Stryker AIM Light Source and SafeLight cable

K173866 · Stryker · OWN · Jan 18, 2018 · Gastroenterology, Urology

Device Facts

Record IDK173866
Device NameStryker AIM Light Source and SafeLight cable
ApplicantStryker
Product CodeOWN · Gastroenterology, Urology
Decision DateJan 18, 2018
DecisionSESE
Submission TypeSpecial
Regulation21 CFR 876.1500
Device ClassClass 2

Indications for Use

The Stryker AIM Light Source and SafeLight Cable are indicated for use to provide real-time endoscopic visible and near-infrared fluorescence imaging. The Stryker AIM Light Source and SafeLight Cable enable surgeons to perform minimally invasive surgery using standard endoscope visible light as well as visual assessment of vessels, blood flow and related tissue perfusion, and at least one of the major extra-hepatic bile ducts (cystic duct, common bile duct and common hepatic duct), using near infrared imaging. Fluorescence imaging of biliary ducts with the Stryker AIM Light Source and SafeLight Cable are intended for use with standard of care white light and, when indicated, intraoperative cholangiography. The devices are not intended for standalone use for biliary duct visualization.

Device Story

Stryker AIM Light Source and SafeLight Cable provide endoscopic illumination for visible and near-infrared (NIR) fluorescence imaging; used during minimally invasive surgery. System components include light source console, light cable, camera control unit, coupler, laparoscope, and imaging agent (indocyanine green, ICG). Operation: ICG administered intravenously; laser light from source illuminates surgical site via laparoscope; ICG absorbs laser light and emits infrared signal; camera captures emission, processes image, and displays on surgical monitor. Used by surgeons in clinical settings. Output assists surgeons in visual assessment of vessels, blood flow, tissue perfusion, and biliary ducts. Device is not for standalone biliary visualization; intended as adjunct to white light and intraoperative cholangiography. Current submission updates labeling to include compatibility with additional ICG kit configurations.

Clinical Evidence

No new clinical data provided. Substantial equivalence is supported by previous clinical, animal, and bench testing conducted for the predicate device (K142310). Current submission relies on bench testing to demonstrate compatibility with the additional ICG kit configurations.

Technological Characteristics

System utilizes laser light source for visible and NIR illumination. Components include light source console, fiber-optic SafeLight cable, and laparoscope. Imaging relies on excitation of intravenously administered ICG dye. System is designed for integration with existing endoscopic camera control units and surgical displays. No changes to materials or energy source from predicate.

Indications for Use

Indicated for patients undergoing minimally invasive surgery requiring real-time endoscopic visible and near-infrared fluorescence imaging for visual assessment of vessels, blood flow, tissue perfusion, and major extra-hepatic bile ducts.

Regulatory Classification

Identification

An endoscope and accessories is a device used to provide access, illumination, and allow observation or manipulation of body cavities, hollow organs, and canals. The device consists of various rigid or flexible instruments that are inserted into body spaces and may include an optical system for conveying an image to the user's eye and their accessories may assist in gaining access or increase the versatility and augment the capabilities of the devices. Examples of devices that are within this generic type of device include cleaning accessories for endoscopes, photographic accessories for endoscopes, nonpowered anoscopes, binolcular attachments for endoscopes, pocket battery boxes, flexible or rigid choledochoscopes, colonoscopes, diagnostic cystoscopes, cystourethroscopes, enteroscopes, esophagogastroduodenoscopes, rigid esophagoscopes, fiberoptic illuminators for endoscopes, incandescent endoscope lamps, biliary pancreatoscopes, proctoscopes, resectoscopes, nephroscopes, sigmoidoscopes, ureteroscopes, urethroscopes, endomagnetic retrievers, cytology brushes for endoscopes, and lubricating jelly for transurethral surgical instruments. This section does not apply to endoscopes that have specialized uses in other medical specialty areas and that are covered by classification regulations in other parts of the device classification regulations.

Special Controls

*Classification* —(1)*Class II (special controls).* The device, when it is an endoscope disinfectant basin, which consists solely of a container that holds disinfectant and endoscopes and accessories; an endoscopic magnetic retriever intended for single use; sterile scissors for cystoscope intended for single use; a disposable, non-powered endoscopic grasping/cutting instrument intended for single use; a diagnostic incandescent light source; a fiberoptic photographic light source; a routine fiberoptic light source; an endoscopic sponge carrier; a xenon arc endoscope light source; an endoscope transformer; an LED light source; or a gastroenterology-urology endoscopic guidewire, is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 876.9.(2) Class I for the photographic accessories for endoscope, miscellaneous bulb adapter for endoscope, binocular attachment for endoscope, eyepiece attachment for prescription lens, teaching attachment, inflation bulb, measuring device for panendoscope, photographic equipment for physiologic function monitor, special lens instrument for endoscope, smoke removal tube, rechargeable battery box, pocket battery box, bite block for endoscope, and cleaning brush for endoscope. The devices subject to this paragraph (b)(2) are exempt from the premarket notification procedures in subpart E of part 807of this chapter, subject to the limitations in § 876.9.

