can correctly use the device system across all intended use environments with the provided instructions and training materials, including patient access during normal operating conditions and emergency situations, and effects arising from the integrated nature of the operating table and robotic surgical arms. (7) Labeling must include: (i) A detailed summary of clinical performance testing conducted with the device, including study population, results, adverse events, and comparisons to any comparator groups identified; (ii) A statement in the labeling that the safety and effectiveness for the representative specific procedures was based on evaluation of the device as a surgical tool and did not include evaluation of outcomes related to the treatment of the patient's underlying disease or condition, unless FDA determines that it can be removed or modified based on clinical performance data submitted to FDA; (iii) Identification of compatible devices; (iv) The list of surgical procedures for which the device has been determined to be safe with clinical justification; (v) Reprocessing instructions for reusable components; (vi) A shelf life for any sterile components; (vii) A description of the device-specific use training program; (viii) A statement that the device is only for distribution to facilities that implement and maintain the device-specific use training program and ensure that users of the device have completed the device-specific use training program; and (ix) A summary of any completed postmarket surveillance data collected as required by special control (2), including updated labeling to accurately reflect outcomes observed in postmarket surveillance. (8) Non-clinical performance testing must demonstrate that the device performs as intended under anticipated conditions of use and must include: (i) Device motion accuracy and repeatability; (ii) System testing; (iii) Instrument reliability; (iv) Crosstalk; (v) Table motion control; (vi) Thermal effects on tissue; (vii) User-device interface performance; (viii) Workspace access testing; and (ix) Performance testing with compatible devices. (9) Software verification, validation, and hazard analysis must be performed. (10) Electromagnetic compatibility and electrical, thermal, and mechanical safety testing must be performed. (11) Performance data must demonstrate the sterility of all patient-contacting device components. (12) Performance data must support the shelf life of the device components provided sterile by demonstrating continued sterility and package integrity over the labeled shelf life. (13) Performance data must validate the reprocessing instructions for the reusable components of the device. (14) Performance data must demonstrate that all patient-contacting components of the device are biocompatible. (15) Performance data must demonstrate that all patient-contacting components of the device are non-pyrogenic. (16) The device manufacturer must submit a report to the FDA annually on the anniversary of initial marketing authorization for the device, until such time as FDA may terminate such reporting, which comprises the following information: (i) Cumulative summary, by year, of complaints and adverse events since date of initial marketing authorization; and (ii) Identification and rationale for changes made to the device, labeling, or device specific use training program, which did not require submission of a premarket notification during the reporting period.

Predicate Devices

Submission Summary (Full Text)

