F-400 SYSTEM

K051585 · Mauna Kea Technologies · GCJ · Sep 6, 2005 · Gastroenterology, Urology

Device Facts

Record IDK051585
Device NameF-400 SYSTEM
ApplicantMauna Kea Technologies
Product CodeGCJ · Gastroenterology, Urology
Decision DateSep 6, 2005
DecisionSESE
Submission TypeTraditional
Regulation21 CFR 876.1500
Device ClassClass 2

Indications for Use

The Mauna Kea Technologies F-400 System is a confocal laser imaging system with fiber optic probes that is intended to allow confocal laser imaging of the internal microstructure of tissues in the anatomical tract, i.e., gastrointestinal or pulmonary, accessed by the endoscope.

Device Story

F-400 system is a confocal microscope utilizing fiber optic probes for in vivo microscopic visual inspection of tissues during endoscopic procedures. System components include a laser scanning unit, proprietary software on a remote computer, medical-grade flat panel display, and miniaturized fiber optic probes. Device integrates with standard endoscopes (working channel ≥ 2.8 mm). Laser scanning technology enables real-time tissue observation. Used in clinical settings by physicians to visualize internal microstructure; output displayed on monitor to assist in clinical assessment. Benefits include real-time, high-resolution imaging of internal tissue structures.

Clinical Evidence

No clinical data assessed. Evidence consists of bench testing, design verification, and validation to FDA-recognized standards including electrical safety, electromagnetic compatibility, laser safety, bioburden, and biocompatibility.

Technological Characteristics

Confocal microscope with fiber optic probe; laser scanning technology; remote computer processing; medical-grade flat panel display. Compatible with endoscopes having ≥ 2.8 mm working channel. Tested to standards for electrical safety, EMC, laser safety, and biocompatibility.

Indications for Use

Indicated for confocal laser imaging of internal tissue microstructure in the gastrointestinal or pulmonary tracts accessed via endoscope.

Regulatory Classification

Identification

An endoscope and accessories is a device used to provide access, illumination, and allow observation or manipulation of body cavities, hollow organs, and canals. The device consists of various rigid or flexible instruments that are inserted into body spaces and may include an optical system for conveying an image to the user's eye and their accessories may assist in gaining access or increase the versatility and augment the capabilities of the devices. Examples of devices that are within this generic type of device include cleaning accessories for endoscopes, photographic accessories for endoscopes, nonpowered anoscopes, binolcular attachments for endoscopes, pocket battery boxes, flexible or rigid choledochoscopes, colonoscopes, diagnostic cystoscopes, cystourethroscopes, enteroscopes, esophagogastroduodenoscopes, rigid esophagoscopes, fiberoptic illuminators for endoscopes, incandescent endoscope lamps, biliary pancreatoscopes, proctoscopes, resectoscopes, nephroscopes, sigmoidoscopes, ureteroscopes, urethroscopes, endomagnetic retrievers, cytology brushes for endoscopes, and lubricating jelly for transurethral surgical instruments. This section does not apply to endoscopes that have specialized uses in other medical specialty areas and that are covered by classification regulations in other parts of the device classification regulations.

Special Controls

*Classification* —(1)*Class II (special controls).* The device, when it is an endoscope disinfectant basin, which consists solely of a container that holds disinfectant and endoscopes and accessories; an endoscopic magnetic retriever intended for single use; sterile scissors for cystoscope intended for single use; a disposable, non-powered endoscopic grasping/cutting instrument intended for single use; a diagnostic incandescent light source; a fiberoptic photographic light source; a routine fiberoptic light source; an endoscopic sponge carrier; a xenon arc endoscope light source; an endoscope transformer; an LED light source; or a gastroenterology-urology endoscopic guidewire, is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 876.9.(2) Class I for the photographic accessories for endoscope, miscellaneous bulb adapter for endoscope, binocular attachment for endoscope, eyepiece attachment for prescription lens, teaching attachment, inflation bulb, measuring device for panendoscope, photographic equipment for physiologic function monitor, special lens instrument for endoscope, smoke removal tube, rechargeable battery box, pocket battery box, bite block for endoscope, and cleaning brush for endoscope. The devices subject to this paragraph (b)(2) are exempt from the premarket notification procedures in subpart E of part 807of this chapter, subject to the limitations in § 876.9.

