K122470 · Pentax Medical · PEA · Apr 10, 2013 · Gastroenterology, Urology
Device Facts
Record ID
K122470
Device Name
PENTAX EPK-I5010 VIDEO PROCESSOR
Applicant
Pentax Medical
Product Code
PEA · Gastroenterology, Urology
Decision Date
Apr 10, 2013
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 876.1500
Device Class
Class 2
Indications for Use
The PENTAX EPK-i5010 Video Processor is intended to be used with the PENTAX camera heads, endoscopes, light sources, monitors and other ancillary equipment for gastrointestinal endoscopic diagnosis, treatment and video observation. The PENTAX EPK-i5010 includes PENTAX i-Scan™, a digital, post-processing imaging enhancement technology. i-Scan is intended to be used as an optional adjunct following traditional white light endoscopy and is not intended to replace histopathological sampling. i-Scan is compatible with PENTAX k-series and i-series gastrointestinal videoscopes and colonovideoscopes.
Device Story
EPK-i5010 is an endoscopic video processor and integrated light source; processes electrical signals from flexible/rigid videoscopes for display on monitors. Features i-Scan digital post-processing technology; provides three modes to enhance mucosal surface texture and blood vessel visibility. User selects modes via scope button, foot pedal, or keyboard. Used in clinical settings by physicians for GI diagnosis and treatment. i-Scan serves as an optional adjunct to white light endoscopy; aids visualization of mucosal structures. Does not replace histopathological sampling. Hardware based on EPK-i5020 model.
Clinical Evidence
Seven clinical studies conducted in Germany, Japan, and Korea involving 975 patients (350 esophageal, 625 colonic). Studies evaluated the PENTAX EPK-i processor with i-Scan technology. No safety issues reported across all studies.
Technological Characteristics
Video processor with integrated light source; compatible with PENTAX k-series and i-series videoscopes. Features i-Scan digital post-processing for image enhancement. Complies with IEC 60601 electrical safety standards. Software validated per IEC 62304.
Indications for Use
Indicated for gastrointestinal endoscopic diagnosis, treatment, and video observation in patients requiring endoscopic procedures. Compatible with PENTAX k-series and i-series gastrointestinal videoscopes and colonovideoscopes. i-Scan technology is an optional adjunct to white light endoscopy; not intended to replace histopathological sampling.
Regulatory Classification
Identification
An endoscope and accessories is a device used to provide access, illumination, and allow observation or manipulation of body cavities, hollow organs, and canals. The device consists of various rigid or flexible instruments that are inserted into body spaces and may include an optical system for conveying an image to the user's eye and their accessories may assist in gaining access or increase the versatility and augment the capabilities of the devices. Examples of devices that are within this generic type of device include cleaning accessories for endoscopes, photographic accessories for endoscopes, nonpowered anoscopes, binolcular attachments for endoscopes, pocket battery boxes, flexible or rigid choledochoscopes, colonoscopes, diagnostic cystoscopes, cystourethroscopes, enteroscopes, esophagogastroduodenoscopes, rigid esophagoscopes, fiberoptic illuminators for endoscopes, incandescent endoscope lamps, biliary pancreatoscopes, proctoscopes, resectoscopes, nephroscopes, sigmoidoscopes, ureteroscopes, urethroscopes, endomagnetic retrievers, cytology brushes for endoscopes, and lubricating jelly for transurethral surgical instruments. This section does not apply to endoscopes that have specialized uses in other medical specialty areas and that are covered by classification regulations in other parts of the device classification regulations.
