K033754 · Wilson-Cook Medical, Inc. · OCY · Dec 30, 2003 · Gastroenterology, Urology
Device Facts
Record ID
K033754
Device Name
TRACER METRO SMART WIRE GUIDE
Applicant
Wilson-Cook Medical, Inc.
Product Code
OCY · Gastroenterology, Urology
Decision Date
Dec 30, 2003
Decision
SESE
Submission Type
Special
Regulation
21 CFR 876.1500
Device Class
Class 2
Indications for Use
Wilson-Cook's Tracer Metro Smart Wire Guide is intended to assist in cannulation of the biliary and pancreatic ducts and to aid in bridging difficult strictures during ERCP.
Device Story
Tracer Metro Smart Wire Guide is a .035" diameter wire guide used during Endoscopic Retrograde Cholangiopancreatography (ERCP). Device assists physicians in cannulating biliary and pancreatic ducts and bridging difficult strictures. It is compatible with Wilson-Cook accessories. Device functions as a mechanical guide wire to facilitate access within the ductal system. Used in clinical settings by gastroenterologists or trained endoscopists. Benefits include improved procedural access to complex anatomical structures.
Clinical Evidence
No clinical data provided; substantial equivalence is based on performance characteristics and biocompatibility testing.
Technological Characteristics
Wire guide, .035" diameter. Compatible with Wilson-Cook accessories. Mechanical device; no energy source or software.
Indications for Use
Indicated for patients undergoing ERCP requiring assistance in cannulation of the biliary and pancreatic ducts or aid in bridging difficult strictures.
Regulatory Classification
Identification
An endoscope and accessories is a device used to provide access, illumination, and allow observation or manipulation of body cavities, hollow organs, and canals. The device consists of various rigid or flexible instruments that are inserted into body spaces and may include an optical system for conveying an image to the user's eye and their accessories may assist in gaining access or increase the versatility and augment the capabilities of the devices. Examples of devices that are within this generic type of device include cleaning accessories for endoscopes, photographic accessories for endoscopes, nonpowered anoscopes, binolcular attachments for endoscopes, pocket battery boxes, flexible or rigid choledochoscopes, colonoscopes, diagnostic cystoscopes, cystourethroscopes, enteroscopes, esophagogastroduodenoscopes, rigid esophagoscopes, fiberoptic illuminators for endoscopes, incandescent endoscope lamps, biliary pancreatoscopes, proctoscopes, resectoscopes, nephroscopes, sigmoidoscopes, ureteroscopes, urethroscopes, endomagnetic retrievers, cytology brushes for endoscopes, and lubricating jelly for transurethral surgical instruments. This section does not apply to endoscopes that have specialized uses in other medical specialty areas and that are covered by classification regulations in other parts of the device classification regulations.
Special Controls
*Classification* —(1)*Class II (special controls).* The device, when it is an endoscope disinfectant basin, which consists solely of a container that holds disinfectant and endoscopes and accessories; an endoscopic magnetic retriever intended for single use; sterile scissors for cystoscope intended for single use; a disposable, non-powered endoscopic grasping/cutting instrument intended for single use; a diagnostic incandescent light source; a fiberoptic photographic light source; a routine fiberoptic light source; an endoscopic sponge carrier; a xenon arc endoscope light source; an endoscope transformer; an LED light source; or a gastroenterology-urology endoscopic guidewire, is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 876.9.(2) Class I for the photographic accessories for endoscope, miscellaneous bulb adapter for endoscope, binocular attachment for endoscope, eyepiece attachment for prescription lens, teaching attachment, inflation bulb, measuring device for panendoscope, photographic equipment for physiologic function monitor, special lens instrument for endoscope, smoke removal tube, rechargeable battery box, pocket battery box, bite block for endoscope, and cleaning brush for endoscope. The devices subject to this paragraph (b)(2) are exempt from the premarket notification procedures in subpart E of part 807of this chapter, subject to the limitations in § 876.9.
Predicate Devices
Wilson-Cook Wire Guide (K9910497)
Submission Summary (Full Text)
{0}------------------------------------------------
Image /page/0/Picture/1 description: The image is a black and white logo for the U.S. Department of Health and Human Services. The logo features a stylized eagle with three stripes forming its wing and tail feathers. The words "DEPARTMENT OF HEALTH & HUMAN SERVICES • USA" are arranged in a circular pattern around the eagle.
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002
Wilson-Cook Medical, Inc. Ms. Marge Walls-Walker Regulatory Affairs Specialist 4900 Bethania Station Road Winston-Salem, NC 27105
JUL 2 7 2015
K033754 Re:
Trade/Device Name: TRACER METRO SMART WIRE GUIDE Modification to Wilson-Cook Wire Guide Regulation Number: 21 CFR 876.1500 Regulation Name: Endoscope and accessories Regulatory Class: II Product Code: OCY Dated (Date on orig SE ltr): November 25, 2003 Received (Date on orig SE ltr): December 1, 2003
Dear Ms. Walls-Walker,
This letter corrects our substantially equivalent letter of December 30, 2003.
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be
{1}------------------------------------------------
Page 2 -
found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638 2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerely yours,
# Benjamin R. Fisher -S
Benjamin R. Fisher, Ph.D. Director Division of Reproductive, Gastro-Renal, and Urological Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
510(k) Number (if known): _K_033 75 4
Device Name: Wilson-Cook Tracer Metro Smart Wire Guide
Indications for Use:
The Tracer Metro Smart Wire Guide is intended to assist in cannulation of the billary and The Tracer Metro Only . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
### (Please DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE-FF NEEDED) Concurrence of CDRH, Office of Device Evaluation (ODE)
**Prescription Use Only**
(Per 21 CFR § 801.109
OR
Over-the-Counter_
Nancyc Brogdon
(Division Sign-Off)
Division of Reproductive, Abdominal,
and Radiological Devices,
510(k) Number K033754
{3}------------------------------------------------
## DEC 3 0 2003
/<033754
#### ATTACHMENT F: 510(k) Summary of Safety and Effectiveness
#### SPONSOR:
CONTACT/SUBMITTER:
DATE OF SUBMISSION:
DEVICE:
Trade Name: Common Name: Classification:
#### PREDICATE DEVICES:
INTENDED USE:
DEVICE DESCRIPTION:
COMPARISON OF CHARACTERISITICS:
PERFORMANCE DATA:
Wilson-Cook Medical 4900 Bethania Station Road Winston-Salem, NC 27105
Marge Walls-Walker Regulatory Affairs Specialist [336] 744-0157 Ex.290
November 25, 2003
Tracer Metro Smart Wire Guide
Tracer Metro Smart Wire Guide Wire Guide Endoscope and/or Accessories, Class II 21 CFR § 876.1500. 78 KOG
Wilson-Cook Wire Guide (k9910497)
Wilson-Cook's Tracer Metro Smart Wire Guide is intended to assist in cannulation of the biliary and pancreatic ducts and to aid in bridging difficult strictures during ERCP.
The proposed Tracer Metro Smart Wire Guide is a modification to existing wire guides currently marketed by Wilson-Cock. The Tracer Metro Smart Wire Guide is .035" in diameter and is compatible with a full range of Wilson-Cook accessories.
We believe the proposed device to be substantially equivalent to currently marketed Wilson-Cook wire guides as cleared per k910497.
We believe the proposed device to be substantially equivalent to the named predicate in terms of performance characteristics tested and biocompatibility.