K142556 · KARL STORZ Endoscopy-America, Inc. · FGB · Oct 23, 2014 · Gastroenterology, Urology
Device Facts
Record ID
K142556
Device Name
Flexible Video-Uretero-Choledochoscope System
Applicant
KARL STORZ Endoscopy-America, Inc.
Product Code
FGB · Gastroenterology, Urology
Decision Date
Oct 23, 2014
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 876.1500
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
The KARL STORZ Flexible Video-Uretero- Choledochoscope System is indicated for endoscopic examination in the urinary tract and can be used percutaneously to examine the interior of the kidney, and using additional accessories, to perform various diagnostic and therapeutic procedures. The Flexible Video-Uretero-Choledochoscope System is also indicated for the examination of bile ducts, and using additional accessories, to perform various diagnostic and therapeutic procedures during cholecystectomy.
Device Story
System comprises flexible video-uretero-choledochoscope and Image 1 SPIES Camera Control Unit (CCU). Device uses integrated LED light and fiber light guides for cavity illumination; CMOS imaging sensor at insertion shaft tip captures video. Sensor signals processed by CCU for display on standard HD monitor. Used by physicians in clinical settings for visualization during diagnostic/therapeutic procedures. No design changes from predicate; update reflects expanded indications and labeling. Benefits include visualization of urinary tract and bile ducts to facilitate minimally invasive interventions.
Clinical Evidence
No clinical data. Substantial equivalence demonstrated via non-clinical performance data, including optical characteristics (ISO 8600), electrical safety (IEC 60601-1), electromagnetic compatibility (IEC 60601-1-2), biocompatibility (ISO 10993), and validation of cleaning/sterilization methods.
Technological Characteristics
Flexible endoscope with CMOS sensor, LED illumination, and fiber light guides. Connects to Image 1 SPIES CCU for HD display. Biocompatible materials per ISO 10993. Tested for electrical safety (IEC 60601-1), EMC (IEC 60601-1-2), and optical performance (ISO 8600).
Indications for Use
Indicated for endoscopic examination of the urinary tract, percutaneous examination of the kidney interior, and examination of bile ducts during cholecystectomy, including use with accessories for diagnostic and therapeutic procedures.
Regulatory Classification
Identification
An endoscope and accessories is a device used to provide access, illumination, and allow observation or manipulation of body cavities, hollow organs, and canals. The device consists of various rigid or flexible instruments that are inserted into body spaces and may include an optical system for conveying an image to the user's eye and their accessories may assist in gaining access or increase the versatility and augment the capabilities of the devices. Examples of devices that are within this generic type of device include cleaning accessories for endoscopes, photographic accessories for endoscopes, nonpowered anoscopes, binolcular attachments for endoscopes, pocket battery boxes, flexible or rigid choledochoscopes, colonoscopes, diagnostic cystoscopes, cystourethroscopes, enteroscopes, esophagogastroduodenoscopes, rigid esophagoscopes, fiberoptic illuminators for endoscopes, incandescent endoscope lamps, biliary pancreatoscopes, proctoscopes, resectoscopes, nephroscopes, sigmoidoscopes, ureteroscopes, urethroscopes, endomagnetic retrievers, cytology brushes for endoscopes, and lubricating jelly for transurethral surgical instruments. This section does not apply to endoscopes that have specialized uses in other medical specialty areas and that are covered by classification regulations in other parts of the device classification regulations.
Special Controls
*Classification* —(1)*Class II (special controls).* The device, when it is an endoscope disinfectant basin, which consists solely of a container that holds disinfectant and endoscopes and accessories; an endoscopic magnetic retriever intended for single use; sterile scissors for cystoscope intended for single use; a disposable, non-powered endoscopic grasping/cutting instrument intended for single use; a diagnostic incandescent light source; a fiberoptic photographic light source; a routine fiberoptic light source; an endoscopic sponge carrier; a xenon arc endoscope light source; an endoscope transformer; an LED light source; or a gastroenterology-urology endoscopic guidewire, is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 876.9.(2) Class I for the photographic accessories for endoscope, miscellaneous bulb adapter for endoscope, binocular attachment for endoscope, eyepiece attachment for prescription lens, teaching attachment, inflation bulb, measuring device for panendoscope, photographic equipment for physiologic function monitor, special lens instrument for endoscope, smoke removal tube, rechargeable battery box, pocket battery box, bite block for endoscope, and cleaning brush for endoscope. The devices subject to this paragraph (b)(2) are exempt from the premarket notification procedures in subpart E of part 807of this chapter, subject to the limitations in § 876.9.