In combination with the general controls of the FD&C Act, the integrated operating table-electromechanical surgical system is subject to the following special controls: 1. (1) Premarket clinical performance testing, or a combination of premarket clinical performance testing and postmarket surveillance (in accordance with special control (2)), must include the following: 1. (i) Objective performance measures (e.g., rate and number of conversions to other surgical modalities, rate of device related adverse events (including tissue injury, hematoma, and increased blood loss), and their severity, cause, and outcomes) must be reported with relevant descriptive comparator performance measures. 2. (ii) The data must demonstrate the performance of the device for providing accurate and precise control of attached surgical instruments in range of clinical conditions relevant to the device's intended use. 3. (iii) The test dataset must include data collected from a patient population representative of the intended patient population under anticipated conditions of use. 2. (2) Data obtained from postmarket surveillance must demonstrate, in consideration of the premarket data obtained in accordance with special control (1), that the device performs in accordance with special control (1), unless FDA determines, based on the totality of the premarket data, that data from postmarket surveillance is not required to demonstrate that the device performs as intended. Such postmarket surveillance must be conducted per a protocol determined appropriate by FDA to demonstrate that the device performs as intended (in consideration of the premarket data obtained in accordance with special control (1)), and must include initiation, enrollment, and reporting requirements to ensure timely periodic updates to FDA on post-market surveillance progress and outcomes. 3. (3) Animal performance testing must evaluate the extent of port site trauma due to repositioning of table during surgical procedures when utilizing robotic minimally invasive and laparoscopic approaches 4. (4) The device manufacturer must develop, and update as necessary, a device-specific use training program that ensures proper device setup/use/shutdown, accurate control of instruments to perform the intended surgical procedures, troubleshooting and handling during unexpected events or emergencies, and safe practices to mitigate use error. 5. (5) The device manufacturer may only distribute the device to facilities that implement and maintain the device-specific use training program and ensure that users of the device have completed the device-specific use training program. (6) Human factors assessment must demonstrate that the user can correctly use the device system across all intended use environments with the provided instructions and training materials, including patient access during normal operating conditions and emergency situations, and effects arising from the integrated nature of the operating table and robotic surgical arms. (7) Labeling must include: (i) A detailed summary of clinical performance testing conducted with the device, including study population, results, adverse events, and comparisons to any comparator groups identified; (ii) A statement in the labeling that the safety and effectiveness for the representative specific procedures was based on evaluation of the device as a surgical tool and did not include evaluation of outcomes related to the treatment of the patient's underlying disease or condition, unless FDA determines that it can be removed or modified based on clinical performance data submitted to FDA; (iii) Identification of compatible devices; (iv) The list of surgical procedures for which the device has been determined to be safe with clinical justification; (v) Reprocessing instructions for reusable components; (vi) A shelf life for any sterile components; (vii) A description of the device-specific use training program; (viii) A statement that the device is only for distribution to facilities that implement and maintain the device-specific use training program and ensure that users of the device have completed the device-specific use training program; and (ix) A summary of any completed postmarket surveillance data collected as required by special control (2), including updated labeling to accurately reflect outcomes observed in postmarket surveillance. (8) Non-clinical performance testing must demonstrate that the device performs as intended under anticipated conditions of use and must include: (i) Device motion accuracy and repeatability; (ii) System testing; (iii) Instrument reliability; (iv) Crosstalk; (v) Table motion control; (vi) Thermal effects on tissue; (vii) User-device interface performance; (viii) Workspace access testing; and (ix) Performance testing with compatible devices. (9) Software verification, validation, and hazard analysis must be performed. (10) Electromagnetic compatibility and electrical, thermal, and mechanical safety testing must be performed. (11) Performance data must demonstrate the sterility of all patient-contacting device components. (12) Performance data must support the shelf life of the device components provided sterile by demonstrating continued sterility and package integrity over the labeled shelf life. (13) Performance data must validate the reprocessing instructions for the reusable components of the device. (14) Performance data must demonstrate that all patient-contacting components of the device are biocompatible. (15) Performance data must demonstrate that all patient-contacting components of the device are non-pyrogenic. (16) The device manufacturer must submit a report to the FDA annually on the anniversary of initial marketing authorization for the device, until such time as FDA may terminate such reporting, which comprises the following information: (i) Cumulative summary, by year, of complaints and adverse events since date of initial marketing authorization; and (ii) Identification and rationale for changes made to the device, labeling, or device specific use training program, which did not require submission of a premarket notification during the reporting period.

