PENTAX Medical EPK-i7010 Video Processor with GI Family

K150618 · Pentax of America, Inc. · PEA · Nov 24, 2015 · Gastroenterology, Urology

Device Facts

Record IDK150618
Device NamePENTAX Medical EPK-i7010 Video Processor with GI Family
ApplicantPentax of America, Inc.
Product CodePEA · Gastroenterology, Urology
Decision DateNov 24, 2015
DecisionSESE
Submission TypeTraditional
Regulation21 CFR 876.1500
Device ClassClass 2
AttributesTherapeutic

Intended Use

The PENTAX Medical EPK-i7010 Video Processor is intended to be used with the PENTAX camera heads, endoscopes, light sources, monitors and other ancillary equipment for gastrointestinal endoscopic diagnosis, treatment and video observation. The PENTAX Medical EPK-i7010 includes a digital post-processing imaging enhancement technology (PENTAX i-Scan™) and an optical imaging enhancement technology (OE). These imaging enhancement technologies are intended to be used as an optional adjunct following traditional white light endoscopy and is not intended to replace histopathological sampling. i-Scan and OE are compatible with PENTAX video gastrointestinal endoscopes.

Device Story

Video processor system with integrated 300W xenon light source; used with endoscopes, camera heads, and monitors for GI diagnosis/treatment. Physician operates system in clinical setting. Inputs: white light illumination captured via endoscope CCD. Processing: digital post-processing (i-Scan) modifies RGB pixel components; optical enhancement (OE) inserts optical filters into light path. Outputs: enhanced video images on monitor. i-Scan/OE serve as adjuncts to white light endoscopy; do not replace histopathology. System includes front operation panel, digital output, electrical zoom, and display options. Benefits: improved visualization of GI mucosa to assist clinical decision-making.

Clinical Evidence

Non-clinical bench and animal testing only. Animal study used porcine GI mucosa to compare white light, PENTAX OE, and Olympus NBI images. Bench testing included optical resolution analysis and artifact analysis comparing PENTAX and Olympus endoscopes. Software verification/validation conducted per FDA guidance.

Technological Characteristics

Video processor with integrated 300W xenon lamp. Features digital (i-Scan) and optical (OE) enhancement. Complies with IEC 60601-1 (safety), IEC 60601-1-2 (EMC), IEC 60601-1-6 (usability), IEC 60601-2-18 (endoscopic equipment), and IEC 62471 (photobiological safety). Software classified as Class B per IEC 62304.

Indications for Use

Indicated for gastrointestinal endoscopic diagnosis, treatment, and video observation in patients requiring endoscopic procedures. Compatible with PENTAX video gastrointestinal endoscopes.

Regulatory Classification

Identification

An endoscope and accessories is a device used to provide access, illumination, and allow observation or manipulation of body cavities, hollow organs, and canals. The device consists of various rigid or flexible instruments that are inserted into body spaces and may include an optical system for conveying an image to the user's eye and their accessories may assist in gaining access or increase the versatility and augment the capabilities of the devices. Examples of devices that are within this generic type of device include cleaning accessories for endoscopes, photographic accessories for endoscopes, nonpowered anoscopes, binolcular attachments for endoscopes, pocket battery boxes, flexible or rigid choledochoscopes, colonoscopes, diagnostic cystoscopes, cystourethroscopes, enteroscopes, esophagogastroduodenoscopes, rigid esophagoscopes, fiberoptic illuminators for endoscopes, incandescent endoscope lamps, biliary pancreatoscopes, proctoscopes, resectoscopes, nephroscopes, sigmoidoscopes, ureteroscopes, urethroscopes, endomagnetic retrievers, cytology brushes for endoscopes, and lubricating jelly for transurethral surgical instruments. This section does not apply to endoscopes that have specialized uses in other medical specialty areas and that are covered by classification regulations in other parts of the device classification regulations.

Special Controls

*Classification* —(1)*Class II (special controls).* The device, when it is an endoscope disinfectant basin, which consists solely of a container that holds disinfectant and endoscopes and accessories; an endoscopic magnetic retriever intended for single use; sterile scissors for cystoscope intended for single use; a disposable, non-powered endoscopic grasping/cutting instrument intended for single use; a diagnostic incandescent light source; a fiberoptic photographic light source; a routine fiberoptic light source; an endoscopic sponge carrier; a xenon arc endoscope light source; an endoscope transformer; an LED light source; or a gastroenterology-urology endoscopic guidewire, is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 876.9.(2) Class I for the photographic accessories for endoscope, miscellaneous bulb adapter for endoscope, binocular attachment for endoscope, eyepiece attachment for prescription lens, teaching attachment, inflation bulb, measuring device for panendoscope, photographic equipment for physiologic function monitor, special lens instrument for endoscope, smoke removal tube, rechargeable battery box, pocket battery box, bite block for endoscope, and cleaning brush for endoscope. The devices subject to this paragraph (b)(2) are exempt from the premarket notification procedures in subpart E of part 807of this chapter, subject to the limitations in § 876.9.