Predicate Devices

Reference Devices

Submission Summary (Full Text)

{0}------------------------------------------------ Image /page/0/Picture/0 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which consists of the letters "FDA" in a blue square, followed by the words "U.S. FOOD & DRUG" in blue, with the word "ADMINISTRATION" underneath. January 18, 2018 Stryker Jay Nayar Senior Regulatory Affairs Specialist 5900 Optical Court San Jose, California 95138 Re: K173866 Trade/Device Name: Stryker AIM Light Source and SafeLight cable Regulation Number: 21 CFR 876.1500 Regulation Name: Endoscope and Accessories Regulatory Class: Class II Product Code: OWN Dated: December 18, 2017 Received: December 20, 2017 Dear Jay Nayar: We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading. If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register. Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050. {1}------------------------------------------------ Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance. For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/MedicalDevices/DeviceRegulationandGuidance/) and CDRH Learn (http://www.fda.gov/Training/CDRHLearn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (http://www.fda.gov/DICE) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100). Sincerely. # Jennifer R. Stevenson -S3 For Binita S. Ashar, M.D., M.B.A., F.A.C.S. Director Division of Surgical Devices Office of Device Evaluation Center for Devices and Radiological Health Enclosure {2}------------------------------------------------ ### Indications for Use 510(k) Number (if known) K173866 Device Name Stryker AIM Light Source and SafeLight cable Indications for Use (Describe) The Stryker AIM Light Source and SafeLight Cable are indicated for use to provide real-time endoscopic visible and near-infrared fluorescence imaging. The Stryker AIM Light Source and SafeLight Cable enable surgeons to perform minimally invasive surgery using standard endoscope visible light as well as visual assessment of vessels, blood flow and related tissue perfusion, and at least one of the major extra-hepatic bile ducts (cystic duct, common bile duct and common hepatic duct), using near infrared imaging. Fluorescence imaging of biliary ducts with the Stryker AIM Light Source and SafeLight Cable are intended for use with standard of care white light and, when indicated, intraoperative cholangiography. The devices are not intended for standalone use for biliary duct visualization. | Type of Use (Select one or both, as applicable) | |-------------------------------------------------| |-------------------------------------------------| X Prescription Use (Part 21 CFR 801 Subpart D) Over-The-Counter Use (21 CFR 801 Subpart C) ### CONTINUE ON A SEPARATE PAGE IF NEEDED. This section applies only to requirements of the Paperwork Reduction Act of 1995. ### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW * The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to: > Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov "An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number." {3}------------------------------------------------ ## Section 7. 510(k) Summary #### General Information 1. | 510(k) Sponsor | Stryker Endoscopy | |-------------------------|---------------------------------------------------------------------| | Address | 5900 Optical Court<br>San Jose, CA 95138 | | FDA Registration Number | 2936485 | | Correspondence Person | Jay Nayar<br>Sr. Regulatory Affairs Specialist<br>Stryker Endoscopy | | Contact Information | Email: jay.nayar@stryker.com<br>Phone: 408-754-2729 | | Date Prepared | December 18, 2017 | #### 2. Proposed Device Proposed Device: | Proprietary Name | Stryker® AIM Light Source and SafeLight Cable | |---------------------|-----------------------------------------------| | Common Name | Light Source, Illuminator | | Classification Name | Confocal optical imaging | | Regulation Number | 21 CFR 876.1500 | | Product Code | OWN | | Regulatory Class | II | #### Predicate Device: 3. | Proprietary Name | Stryker® Infrared Fluorescence (IRF) Light Source and SafeLight<br>Cable | |---------------------------------------------------------------|--------------------------------------------------------------------------| | Premarket Notification | K142310 | | Common Name | Light Source, Illuminator | | Classification Name | Confocal Optical Imaging | | Regulation Number | 21 CFR 876.1500 | | Product Code | OWN | | Regulatory Class | II | | This device has not been subjected to a design-related recall | | #### Reference Device 4. | Proprietary Name | Stryker® Infrared Illuminating System (IRIS)[Consisting of AIM Light<br>Source and IRIS Ureteral Kit ] | |---------------------------------------------------------------|--------------------------------------------------------------------------------------------------------| | Premarket Notification | K151243 | | Common Name | Light Source, Illuminator | | Classification Name | Fiberoptic light ureteral catheter, Confocal optical imaging | | Regulation Number | 21 CFR 876.4020, 21 CFR 876.1500, | | Product Code | FCS, OWN | | Regulatory Class | II | | This device has not been subjected to a design-related recall | | {4}------------------------------------------------ #### Device Description ട്. This is a Special 510(k) submission for a proposed labeling change that impacts the Stryker® IRF Light Source and SafeLight Cable (referred to as Stryker AIM Light Source hereafter) which was cleared under K142310. The Stryker AIM Light Source is part of the Stryker® Infrared Fluorescence (IRF) Imaging System, which is an endoscopic illumination and imaging system for real-time high definition (HD) visible light and near-infrared dye fluorescence imaging of indocvanine green (ICG) during minimally invasive surgery. The Stryker® Infrared Fluorescence (IRF) Imaging System consists of the following main components: - A light source console and a light cable for outputting light within a visible