Predicate Devices
DUR-Digital Ureteroscope and Choledochoscope System (DUR-D) (K060269)
KARL STORZ Flexible Video-Uretero-Renoscope System (K141250)
Related Devices
K060269 — ACMI DUR-DIGITAL URETEROSCOPE AND CHOLEDOCHOSCOPE SYSTEM · Acmi Corporation · Mar 31, 2006
K141250 — FLEXIBLE VIDEO-URETERO-RENOSCOPE SYSTEM · KARL STORZ Endoscopy-America, Inc. · Jun 11, 2014
K131369 — FLEXIBLE VIDEO-URETERO-RENOSCOPE SYSTEM · KARL STORZ Endoscopy-America, Inc. · Nov 26, 2013
K241532 — Flexible Video-Choledo-Cysto-Ureteroscope System (PL-2100) · Shanghai SeeGen Photoelectric Technology Co., Ltd. · Oct 22, 2024
Submission Summary (Full Text)
{0}------------------------------------------------
Image /page/0/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of the department's name arranged in a circular fashion around a symbol. The symbol is a stylized representation of three human profiles facing to the right, with flowing lines beneath them, possibly representing water or movement. The text reads "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA".
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002
October 23, 2014
Karl Storz Endoscopy-America, Inc. Mike Samuels Regulatory Affairs Specialist 2151 E. Grand Avenue El Segundo, CA 90245
Re: K142556
> Trade/Device Name: Flexible Video-Uretero-Choledochoscope System Regulation Number: 21 CFR 876.1500 Regulation Name: Endoscope and Accessories Regulatory Class: Class II Product Codes: FGB, FBN Dated: September 10, 2014 Received: September 11, 2014
Dear Mike Samuels,
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical devicerelated adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in
{1}------------------------------------------------
the quality systems (OS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerely yours,
Herbert P.
Lerner -S
for Benjamin Fisher, Ph.D.
Director Division of Reproductive, Gastro-Renal, and Urological Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
## Indications for Use
510(k) Number (if known) K142556
Device Name
Flexible Video-Uretero-Choledochoscope System
#### Indications for Use (Describe)
The KARL STORZ Flexible Video-Uretero- Choledochoscope System is indicated for endoscopic examination in the urinary tract and can be used percutaneously to examine the interior of the kidney, and using additional accessories, to perform various diagnostic and therapeutic procedures. The Flexible Video-Uretero-Choledochoscope System is also indicated for the examination of bile ducts, and using additional accessories, to perform various diagnostic and therapeutic procedures during cholecystectomy.
Type of Use (Select one or both, as applicable)
2 Prescription Use (Part 21 CFR 801 Subpart D)
_ Over-The-Counter Use (21 CFR 801 Subpart C)
### PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON A SEPARATE PAGE IF NEEDED.
#### FOR FDA USE ONLY
Concurrence of Center for Devices and Radiological Health (CDRH) (Signature)
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Druq Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
Image /page/3/Picture/0 description: The image shows the logo for Karl Storz Endoskope. The logo is in blue and features the word "STORZ" in large, bold letters. Below the word "STORZ" is the text "KARL STORZ - ENDOSKOPE" in a smaller font.
# 510(k) SUMMARY OF SAFETY AND EFFECTIVENESS
This summary of 510(k) safety and effectiveness information is being submitted in accordance with the requirements of the Safe Medical Devices Act (SMDA) of 1990 and 21 CFR 807.92. All data included in this document is accurate and complete to the best of KSEA's knowledge.