{0}------------------------------------------------ Image /page/0/Picture/0 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). On the left side of the image is the Department of Health & Human Services logo. To the right of that is the FDA logo, which is a blue square with the letters "FDA" in white. To the right of the blue square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue. March 20, 2020 enlightenVue. Inc. Giacomo Basadonna 1111 Race Street, Suite 6a Denver, Colorado 80206 Re: K192300 Trade/Device Name: enlightenVue Microendoscopy System Regulation Number: 21 CFR 876.1500 Regulation Name: Endoscope and Accessories Regulatory Class: Class II Product Code: GCJ Dated: February 21, 2020 Received: February 25, 2020 Dear Giacomo Basadonna: We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading. If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register. Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's {1}------------------------------------------------ requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems. For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100). Sincerely, Long Chen, Ph.D. Assistant Director DHT4A: Division of General Surgery Devices OHT4: Office of Surgical and Infection Control Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health Enclosure {2}------------------------------------------------ # Indications for Use 510(k) Number (if known) K192300 Device Name enlightenVue Microendoscopy System ### Indications for Use (Describe) The enlighten Vue Microendoscopy System is intended for visualization of body cavities, hollow organs, and canals. The surgiVue microendoscope is designed to be introduced through natural body cavities or surgical incisions through introducers, trocars, needles, sheaths, or other deving inside diameters larger than the outer diameter of the microendoscope. | Type of Use (Select one or both, as applicable) | | |-------------------------------------------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------| | <div> <span> </span> <span> × </span> <span> </span> <span> Prescription Use (Part 21 CFR 801 Subpart D) </span> </div> | <div> <span> Over-The-Counter Use (21 CFR 801 Subpart C) </span> </div> | ## CONTINUE ON A SEPARATE PAGE IF NEEDED. This section applies only to requirements of the Paperwork Reduction Act of 1995. ## *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.* The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to: > Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov "An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number." {3}------------------------------------------------ # 510(k) Summary (K192300) for the enlightenVue Microendoscopy System (per 21CFR 807.87(h)) #### 1. SUBMITTER/510(K) HOLDER enlightenVue 1111 Race Street, Suite 6a Denver, CO 80206 Contact Person: Giacomo Basadonna M.D. Telephone: 202-427-5177 Date Prepared: March 19, 2020 #### 2. DEVICE NAME Proprietary Name: enlightenVue Microendoscopy System Common/Usual Name: Endoscope Classification Name: Endoscope #### 3. PREDICATE DEVICE Galileo Disposable Endoscope (K981928) #### 4. DEVICE DESCRIPTION The enlightenVue Microendoscopy System is comprised of the surgiVue microendoscope and surgiTrac laser light source. This system is used to provide imaging capability of internal anatomy for both diagnostic and interventional procedures. The small laser light source focuses light onto a small fiber in the form of white light. The surgiTrac interfaces with the disposable surgiVue microendoscope and a PC to allow for illumination of the captured image. 5.0 {4}------------------------------------------------ #### 5. INTENDED USE The enlightenVue Microendoscopy System is intended for visualization of body cavities, hollow organs, and canals. The surgiVue microendoscope is designed to be introduced through natural body cavities or surgical incisions through introducers, trocars, needles, sheaths, or other devices with lumens having inside diameters larger than the outer diameter of the microendoscope. #### TECHNOLOGICAL CHARACTERISTICS AND SUBSTANTIAL EQUIVALENCE 6. The enlightenVue Microendoscopy System is substantially equivalent in function, and intended use to the predicate, the Galileo Disposable Endoscopes, subject of K981928. The intended use of the enlightenVue Microendoscopy System and the predicate device is identical in that they are all indicated for visualization of body cavities, hollow organs, and canals. They are designed to be introduced through natural body cavities or surgical incisions through introducers, trocars, needles, sheaths, or other devices with lumens having inside diameters larger than the outer diameter of the endoscopes. The technological characteristics of the enlightenVue Microendoscopy System and the predicate device are identical in that they both offer a channel for either viewing body cavities, tissues, organs, or canals and an optional channel for passing instruments or fluids into the desired anatomical sites. Both devices are provided sterile and intended for single use only. The devices are different in that the subject device includes an expandable tip, two accessory channels, a light source, and a camera. A reference device is used for comparison to validate the technological differences. Table 1, on the next page, compares the subject device with the predicate. {5}------------------------------------------------ | PRODUCT<br>CHARACTERISTICS | ENLIGHTENVUE<br>MICROENDOSCOPY<br>SYSTEM | PREDICATE<br>GALILEO<br>DISPOSABLE<br>ENDOSCOPE | |----------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-------------------------------------------------| | Intended Use | The intended use of the enlightenVue<br>Microendoscopy System is intended for the<br>visualization of body cavities, hollow organs, and<br>canals. It is also designed to be introduced through<br>natural body orifices cavities or surgical incisions<br>through introducers, trocars, needles, sheaths, or<br>other devices with lumens having inside diameters<br>larger than the outer diameter of the endoscope | | | Product Code | GCJ | GCJ | | Classification | Class II | Class II | | Materials | surgiVue: Polypropylene<br>Impact Copolymer,<br>Polyimide, Pebax 7233,<br>Pebax 5533, Pebax 3533,<br>Nylon, Polysulfone,<br>Polyester,<br>Polymethylmethacrylate<br>Fluorinated Polymer,<br>Stainless Steel,<br>Polycarbonate, Polyolefin,<br>Pebax 4033, ABS,<br>Polyvinylidene Fluoride | N/A | | Reusable or Disposable | Disposable | Disposable | | Light Source | Class 3R Laser | External | | Field of View | 120° | 40°-70° | | Direction of View | 90° | 0°-30° | | Biocompatible | Yes | Yes | | Sterilization Method | EtO | EtO | | Accessories | Accessory tools with a<br>maximum diameter of<br>1.0mm can be used in<br>conjunction with the white<br>instrument channel female<br>luer. Active endoscopic<br>accessories, such as laser<br>probes and electrosurgical<br>equipment may not be used<br>in conjunction with the<br>enlightenVue<br>Microendoscopy System. | N/A | | Regulatory Status | K192300 | K081028 | ## Table 1: Side-by-Side Comparison of enlightenVue Microendoscopy System with Predicate Device {6}------------------------------------------------ #### 7. PERFORMANCE TESTING Biocompatibility, packaging, mechanical and functional testing demonstrate that the enlightenVue Microendoscopy System is biocompatible and mechanically and functionally similar to the predicate device. The following tests have been performed: - Accelerated Aging, 6 and 12 months ● - Real-Time Aging, 3 months ● - Shelf life Verification ● - Cytotoxicity Test ● - Irritation Test - Sensitization Test - . Verification Test: PEMS System, Video System - Verification Test: PEMS System ● - Verification Test: PEMS System, FPGA ● - Dimensional Verification ● - 0 Cable Length - o Working Length - Working Diameter o - Secondary Channel Color and Symbol O - Working Channel Color and Symbol o - Handpiece Gap O - o Heat Shrink - Sharp Edges o - Verification Test: surgiTrac ● - Physical Characteristics o - Electrical Characteristics o - Safety Features о - Functional Characteristics о - Video System Communications O - Video Stream o - Functional Verification: surgiVue Microendoscope ● - Cameral Specifications o - о Endoscope functionality - Endoscope Strength and Durability o - Serialization Check Test O - Radiant Power Distribution Test O - о Secondary Channel Camera Rinse Test - o Water Ingress Test - Usability Testing ● {7}------------------------------------------------ - . Laser Safety Testing Evaluations were performed to determine the material and mechanical characteristics of the enlightenVue Microendoscopy System according to FDA recognized standards. In conclusion, the functional verification and validation testing have been performed which demonstrate that enlightenVue Microendoscopy System functions as intended and is safe and effective for its intended use. #### CONCLUSION 8. The subject device is substantially equivalent to the predicate device. These conclusions are based upon the facts that the subject device is identical in intended use to the predicate, extensive validation testing was performed, and a reference device is used for comparison to validate different technological characteristics - the camera and light source - to the subject device. Any differences do not raise new types of questions of safety and effectiveness.
Innolitics
Guidance Recalls

Panel 1

/
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...