CELLVIZIO WITH CONFOCAL MINIPROBES, MODEL F-600; COLOFLEX AND GASTROFLEX

K061666 · Mauna Kea Technologies · GCJ · Aug 24, 2006 · Gastroenterology, Urology

Device Facts

Record IDK061666
Device NameCELLVIZIO WITH CONFOCAL MINIPROBES, MODEL F-600; COLOFLEX AND GASTROFLEX
ApplicantMauna Kea Technologies
Product CodeGCJ · Gastroenterology, Urology
Decision DateAug 24, 2006
DecisionSESE
Submission TypeSpecial
Regulation21 CFR 876.1500
Device ClassClass 2

Indications for Use

The Cellvizio® (-GI, -LUNG) with Confocal Miniprobe™ (Coloflex, Gastroflex, Alveoflex) is a confocal laser imaging system with fiber optic probes that is intended to allow confocal laser imaging of the internal microstructure of tissues in the anatomical tract, i.e., gastrointestinal or pulmonary, accessed by the endoscope.

Device Story

Cellvizio F-600 is a confocal laser imaging system for in vivo microscopic tissue visualization. Input: laser light transmitted through miniaturized fiber optic probes (Coloflex, Gastroflex, Alveoflex) inserted into the working channel (≥2.8 mm) of standard endoscopes. Operation: laser scanning unit generates light; fiber bundle transmits light to tissue and collects backscattered signal; proprietary software on remote computer processes signals into real-time images. Output: microscopic tissue images displayed on a flat panel monitor. Used in clinical settings by physicians during endoscopic procedures. Provides real-time visual feedback of internal microstructure to assist in clinical assessment of tissue.

Clinical Evidence

No clinical data provided; substantial equivalence is based on technological comparison and design validation of the F-600 system relative to the predicate F-400 system.

Technological Characteristics

Confocal laser scanning system; fiber optic probe-based imaging; compatible with standard endoscopes (≥2.8 mm working channel). Components: Laser Scanning Unit, remote computer, flat panel display, and miniaturized fiber optic probes. Connectivity: remote computer processing. Software-based image reconstruction.

Indications for Use

Indicated for patients undergoing endoscopic procedures requiring visualization of internal tissue microstructure within the gastrointestinal or pulmonary tracts.

Regulatory Classification

Identification

An endoscope and accessories is a device used to provide access, illumination, and allow observation or manipulation of body cavities, hollow organs, and canals. The device consists of various rigid or flexible instruments that are inserted into body spaces and may include an optical system for conveying an image to the user's eye and their accessories may assist in gaining access or increase the versatility and augment the capabilities of the devices. Examples of devices that are within this generic type of device include cleaning accessories for endoscopes, photographic accessories for endoscopes, nonpowered anoscopes, binolcular attachments for endoscopes, pocket battery boxes, flexible or rigid choledochoscopes, colonoscopes, diagnostic cystoscopes, cystourethroscopes, enteroscopes, esophagogastroduodenoscopes, rigid esophagoscopes, fiberoptic illuminators for endoscopes, incandescent endoscope lamps, biliary pancreatoscopes, proctoscopes, resectoscopes, nephroscopes, sigmoidoscopes, ureteroscopes, urethroscopes, endomagnetic retrievers, cytology brushes for endoscopes, and lubricating jelly for transurethral surgical instruments. This section does not apply to endoscopes that have specialized uses in other medical specialty areas and that are covered by classification regulations in other parts of the device classification regulations.

Special Controls

*Classification* —(1)*Class II (special controls).* The device, when it is an endoscope disinfectant basin, which consists solely of a container that holds disinfectant and endoscopes and accessories; an endoscopic magnetic retriever intended for single use; sterile scissors for cystoscope intended for single use; a disposable, non-powered endoscopic grasping/cutting instrument intended for single use; a diagnostic incandescent light source; a fiberoptic photographic light source; a routine fiberoptic light source; an endoscopic sponge carrier; a xenon arc endoscope light source; an endoscope transformer; an LED light source; or a gastroenterology-urology endoscopic guidewire, is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 876.9.(2) Class I for the photographic accessories for endoscope, miscellaneous bulb adapter for endoscope, binocular attachment for endoscope, eyepiece attachment for prescription lens, teaching attachment, inflation bulb, measuring device for panendoscope, photographic equipment for physiologic function monitor, special lens instrument for endoscope, smoke removal tube, rechargeable battery box, pocket battery box, bite block for endoscope, and cleaning brush for endoscope. The devices subject to this paragraph (b)(2) are exempt from the premarket notification procedures in subpart E of part 807of this chapter, subject to the limitations in § 876.9.