Special Controls
*Classification* —(1)*Class II (special controls).* The device, when it is an endoscope disinfectant basin, which consists solely of a container that holds disinfectant and endoscopes and accessories; an endoscopic magnetic retriever intended for single use; sterile scissors for cystoscope intended for single use; a disposable, non-powered endoscopic grasping/cutting instrument intended for single use; a diagnostic incandescent light source; a fiberoptic photographic light source; a routine fiberoptic light source; an endoscopic sponge carrier; a xenon arc endoscope light source; an endoscope transformer; an LED light source; or a gastroenterology-urology endoscopic guidewire, is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 876.9.(2) Class I for the photographic accessories for endoscope, miscellaneous bulb adapter for endoscope, binocular attachment for endoscope, eyepiece attachment for prescription lens, teaching attachment, inflation bulb, measuring device for panendoscope, photographic equipment for physiologic function monitor, special lens instrument for endoscope, smoke removal tube, rechargeable battery box, pocket battery box, bite block for endoscope, and cleaning brush for endoscope. The devices subject to this paragraph (b)(2) are exempt from the premarket notification procedures in subpart E of part 807of this chapter, subject to the limitations in § 876.9.
In combination with the general controls of the FD&C Act, the integrated operating table-electromechanical surgical system is subject to the following special controls:
1. (1) Premarket clinical performance testing, or a combination of premarket clinical performance testing and postmarket surveillance (in accordance with special control (2)), must include the following:
1. (i) Objective performance measures (e.g., rate and number of conversions to other surgical modalities, rate of device related adverse events (including tissue injury, hematoma, and increased blood loss), and their severity, cause, and outcomes) must be reported with relevant descriptive comparator performance measures.
2. (ii) The data must demonstrate the performance of the device for providing accurate and precise control of attached surgical instruments in range of clinical conditions relevant to the device's intended use.
3. (iii) The test dataset must include data collected from a patient population representative of the intended patient population under anticipated conditions of use.
2. (2) Data obtained from postmarket surveillance must demonstrate, in consideration of the premarket data obtained in accordance with special control (1), that the device performs in accordance with special control (1), unless FDA determines, based on the totality of the premarket data, that data from postmarket surveillance is not required to demonstrate that the device performs as intended. Such postmarket surveillance must be conducted per a protocol determined appropriate by FDA to demonstrate that the device performs as intended (in consideration of the premarket data obtained in accordance with special control (1)), and must include initiation, enrollment, and reporting requirements to ensure timely periodic updates to FDA on post-market surveillance progress and outcomes.
3. (3) Animal performance testing must evaluate the extent of port site trauma due to repositioning of table during surgical procedures when utilizing robotic minimally invasive and laparoscopic approaches
4. (4) The device manufacturer must develop, and update as necessary, a device-specific use training program that ensures proper device setup/use/shutdown, accurate control of instruments to perform the intended surgical procedures, troubleshooting and handling during unexpected events or emergencies, and safe practices to mitigate use error.
5. (5) The device manufacturer may only distribute the device to facilities that implement and maintain the device-specific use training program and ensure that users of the device have completed the device-specific use training program.
(6) Human factors assessment must demonstrate that the user can correctly use the device system across all intended use environments with the provided instructions and training materials, including patient access during normal operating conditions and emergency situations, and effects arising from the integrated nature of the operating table and robotic surgical arms.
(7) Labeling must include:
(i) A detailed summary of clinical performance testing conducted with the device, including study population, results, adverse events, and comparisons to any comparator groups identified;
(ii) A statement in the labeling that the safety and effectiveness for the representative specific procedures was based on evaluation of the device as a surgical tool and did not include evaluation of outcomes related to the treatment of the patient's underlying disease or condition, unless FDA determines that it can be removed or modified based on clinical performance data submitted to FDA;
(iii) Identification of compatible devices;
(iv) The list of surgical procedures for which the device has been determined to be safe with clinical justification;
(v) Reprocessing instructions for reusable components;
(vi) A shelf life for any sterile components;
(vii) A description of the device-specific use training program;
(viii) A statement that the device is only for distribution to facilities that implement and maintain the device-specific use training program and ensure that users of the device have completed the device-specific use training program; and
(ix) A summary of any completed postmarket surveillance data collected as required by special control (2), including updated labeling to accurately reflect outcomes observed in postmarket surveillance.
