Evis Exera III Duodenovideoscope Olympus TJF-Q190V
K220587 · Olympus Medical Systems Corp. · FDT · May 11, 2022 · Gastroenterology, Urology
Device Facts
Record ID
K220587
Device Name
Evis Exera III Duodenovideoscope Olympus TJF-Q190V
Applicant
Olympus Medical Systems Corp.
Product Code
FDT · Gastroenterology, Urology
Decision Date
May 11, 2022
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 876.1500
Device Class
Class 2
Attributes
Therapeutic
Intended Use
This instrument has been designed to be used with an Olympus video system center, light source, documentation equipment, monitor, EndoTherapy accessories (such as a biopsy forceps), and other ancillary equipment for endoscopy and endoscopic surgery within the duodenum.
Device Story
Flexible duodenovideoscope for endoscopic procedures in duodenum; utilizes CCD chip at distal end to capture images; light source provides white light and narrow band imaging (NBI) via light guide; video processor converts electrical signals to video for clinician display; features instrument channel for EndoTherapy accessories and forceps elevator for accessory manipulation; includes sterile, single-use distal cover (MAJ-2315) to cover distal end and forceps elevator; operated by physicians in clinical settings; provides visualization to assist in diagnosis and surgical intervention.
Clinical Evidence
No clinical data were collected. Substantial equivalence supported by bench testing, sterilization validation, shelf-life testing, biocompatibility evaluation, and human factors evaluation.
Technological Characteristics
Flexible endoscope with CCD imaging sensor; light guide for white light and NBI; instrument channel with forceps elevator; sterile single-use distal cover (MAJ-2315). Compatible with Olympus CV-190/CLV-190 systems. No changes to optical/electrical performance or patient-contacting materials.
Indications for Use
Indicated for endoscopy and endoscopic surgery within the duodenum using Olympus video systems and ancillary equipment.
Regulatory Classification
Identification
An endoscope and accessories is a device used to provide access, illumination, and allow observation or manipulation of body cavities, hollow organs, and canals. The device consists of various rigid or flexible instruments that are inserted into body spaces and may include an optical system for conveying an image to the user's eye and their accessories may assist in gaining access or increase the versatility and augment the capabilities of the devices. Examples of devices that are within this generic type of device include cleaning accessories for endoscopes, photographic accessories for endoscopes, nonpowered anoscopes, binolcular attachments for endoscopes, pocket battery boxes, flexible or rigid choledochoscopes, colonoscopes, diagnostic cystoscopes, cystourethroscopes, enteroscopes, esophagogastroduodenoscopes, rigid esophagoscopes, fiberoptic illuminators for endoscopes, incandescent endoscope lamps, biliary pancreatoscopes, proctoscopes, resectoscopes, nephroscopes, sigmoidoscopes, ureteroscopes, urethroscopes, endomagnetic retrievers, cytology brushes for endoscopes, and lubricating jelly for transurethral surgical instruments. This section does not apply to endoscopes that have specialized uses in other medical specialty areas and that are covered by classification regulations in other parts of the device classification regulations.
Special Controls
*Classification* —(1)*Class II (special controls).* The device, when it is an endoscope disinfectant basin, which consists solely of a container that holds disinfectant and endoscopes and accessories; an endoscopic magnetic retriever intended for single use; sterile scissors for cystoscope intended for single use; a disposable, non-powered endoscopic grasping/cutting instrument intended for single use; a diagnostic incandescent light source; a fiberoptic photographic light source; a routine fiberoptic light source; an endoscopic sponge carrier; a xenon arc endoscope light source; an endoscope transformer; an LED light source; or a gastroenterology-urology endoscopic guidewire, is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 876.9.(2) Class I for the photographic accessories for endoscope, miscellaneous bulb adapter for endoscope, binocular attachment for endoscope, eyepiece attachment for prescription lens, teaching attachment, inflation bulb, measuring device for panendoscope, photographic equipment for physiologic function monitor, special lens instrument for endoscope, smoke removal tube, rechargeable battery box, pocket battery box, bite block for endoscope, and cleaning brush for endoscope. The devices subject to this paragraph (b)(2) are exempt from the premarket notification procedures in subpart E of part 807of this chapter, subject to the limitations in § 876.9.