Predicate Devices

Reference Devices

Related Devices

Submission Summary (Full Text)

{0}------------------------------------------------ Image /page/0/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a circular seal with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" around the perimeter. Inside the circle is a stylized image of three human profiles facing to the right, with flowing lines connecting them. Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002 November 24, 2015 Pentax of America, Inc. Krishna Govindarajan Senior Manager, Regulatory Affairs 3 Paragon Drive Montvale, NJ 07645 Re: K150618 Trade/Device Name: Pentax Medical EPK-i7010 Video Processor with GI Family Regulation Number: 21 CFR 876.1500 Regulation Name: Endoscope and Accessories Regulatory Class: Class II Product Code: PEA Dated: October 19, 2015 Received: October 20, 2015 Dear Krishna Govindarajan, We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading. If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register. Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical devicerelated adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in {1}------------------------------------------------ the quality systems (OS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050. If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance. You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Sincerely yours. # Herbert P. Lerner -S for Benjamin R. Fisher, Ph.D. Director Division of Reproductive, Gastro-Renal, and Urological Devices Office of Device Evaluation Center for Devices and Radiological Health Enclosure {2}------------------------------------------------ ## Indications for Use 510(k) Number (if known) K150618 #### Device Name Pentax Medical EPK-i7010 Video Processor with GI Family Indications for Use (Describe) The PENTAX Medical EPK-i7010 Video Processor is intended to be used with the PENTAX camera heads, endoscopes, light sources, monitors and other ancillary equipment for gastrointestinal endoscopic diagnosis, treatment and video observation. The PENTAX Medical EPK -7010 includes a digital post-processing imaging enhancement technology (PENTAX i-Scan™) and an optical imaging enhancement technology (OE).These imaging enhancement technologies are intended to be used as an optional adjunct following traditional white light endoscopy and is not intended to replace histopathological sampling. i-Scan and OE are compatible with PENTAX video gastrointestinal endoscopes. | Type of Use (Select one or both, as applicable) | |-------------------------------------------------| |-------------------------------------------------| X Prescription Use (Part 21 CFR 801 Subpart D) | Over-The-Counter Use (21 CFR 801 Subpart C) #### CONTINUE ON A SEPARATE PAGE IF NEEDED. This section applies only to requirements of the Paperwork Reduction Act of 1995. #### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.* The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to: > Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov "An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number." {3}------------------------------------------------ Image /page/3/Picture/0 description: The image contains the logo for Pentax Medical. The word "PENTAX" is in red, bold, sans-serif font. Below the word "PENTAX" is a gray line. Below the gray line is the word "MEDICAL" in gray, bold, sans-serif font. ## 510(k) Summary The following summary is provided in accordance with 21 CFR 807.92: ### I. SUBMITTER PENTAX of America, Inc., HOYA Corporation PENTAX Division 3 Paragon Drive Montvale, New Jersey 07645-1782 Phone: 201-251-2300 x 2125 Fax: 201-799-4117 Contact Person: Krishna Govindarajan Date Prepared: October 16, 2015 #### II. DEVICE | Name of Device: | PENTAX Medical EPK-i7010 Video Processor with GI Family | |-----------------------|----------------------------------------------------------------------------| | Common or Usual Name: | Endoscopic Video Processor and Light Source | | Classification Name: | Endoscopic video imaging system/component, | | Classification Name: | Gastroscopes (Flexible or rigid) and accessories (21 CFR Part<br>876.1500) | | Regulatory Class: | Class II | | Product Code: | PEA | #### III. PREDICATE DEVICE The PENTAX Medical EPK-i5010 Video Processor (K122470; dated April 22, 2013) is the primary predicate for this submission. This predicate device has not been subject to a design-related recall. The OLYMPUS EVIS EXERA III Video System CV - 190 Video System Center and CLV -190, Xenon Light Source (K112680) is the reference device for this submission. #### IV. DEVICE DESCRIPTION The PENTAX Medical EPK-i7010 video processor consists of a video system, integrated light source, monitor, and ancillary equipment. This processor is intended for endoscopic diagnostic, treatment and video observation. The PENTAX Medical EPK-i7010 video processor contains two types of contrast enhancement techniques: PENTAX i-Scan technology, and optical enhancement (OE) technology. {4}------------------------------------------------ Image /page/4/Picture/0 description: The image contains