Predicate Devices

Reference Devices

Submission Summary (Full Text)

{0}------------------------------------------------ Ko61664 #### Mauna Kea Technologies Special 510(k): F-600 System #### Premarket Notification Special 510(k) Summary 10.0 In accordance with the Safe Medical Devices Act of 1990 and 21 CFR 807.92, a Premarket Notification 510(k) Summary for the F-600 system Premarket Notification is provided below ### 10.1 Submitter Information Company Name and Address: Mauna Kea Technologies 9, rue d'Enghien 75010 Paris France Contact Name: Fouad Tarabah, PharmD. Director of Quality and Regulatory Affairs Mauna Kea Technologies Telephone: +33 1 70 08 09 61 Fax: +33 1 48 24 12 18 e-mail: fouad@maunakeatech.com Date Prepared: May 2006 #### 10.2 Name of Device Proprietary Name: Cellvizio® (-GI, -LUNG) with Confocal Miniprobe™ (Coloflex, Gastroflex, Alveoflex) Classification Name: Endoscope and/or Accessories ## 10.3 Predicate Device(s) Information Pentax Confocal Laser System, K042740 Pentax EC-3870CILK Confocal Video Colonoscope, K042741 superDimension Bronchus, K042438 #### 10.4 Device Description The Cellvizio® (-GI, -LUNG) with Confocal Miniprobe™ (Coloflex, Gastroflex, Alveoflex) is a confocal microscope with a fiber optic probe which allows in vivo visual inspection of tissues with a microscopic resolution during an endoscopic procedure. The F-600 system has been designed to allow real- {1}------------------------------------------------ 61666 #### Mauna Kea Technologies Special 510(k): F-600 System time observations of tissues. The device is based on a common laser scanning technology adapted for imaging through a bundle of optical fibers and is thus composed of several elements: a Laser Scanning Unit, proprietary software running on a remote computer, a flat panel display and Miniaturized Fiber Optic Probes. The F-600 system can be used with any legally marketed endoscope with a working channel of 2.8 mm or greater. ### 10.5 Intended Use The Cellvizio® (-GI, -LUNG) with Confocal Miniprobe™ (Coloflex, Gastroflex, Alveoflex) is a confocal faser system that is intended to allow confocal laser imaging of the internal microstructure of tissues in the anatomical tract, i.e., gastrointestinal or respiratory, accessed by the endoscope. ### 10.6 Comparison to Predicate Device(s) The comparison to the predicate devices was based on a review of the F-400 system information included in the company's 510(k) Premarket Notification K051585 and information concerning the predicate devices that was available in the public domain. The F-400 system is the Mauna Kea Technologie's legally marketed device (K051585) which serves as the unmodified device for the Cellvizio® (-GI, -LUNG) with Confocal Miniprobe™ (Coloflex, Gastroflex, Alveoflex) special 510(k) submission. {2}------------------------------------------------ Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850 # AUG 2 4 2006 Mauna Kea Technologies % Fouad Tarabah Director of Quality and Regulatory Affairs 9, Rue D'Enghien Paris, France 75010 Re: K061666 Trade/Device Name: Cellvizio® (-GI, -LUNG) with Confocal Miniprobe™ (Coloflex, Gastroflex, Alveoflex) Regulation Number: 21 CFR 876.1500 Regulation Name: Endoscope and accessories Regulatory Class: II Product Code: GCJ Dated: June 20, 2006 Received: June 25, 2006 Dear Fouad Tarabah: We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register. Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product {3}------------------------------------------------ Page 2 -- Fouad Tarabah radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050. This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market. If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Office of Compliance at (240) 276-0115. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (240) 276-3150 or at its Internet address http://www.fda.gov/cdrh/industry/support/index.html. Sincerely vours. Vabare Preelup Mark N. Melkerson Director Division of General, Restorative and Neurological Devices Office of Device Evaluation Center for Devices and Radiological Health Enclosure {4}------------------------------------------------ Mauna Kea Technologies Special 510(k): F-600 System # ODE Indications Statement Special 510(k) Number (if known): K061666 Device Name: Cellvizio® (-GI, -LUNG) with Confocal Miniprobe™ (Coloflex, Gastroflex, Alveoflex) Indications for Use: The Cellvizio® (-GI, -LUNG) with Confocal Miniprobe™ (Coloflex, Gastroflex, Alveoflex) is a confocal laser imaging system with fiber optic probes that is intended to allow confocal laser imaging of the internal microstructure of tissues in the anatomical tract, i.e., gastrointestinal or pulmonary, accessed by the endoscope. Prescription Use: ____________________________________________________________________________________________________________________________________________________________ AND/OR Over-the-Counter Use: (Part 21 CFR 801 Subpart D) (21 CFR 807 Subpart C) # (PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NEEDED) Concurrence of CDRH, Office of Device Evaluation (ODE) Tarbar Buchen for wwn (Division Sign-Off) Division of General, Restorative, and Neurological Devices 510(k) Number K061666 F-600 System Special 510(k) July 17, 2006 Confidential Page 10 of 63
Innolitics

Panel 1

/
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...