(8) Non-clinical performance testing must demonstrate that the device performs as intended under anticipated conditions of use and must include:
(i) Device motion accuracy and repeatability;
(ii) System testing;
(iii) Instrument reliability;
(iv) Crosstalk;
(v) Table motion control;
(vi) Thermal effects on tissue;
(vii) User-device interface performance;
(viii) Workspace access testing; and
(ix) Performance testing with compatible devices.
(9) Software verification, validation, and hazard analysis must be performed.
(10) Electromagnetic compatibility and electrical, thermal, and mechanical safety testing must be performed.
(11) Performance data must demonstrate the sterility of all patient-contacting device components.
(12) Performance data must support the shelf life of the device components provided sterile by demonstrating continued sterility and package integrity over the labeled shelf life.
(13) Performance data must validate the reprocessing instructions for the reusable components of the device.
(14) Performance data must demonstrate that all patient-contacting components of the device are biocompatible.
(15) Performance data must demonstrate that all patient-contacting components of the device are non-pyrogenic.
(16) The device manufacturer must submit a report to the FDA annually on the anniversary of initial marketing authorization for the device, until such time as FDA may terminate such reporting, which comprises the following information:
(i) Cumulative summary, by year, of complaints and adverse events since date of initial marketing authorization; and
(ii) Identification and rationale for changes made to the device, labeling, or device specific use training program, which did not require submission of a premarket notification during the reporting period.
{1}------------------------------------------------
Image /page/0/Picture/0 description: The image shows the logo for Pentax Medical Company. The logo includes the word "PENTAX" in large, bold letters. Above the word "PENTAX" is the code "K122470". Below the word "PENTAX" is the text "PENTAX Medical Company A Division of PENTAX of America, Inc."
# 510(K) Summary
APR 1 0 2013
The following summary is provided in accordance with 21 CFR 807.92:
Date: April 08, 2013
#### Sponsor Information:
PENTAX Medical Company, A Division of PENTAX America, Inc. 3 Paragon Drive Montvale, New Jersey 07645-1782
#### Contact Person:
| Name: | Krishna Govindarajan |
|------------|----------------------------------------|
| Title: | Regulatory Manager |
| Telephone: | (201) 251-2300 x 2125 |
| Facsimile: | (201) 251-4117 |
| E-Mail: | Krishna.govindarajan@pentaxmedical.com |
#### Manufacturer:
HOYA Corporation PENTAX Yamagata Factory 4-1 Hinode-Cho Nagai-Shi, Japan, 993-0012
FDA Establishment Registration #3003782606
#### Identification of the Proposed Device:
| Proprietary/Trade Name: | PENTAX EPK-i5010 Video Processor |
|-------------------------|------------------------------------------------------------------------|
| Common/Usual Name: | Endoscopic Video Processor and Light Source |
| Classification Name: | Endoscopic video imaging system/component,<br>gastroenterology-urology |
| Regulation Number: | 21 CFR Part 876.1500 |
| Regulation Description: | Endoscope and accessories |
| Medical Specialty: | Gastroenterology/Urology |
| Regulatory Class: | Class II |
| Product Code: | FET and GCT |
# Device Description:
The PENTAX EPK-i5010 video processor consists of a video system, integrated light source, monitor, and ancillary equipment. This processor is intended for endoscopic diagnostic, treatment and video observation. It is intended to process electrical signals from a video endoscope.
{2}------------------------------------------------
Image /page/1/Picture/0 description: The image shows the logo for PENTAX Medical Company. The logo is in black and white and features the word "PENTAX" in large, bold letters. Above the word "PENTAX" is the code "K122470". Below the word "PENTAX" is the text "PENTAX Medical Company A Division of PENTAX of America, Inc."