light spectrum ● as well as near-infrared light spectrum (focus of this submission). - A camera control unit for processing near-infrared and visible light images, and a coupler ● that is attached to the laparoscope and a camera head (cleared under K132785). - . A laparoscope for visible light and near-infrared light illumination and imaging (cleared under K142310) - An imaging agent kit containing ICG (Kit containing IC-Green® drug and Aqueous . Solvents, manufactured by Akorn (NDA 011525), cleared under K142310) There are no proposed changes to any other components of the Stryker® Infrared Fluorescence (IRF) system. The Stryker AIM Light Source uses an illuminator with a laser light source to illuminate the area of surgery. The ICG is administered intravenously prior to image obtainment. A laser light is illuminated to the site of the surgery using a laparoscope. Upon the absorption of laser light, the ICG in the vessels is excited and emits infrared light at a different wave length. The camera system captures the infrared emission, processes the image and displays it on a surgical display. The proposed labeling change is to allow the Stryker AIM Light Source to work with two additional kit configurations of the same imaging agent (ICG). Specifically, both device user manual and ICG kit Instruction for Use are updated to adequately cross-label and cross-reference the compatible device/ICG kit(s) per FDA guidance . It is to be noted that Stryker Endoscopy is not the manufacturer of the Imaging agent kits or the ICG drug; the proposed change in labeling does not necessitate any change in the imaging drug formulation, dosage, rate, or route of administration, #### 6. Imaging Agent Description The proposed versions of ICG kits were previously cleared by Novadaq Technologies Inc under K150956 and K161792. There are two configurations of the ICG kits, one includes IC-Green from Akorn (NDA 011525), and the other one includes Indocyanine Green for Injection, USP from Diagnostic Green (formerly Pulsion, ANDA 040811). Indocyanine Green for Injection, USP is the <sup>1</sup> Guidance for Industry: New Contrast Indication Considerations for Devices and Approved Drug and Biological Products. {5}------------------------------------------------ generic version of IC-Green. The configurations are: - . PINPOINT PAQ® AIM (containing IC-Green®, NDA 011525); - PINPOINT PAQ® AIM (containing Indocyanine Green for Injection, USP, ANDA 040811); Each PINPOINT PAQ® AIM contains sufficient amounts of ICG, diluents, syringes, and Instructions for Use for guidance for the end user. Each configuration of the PINPOINT PAQ® AIM has a unique part number assigned to it, and a corresponding Instructions for Use insert. The Instructions for Use also have unique part numbers for ease of assembly of the PAQs and guidance for the end user regarding compatible imaging devices. #### 7. Indications for Use The Stryker® AIM Light Source and SafeLight Cable are indicated for use to provide realtime endoscopic visible and near-infrared fluorescence imaging. The Stryker® AIM Light Source and SafeLight Cable enable surgeons to perform minimally invasive surgery using standard endoscope visible light as well as visual assessment of vessels, blood flow and related tissue perfusion, and at least one of the major extra-hepatic bile ducts (cystic duct, common bile duct and common hepatic duct), using near infrared imaging. Fluorescence imaging of biliary ducts with the Stryker® AIM Light Source and SafeLight Cable are intended for use with standard of care white light and, when indicated, intraoperative cholangiography. The devices are not intended for standalone use for biliary duct visualization. Please note this is the same indications for use as the previously cleared one for the Stryker AIM Light Source under K142310. #### 8. Comparison of Technological Characteristics with the Predicate Device The intent of this premarket submission is to modify the labeling of the Stryker® AIM Light Source to show compatibility with two additional ICG kits. The ICG kits consist of the imaging agent (indocyanine green or ICG), diluents, syringes and device specific Instruction For Use. All proposed modifications have been assessed through risk analysis, and where necessary, verification and validation activities were performed. These activities did not raise new questions of safety or effectiveness. Given this, the proposed Stryker® AIM Light Source is substantially equivalent to the respective legally marketed predicate device (cleared under K142310) in regard to intended use, fundamental scientific technology, operational principles, and performance attributes. #### 9. Performance Testing The modification proposed within this submission for the Stryker AIM Light source does {6}------------------------------------------------ not impact the performance characteristics that were cleared in the previous submissions. Bench testing was conducted to demonstrate the compatibility of the Stryker AIM Light Source with the PINPOINT PAQ® AIM . Please note clinical studies, animal studies, human experience, bench testing and in vitro testing were conducted to evaluate the safety and performance of the Stryker® IRF Light Source and SafeLight Cable which was cleared under K142310. Additional clinical testing is not required to demonstrate substantial equivalence for this submission as the proposed modification is primarily to the labeling to satisfy the cross labeling requirements per FDA guidance2. ### 10. Conclusion The information presented within this special premarket submission demonstrates that the Stryker AIM Light source is compatible with the PINPOINT PAQ® AIM, and is substantially equivalent to the Stryker® Infrared Fluorescence (IRF) Light Source and SafeLight Cable (cleared under K142310). <sup>2</sup> Guidance for Industry: New Contrast Indication Considerations for Devices and Approved Drug and Biological Products.
Innolitics

Panel 1

/
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...