| Applicant: | Karl Storz Endoscopy-America, Inc<br>2151 E. Grand Avenue<br>El Segundo, CA 90245 |
|-----------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Contact: | Mike Samuels |
| | Regulatory Affairs Specialist |
| | Phone: (424) 218-8288 |
| | Fax: (424) 218-8519 |
| Date of Preparation: | September 10, 2014 |
| Device Identification: | Trade Name: Flexible Choledochoscope |
| | Common Name: Flexible Video-Uretero-Choledochoscope<br>System |
| | Classification Name: Choledochoscope and Accessories,<br>Flexible/ Rigid |
| Product Code: | FGB and FBN |
| Regulation: | 21 CFR part 876.1500 |
| Predicate Device(s): | The primary predicate device is the DUR-Digital<br>Ureteroscope and Choledochoscope System (DUR®-D),<br>K060269, manufactured by ACMI Corporation. The<br>secondary predicate is the KARL STORZ Flexible Video-<br>Uretero-Renoscope System, K141250. |
| Device Description: | The Flexible Video-Uretero-Choledochoscope System is used<br>for visualization purposes during diagnostic and therapeutic<br>procedures. The system components are the Flexible Video-<br>Uretero-Choledochoscope and the Image 1 SPIES Camera<br>Control Unit (CCU). The Flexible Video-Uretero-<br>Choledochoscope uses an LED light integrated in the handle<br>and fiber light guides to illuminate the cavity under<br>examination. The video image is produced by a<br>complementary metal-oxide-semiconductor (CMOS)<br>imaging sensor located at the tip of the insertion shaft. The<br>imaging sensor transfers the video signal to the Image 1 |
| | SPIES CCU via electronics in the handle. The Image1 HD<br>CCU processes the sensor images and displays them on a<br>standard HD display monitor. |
| | This change is to expand the indications for the currently<br>cleared secondary predicate Flexible Video-Uretero-<br>Renoscope System (K141250) to allow it to also be<br>marketed as a Choledochoscope used for examination of the<br>bile duct and to allow the ability to percutaneously access<br>the abdominal cavity for examination of the kidney. No<br>changes were made to the device design as part of this<br>update. Only the model number and applicable labeling<br>were updated. |
| Indications For Use: | The Flexible Video-Uretero-Choledochoscope System is<br>indicated for endoscopic examination in the urinary tract<br>and can be used percutaneously to examine the interior of<br>the kidney, and using additional accessories, to perform<br>various diagnostic and therapeutic procedures. The Flexible<br>Video-Uretero-Choledochoscope System is also indicated for<br>the examination of bile ducts, and using additional<br>accessories, to perform various diagnostic and therapeutic<br>procedures during cholecystectomy. |
| Technological<br>Characteristics: | The Karl Storz Flexible Video-Uretero-Choledochoscope<br>System has the same indications for use as the originally<br>cleared primary predicate ACMI Corporation DUR-Digital<br>Ureteroscope and Choledochoscope System (DUR®-D),<br>K06026. The Flexible Video-Uretero-Choledochoscope<br>System and the secondary predicate Flexible Video-Uretero-<br>Renoscope System (K141250), both manufactured by KARL<br>STORZ, are exactly the same and share the same<br>fundamental technology and physical characteristics. The<br>methods of operation, design and materials used are either<br>identical or substantially equivalent to existing legally<br>marketed predicate devices. |
| Non-Clinical<br>Performance Data: | The Flexible Video-Uretero-Choledchoscope System and the<br>secondary predicate Flexible Video-Uretero-Renoscope<br>System (K141250), both manufactured by KARL STORZ, are<br>exactly the same and share the same fundamental<br>technology and physical characteristics. Therefore, the<br>performance data described in the cleared Flexible Video-<br>Uretero-Renoscope System, K141250, is also applicable to<br>the Flexible Video-Uretero-Choledochoscope System. No<br>changes were made to the device design as part of this<br>update. Only the model number and applicable labeling<br>were updated. The KARL STORZ secondary predicate<br>Flexible Video-Uretero-Renoscope System has been<br>evaluated according to ISO 14971 risk management process<br>and the system was successfully tested for its functions and |
| | performance; including verification of optical characteristics<br>per ISO 8600 (image quality, illumination). Safety testing<br>was performed including electrical safety IEC 60601-1,<br>electromagnetic compatibility per IEC 60601-1-2, and<br>biocompatibility of the patient contacting materials per ISO<br>10993. Additional validations were conducted for the<br>manual cleaning method, sterilization process. |
| Clinical<br>Performance Data: | Clinical testing was not required to demonstrate substantial<br>equivalence to the predicate devices. |
| Conclusion: | The Flexible Video-Uretero-Choledochoscope System is<br>substantially equivalent to its predicate devices. The non-<br>clinical testing demonstrates that the device is as safe, as<br>effective and performs as well as or better than the legally<br>marketed devices. |
{4}------------------------------------------------
{5}------------------------------------------------
Panel 1
/
Sort by
Ready
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.