Predicate Devices
EVIS EXERA III DUODENOVIDEOSCOPE OLYMPUS TJF TYPE Q190V (K202661)
K202661 — Evis Exera III Duodenovideoscope Olympus TJF-Q190V · Olympus Medical Systems Corp. · Dec 8, 2020
K251867 — EVIS EXERA III DUODENOVIDEOSCOPE OLYMPUS TJF-Q190V (TJF-Q190V) · Olympus Medical Systems Corp. · Sep 19, 2025
K193182 — Evis Exera III Duodenovideoscope Olympus TJF-Q190V · Olympus Medical Systems Corp. · Jan 17, 2020
K250701 — Evis Exera III Duodenovideoscope Olympus TJF-Q190V (TJF-Q190V) · Olympus Medical Systems Corp. · Jun 5, 2025
K080403 — XTJF TYPE Q160VF1 · Olympus Medical Systems Corporation · May 20, 2008
Submission Summary (Full Text)
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which is a blue square with the letters "FDA" in white. To the right of the blue square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue.
May 11, 2022
Olympus Medical Systems Corp. % Wendy Perreault, Regualtory Consultant Olympus Corporation of the Americas 3500 Corporate Parkway. P.O. Box 610 Center Valley, PA 18034-0610
Re: K220587
> Trade/Device Name: Evis Exera III Duodenovideoscope Olympus TJF-Q190V Regulation Number: 21 CFR 876.1500 Regulation Name: Endoscope and accessories Regulatory Class: Class II Product Code: FDT, NWB Dated: February 25, 2022 Received: March 1, 2022
Dear Wendy Perreault:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's
{1}------------------------------------------------
requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
Shanil P. Haugen, Ph.D. Assistant Director DHT3A: Division of Renal, Gastrointestinal, Obesity and Transplant Devices OHT3: Office of GastroRenal, ObGyn, General Hospital and Urology Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
## Indications for Use
510(k) Number (if known)
K220587
Device Name
EVIS EXERA III DUODENOVIDEOSCOPE OLYMPUS TJF-QI90V
Indications for Use (Describe)
This instrument has been designed to be used with an Olympus video system center, light source, documentation equipment, monitor, EndoTberapy accessories (such as a biopsy forceps), and other ancillary equipment for endoscopy and endoscopic surgery within the duodenum.
| Type of Use (Select one or both, as applicable) |
|-------------------------------------------------|
|-------------------------------------------------|
| <div> <span> <span style="font-size:16px">❌</span> Prescription Use (Part 21 CFR 801 Subpart D) </span> </div> | <div> <span> <span style="font-size:16px">❍</span> Over-The-Counter Use (21 CFR 801 Subpart C) </span> </div> |
|------------------------------------------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------|
|------------------------------------------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------|
CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
***DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.***
The burden time for this collection of information is estimated to average 79 hours per response, including the
time to review instructions, search existing data sources, gather and maintain the data needed and complete
and review the collection of information. Send comments regarding this burden estimate or any other aspect
of this information collection, including suggestions for reducing this burden, to:
Department of Health and Human Services
Food and Drug Administration
Office of Chief Information Officer
Paperwork Reduction Act (PRA) Staff
PRAStaff@fda.hhs.gov
*"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of
information unless it displays a currently valid OMB number."*
Form Approved: OMB No. 0910-0120 Expiration Date: 06/30/2023 See PRA Statement below.
{3}------------------------------------------------
Image /page/3/Picture/0 description: The image shows the word "OLYMPUS" in large, bold, blue letters. The letters are slightly shadowed, giving them a three-dimensional appearance. A thin, horizontal yellow line is located beneath the word.
K220587 Page 1 of 4
## 510(k) Summary: K220587
#### Company Information 1.
- Applicant: OLYMPUS MEDICAL SYSTEMS CORP. ● 2951 Ishikawa-cho, Hachioji-shi, Tokyo, Japan 192-8507 Establishment Registration No: 8010047
- Wendy Perreault ● Official c/o Olympus Corporation of the Americas Correspondent: 3500 Corporate Parkway P.O. Box 610 Center Valley, PA 18034-0610 Phone: 404.542.5854 E-mail: wendy.perreault(@olympus.com
- Manufacturing Site: Aizu Olympus Co., Ltd., ● 3-1-1 Niiderakita, Aizuwakamatsu-shi, Fukushima 965-8520, Japan
- May 2, 2022 Date Prepared: ●
### Product Information 2.
- EVIS EXERA III DUODENOVIDEOSCOPE ● Trade Name: OLYMPUS TJF-Q190V Duodenoscope and accessories Common Name: ● Endoscope and Accessories ● Classification Name: Regulation Number: 876.1500 ● ● Regulation Name: Endoscope and Accessories
{4}------------------------------------------------
Image /page/4/Picture/0 description: The image shows the text "K220587" on the first line and "Page 2 of 4" on the second line. The text is in a simple, sans-serif font and is likely part of a document or report. The page number indicates that this is the second page of a four-page document. The text is black on a white background.