The PENTAX EPK-i5010 offers an optional digital, post-processing imaging enhancement technology called the PENTAX i-Scan™ modes 1, 2, and 3, which is intended to give the user an enhanced view of the texture of the mucosal surface and blood vessels. i-Scan 1 provides the user with a view that sharpens surface vessels and enhances surface texture of the mucosa. i-Scan 2 provides the user with increased visibility of blood vessels while also providing the same enhancements to the mucosa achieved in i-Scan 1. i-Scan 3 provides the user with increased visibility of blood vessels including dimly illuminated far-field regions while also providing the same enhancement to the mucosa achieved in i-Scan 1. The user can select either white light image or i-Scan modes by pressing a pre-programmed button on the scope, by using a pre-programmed foot pedal or by pressing a keyboard button. i-Scan is intended to be used as an optional adjunct following traditional white light endoscopy and is not intended to replace histopathological sampling.
The EPK-i5010 video processor incorporates the hardware of the EPK-i5020 video processor model (K 113873) and the PENTAX i-Scan™ technology.
The EPKi-5010 is compatible with PENTAX flexible and rigid k-Series and i-Series videoscopes. The subject premarket notification is specific for gastrointestinal videoscopes and colonovideoscopes.
PENTAX Medical Company considers video endoscope families, defined as a grouping of endoscopes with the same: Intended diagnostic/ therapeutic affect for given anatomy (with respect to the indication for procedure), method(s) of introduction (trans oral, trans rectal, etc). Technical characteristics (design, materials, etc), and range of descriptive characteristics (specifications). Changes within the families are considered not-significant changes and mechanics for the description and grouping of endoscope families include 510(k) consideration for changes.
# , Intended Use:
The PENTAX EPK-i5010 Video Processor is intended to be used with the PENTAX camera heads, endoscopes, light sources, monitors and other ancillary equipment for gastrointestinal endoscopic diagnosis, treatment and video observation.
The PENTAX EPK-i5010 includes PENTAX i-Scan™, a digital, post-processing imaging enhancement technology. i-Scan is intended to be used as an optional adiunct following traditional white light endoscopy and is not intended to replace histopathological sampling. i-Scan is compatible with PENTAX k-series and i-series gastrointestinal videoscopes and colonovideoscopes.
#### Technology Characteristics:
The EPK-i5010 Video Processor has the same fundamental scientific technology as commercially available endoscope video processor and the substantially equivalent predicate devices.
{3}------------------------------------------------
Image /page/2/Picture/0 description: The image shows the logo for Pentax Medical Company. The logo includes the text "K122470" at the top, followed by the word "PENTAX" in large, bold letters. Below that, it says "PENTAX Medical Company" and "A Division of PENTAX of America, Inc." in a smaller font.
### Substantial Equivalence Determination:
#### 1. Summary of Nonclinical Tests
The PENTAX EPK-i5010 Video Processor complies with voluntary standards as detailed in this premarket notification submission. The following quality assurance measures were applied to the design and development:
- Animal Model Testing ●
- . Image Enhancement Analysis
- Endoscope Optical Testing ●
- Risk Management (Analysis & Mitigation) ●
- Requirements Reviews .
- Design Reviews .
- Unit Testing (Software & Hardware Verification) .
- Integration testing (Software & Hardware Verification) 0
- t Performance testing (Software & Hardware Verification)
- Safety testing (System level Verification & Validation) .
- t Simulated use testing (System level QA Validation)
- Electrical testing according to IEC 60601 ●
- Software validation performed in accordance with IEC 62304 .
#### 2. Summary of Clinical Tests
A total of seven (7) clinical studies are presented in this review to help establish the safety of the device. The studies were conducted in Germany (3), Japan (1) and Korea (3). An overall total of 975 patients were evaluated in the 7 clinical studies. All of the 7 studies used the PENTAX EPK-i processor with i-Scan technology.
Four of the studies were conducted on the esophagus totaling 350 patients and three of the studies were conducted on the colon, totaling 625 patients. It should be noted that out of the 975 patients in the seven (7) clinical studies conducted with the PENTAX EPK-i video processor there were no safety issues reported.