the logo for Pentax Medical. The word "PENTAX" is written in large, red, bold letters. Below the word "PENTAX" is a gray line, and below the line is the word "MEDICAL" written in gray, bold letters. White light is captured from a 300 Watt xenon lamp housed in the PENTAX Medical EPKi7010 video processor. All visualization is done with the white light mode first. White light (BGR) illuminates the tissue and transfers the captured light through the video scope or a charged coupled device (CCD). Note that the white light visualization mode is always used first by the physician. For i-Scan image enhancement, the modification of the combination of RGB components for each pixel occurs when the i-Scan function is turned on in the PENTAX Medical EPKi7010 video processor. The resulting i-Scan image is then displayed on the observation monitor. For OE image enhancement, one of the two optical filters corresponding to Mode1 and Mode2 are inserted into illumination light path when the OE function is turned on in the PENTAX Medical EPK-i7010 video processor. The resulting OE image is then displayed on the observation monitor. The table below are the list of accessories and compatible devices that are used with the EPK-i7010 Video Processor. | Accessories Name | Intended Use | Model Number | |-------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-------------------------------------| | PENTAX Condenser<br>Earth Cable | Condenser earth cable OL-Z4 is intended<br>to reduce high- frequency noise which is<br>generated during high- frequency electro<br>cautery device use together with Pentax<br>endoscopes. | OL-Z4 | | PENTAX Foot Switch | Foot Switch OS-A61 is used to remotely<br>control processor functions. | OS-A61 | | PENTAX Keyboard | PENTAX medical keyboard OS-A83 is<br>intended to be used with compatible<br>PENTAX video processor EPK-i7010.<br>OS-A83 keyboard for input data entry and<br>operate the video processor by assigned<br>function keys. | OS-A83 | | PENTAX White Balance<br>Adjuster | White Balance Adjuster OS- A43H is a<br>white tube used as the object of white<br>balance feature. | OS-A43H | | PENTAX Medical<br>DispoCap Air<br>(MEDIVATORS Endo<br>SmartCap Tubing -<br>K093665) | PENTAX Medical DispoCap tubing is<br>intended to be used with an air or CO2<br>source along with a sterile water source to<br>supply air or CO2 and sterile water to an<br>gastrointestinal endoscope during<br>endoscopic procedures. | 100160P<br>(DispoCap Air<br>Tubing) | #### Table 1: List of Accessories {5}------------------------------------------------ Image /page/5/Picture/0 description: The image shows the logo for Pentax Medical. The word "PENTAX" is in red, bold, sans-serif letters. Below "PENTAX" is a gray line. Below the line, the word "MEDICAL" is in gray, bold, sans-serif letters. | Compatible Devices | Manufacturer | Model Name | |---------------------|-------------------------------------|-------------------| | Printer | Sony Business Solutions Corporation | UP-55MD | | | Sony Business Solutions Corporation | UP-25MD | | External Hard Drive | Western Digital Corporation | WDBAAU0020HBK | | HD Video Recorder | Sony Business Solutions Corporation | HVO-1000MD | | USB Flash Memory | Transcend Information Inc. | TS32GJF600 | | LCD Monitor | NDS Surgical Imaging | 26"Radiance G2 HB | #### Table 2: List of Compatible Devices The EPK-i7010 is compatible with PENTAX video gastrointestinal endoscopes. ## V. INDICATIONS FOR USE The PENTAX Medical EPK-i7010 Video Processor is intended to be used with the PENTAX camera heads, endoscopes, light sources, monitors and other ancillary equipment for gastrointestinal endoscopic diagnosis, treatment and video observation. The PENTAX Medical EPK-i7010 includes a digital post-processing imaging enhancement technology (PENTAX i-Scan™) and an optical imaging enhancement technology (OE). These imaging enhancement technologies are intended to be used as an optional adjunct following traditional white light endoscopy and is not intended to replace histopathological sampling. i-Scan and OE are compatible with PENTAX Medical video gastrointestinal endoscopes. #### VI. COMPARISON OF TECHNOLOGICAL CHARACTERISTICS WITH THE PREDICATE DEVICE The PENTAX Medical EPK-i7010 Video Processor with GI Family (subject device) has the same intended use, design, fundamental operating principle, and scientific technology as the PENTAX Medical EPK-i5010 Video Processor (K122470; dated April 22, 2013) (primary predicate device). Both devices also have the same indications for use; namely, for use in gastrointestinal endoscopic diagnosis, treatment and video observation in the upper and lower gastrointestinal anatomy. The devices differ slightly with regard to design and technological characteristics. Performance data, specifically optical bench and animal testing, is provided to support substantial equivalence of the devices. The EPK-i7010 Video Processor includes additional functionalities that are not available in the EPK-i5010, such as Optical Enhancement (OE) technology, Digital output, Electrical Zoom, Front Operation Panel, Language, Input method, Freeze