Image /page/4/Picture/1 description: The image shows the word "OLYMPUS" in large, bold, blue letters. The letters are slightly elongated and have a glossy appearance. A thin, horizontal, yellow line runs beneath the word, adding a subtle contrast to the overall design. The registered trademark symbol is visible to the right of the word.
| • Regulatory Class: | II |
|-------------------------|-------------------------------------------------------------------------------------------------------|
| • Product Code(s): | FDT (Duodenoscope, Accessories, Flexible/Rigid)<br>NWB (Endoscope, accessories, narrow band spectrum) |
| • Classification Panel: | Gastroenterology/Urology |
## Predicate Device 3.
The subject device is equivalent to the predicate device listed below in Table 1.
| Device name | 510(k) Submitter | 510(k) No. |
|-------------------------------------|----------------------------------|------------|
| EVIS EXERA III<br>DUODENOVIDEOSCOPE | OLYMPUS MEDICAL<br>SYSTEMS CORP. | K202661 |
| OLYMPUS TJF TYPE Q190V | | |
## Table 1: Predicate device on TJF-O190V
## Device Description 4.
The EVIS EXERA III DUODENOSCOPE TJF-Q190V has been designed to be used with an Olympus video system center, light source, documentation equipment, monitor, EndoTherapy accessories (such as a biopsy forceps), and other ancillary equipment for endoscopy and endoscopic surgery within the duodenum. The TJF-Q190V is compatible with Olympus system "Video System Center OLYMPUS CV-190 and XENON LIGHT SOURCE OLYMPUS CLV-190 (K112680)."
The subject device consists of a flexible insertion section, control section and endoscope connector section with equipped charge-coupled device (CCD) chip which delivers images.
The light from the light source travels through the light guide to the light guide lens at the distal end. The light source can offer both white light for normal observation and narrow band imaging (NBI). The CCD chip transduces the incident light from the objective lens to electrical signal. The video processor transduces the electrical signal to video signal.
There is an instrument channel located inside of the flexible insertion section. EndoTherapy accessories can be inserted through the instrument channel. A forceps elevator is located at the distal end of the insertion to elevate EndoTherapy
{5}------------------------------------------------
Image /page/5/Picture/1 description: The image shows the word "OLYMPUS" in large, bold, blue letters. The letters are slightly stylized, with a subtle gradient effect. A thin, horizontal yellow line is underneath the word. The registered trademark symbol is located to the right of the word.
accessories for endoscopic treatment.
A sterile, single-use Distal Cover (MAJ-2315) has been designed to be attached to the OLYMPUS TJF-Q190V to cover the distal end of the insertion tube and fit around the forceps elevator. MAJ-2315 is to be discarded after clinical use. MAJ-2315 and TJF-Q190V were previously cleared under 510(k)s K193182 and K202661.
# 5. Indications for Use
This instrument has been designed to be used with an Olympus video system center, light source, documentation equipment, monitor, EndoTherapy accessories (such as a biopsy forceps), and other ancillary equipment for endoscopy and endoscopic surgery within the duodenum.
## Comparison of Technological Characteristics 6.
Compared to the predicate device, the changes are limited to design modifications to the sterile, single-use Distal Cover MAJ-2315 used with the TJF-Q190V and labeling updates.
There are no other changes to either the MAJ-2315 or TJF-Q190V devices, including changes to indications for use, conditions of use, compatible components and accessories to be marketed/used with the device, or patient-contacting materials. There are no changes to the device specifications for TJF-Q190V, including optical or electrical performance.
### 7. Summary of Non-clinical Performance Data
Verification/validation activities were performed subsequent to a risk assessment evaluation of the device modifications per the Olympus Quality Management System. Results of the following testing demonstrate that the changes to the Distal Cover MAJ-2315 do not adversely affect device performance:
- · Performance Testing Bench
- · Sterilization Validation and Shelf-Life Testing
- · Biocompatibility Evaluation
- Human Factors Evaluation
{6}------------------------------------------------
K220587 Page 4 of 4
Image /page/6/Picture/1 description: The image contains the word "OLYMPUS" in large, bold, blue letters. The font is sans-serif and the letters are closely spaced together. A small registered trademark symbol is located to the right of the word.
### 8. Summary of Clinical Performance Data
No clinical data were collected.
#### 9. Conclusion
Based on the results of the comparison of the indications for use, technological characteristics, and performance testing of the subject and predicate device, the modified Distal Cover MAJ-2315 used with the TJF-Q190V duodenoscope raises no new issues of safety and effectiveness and the device is substantially equivalent to the predicate device and is as safe, as effective, and performs as well as or better than the predicate device.
Panel 1
/
Sort by
Ready
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.