Predicate Devices

Submission Summary (Full Text)

{0}------------------------------------------------ Image /page/0/Picture/0 description: The image shows the text "SEP - 6 2005" at the top, followed by the text "K051585". Below this is a black circle with the letters "MKT" inside. To the right of the circle is the text "Mauna Kea Technologies". The image appears to be a stamp or logo with a date and company name. ## E.0 Premarket Notification 510(k) Summary ### E.1 Submitter Information Company Name and Address: Mauna Kea Technologies 9, rue d'Enghien 75010 Paris FRANCE Contact Name: Alexandre Loiseau, Ph.D. (Official Correspondent) President Mauna Kea Technologies Telephone: +33 1 48 24 06 21 Fax: +33 1 48 24 12 18 Date Prepared: June 9, 2005 ; revised August 16, 2005 ## E.2 Name of Device Proprietary Name: F-400 System Classification Name: Endoscope and/or Accessories ### E.3 Predicate Device(s) Information Pentax Confocal Laser System, K042740 Pentax EC-3870CILK Confocal Video Colonoscope, K042741 superDimension Bronchus, K042438 Mauna Kea Technologies SAS Société par actions simplifiée au capital de 163.157,34 Euros -431 268 028, RCS Paris. Siège social : 9, rue d'Enghien - 75010 Paris - France Tél : 01 48 24 03 45 - Fax : 01 48 24 12 18 F-400 System 510(k) Summary - Appendix E June 9, 2005 - Revised August 16, 2005 Confidential Page E-2 of E-4 {1}------------------------------------------------ Image /page/1/Picture/0 description: The image shows the letters "MKT" in white text inside of a black circle. To the right of the circle, the letter "M" is visible. The image is simple and contains a logo or abbreviation. # Kajis Sis Mauna Kea Technologies ### Device Description E.4 The F-400 system is a confocal microscope with a fiber optic probe which allows in vivo visual inspection of tissues with a microscopic resolution during an endoscopic procedure. The F-400 system has been designed to allow real-time observations of tissues. The device is based on a common laser scanning technology adapted for imaging through a bundle of optical fibers and is thus composed of several elements: a Laser Scanning Unit, proprietary software running on a remote computer, a medical grade flat panel display and Miniaturized Fiber Optic Probes. The F-400 system can be used with any legally marketed endoscope with a working channel of 2.8 mm or greater. ## E.5 Intended Use The Mauna Kea technologies F-400 system is a confocal laser imaging system that is intended to allow confocal laser imaging of the internal microstructure of tissues in the anatomical tract, i.e., gastrointestinal or respiratory, accessed by the endoscope. #### E.6 Comparison to Predicate Device(s) The comparison to the predicate devices was based on a review of the F-400 system information included in the company's 510(k) Premarket Notification and information concerning the predicate devices that was available in the public domain. A product brochure for the Pentax Confocal Laser System, K042740, and Pentax EC-3870CILK Confocal Video Colonoscope, K042741 was reviewed and information for the superDimension / Bronchus was obtained from the superDimension web site, www.superdimension.com. Copies of reviewed information are included in the F-400 System 510(k) Premarket Notification. The comparison reviewed general technological considerations as well as specific performance parameters for all predicate devices where information was available. Neither animal nor clinical data were assessed. Accepted scientific methods were followed to conduct non-clinical testing, i.e., design verification and validation and testing to recognized standards. The purpose of this testing served to evaluate whether safety or effectiveness had been adversely affected by technological differences between the F-400 system and the predicate devices. > Mauna