Scan, Contrast, Twin Mode, and several additional video image display options. However, these additional functionalities do not affect the final product performance and did not raise any questions of safety or effectiveness as demonstrated by the software verification and validation testing. PENTAX Medical EPK-i7010 Video Processor with GI Family 510(k) Summary {6}------------------------------------------------ Image /page/6/Picture/0 description: The image shows the logo for Pentax Medical. The word "PENTAX" is in large, red, bold letters. Below that is a gray line, and below that is the word "MEDICAL" in gray, bold letters. The PENTAX Medical EPK-i7010 Video Processor with GI Family (subject device) has the same intended use and fundamental operating principle including the Optical Enhancement (OE) technology, compared to the commercially available The OLYMPUS EVIS EXERA III Video System CV - 190 Video System Center and CLV - 190, Xenon Light Source (K112680) (reference device). The devices differ slightly with regard to design and technological characteristics. Performance data, specifically optical bench and animal testing, is provided to support substantial equivalence of the devices. ## VII. PERFORMANCE DATA The following performance data are provided in support of the substantial equivalence determination. #### Electrical Safety and electromagnetic compatibility (EMC) The acceptable level of Electromagnetic Compatibility (EMC) and Electrical Safety (ES) for the PENTAX Medical EPK-i7010 Video Processor with GI Family was confirmed by testing in accordance with the following standards: - 1. IEC 60601-1:2005+A1:2012, Medical electrical equipment Part 1: General requirements for basic safety and essential performance - 2. IEC 60601-1-2:Edition 3:2007, Medical electrical equipment Part 1-2: General requirements for basic safety and essential performance - Collateral standard: Electromagnetic compatibility - Requirements and tests - 3. IEC 60601-1-6 Edition 3.0:2010, Medical electrical equipment Part 1-6: General requirements for basic safety and essential performance - Collateral standard: Usabilitv - 4. IEC 60601-2-18 Edition 3.0:2009, Medical electrical equipment Part 2-18: Particular requirements for the basic safety and essential performance of endoscopic equipment #### Software Verification and Validation Testing Software verification and validation testing were conducted and documentation was provided as recommended by FDA's Guidance for Industry and FDA Staff, "Guidance for the Content of Premarket Submissions for Software Contained in Medical Devices." The software is classified as CLASS B under the Software Safety Classification per IEC 62304:2006, Medical device software life cycle processes) and the software level of concern is "Moderate" based on the FDA's Guidance for Industry and FDA Staff, "Guidance for the Content of Premarket Submissions for Software Contained in Medical Devices." All testing of the software was conducted in compliance with the following standards: - 1. ISO 14971 Second Edition:2007, Medical devices Application of risk management to medical devices - 2. IEC 62304 First Edition:2006, Medical device software Software life cycle processes - 3. IEC 62471 First Edition:2006, Photobiological safety of lamps and lamp systems PENTAX Medical EPK-i7010 Video Processor with GI Family 510(k) Summary {7}------------------------------------------------ Image /page/7/Picture/0 description: The image shows the logo for Pentax Medical. The word "PENTAX" is in red, bold letters above a gray line. Below the line, the word "MEDICAL" is in gray, bold letters. #### Optical Performance Testing (Bench and Animal non-clinical testing) #### Animal Study A library of images [white light endoscopic (WLE), PENTAX Optical Enhancement (OE), and Olympus Narrow Band Imaging (NBI)] were obtained from the porcine gastrointestinal mucosa. The images were used for image evaluation and quantitative data analysis. #### Bench Testing Optical Enhancement (OE) and Artifact Analysis were performed with the images from the porcine gastrointestinal location that were gathered using two PENTAX Medical gastrointestinal endoscopes, one high definition and one standard definition, along with an Olympus gastrointestinal endoscopes. In addition. PENTAX Medical and Olympus gastrointestinal endoscopes were compared and the effects of processing features on Limited Resolution optical bench testing were conducted. The animal and optical bench test image data gathered with gastrointestinal endoscopes establish the equivalence of the subject and predicate device. #### VIII. CONCLUSIONS The data submitted support the safety of the device and the hardware and software verification and validation demonstrate that the PENTAX Medical EPK-i7010 Video Processor performs as intended in the specified use conditions. The optical data analysis demonstrate that the PENTAX Medical EPK-i7010 Video Processor performs comparably to the predicate device that is currently marketed for the same indication for use.
Innolitics

Panel 1

/
Sort by
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...