#### 3. Predicate Device:
PENTAX EPK-i5010 Video Processor is substantially equivalent to the following devices:
| Legally Marketed Device(s) | 510(k) # |
|----------------------------------|----------|
| EVIS EXERA III VIDEO SYSTEM | K112680 |
| PENTAX EPK-i5020 Video Processor | K113873 |
The subject device is also substantially equivalent to other legally marketed video processors for use with endoscopes including, but not limited to the Olympus predicate K100584. The EPK-i5010 video processor incorporates the hardware of the EPK-i5020 video processor model and the optional i-Scan™ technology. All other functions and specifications between the two processors are substantially identical.
The similarities and differences between the Olympus EVIS EXERA II and III Video System and the PENTAX EPK-i5010 Video Processor can be seen in the Table 1.
{4}------------------------------------------------
Image /page/3/Picture/0 description: The image shows the logo for Pentax Medical Company. The logo includes the text "K122470" at the top, followed by the word "PENTAX" in large, bold letters. Below that, it says "PENTAX Medical Company" and "A Division of PENTAX of America, Inc."
When compared to the predicate devices, the EPK-i5010 does not incorporate any significant changes in the Intended Use, in the Method of Operation, Material or Design that could be considered to affect the safety or effectiveness of the Medical Device.
{5}------------------------------------------------
Image /page/4/Picture/1 description: The image shows the logo for the U.S. Department of Health and Human Services. The logo features the department's name arranged in a circular fashion around a stylized emblem. The emblem consists of a stylized human figure with outstretched arms, overlaid with three parallel lines that curve upwards, resembling a stylized bird or wing.
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002
April 22, 2013
PENTAX of America, Inc. PENTAX Medical Company % Robert Schiff, Ph.D., RAC, CQA, FRAPS President Schiff & Company, Inc. 1129 Bloomfield Avenue WEST CALDWELL NJ 07006
Re: K122470 Trade/Device Name: PENTAX EPK-i5010 Video Processor Regulation Number: 21 CFR§ 876.1500 Regulation Name: Endoscope and accessories Regulatory Class: II Product Code: PEA Dated: March 6, 2013 Received: March 7, 2013
Dear Dr. Schiff:
This letter corrects our substantially equivalent letter of April 10, 2013.
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{6}------------------------------------------------
# Page 2 - Robert Schiff. Ph.D., RAC, COA, FRAPS
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements. including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Small Manufacturers, International and Consumer Assistance at its tollfree number (800) 638 2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21
CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportalProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerely yours,
# Herbert P. Lerner -S
for
Benjamin R. Fisher, Ph.D. Director Division of Reproductive, Gastro-Renal, and Urological Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{7}------------------------------------------------
# Indications for Use
510(k) Number (if known): K122470
Device Name: PENTAX EPK-i5010 Video Processor
# Indications for Use:
The PENTAX EPK-i5010 Video Processor is intended to be used with the PENTAX camera heads, endoscopes, light sources, monitors and other ancillary equipment for gastrointestinal endoscopic diagnosis, treatment and video observation.
The PENTAX EPK-i5010 includes PENTAX i-Scan™ a digital, post-processing imaging enhancement technology. i-Scan is intended to be used as an optional adjunct following traditional white light endoscopy and is not intended to replace histopathological sampling. i-Scan is compatible with PENTAX k-series and i-series gastrointestinal videoscopes and colonovideoscopes.
Prescription Use × (Part 21 CFR 801 Subpart D) AND/OR
Over-The-Counter Use (21 CFR 807 Subpart C)
(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
# Herbert P. Lerner -S
(Division Sign-Off) Division of Reproductive. Gastro-Renal, and Urological Devices K122470 510(k) Number ________________________________________________________________________________________________________________________________________________________________
Page 1 of 1
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.