Kea Technologies SAS Société par actions simplifiée au capital de 163.157,34 Euros -431 268 028, RCS Paris. Siège social : 9, rue d'Enghien - 75010 Paris - France Tél : 01 48 24 03 45 - Fax : 01 48 24 12 18 F-400 System 510(k) Summary - Appendix E June 9, 2005 - Revised August 16, 2005 Confidential Page E-3 of E-4 {2}------------------------------------------------ cs'jj's Image /page/2/Picture/1 description: The image contains the logo for Mauna Kea Technologies. The logo consists of a black circle with the letters "MKT" in white, followed by the words "Mauna Kea Technologies" in black. The text is in a sans-serif font and is aligned horizontally. Testing to FDA recognized standards was performed in the areas of electrical safety, electromagnetic compatibility, laser safety, endoscopic equipment, bioburden, soft-controlled medical devices, risk analysis, biocompatibility and packaging. In conclusion, the data from the non-clinical testing, i.e., design verification and validation and testing to FDA recognized standards demonstrated that the F-400 system is as safe and effective as all of the above mentioned predicate devices. > Mauna Kea Technologies SAS Société par actions simplifiée au capital de 163.157,34 Euros 431 268 028, RCS Paris. Siège social : 9, rue d'Enghien - 75010 Paris - France Tél : 01 48 24 03 45 - Fax : 01 48 24 12 18 F-400 System 510(k) Summary - Appendix E June 9, 2005 - Revised August 16, 2005 Confidential Page E-4 of E-4 {3}------------------------------------------------ Image /page/3/Picture/1 description: The image shows the seal of the U.S. Department of Health and Human Services. The seal features a stylized eagle with three lines representing its wings and body. The eagle is encircled by the words "DEPARTMENT OF HEALTH AND HUMAN SERVICES, USA" in a circular arrangement. The seal is black and white and appears to be a simple, official emblem. SEP -6 2005 Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850 Alexandre Loiseau, Ph.D. President Mauna Kea Technologies 9, rue d'Enghien 75010 Paris France Re: K051585 Trade/Device Name: F-400 System Regulation Number: 21 CFR 876.1500 Regulation Name: Endoscope and accessories Regulatory Class: II Product Code: GCJ Dated: August 16, 2005 Received: August 17, 2005 Dear Dr. Loiseau: We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register. Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050. {4}------------------------------------------------ This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market. If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Office of Compliance at (240) 276-0115. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its Internet address http://www.fda.gov/cdrh/dsma/dsmamain.html Sincerely yours, Saubare Buchholz for Mark N. Melkerson Acting Director Division of General, Restorative and Neurological Devices Office of Device Evaluation Center for Devices and Radiological Health Enclosure . {5}------------------------------------------------ ## Indications Statement 510(k) Number (if known): K# 051585 Device Name: F-400 System Indications for Use: The Mauna Kea Technologies F-400 System is a confocal laser imaging system with fiber optic probes that is intended to allow confocal laser imaging of the internal microstructure of tissues in the anatomical tract, i.e., gastrointestinal or pulmonary, accessed by the endoscope. Prescription Use: __ X (Part 21 CFR 801 Subpart D) AND/OR Over-the-Counter Use: (21 CFR 807 Subpart C) (PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NEEDED) Concurrence of CDRH, Office of Device Evaluation (ODE) Saubere Boucher for MXM (Division Sign-C Division of General, Restorative, and Neurological Devices **510(k) Number** K051525
Innolitics
Guidance Recalls

Panel 1

/
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...